Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Control analyzer/temperature

Polyakov 1997). Because the second-order Doppler shift is not the only factor controlling Mossbauer absorption frequencies, it is generally necessary to process data taken at a variety of temperatures, and to make a number of assumptions about the invariance of other factors with temperature and the form and properties of the vibrational density of states of the Mossbauer atom. Principles involved in analyzing temperature dependencies in Mossbauer spectra are extensively discussed in the primary literature (Hazony 1966 Housley and Hess 1966 Housley and Hess 1967) and reviews (e.g., Heberle 1971). [Pg.91]

One of the most impressive developments during the past year or so with traditional drift tubes is the introduction of an IMS analyzer with authentic twin drift tubes where the sample is ionized in a single reaction region and positive and negative ions are extracted and characterized in two separated drift tubes placed at appropriate polarity. In this design, the two drift tubes can be individually controlled in temperature although the ion source is common to both drift tubes. The analyzer... [Pg.187]

The density of CO2 in the absorption cell, however, is a function of both concentration and bulk air density. In normal process analyzers, where temperature and pressure within the absorption cell are controlled, measurements can be easily referred to gas density by a simple calibration curve. In an open path system, changes in bulk air density must be measured. Indeed, one of the major problems faced in testing the sensor was the development of test facilities where we could control the temperature, pressure and CC>2 more accurately than the sensor could measure. Even the small changes in building pressure associated with ventilation system fluctuations resulted in output signal changes three to four times the sensor signal to noise level. In operation, pressure and temperature near the open cell are measured and used to calculate gas density. [Pg.229]

Since we have two control degrees of freedom, our objectives in distillation are to control the amount of LK impurity in the bottoms product ( b.lk) and the amount of HK impurity in the distillate ( 5>Hk) Controlling these compositions directly requires that we have composition analyzers to measure them. Instead of doing this, it is often possible to achieve fairly good product quality control by controlling the temperature on some tray in the column and keeping one manipulated variable constant. Quite often the best variable to fix is the reflux flowrate, but other possibilities include holding heat input or reflux ratio constant. [Pg.205]

The Microtox equipment includes a self-calibrating analyzer which incorporates a photomultiplier tube, a data collection and reduction system, and software. The temperature-controlled analyzer maintains the test organisms and samples at a standard temperature of 15°C. It also detects the light intensity at 490 nm, the wavelength emitted by the bacteria. [Pg.1694]

The experiments were conducted in a fully automated flow reactor. The gas compositions were analyzed by a GC (see [8] for details). The reactor was fed with a mixture of propane, propene, hydrogen and nitrogen to achieve differential conditions. This was desired to avoid changes in the gas compositions along the bed leading to a non-uniform coke profile, and also to reduce the computational effort. The gas flow rates were controlled by mass flow controllers. The temperatures were 780,800 and 820 K, and the total pressure was 1.5 bar. [Pg.674]

The reactor was a tube of 446 stainless steel with an internal diameter of % in. and of 20 in. total length. The catalyst bed was the central 2 in., and a furnace surrounding the tube controlled the temperature to 1°. A sliding thermocouple in a central well of 3 -in. o.d. measured the bed temperature. The catalyst bed volume was 10 cc., and the rest of the reactor was filled with stainless steel spacers to reduce the dead space. Steam was heated to 650° in a preheater and mixed with the reacting gas or air at the top of the reactor, which was operated at atmospheric pressure. The supply of gas and steam was measured by flow meters and controlled by solenoid valves activated by a timer so that the catalyst could be fed on hourly or half-hourly cycles with steam and hydrocarbon or steam and air. In this way, any carbon laydown could be measured by analyzing for carbon dioxide in the air from the second (regeneration) cycle. [Pg.244]

In brief, an increase in heavy nonkeys concentration will "fool a bottom section temperature controller into letting more light keys out of the bottom, but will have little impact on a top section temperature controller. Conversely, an increase in light nonkeys in the feed will fool a top section temperature controller into letting more heavy non-keys out of the top, but will barely affect a bottom section temperature controller. In two different troublesome cases (239, 378), the top section temperature controller of an isobutane-normal butane splitter was frequently fooled into letting normal butane into the top product each time propane concentration in the column feed suddenly rose. In one of these (378), the problem was cured by using an analyzer/temperature control (see Sec. 18.3). The author is familiar with several similar experiences of temperature controller fooling. [Pg.559]

Fourroux et al. (122) found the same oscillation period when comparing the response of an analyzer/temperature control to that of an analyzer only control. This can be expected if the temperature controller is tuned for slow response, because the analyzer will adjust the temperature controller set point faster than the temperature controller can react. In the imusual case of slow temperature control, an analyzer/temperature control is therefore best avoided (406). The analyzer/temperature control relies on fast temperature control for its success. [Pg.576]

Stanton and Bremer (378) demonstrated the superiority of the response of the analyzer/temperature control system over both a temperature control and a direct analyzer control in a 72-tray deisobutanizer. A temperature control alone produced a product purity offset due to variations in nonkeys an analyzer-only control had a long lineout time and was sensitive to feed flow changes. An analyzer/temperature control gave a fast response and eliminated all these ill effects. Other favorable experiences with analyzer/temperature control have also been reported (25, 89, 203, 287, 379). Two references (25, 379) contain in-depth descriptions of tuning considerations and of other accessories that can improve system operation, particularly if the analyzer control is performed through a computer system. [Pg.576]

An ana rzer-temperature cascade can give better control than temperature or direct analyzer control Sampling at the accumulator outlet can be troublesome with direct analyzer control Temperature control can be troublesome in the presence of nonk. Direct anab zer control can be troublesame in the presence of feed disturbances. [Pg.760]

Fig. 26 Diagram showing Ca-looping sonoprocessing. 1 Compressed gas used for carbonation (15 vol.% CO2/ 85 vol.% N2) 2 compressed gas used for calcinations (dry air) 3 mass flow controllers 4 temperature controller 5 furnace 6 quartz reactor 7 sound waveguide 8 elastic membrane 9 microphone 10 loudspeaker 11 differential pressure transducer 12 particulate filter 13 mass flow meter 14 gas analyzer 15 signal amplifier 16 signal generator 17 oscilloscope 18 air cooling system. Reproduced with permission from [133], Copyright 2013 American Chemical Society... Fig. 26 Diagram showing Ca-looping sonoprocessing. 1 Compressed gas used for carbonation (15 vol.% CO2/ 85 vol.% N2) 2 compressed gas used for calcinations (dry air) 3 mass flow controllers 4 temperature controller 5 furnace 6 quartz reactor 7 sound waveguide 8 elastic membrane 9 microphone 10 loudspeaker 11 differential pressure transducer 12 particulate filter 13 mass flow meter 14 gas analyzer 15 signal amplifier 16 signal generator 17 oscilloscope 18 air cooling system. Reproduced with permission from [133], Copyright 2013 American Chemical Society...
DNA thermodegradation is analyzed by incubating 0.5 xg of DNA in 20 p.1 incubation mixture covered with 150 p,l of HaO-saturated paraffin oil to prevent evaporation. The pH and temperature of the incubation mixtures are controlled through temperature and pH probes. The pH determined by probes can be slightly different from the pH calculated for pH-temperature compensation, depending on salt present or buffer strength. [Pg.208]

The instrument for determination of shear moduli was a Rheometric Scientific dynamic mechanical thermal analyzer, model DMTA V. Round shear sandwich geometry was used. The instrument was inverted so that the sandwich fixtures and sample were in water. A water-jacketed 1000 mL Pyrex cylinder supplied by Rheometric Scientific allowed control of temperature with a circulating temperature bath. A sinusoidal linear shear was applied by moving a flat plate between two identical disk-shaped samples over a specified range of frequencies. The two identical disk-shaped samples were sandwiched between the moving plate and two 12 mm diameter plates (called studs) fastened to a frame. The dynamic frequency sweeps to obtain the loss shear modulus, G", from 0.16 Hz to 318 Hz were reported in log scale. For our purposes the applied initial static force was 0.05 N. Sample size was 12 mm diameter and 0.7 mm thickness. The sample was equilibrated at 40°C for 12 hours prior to starting the series of measurements. Shear moduli were measured from high to low temperature with an equilibration time of 2 hours at each temperature. The sequence of measurements was 40°C, 30°C, 25°C, and 15°C. [Pg.603]


See other pages where Control analyzer/temperature is mentioned: [Pg.747]    [Pg.339]    [Pg.122]    [Pg.101]    [Pg.575]    [Pg.204]    [Pg.136]    [Pg.125]    [Pg.43]    [Pg.66]    [Pg.85]    [Pg.220]    [Pg.282]    [Pg.43]    [Pg.571]    [Pg.486]    [Pg.320]    [Pg.352]    [Pg.548]    [Pg.918]    [Pg.402]    [Pg.228]    [Pg.249]    [Pg.1227]    [Pg.35]    [Pg.923]    [Pg.114]    [Pg.181]    [Pg.751]    [Pg.570]    [Pg.574]    [Pg.189]    [Pg.126]    [Pg.263]   
See also in sourсe #XX -- [ Pg.559 , Pg.574 , Pg.576 , Pg.668 , Pg.677 ]




SEARCH



Analyzer controller

Control analyzer

Temperature control

Temperature control controllers

Temperature controller

Temperature-controlled

© 2024 chempedia.info