Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Temperature controls

Temperature Control.—For moderate accuracy, the temperature of the column should be controlled to within 0.2 K if the solute vapour pressure is known to be better than 1 per cent. A good air thermostat is capable of this control. It is, however, much easier to control the temperature of the water bath, and a simple on-off relay in conjunction with a toluene regulator and backing heater is capable of controlling the temperature to 0.01 K. This is the order of temperature control required for the accurate determination of activity coefficients. [Pg.55]

Because it is necessary to have the column completely thermostatted and because the injector cannot usually be immersed in a water bath, it is convenient to have the bath mounted in a frame on pulleys so that it can be moved vertically. In this way the column can be replaced without affecting other parts of the flow line and, furthermore, the column can be taken out of the bath at the end of a run, thus reducing unnecessary loss of solvent. [Pg.55]

Temperature-control problems are really heat transfer problems, whether the mechanism is radiation, conduction, or convection. Al- [Pg.74]

Because most heat transfer processes have variable parameters-heat transfer coefficient, dead time, etc.-which vaiy with flow, care has been taken to choose an example free of these complications, to better introduce the subject. The example chosen is that of a stirred tank reactor cooled by a constant flow of liquid circulating through its jacket. [Pg.75]

The temperature controller, as shown in Fig. 3.6, adds cold water to the circulating coolant, in order to remove the heat of reaction. There are five important dynamic elements in the process  [Pg.75]

Because all the heat leaving the reactor flows through the walls and into the coolant, the capacities of reactants, walls, and coolant interact. But in view of the slight heat capacity of the bulb, its time constant does not significantly interact with the others. Basically the process is four-capacity plus dead-time. [Pg.75]

To determine the values of the time constants, an unsteady-state heat balance must be written across each heat transfer surface. The equation [Pg.75]

For controlling the temperature of baths or flowing liquids and of various portions of an apparatus (such as column jackets, tube heaters, etc.) contact thermometers are largely used they are obtainable with standard ground joints (e.g., NS 14.5). Contact thermometers can be obtained either with contacts set to definite temperatures or with variable temperature. In the variable type (Fig. 366) the upper thermometer scale is used to set the required temperature, while the lower one carries the electrical contact. A contact thermometer with a spiral bulb with a large surface area (Fig. 367a) [Pg.445]

Contact thermometers of a maximum-minimum type are also manufactured these make contact both on exceeding an upper limit and on dropping below a lower limit. By means of a fine adjusting screw the temperatures may be set with an accuracy of 0.1 °C [23]. Temperature control may be arranged to follow a variable reference temperature by moving the thermometer contact with a synchronous motor. [Pg.446]

Quite satisfactory experience has been gathered with contact thermometers having a special three-pole plug and sliding ground joint, the latter ensuring that immersion can be continuously varied [24]. [Pg.446]

Between the contact thermometer [25] and the controlled apparatus a relay is placed, since a contact thermometer is not capable of carrying a normal heating current. Relays with vertical mercury switches have proved the moat satisfactory in laboratory use, as they contain no parts subject to wear, such as bearings, levers or flexible leads [26]. [Pg.446]

Transistor relays may also be used in coimection with contact manometers ( /. chap. 8.3.1 and 8.3.1.1). They are commercially available as a combination of standing or supported device. They operate on the normally open and normally closed principles where a circuit is switched on and off, respectively, on making a contact. These relays are further employed for photoelectric circuits, e.g. in the level control of liquids and in vacuum control (c/. chap, 8.6 and 8.3.1.1). As safety relays thev switch off the circuit if the contact thermometer fails. [Pg.446]

Today s extruder temperature control systems are capable of maintaining temperature within a range of 1 °E Precision of this level is acceptable for practically all extrusion applications however, when variation exceeds just a few degrees, variations in product output can result. Therefore, it is important that accurate temperature control circuits are employed. [Pg.37]

It is common for an extrusion line to be separated into several temperature control zones. The number of zones depends on the length of the barrel, the type of adapter or transfer line to the die, and the size and complexity of the die. An extruder may have as few as three or well over ten zones. Each zone, or circuit, contains up to four of the following components temperature controller, temperature sensor, heating unit, and cooling unit. Usually only barrel temperature zones utilize a cooling unit. [Pg.37]

Simulator Exercise The temperature controllers have an on/off switch and allow set-point adjustment between 0 and 500 °E Heating and cooling time is compressed for user-friendliness. [Pg.37]

Another component in the circuit is a temperature sensor. The temperature sensor measures the actual temperature in the zone, converting it to a signal that is sent to the controller. The sensor may be in contact with metal or polymer. There are various types of sensors, including thermistors and infrared detectors, but the most common type is the thermocouple. [Pg.38]

Thermocouples are primarily used to measure the metal temperature in a particular control zone. Usually a well is drilled into the metal component to locate the couple junction as close as possible to the interior surface (where polymer is flowing). When installing a thermocouple, it is good practice to make sure the probe tip is in contact with the base of the well before securing the connection. [Pg.39]

Retention, particularly on reversed phases, is temperature dependent. Increasing temperature reduces the retention for most compounds (with a few exceptions, notably for polyethylene glycols). [Pg.61]

This means that in laboratories with temperature fluctuations (particularly during the summer), the columns should be kept at constant temperature in a column oven. [Pg.61]

Improved peak shapes at higher temperatures are often related to the fact that the diffusion coefficient D increases with increasing temperature, as shown by the Stokes-Einstein equation (Equation 3.3), improving the mass transfer in both phases. The viscosity decreases at the same time  [Pg.61]

Since the B-term of the van Deemter equation (Equation 1.2) is of little significance in HPLC, the increased diffusion coefficient with increasing temperature will reduce the C-term more than the increase of the B-term. [Pg.61]

Since elevated temperatures reduce the viscosity of the mobile phase and thereby reduce the backpressure, higher speed and improved resolution can often be obtained by selecting a column temperature above room temperature. [Pg.62]

As noted earlier, control of the column s temperature is critical to attaining a good separation in gas chromatography. For this reason the column is located inside a thermostated oven. In an isothermal separation the column is maintained at a constant temperature, the choice of which is dictated by the solutes. Normally, the tern- [Pg.568]

There are three general methods for maintaining a RV at a particular temperature thermostats, furnaces and vapour baths . [Pg.8]

At temperatures close to 20 20°, water is a suitable liquid for a thermostat. At higher temperatures it is more convenient to use a fluid such as silicone oil, or above 200°, molten metals. A typical furnace is shown in Fig. 5. A silica former is wound with nichrome wire such that the pitch decreases from one end to the centre and then increases again to the other end. Since heat losses are greatest at the ends this provides a rough correction. An inconel tube also evens out the tempera- [Pg.8]

EXPERIMENTAL METHODS FOR SLOW REACTIONS Reaction Furnace Circuit [Pg.10]

Temperature measurement is almost invariably made using thermocouples . The latter must be constructed from fine wire °and have a fast response such that even a very small temperature change may be measured precisely.  [Pg.10]

For maximum accuracy the calibrated volumes and the manifold should be maintained at constant temperature by using a water or constant-temperature air jacket. The liquid-nitrogen level should be maintained as near constant as possible to prevent any variation in the effective volume of the sample cell. [Pg.152]

Because the shape of the isotherm establishes the quantity adsorbed at various relative pressures, it is often difficult to predict a priori the exact dosing quantity to admit to the calibrated volumes in order to obtain the desired equilibrium pressure. If the first dose of adsorbate equilibrates at a relative pressure higher than desired because of insufficient adsorption, it is best to re-evacuate the sample and admit a smaller initial dose. Thereafter, smaller dose requirements can be anticipated for subsequent data points. Conversely, if the first dose equilibrates at a lower relative pressure than desired, it is necessary only to dose a second time to attain a higher equilibrium pressure. [Pg.152]

Here is the adsorption potential within the constriction. Equation [Pg.153]

For a surface containing area in pores which possess constrictions leading into wide inner bodies, one can approximate the ratio of the adsorption rates of the more accessible parts of the surface and the portion of the surface in the wide inner body. If the adsorption potential in a constriction is 2 kcal more than that on an open surface, the ratio of adsorption rates at liquid-nitrogen temperature, assuming the same preexponential value A, will be [Pg.154]

The consequences of this type of activated physical adsorption is not only that the quantity adsorbed can lie off the isotherm but also that the measured quantity of adsorption is far less than the equilibrium value. No experiments have been conducted to illustrate whether or not the quantity adsorbed lies within the hysteresis loop. The occasional failure of the vacuum volumetric method to agree with the dynamic method, which is not subject to any pressure overshoot, may in part be attributed to this phenomenon. [Pg.154]

Up to this point, we have discussed the importance of fast mixing in conducting extremely fast reactions in a highly controlled manner. Another important issue should be considered when extremely fast reactions are conducted on a preparative scale. [Pg.87]

Fast reactions are often highly exothermic. The energy diagram shown in [Pg.87]

Activation of a reactant molecule (to a reactant ) leads to a decrease in the activation energy of the reaction (from AG to AG ), which leads to a higher reaction rate. However, activation of a reactant molecule leads to an increase in the reaction free energy (from AG to AG ), which often makes the reaction more exothermic. An exothermic reaction may lead to a temperature increase in the reaction environment if heat transfer is not efficient. Such a case might be problematic. The product may undergo further reactions or decomposition, which leads to the formation of undesired by-products. Therefore, the selectivity of a reaction, in general, decreases with an increase in the temperature. [Pg.88]

In the case of oyostats, liquid nitrogen or helium is used to cool the sample and a heater coil is used to raise the temperature of the sample relative to the temperature that would otherwise be set by the nitrogen or helium. It is usually preferable that the cryogen does not come into contact with the sample to ensure that there is no contamination or condensation on the sample which may, in some cases, affect the results. It is also preferable for the sample chamber not to be filled with air to avoid possible oxidation problems. For this reason it is often the case that the sample is placed in a separate chamber in the cryostat in an inert gas environment. [Pg.186]

There are often two temperatnre sensors in the system, one is positioned close to the heater coil and provides very stable control of temperature since there is immediate feedback of any temperature changes in the coil. The other is positioned [Pg.186]

Sample temperature sensor in mert gas sample chamber [Pg.186]

Heating equipment (furnaces and drying cabinets) often contains rheostats that can be used to roughly maintain the required temperature. [Pg.26]

Autotransformers are also widely used in laboratories. They allow the temperature to he controlled within a narrower interval. Let us consider a type JIATP-1 autotransformer (Fig. 12). Its front panel has six terminals. It is connected to 220-V mains via terminals 1 and 3, [Pg.26]

The upper panel of the autotransformer has a scale graduated from 0 to 250 V, and a handle provided with a pointer. Turning of the handle changes the voltage of the current in the furnace winding and thus sets the required temperature. Connect a furnace to the mains via an autotransformer as follows place the pointer on the upper autotransformer scale opposite the zero graduation, next connect the furnace to it and connect the autotransformer to the mains in accordance with their voltage. [Pg.26]

In a molecular dynamics calculation, you can add a term to adjust the velocities, keeping the molecular system near a desired temperature. During a constant temperature simulation, velocities are scaled at each time step. This couples the system to a simulated heat bath at Tq, with a temperature relaxation time ofT. The velocities are scaled by a factor X, where [Pg.72]

If the coupling parameter (the Bath relaxation constant in HyperChem), t, is too tight ( 0.1 ps), an isokinetic energy ensemble results rather than an isothermal (microcanonical) ensemble. The trajectory is then neither canonical or microcanon-ical. You cannot calculate true time-dependent properties or ensemble averages for this trajectory. You can use small values of T for these simulations  [Pg.72]

If the Bath relaxation constant, t, is greater than O.I ps, you should be able to calculate dynamic properties, like time correlation functions and diffusion constants, from data in the SNP and/or CSV files (see Collecting Averages from Simulations on page 85). [Pg.72]

Note This method of temperature regulation does not give all properties of the canonical ensemble. In particular, you cannot calculate Cy, heat capacity at constant volume. [Pg.72]

Berendsen, H.J.C., Postma, J.P.M. van Gunsteren, W.R DiNola, A. Haak, J.R. Molecular Dynamics with coupling to an external bath J. Chem. Phys. 81 3684, 1984. [Pg.72]

The dynamic behavior of an extruder is significantly determined by the temperature control system on the extruder. It is, therefore, important to understand the basic characteristics of the various temperature control systems. Most control systems are closed-loop or feedback systems. The variable to be controlled is measured and this information is sent to a control unit. From the control unit a signal is sent to an actuator that adjusts the process such that the control variable is as close as possi- [Pg.116]

In addition to the measnrement of temperature, it is often necessary to maintain a constant temperature. The importance of this type of control in experimental physical chemistry is illustrated by the fact that 30 of the 48 experiments described in this book require temperature control of some kind. Many physical quantities such as rate constants, equi-librinm constants, and vapor pressnres are sensitive functions of temperatnre and mnst be measmed at a known temperatnre that is held constant to within 0.1 K or better. Certain physical techniques ate even more demanding for example, the measurement of the coexistence cnrve in Exp. 16 reqnires that the temperature be controlled to within 0.02 K. [Pg.576]

There are two basic methods of achieving a constant temperature [Pg.577]

Phase equilibrium is maintained at constant pressure between two phases of a pure substance or three phases of a two-component system (eutectic). [Pg.577]

A temperature sensor provides a feedback signal to control the input of heat (or refrigeration cooling in some cases) in order to maintain the temperature close to any arbitrary desired value. [Pg.577]

Method 1 is the simplest approach method 2 is more flexible and generally more useful. [Pg.577]

Current versions of resistively heated pyrolyzers incorporate small computers to control and monitor the filament temperature. These computers may be used to control the voltage used, adjust for changes in resistance as the filament heats, and compensate for differences when broken filaments are replaced. In addition, instruments have been designed that include photodiodes, which are used by the computer to measure the actual temperature of the filament during a run. Other instruments include a small thermocouple welded directly to the filament for temperature readout, or use the computer to measure the resistance of the filament itself and make adjustments as needed during a program. [Pg.39]

Since the temperature and the rate of heating the filament are completely variable, the instrument may control these parameters independently, as single steps or multiple steps. This gives the analyst the control to select any final pyrolysis temperature and to heat the sample to this temperature at any desired rate. Instruments are commercially available that heat as slowly as 0.01°C/minute and as rapidly as 30,000°C/second. [Pg.39]

By contrast with ideal models, practical reactors must consider many factors other than variations in temperature, concentration and residence time. Consider the temperature control of the reactor first. [Pg.121]

In the first instance, adiabatic operation of the reactor should be considered since this leads to the simplest and cheapest reactor design. If adiabatic operation produces an unacceptable rise in temperature for exothermic reactions or an unacceptable fall in temperature for endothermic reactions, this can be dealt with in a number of ways  [Pg.121]

As an example, consider the production of ethylene oxide, which uses a silver-supported catalyst1  [Pg.122]

A parallel reaction occurs leading to a selectivity loss  [Pg.122]

As another example, consider the application of a fixed-bed tubular reactor for the production of methanol. Synthesis gas (a mixture of hydrogen, carbon monoxide and carbon dioxide) is reacted over a copper-based catalyst2. The main reactions are [Pg.122]

As well as the thermocouple system for measurement, a second, entirely separate, thermocouple system is provided to sense the furnace temperature and is connected to the furnace control circuits. The same thermocouple is not used for the two purposes because the criteria for them are very different. The measuring couple has to be positioned as near to the sample as possible. Sometimes this is just below the sample (see [Pg.16]

On the other hand the furnace thermocouple has to be able to respond rapidly to furnace temperature. If there is a lag in time between furnace power being turned up and temperature rise being detected, then the system will tend to go into temperature swings instead of a steady linear rise. For this reason the furnace measuring couple is positioned as near as possible to the source of heat, which is the resistance wire winding. The couple has to be electrically insulated from the winding, but is at least embedded in the furnace cement coating on the furnace. [Pg.17]

Phthalic anhydride Naphthalene is oxidized by air to phthalic anhydride in a bubbling fluidized reactor. Even though the naphthalene feed is in liquid form, the reaction is highly exothermic. Temperature control is achieved by removing heat through vertical tubes in the bed to raise steam [Graham and Way, Chem. Eng. Prog. 58 96 (January 1962)]. [Pg.17]

Acrylonitrile Acrylonitrile is produced by reacting propylene, ammonia, and oxygen (air) in a single fluidized bed of a complex catalyst. Known as the SOHIO process, this process was first operated commercially in 1960. In addition to acrylonitrile, significant quantities of HCN and acetonitrile are produced. This process is also exothermic, and temperature control is achieved by raising steam inside vertical tubes immersed in the bed [Veatch, Hydrocarbon Process. Pet. Refiner 41 18 (November 1962)]. [Pg.17]

Polyethylene The first commercial fluidized-bed polyethylene plant was constructed by Union Carbide in 1968. Modern units operate at a temperature of approximately 100°C and a pressure of [Pg.17]

Although typically not thought of as a platform for more than presenting a flat surface, platens are being exploited in several ways to enhance overall CMP performance. Included in these approaches are temperature control, slurry delivery routing, and pressure control. [Pg.24]

Second, friction during CMP generates heat that can affect the reaction kinetics as well as soften the polish pad. On most tools, heat transfer away [Pg.24]

Temperature control is accomplished in one of three general ways. One method is by controlling the temperature of the platen, usually by means of an integral channel in the platen through which temperature-controlled heat transfer fluid flows. Second, the temperature of the slurry itself can be regulated prior to being dispensed onto the platen. Finally, a means of heating the backside of the wafer can be built into the carrier [42,43]. [Pg.25]

Most separations in liquid chromatography are performed at room temperature for convenience and because ambient temperatures provide reasonable column efficiency for low molecular mass solutes. Elevated temperatures are commonly used in ion-exchange chromatography to improve mass transfer kinetics and in size-exclusion chromatography to provide adequate solubility for polymers in useful mobile phases. Wider interest in temperature control and high-temperature separations in general results from improved precision of retention measurements (section 1.1.1), greater column efficiency (section 1.5.2), the use of temperature as a variable for method development (section 4.4.4), and shorter separation times due to the more favorable use of the column inlet pressure [70,71]. [Pg.449]

Column thermostats are required to maintain a constant temperature ( 0.1 °C) in both time and space. Differences in temperature between the mobile phase entering [Pg.449]

Heat is usually applied in various amounts and in different locations, whether in a metal plasticating barrel (extrusion, injection molding, etc.) or in a metal mold/die (compression, injection, thermoforming, extrusion, etc.). With barrels a thermocouple is usually embedded in the metal to send a signal to a temperature controller. In turn, it controls the electric power output device regulating the power to the heater bands in different zones of the barrel. The placement of the thermocouple temperature sensor is extremely important. The heat flow in any medium sets up a temperature gradient in that medium, just as the flow of water in a pipe sets up a pressure drop, and the flow of electricity in a wire causes a voltage drop. [Pg.15]

Barrels are made of steel, which is not a particularly good conductor of heat (being ten times worse than copper). Thus there is a gradient in the steel barrel from the outside of the barrel to the inside next to the plastic. In 31 2 in. (88.9 mm) and 4 in. (114.3 mm) extruder barrels, these gradients or differences in temperature can routinely be 75 to 100°F (23.9 to 32.8°C) or more, as the zone heaters pump in heat or zone coolers take excess heat out. Yet, for years users routinely accepted extruders with sensors mounted in very shallow wells, or, even worse, mounted in the heating/cooling jacket. [Pg.15]

Consider a barrel with a shallow well for its sensor. Assume a perfect temperature controller set at 400°F (204°C). There is a 75°F gradient from the outside to the inside of the barrel thus the actual temperature down near the plastic would be 325°F with the sensor set at 400°F. If the extruder started to generate too much heat, the temperature could reach 475°F before the sensor detected the increase. With this on-off control action, even with the controller set at 400°F the plastic temperature variation is 150°F. The result could be poor product performance and increased cost to process the plastic. [Pg.15]

The DUO-Sense process (Holton/Harrel Inc., U.S. Patent 4,272,466 June 9, 1981) solved this problem, retaining the advantages of both deep and [Pg.15]

These on-off controllers are unsatisfactory for a loading having a long time constant, such as an extruder barrel, a die adapter, a die, and so forth. The temperature will oscillate violently at an amplitude that is set not by the characteristic of the controller, but by the delay in the load, as reviewed in Fig. 1-7. To reduce this variation, a proportional control was developed. It is similar to the on-off, but operates in between full on and off, with its output proportional to the deviation of temperature from the set point value (Fig. 1-8). Variations still exist with this system, but they are less than those of the on-off control. [Pg.16]

Forming requires thorough, fast, and uniform radiant heat from the surface to the core to the surface of the sheet or film. As a general guide, to achieve these conditions, sheet plastics over 0.040 in. (1.02 mm) should use sandwich-type (under and over the sheet) heater banks. To ensure that sufficient heat [Pg.495]

The cycle time is controlled by the heating and cooling rates, which in turn depend on the following factors the temperature of the heaters and the cooling medium, the initial temperature of the sheet, the effective heat transfer coefficient, the sheet thickness, and thermal properties of the sheet. [Pg.495]

Different plastics absorb radiant heat more efficiently at various wavelengths, which in turn are effected by the temperature of the emitting heater. Thus it requires that the proper wavelength be used for what the [Pg.495]

The alternative path of evolution was to the continuous solution process, first demonstrated with the tower process by I. G. Farben and implemented by Dow and others as either towers or tanks filled with heat transfer tubes. These [Pg.46]

Because of the rate limitations of the tower and tube-tank processes that were primarily heat transfer constraints, further developments in the continuous solution process for crystal polystyrene (GP) were aimed at improving heat transfer. One obvious solution was to incorporate agitation of some type in the reactor. Although at Dow the incorporation of agitation in the reactors came about with the development of rubber-modified polystyrene [11], and this aspect will be discussed in a later section, agitation also significantly raises the heat transfer [Pg.47]

The use of boiling heat transfer raises the maximum conversion rates that can be controlled significantly beyond that of the agitated towers filled with heat transfer tubes. The main limitation that occurs is the removal of the vapor bubbles from the polymer solution. As the viscosity of the polystyrene solution increases rapidly with conversion, this becomes most limiting when the viscosity of the polystyrene solution exceeds 1000P (dPa/s). Below this viscosity, conversion rates of 40%/h can be controlled, but above this viscosity, the polymer mass foams up into the condenser and temperature control is lost, so that the maximum conversion rate decreases rapidly at high polymer concentrations. [Pg.48]

Similar to any oxidation reaction, the MMF oxidation is an exothermic reaction. [Pg.323]

the first technical challenge in the design of an oxidation reactor deals with the methodology to remove the generated heat. Although the heat in the production of FDCA from HMF and MMF ( 850 kj/mol) is considerably less than that of TA from para-xylene ( 1300 kj/mol) [40], still the amount of heat is so high that very efficient cooling is required to avoid thermal runaways. [Pg.323]

Since the heat of evaporation of acetic acid of -1-23.7 kJ/mol is low compared with the overall heat formed in the oxidation ( 1300kJ/mol for TA), it is clear that considerable amounts of acetic acid need to be evaporated in order to maintain thermal stability. The amount of acetic acid that can be evaporated is directly related to the vapor pressure in the effluent gas flow. Since the vapor pressure is only a function of temperature, the actual operation temperature in the oxidation reactor is determined by the temperature required to obtain the required vapor pressure for safe heat removal in the reactor. Though counterintuitive, it means that higher operation temperatures are typically safer to operate for this catalyst system. For the TA process, this leads to a minimum operation temperature of around 180 °C. Higher temperatures are even better from a heat integration perspective but yield more unwanted by-products and overoxidation to CO2. [Pg.323]

The safe industrial oxidation of furanics using the Co/Mn/Br catalyst system needs a similar heat removal system as FDCA, like TA, is a very insoluble diacid, preventingthe use of jacketed cooling. In contrast to para-xylene, HMF, MMF, and AcMF are already partially oxidized on the benzylic positions, and consequently less heat is formed in their oxidation to FDCA, and reactor temperature may be [Pg.323]

1) Calculated based on heat of formation simulated by Aspen Plus and heat of combustion measured experimentally. [Pg.323]

A method which has given considerable satisfaction in the author s laboratory will be described. The method is shown in Fig. 2. The sample tube is surrounded by a slowly rising layer of hydrogen gas, which is, in turn surrounded by a glass tube and a double-walled evacuated tube. The annular space between the tube containing the hydrogen layer and the double-walled tube carries the refrigerating gas stream. [Pg.31]

The calibration of a magnetic susceptibility balance may be done by use of substances of known susceptibility. Two substances which have [Pg.31]

A convenient method for obtaining controlled temperatures in the range —175 to +25  [Pg.31]

In all calibrations with solid substances of supposedly known susceptibility the greatest care must be taken to avoid contamination with minute traces of ferromagnetic impurities, which may have a disastrous effect on the accuracy of any contemplated measurements. [Pg.32]

If the susceptibility seems to rise with increasing field the probability is that the calibrating agent contains a ferromagnetic impurity. [Pg.32]


However, the laboratory data seem to indicate that a constant concentration in the reactor to maintain 63 percent sulfuric acid would be beneficial. Careful temperature control is also important. These two factors would suggest that a continuous well-mixed reactor is appropriate. There is a conflict. How can a well-defined residence time be maintained and simultaneously a constant concentration of sulfuric acid be maintained ... [Pg.52]

Using a batch reactor, a constant concentration of sulfuric acid can be maintained by adding concentrated sulfuric acid as the reaction progresses, i.e., semi-batch operation. Good temperature control of such systems can be maintained, as we shall discuss later. [Pg.52]

Generally speaking, temperature control in fixed beds is difficult because heat loads vary through the bed. Also, in exothermic reactors, the temperature in the catalyst can become locally excessive. Such hot spots can cause the onset of undesired reactions or catalyst degradation. In tubular devices such as shown in Fig. 2.6a and b, the smaller the diameter of tube, the better is the temperature control. Temperature-control problems also can be overcome by using a mixture of catalyst and inert solid to effectively dilute the catalyst. Varying this mixture allows the rate of reaction in different parts of the bed to be controlled more easily. [Pg.56]

Figure 13.5 shows a flowsheet for the manufacture of phthalic anhydride by the oxidation of o-xylene. Air and o-xylene are heated and mixed in a Venturi, where the o-xylene vaporizes. The reaction mixture enters a tubular catalytic reactor. The heat of reaction is removed from the reactor by recirculation of molten salt. The temperature control in the reactor would be diflficult to maintain by methods other than molten salt. [Pg.332]

Hence, it is necessary to correct the temperature change observed to the value it would have been if there was no leak. This is achieved by measuring the temperature of the calorimeter for a time period both before and after the process and applying Newton s law of cooling. This correction can be reduced by using the teclmique of adiabatic calorimetry, where the temperature of the jacket is kept at the same temperature as the calorimeter as a temperature change occurs. This teclmique requires more elaborate temperature control and it is prunarily used in accurate heat capacity measurements at low temperatures. [Pg.1901]

Demand for temperature controlled troughs came from the material scientists who worked witli large molecules and polymers tliat establish viscous films. Such troughs allow a deeper understanding of tire distinct phases and tire transitions in LB films and give more complete pressure-area isotlienns (see d) below). [Pg.2611]

However, it is common practice to sample an isothermal isobaric ensemble NPT, constant pressure and constant temperature), which normally reflects standard laboratory conditions well. Similarly to temperature control, the system is coupled to an external bath with the desired target pressure Pq. By rescaling the dimensions of the periodic box and the atomic coordinates by the factor // at each integration step At according to Eq. (46), the volume of the box and the forces of the solvent molecules acting on the box walls are adjusted. [Pg.368]

A shallow metal vessel containing sand, the so-called sand bath, heated by means of a flame, was formerly employed for heating flasks and other glass apparatus. Owing to the low heat conductivity of sand, the temperature control is poor the use of sand baths is therefore not... [Pg.59]

The advantages of the above air bath are (1) simplicity and cheapness of construction (2)ease of temperature control (3) rapidity of cooling of the contents of the flask effected either by removing the asbestos covers or by completely removing the air bath and (4) the contents of the flask may be inspected by removing the asbestos covers. [Pg.60]

The controlled thermal decomposition of dry aromatic diazonium fluoborates to yield an aromatic fluoride, boron trifluoride and nitrogen is known as the Schiemann reaction. Most diazonium fluoborates have definite decomposition temperatures and the rates of decomposition, with few exceptions, are easily controlled. Another procedure for preparing the diazonium fluoborate is to diazotise in the presence of the fluoborate ion. Fluoboric acid may be the only acid present, thus acting as acid and source of fluoborate ion. The insoluble fluoborate separates as it is formed side reactions, such as phenol formation and coupling, are held at a minimum temperature control is not usually critical and the temperature may rise to about 20° without ill effect efficient stirring is, however, necessary since a continuously thickening precipitate is formed as the reaction proceeds. The modified procedure is illustrated by the preparation of -fluoroanisole ... [Pg.594]

Dissolve 46-5 g. (45-5 ml.) of aniUne in a mixture of 126 ml. of concentrated hydrochloric acid and 126 ml. of water contained in a 1-htre beaker. Cool to 0-5° in a bath of ice and salt, and add a solution of 36-5 g. of sodium nitrite in 75 ml. of water in small portions stir vigorously with a thermometer (1) and maintain the temperature below 10°, but preferably at about 5° by the addition of a httle crushed ice if necessary. The diazotisation is complete when a drop of the solution diluted with 3-4 drops of water gives an immediate blue colouration with potassium iodide - starch paper the test should be performed 3-4 minutes after the last addition of the nitrite solution. Prepare a solution of 76 g. of sodium fluoborate (2) in 150 ml. of water, cool, and add the chilled solution slowly to the diazonium salt solution the latter must be kept well stirred (1) and the temperature controlled so that it is below 10°. Allow to stand for 10 minutes with frequent stirring. Filter... [Pg.609]

Hydroxyquinoline ( oxine ). The technique adopted in this preparation is based upon the fact that, in general, the reactants glycerol, amine, nitro compound and sulphuric acid can be mixed with temperature control, and then maintained at any convenient temperature below 120° without any appreciable chemical reaction taking place. A pre-mix of the amine, glycerol and sulphuric acid, maintained at a temperature which keeps it fluid (60-90°), is added in portions to a reaction vessel containiug the nitro compound and warmed with stirring to 140-170° at which temperature the Skraup reaction takes place. [Pg.830]

Fig. 1. Addition of the reagent with temperature control and introduction of nitrogen. Fig. 1. Reaction vessel suitable for conversions in liquid ammonia. Fig. 1. Addition of the reagent with temperature control and introduction of nitrogen. Fig. 1. Reaction vessel suitable for conversions in liquid ammonia.
A traditional method for such reductions involves the use of a reducing metal such as zinc or tin in acidic solution. Examples are the procedures for preparing l,2,3,4-tetrahydrocarbazole[l] or ethyl 2,3-dihydroindole-2-carbox-ylate[2] (Entry 3, Table 15.1), Reduction can also be carried out with acid-stable hydride donors such as acetoxyborane[4] or NaBHjCN in TFA[5] or HOAc[6]. Borane is an effective reductant of the indole ring when it can complex with a dialkylamino substituent in such a way that it can be delivered intramolecularly[7]. Both NaBH -HOAc and NaBHjCN-HOAc can lead to N-ethylation as well as reduction[8]. This reaction can be prevented by the use of NaBHjCN with temperature control. At 20"C only reduction occurs, but if the temperature is raised to 50°C N-ethylation occurs[9]. Silanes cun also be used as hydride donors under acidic conditions[10]. Even indoles with EW substituents, such as ethyl indole-2-carboxylate, can be reduced[ll,l2]. [Pg.145]

If there is no external temperature control (using a simulated constant temperature bath), molecular dynamics simulations are constant energy. [Pg.77]

For a constant temperature simulation, a molecular system is coupled to a heat bath via a Bath relaxation constant (see Temperature Control on page 72). When setting this constant, remember that a small number results in tight coupling and holds the temperature closer to the chosen temperature. A larger number corresponds to weaker coupling, allowing more fluctuation in temper-... [Pg.77]

A pH electrode is normally standardized using two buffers one near a pH of 7 and one that is more acidic or basic depending on the sample s expected pH. The pH electrode is immersed in the first buffer, and the standardize or calibrate control is adjusted until the meter reads the correct pH. The electrode is placed in the second buffer, and the slope or temperature control is adjusted to the-buffer s pH. Some pH meters are equipped with a temperature compensation feature, allowing the pH meter to correct the measured pH for any change in temperature. In this case a thermistor is placed in the sample and connected to the pH meter. The temperature control is set to the solution s temperature, and the pH meter is calibrated using the calibrate and slope controls. If a change in the sample s temperature is indicated by the thermistor, the pH meter adjusts the slope of the calibration based on an assumed Nerstian response of 2.303RT/F. [Pg.492]

The previous discussion demonstrates that measurement of precise isotope ratios requires a substantial amount of operator experience, particularly with samples that have not been examined previously. A choice of filament metal must be made, the preparation of the sample on the filament surface is important (particularly when activators are used), and the rate of evaporation (and therefore temperature control) may be crucial. Despite these challenges, this method of surface ionization is a useful technique for measuring precise isotope ratios for multiple isotopes. Other chapters in this book discuss practical details and applications. [Pg.52]

For LC, temperature is not as important as in GC because volatility is not important. The columns are usually metal, and they are operated at or near ambient temperatures, so the temperature-controlled oven used for GC is unnecessary. An LC mobile phase is a solvent such as water, methanol, or acetonitrile, and, if only a single solvent is used for analysis, the chromatography is said to be isocratic. Alternatively, mixtures of solvents can be employed. In fact, chromatography may start with one single solvent or mixture of solvents and gradually change to a different mix of solvents as analysis proceeds (gradient elution). [Pg.249]

In practice, such a fractionation experiment could be carried out by either lowering the temperature or adding a poor solvent. In either case good temperature control during the experiment is important. Note that the addition of a poor solvent converts the system to one containing three components, so it is apparent that the two-component Flory-Huggins model is at best only qualitatively descriptive of the situation. A more accurate description would require a... [Pg.540]

In both of these pieces of apparatus, isothermal operation and optimum membrane area are obtained. Good temperature control is essential not only to provide a value for T in the equations, but also because the capillary attached to a larger reservoir behaves like a thermometer, with the column height varying with temperature fluctuations. The contact area must be maximized to speed up an otherwise slow equilibration process. Various practical strategies for presetting the osmometer to an approximate n value have been developed, and these also accelerate the equilibration process. [Pg.550]

Va2o-64, Self-Reactive Solid Type C, Temperature Controlled (2,2 -a2odi(iso-butyronitrile)). [Pg.224]


See other pages where Temperature controls is mentioned: [Pg.52]    [Pg.54]    [Pg.55]    [Pg.93]    [Pg.850]    [Pg.486]    [Pg.1563]    [Pg.72]    [Pg.72]    [Pg.77]    [Pg.400]    [Pg.469]    [Pg.499]    [Pg.499]    [Pg.174]    [Pg.106]    [Pg.72]    [Pg.72]    [Pg.94]    [Pg.568]    [Pg.549]    [Pg.629]    [Pg.966]    [Pg.80]    [Pg.126]   
See also in sourсe #XX -- [ Pg.7 ]

See also in sourсe #XX -- [ Pg.1072 ]

See also in sourсe #XX -- [ Pg.170 ]

See also in sourсe #XX -- [ Pg.1326 ]

See also in sourсe #XX -- [ Pg.230 ]

See also in sourсe #XX -- [ Pg.2 ]

See also in sourсe #XX -- [ Pg.69 ]

See also in sourсe #XX -- [ Pg.287 , Pg.328 ]

See also in sourсe #XX -- [ Pg.234 ]

See also in sourсe #XX -- [ Pg.542 , Pg.544 , Pg.553 ]

See also in sourсe #XX -- [ Pg.6 , Pg.73 , Pg.85 , Pg.144 ]

See also in sourсe #XX -- [ Pg.24 ]

See also in sourсe #XX -- [ Pg.229 ]

See also in sourсe #XX -- [ Pg.36 ]

See also in sourсe #XX -- [ Pg.322 , Pg.336 , Pg.338 ]

See also in sourсe #XX -- [ Pg.42 , Pg.45 , Pg.46 ]

See also in sourсe #XX -- [ Pg.114 , Pg.120 , Pg.134 , Pg.141 ]

See also in sourсe #XX -- [ Pg.17 ]

See also in sourсe #XX -- [ Pg.260 , Pg.261 , Pg.269 ]

See also in sourсe #XX -- [ Pg.416 ]

See also in sourсe #XX -- [ Pg.79 ]

See also in sourсe #XX -- [ Pg.7 ]

See also in sourсe #XX -- [ Pg.576 , Pg.577 , Pg.578 , Pg.579 , Pg.580 , Pg.581 , Pg.582 , Pg.583 , Pg.584 ]

See also in sourсe #XX -- [ Pg.233 , Pg.238 , Pg.375 , Pg.376 , Pg.377 ]

See also in sourсe #XX -- [ Pg.37 ]

See also in sourсe #XX -- [ Pg.320 ]

See also in sourсe #XX -- [ Pg.637 ]

See also in sourсe #XX -- [ Pg.44 , Pg.45 , Pg.47 , Pg.48 ]

See also in sourсe #XX -- [ Pg.42 , Pg.45 , Pg.46 ]

See also in sourсe #XX -- [ Pg.278 , Pg.281 ]

See also in sourсe #XX -- [ Pg.42 , Pg.45 , Pg.46 ]

See also in sourсe #XX -- [ Pg.283 ]

See also in sourсe #XX -- [ Pg.42 , Pg.45 , Pg.46 ]

See also in sourсe #XX -- [ Pg.84 ]

See also in sourсe #XX -- [ Pg.87 , Pg.88 , Pg.89 , Pg.90 , Pg.91 , Pg.92 , Pg.93 , Pg.94 , Pg.95 ]

See also in sourсe #XX -- [ Pg.686 ]

See also in sourсe #XX -- [ Pg.326 , Pg.446 , Pg.448 , Pg.449 ]

See also in sourсe #XX -- [ Pg.29 ]

See also in sourсe #XX -- [ Pg.154 ]

See also in sourсe #XX -- [ Pg.59 ]

See also in sourсe #XX -- [ Pg.445 ]

See also in sourсe #XX -- [ Pg.252 ]

See also in sourсe #XX -- [ Pg.642 ]

See also in sourсe #XX -- [ Pg.16 , Pg.70 ]

See also in sourсe #XX -- [ Pg.614 ]

See also in sourсe #XX -- [ Pg.3 , Pg.21 ]

See also in sourсe #XX -- [ Pg.28 , Pg.29 ]

See also in sourсe #XX -- [ Pg.82 , Pg.208 ]

See also in sourсe #XX -- [ Pg.474 , Pg.475 ]

See also in sourсe #XX -- [ Pg.3 , Pg.30 , Pg.32 , Pg.36 , Pg.46 , Pg.75 , Pg.84 , Pg.85 ]

See also in sourсe #XX -- [ Pg.329 ]

See also in sourсe #XX -- [ Pg.16 ]

See also in sourсe #XX -- [ Pg.102 , Pg.118 ]

See also in sourсe #XX -- [ Pg.20 ]

See also in sourсe #XX -- [ Pg.186 ]

See also in sourсe #XX -- [ Pg.52 ]

See also in sourсe #XX -- [ Pg.35 , Pg.40 , Pg.82 ]

See also in sourсe #XX -- [ Pg.381 ]

See also in sourсe #XX -- [ Pg.122 , Pg.126 , Pg.159 ]

See also in sourсe #XX -- [ Pg.1976 ]

See also in sourсe #XX -- [ Pg.263 , Pg.265 ]

See also in sourсe #XX -- [ Pg.205 ]

See also in sourсe #XX -- [ Pg.226 , Pg.231 , Pg.235 ]

See also in sourсe #XX -- [ Pg.567 , Pg.579 ]

See also in sourсe #XX -- [ Pg.13 , Pg.41 , Pg.43 , Pg.57 , Pg.178 , Pg.180 ]

See also in sourсe #XX -- [ Pg.74 , Pg.75 , Pg.76 , Pg.77 , Pg.78 , Pg.79 ]

See also in sourсe #XX -- [ Pg.154 , Pg.265 , Pg.345 , Pg.383 , Pg.384 , Pg.514 ]

See also in sourсe #XX -- [ Pg.471 , Pg.479 , Pg.481 , Pg.484 ]

See also in sourсe #XX -- [ Pg.145 ]

See also in sourсe #XX -- [ Pg.8 ]

See also in sourсe #XX -- [ Pg.22 ]

See also in sourсe #XX -- [ Pg.12 , Pg.22 , Pg.30 , Pg.50 , Pg.79 , Pg.80 , Pg.83 , Pg.84 , Pg.85 , Pg.86 , Pg.124 ]

See also in sourсe #XX -- [ Pg.20 ]




SEARCH



Acid Temperature Control and Heat Recovery

Acid temperature control

Acid temperature control Goal Seek

Acid temperature control H2O enthalpy

Acid temperature control absorption tower

Acid temperature control adjustments

Acid temperature control advantages

Acid temperature control affected by specified mass

Acid temperature control affected by volume% SO3 in input

Acid temperature control alloys

Acid temperature control calculation

Acid temperature control commercial systems

Acid temperature control composition

Acid temperature control construction materials

Acid temperature control exothermic reactions

Acid temperature control exterior

Acid temperature control final acidmaking

Acid temperature control flow rates

Acid temperature control heat production rates

Acid temperature control influenced

Acid temperature control inputs

Acid temperature control interior

Acid temperature control matrix

Acid temperature control mixing

Acid temperature control mixing heat

Acid temperature control outputs

Acid temperature control photographs

Acid temperature control schematic

Acid temperature control steam production

Acid temperature control target temperatures

Acid temperature control water cooling

Advanced temperature control

Affinity techniques temperature control

Alcoholic fermentation temperature control

Alcoholic fermentation temperature control during

Alternative Temperature Control Structures

Analog-proportional temperature controlle

Analysis for Dual Temperature Control

Application of temperature control to asymmetric protonation

Averaged kinetic temperature controlling

Barrel control temperature

Bearing temperature control

Binding affinity temperature control

Board temperature controller

Body Control Temperatures

Boilers steam temperature controls

CO2 temperature control

Cascade temperature controller

Case Study Temperature Control

Closed-loop temperature control system

Column Oven Temperature Control

Columns temperature control

Compression molding temperature control

Condensate temperature control

Contour-Depending Temperature Control

Contour-dependent temperature control

Control (Cont temperature

Control Micro temperature

Control analyzer/temperature

Control average temperature

Control based on temperature

Control differential temperature

Control double differential temperature

Control gas temperature controlled

Control of Gas Temperature by Bypassing

Control of Temperature and Stirring

Control of bed temperature

Control of product gas temperature method

Control of reactor temperature

Control of temperature

Control of temperature and

Control systems temperature measurement

Control unit mould temperature

Controlled room temperature

Controlled temperature oscillatory rheometry

Controlled-temperature furnace

Controllers temperature controller board

Controlling Transport Number of Anions through Anion Exchange Membranes with Temperature

Controlling Vibration and Temperature Rise

Controlling the Growth Speed Evaporation Rate and Temperature Dependence

Corrosion testing continued temperature control

Criteria For Selection of Best Temperature Control Tray

Criteria for Locating the Control Temperature

Crystallization controlled temperature

Density cell, temperature control

Dual output temperature controllers

Dual sensor temperature control

Dual temperature control

Dual-Temperature Control Strategy

Dyebath temperature control

Dynamic Temperature Control

Effect of Nonkeys on Temperature Control

Effect of Pressure on Temperature Control

Effective control of temperature

Efficient Temperature Control

Emulsification temperature-controlled

Endothermic Processes Control by the Average or Maximum Ion Temperature

Enthalpy management and temperature control

Enzyme activity temperature control

Evaporator temperature control

Examples temperature controller

Exothermic reactor temperature control loop

Experimentation temperature control

Experimentation, effective diffusivity temperature control

Extruder temperature control

Feed temperature control

Feedback controlled temperature filament

Feedback temperature control

Flat temperature control

Flowsheets bypass temperature control after

Fluorescence spectroscopy temperature control

Furnace automatic temperature control

Furnaces temperature control regimes

Gas composition control target temperature

Hard temperature control

Heat exchangers temperature control

Heat transfer and temperature control

High performance liquid chromatography temperature control

Horizontal chambers temperature control

Hypochlorite temperature control

Importance of Temperature Control

Inductive mold temperature control

Industrial hydrogenation reactor temperature control

Installing Temperature and Composition Controllers

Isobutylene polymerizations, controlled high temperature

Isoperibolic Temperature Control

Kinetic methods, advantages temperature control

Level and Steam Temperature Controls

Level, pressure, temperature and composition control

Liners temperature control

Liquid sulfur product temperature control

Lyophilization, temperature control

Main Steam Temperature Control System

Maturation temperature/time control

Melt flow temperature control

Melting temperature control

Methodologies controlling temperature

Microscopy, temperature control

Microwave sintering temperature control

Mixer Temperature Control Systems

Molds temperature controls

Morphology Control by Substrate Temperature

Mould Temperature control system

Mould temperature control

No Temperature Control

Nuclear magnetic resonance temperature control

On temperature control

Oven temperature control

Oven temperature control chromatography

Oven temperature control considerations

Oven temperature control isothermal operation

Oven temperature control separations

Oven temperature control, subambient

PID temperature control

PID temperature controller

Parallel temperature control

Partition coefficient temperature control

Petroleum oils temperature control

Plasticator barrel temperature control

Platen temperature control

Polymer composition, temperature control

Polymerase chain reaction temperature control

Polymerization temperature control

Poor Barrel Zone Temperature Control

Pre-Tuned Temperature Controllers

Precise Temperature Control in Microflow Systems

Precision temperature control

Pressure and temperature controls

Pressure-compensated temperature control

Process control, automatic temperature

Process monitoring/control technologies temperature

Process temperature control

Process-control parameters from time-temperature superposition

Product temperature, controlling

Proportional integral derivative, temperature control

Proportional temperature controller

Pulsed temperature control

Quality control tests temperature

Rate control temperature effect

Reactor Temperature Control Using Feed Manipulation

Reactor temperature control

Reboiler temperature control

Recirculating temperature control

Riser temperature control

Riser temperature control (catalyst

Room temperature control system

SEMIEX - Temperature Control for Semi-Batch Reactor

Sample temperature control

Secondary Circulation Loop Temperature Control Systems

Selection of Temperature Control Trays

Self-tuning temperature controllers

Semibatch reactors, temperature control

Series temperature control

Silanization temperature control

Simulated annealing temperature control

Single contact acidmaking temperature control

Single temperature control

Single-Temperature Control Strategy

Soft temperature control

Spoilage temperature control

Strategies of Temperature Control

Stroke temperature control

Subambient temperature controls

Sulfur burning temperature control

Surface temperature biological control

TEMPCONT - Control of Temperature in a Water Heater

Tanks temperature control

Technical documentation temperature control

Temperature Calculation Using a Control Volume Technique

Temperature Characteristics with On-Off Control

Temperature Control Above the Load(s)

Temperature Control Below the Load(s)

Temperature Control Channels

Temperature Control Inside the Capillary

Temperature Control Loop and Geometric Mean Circuitry

Temperature Control Loop and Sensor Resistance Readout

Temperature Control Required

Temperature Control Strategies

Temperature Control Test Procedure

Temperature Control Thermometers Below

Temperature Control and Cooling

Temperature Control by Reactant Feed

Temperature Control in Chemical Microstructured Reactors

Temperature Control in Microfluidic Systems

Temperature Control of Industrial Reactors

Temperature Control of Injection Molds

Temperature Control of the Sample

Temperature Control with Boilup (Steam Flow Rate)

Temperature Control with Bottoms Flow Rate

Temperature Control with Distillate Flow Rate

Temperature Control with Reflux Flow Rate

Temperature Controlled Reaction

Temperature Gradient Generation and Control

Temperature Measurement, Furnaces, and Feedback Control

Temperature and Atmosphere Control

Temperature and Auxiliary Controls

Temperature and Composition Control

Temperature and Composition Controllers

Temperature control (distillation

Temperature control (furnace) calibration

Temperature control - chambers

Temperature control Friedel—Crafts reaction

Temperature control cascade loop

Temperature control circulating type

Temperature control condensers

Temperature control continuous

Temperature control controllers

Temperature control controllers

Temperature control cooling

Temperature control decomposition

Temperature control during

Temperature control during fermentation

Temperature control exothermicity

Temperature control finger

Temperature control for semi-batch reactor

Temperature control free-radical polymerization

Temperature control heat transfer

Temperature control hemoglobin

Temperature control in distillation

Temperature control in reactors

Temperature control in reactors with gradients

Temperature control in semi-batch reactors

Temperature control isothermal

Temperature control liquid baths

Temperature control liquid nitrogen

Temperature control loop

Temperature control melting-point baths

Temperature control microflow systems

Temperature control packaging

Temperature control radical initiators

Temperature control surrounding specimen

Temperature control system

Temperature control system fluid

Temperature control thermostat blocks

Temperature control total-immersion tests

Temperature control tray location

Temperature control tray selection

Temperature control tray selection location

Temperature control tube

Temperature control unit

Temperature control vapor

Temperature control water

Temperature control zones

Temperature control, automatic

Temperature control, birefringence

Temperature control, carbon dioxide

Temperature control, enzyme

Temperature control, gas

Temperature control, gas bypass

Temperature control, gas calculations

Temperature control, gas devices

Temperature control, gas flowsheet

Temperature control, gas industrial

Temperature control, gas inefficiency explained

Temperature control, gas percent bypass

Temperature control, gas photograph

Temperature control, gas principle

Temperature control, gas theoretical

Temperature control, gas up and down

Temperature control, kinetic capillary

Temperature control, spin coating

Temperature controlled AFM

Temperature controlled cell

Temperature controller

Temperature controller

Temperature controller rate term

Temperature controller resolution

Temperature controls decomposition furnaces

Temperature dependence controlled chain lengths

Temperature extremes working environment control

Temperature increase, control

Temperature measurement and control

Temperature microbial control

Temperature programming flow control

Temperature rate control

Temperature recorder controller

Temperature, color control

Temperature, control coefficient

Temperature, control measurement

Temperature, transparency control

Temperature-controlled

Temperature-controlled

Temperature-controlled chromatography

Temperature-controlled environment

Temperature-controlled mobile film

Temperature-controlled pyrolysis

Temperature-controlled release

Temperature-controlled residuum oil

Temperatures furnace temperature control regimes

Temperatures, acid and gas control

Temperatures, industrial Temperature control, acid

The influence of enthalpy management and temperature control

Thermal Balance and Temperature Control

Thermal comfort body control temperatures

Thermoforming temperature control

Thin-layer chromatography temperature control

Time constant of temperature control

Time-temperature control

Transport temperature control

Tray temperature control

Two-Temperature Control Structure

Two-temperature control

Variable temperature control

Vessel with temperature control

Vinyl monomers, temperature-controlled free

Vinyl monomers, temperature-controlled free radical polymerization

Wetting, solder temperature control

© 2024 chempedia.info