Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Temperature control thermostat blocks

All modern heat flow calorimeters have twin cells thus, they operate in the differential mode. As mentioned earlier, this means that the thermopiles from the sample and the reference cell are connected in opposition, so that the measured output is the difference between the respective thermoelectric forces. Because the differential voltage is the only quantity to be measured, the auxiliary electronics of a heat flux instrument are fairly simple, as shown in the block diagram of figure 9.3. The main device is a nanovoltmeter interfaced to a computer for instrument control and data acquisition and handling. The remaining electronics of a microcalorimeter (not shown in figure 9.3) are related to the very accurate temperature control of the thermostat and, in some cases, with the... [Pg.141]

All detectors must be thermostatted at temperatures that are sufficient to keep sample and column bleed from condensing. Thermal conductivity and electron capture detectors require very accurate temperature control. If the TCD or ECD block temperatures are allowed to drift, so will the detector baseline. The sensitivity of the TCD increases with an increase in temperature differential between the block and the filaments. For this reason it is ad-... [Pg.338]

Temperature control during irradiation was effected by placing the tubes in holes in a thermostatted aluminum block. [Pg.248]

Most problems with solid fat content determination arise from inconsistencies in sample preparation. As noted above, the solids content is dependent on the temperature history of the sample, so deviations from the tempering protocols can lead to inaccuracies. The time required to measure a sample is not expected to lead to extensive crystallization or melting, but care should be taken to start the measurement as soon as possible after the oil is removed from the tempering block. Several instruments can be fitted with a thermostatic control for the sample in the coil, and this should be used if it is reliable. It is recommended that the NMR tubes be capped, particularly for the long tempering times required for stabilizing fats. [Pg.573]

Figure 13 illustrates a multiple thermocouple installation connected to a single indicator. Compensating lead wires are carried from the couples to a conveniently located cold-junction box. The temperature of this box is thermostatically controlled. From the cold-junction box copper wires are carried to the terminal block and selective switch illustrated. A common return has been employed between the cold-junction box and the switchboard. In general it is preferable to use individual return wires for each couple. The switchboard illustrated is designed for six couples. By pressing one of the buttons shown any desired couple is connected directly to the indicator. [Pg.436]

Tonometry may be achieved by simple homemade assem-bhes or by cormnerdaUy available equipment. Some commercially available tonometers use a thin-film technique. It consists of a glass or plastic cup fitted on a shaft and enclosed in a humidified chamber whose temperature is maintained at 37 C. A few milliliters of blood is placed in the cup, and gas flow is initiated to continuously flush the inside surface of the cup with the humidified gas. A controller unit causes the cup to rotate rapidly and periodically in short bursts, so that the blood in the cup is thrown in a thin layer over the inside walls. Another form of tonometry is the bubble technique. It uses a syringe that is specially constructed to allow gas to be introduced and humidified through the plunger. During tonometry, the syringe is laid in a thermostatically controlled aluminum heat block. Additional detail on tonometry and its applications can be found in a previous edition of this textbook. Reference conditions for tonometry have been recommended by a committee of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). ... [Pg.1006]

Other thermal zones, which should be thermostatted separately from the column oven, include the injector and detector modules. These are generally insulted metal blocks fitted with cartridge heaters and controlled by sensors located in a feedback loop with the power supply. Detector blocks are usually maintained at a temperature selected to minimize detector contamination and to optimize the detector response to different sample types. The requirements for injectors may be different depending on their design, and may include provision for temperature programmed operation. [Pg.177]

Aging is carried out in thermostatically controlled ovens, which may be of the usual cabinet type or consist of a number of cells in a heated block. For plastics, relatively little consideration appears to have been given to details of the oven other than control of temperature, but for rubbers other factors that can influence results are carefully specified, and it would seem reasonable to suppose that these could be significant for plastics also. [Pg.268]

A 250-watt cartridge heater 3/8" in diameter and 3" long is used to heat the heater block. A 250-600 degree adjustable thermostat is used to control the temperature of the cylinder block and is mounted to the side of the block. As mentioned earlier in the book, the cartridge heater and thermostat are not going to be found at the hardware store. For information on how you can purchase these items check sources at the back of tlie book. [Pg.12]

Blachnik et al. (1996) described a calorimeter that involves the dropping of samples with temperatures up to 1400 K into a copper block inside a brass vessel in a water thermostat at 298.15 K. The furnace on the top of the calorimeter consists of three concentric alumina tubes with five independent controlled heaters. The temperature of the furnace is declared to be isothermal with an uncertainty of 0.3 K. The water temperature is said to be constant with an uncertainty of 1.4 mK within 1 h and a fluctuation of the calorimeter block temperature of 0.7 mK. The calorimeter was calibrated electrically as well as with an a-Al203 sample... [Pg.165]

The bottom sketch in Fig. 5.2 represents a drop calorimeter. As in the liquid calorimeter, the mode of operation is isoperibol. The surroundings are at (almost) constant temperature and are linked to the sample via a controlled heat leak. The recipient is chosen to be a solid block of metal. Because it uses no liquid, the calorimeter is called an aneroid calorimeter. The use of the solid recipient eliminates losses due to evaporation and stirring, but causes a less uniform temperature distribution and necessitates a longer time to reach steady state. The sample is heated to a constant temperature in a thermostat (not shown) above the calorimeter and then dropped into the calorimeter, where the heat is exchanged. The temperature rise of the block is used to calculate the average heat capacity. [Pg.225]


See other pages where Temperature control thermostat blocks is mentioned: [Pg.549]    [Pg.235]    [Pg.267]    [Pg.584]    [Pg.89]    [Pg.188]    [Pg.152]    [Pg.104]    [Pg.356]    [Pg.157]    [Pg.69]    [Pg.279]    [Pg.299]    [Pg.110]    [Pg.78]    [Pg.165]    [Pg.99]    [Pg.94]    [Pg.78]    [Pg.352]    [Pg.46]    [Pg.55]    [Pg.173]    [Pg.209]    [Pg.90]    [Pg.1874]    [Pg.236]    [Pg.273]    [Pg.33]    [Pg.486]   
See also in sourсe #XX -- [ Pg.584 ]




SEARCH



Block temperature

Temperature Thermostats

Temperature control

Temperature control controllers

Temperature controller

Temperature-controlled

Thermostatic control

Thermostatting

© 2024 chempedia.info