Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Temperature Thermostats

If understanding the behavior of the system as a function of temperature is the aim of your study, then it is important to be able to control the temperature of your system. The temperature of the system is related to the time average of the kinetic energy, which generally can be calculated by [Pg.221]

Molecular Dynamics Simulation From Ab lnitio to Coarse Grained  [Pg.222]


The glass tubes contain mercury and are firmly fixed in the ebonite cover of the cell so that the distance between the electrodes may not change during the experiment. Contact with the platinum electrodes is made by dipping the copper wires of the circuit in the mercury contained in the tubes. The coating of the electrodes with platinum black is carried out in order to inhibit polarization. When examining certain electrolytes, platinized electrodes cannot be used since platinum black may catalyze the decomposition or the oxidation of the electrolyte. The cell is maintained at constant temperature thermostatically, as conductivity increases rather significantly with temperature. [Pg.610]

Define the following expressions empirical method, metastable equilibrium, kindling temperature, thermostat, interface, dynamic steady state, unimolecular reaction, bimolecular reaction, homogeneous reaction, heterogeneous reaction. [Pg.412]

There are three general methods for maintaining a RV at a particular temperature thermostats, furnaces and vapour baths . [Pg.8]

Fig. 7. General description of the electrodes and measuring cell C.T., cooling thermostat T.C., thermocouple P.S., porous stone R.T., room temperature thermostat M.S., magnetic stirring. From Larroque et al. (1976). Reprinted with permission of Analytical Biochemistry. Copyright by Academic Press. Fig. 7. General description of the electrodes and measuring cell C.T., cooling thermostat T.C., thermocouple P.S., porous stone R.T., room temperature thermostat M.S., magnetic stirring. From Larroque et al. (1976). Reprinted with permission of Analytical Biochemistry. Copyright by Academic Press.
Fig. 26. Apparatus for isoelectric focusing in a thermal pH gradient. T, high-temperature thermostat Ts, low-temperature cryostat isolated with polystyrene (PS) P, plexiglass block G, gel Ei and E2, electrolyte vessels U, tube adaptor SM, stirring magnet H, helix PP, double peristaltic pump. Fig. 26. Apparatus for isoelectric focusing in a thermal pH gradient. T, high-temperature thermostat Ts, low-temperature cryostat isolated with polystyrene (PS) P, plexiglass block G, gel Ei and E2, electrolyte vessels U, tube adaptor SM, stirring magnet H, helix PP, double peristaltic pump.
Fig. 27. Apparatus for electrophoresis at subzero temperatures. Tj, low-temperature thermostat G, electrophoresis gel (4 tubes) E( and Ej, electrolyte vessels. Fig. 27. Apparatus for electrophoresis at subzero temperatures. Tj, low-temperature thermostat G, electrophoresis gel (4 tubes) E( and Ej, electrolyte vessels.
The dehumidification performance of a cooling coil system can be characterized in several ways. The most obvious indicator is the latent capacity, defined as the difference between the total and sensible capacities. The latent capacity increases with reduced airflow and higher entering wet-bulb temperature. However, the sensible capacity changes with these variables as well. As the operation of a DX system is almost always controlled by a dry bulb temperature thermostat, greater latent capacity alone does not necessarily mean lower humidity in a building. [Pg.617]

A noisy baseline of an RI detector can be caused by inadequate mobile phase degassing or temperature thermostatting. However, low light energy can also be caused by a contaminated detector flow cell or high UV absorbance of the mobile phase (Figure 10.7d). [Pg.252]

Note. All instruments had the capability to make measurements in the visible or ultraviolet region of the spectrum, all had temperature thermostatting, all had the capability to interface to a host laboratory computer with bidirectional communications and to retransmit data, all had direct-tube sampling and bar-code identification of the specimen, all had reagent identification to the instrument, and usually with bar codes. [Pg.155]

Turn the temperature thermostat down by 2°F (1°C) to reduce your heating consumption by about ten percent. [Pg.23]

On/ojf control In on/off control, if the e f) is smaller than 0, the controller may activate the plant otherwise the controller stays still. Most household temperature thermostats follow this model. [Pg.160]

Check working of oil temperature thermostat every shift. [Pg.77]

One method of cooling the calorimeter system is to put it inside a low-temperature thermostat (cryostat see later). This method is often used in low-temperature calorimetry and works down to temperatures well below 1K. [Pg.138]

It should be mentioned that calorimeters with the surroundings kept at constant temperature (thermostat) are often named isothermal calorimeters in the literature. This is, however, not correct because the sample and the sample container temperature during the reaction are not constant and furthermore may be very different from the thermostat temperature until heat has been exchanged. Such calorimeters operate under isoperibol conditions (see Section 5.2) we present them in Section 7.9. [Pg.146]

Proper ventilation of these as well as other vented batteries, although not as much a problem here as with other battery systems, is required to avoid the accumulation of hazardous hydrogen, especially during charge. For larger installations, forced air or fans may also be required to prevent undesirable temperature buildups. When close voltage regulation is required at cold temperatures, thermostatically controlled heaters are often used with the batteries. [Pg.1005]

The accessories usually attached to commercially available NIR spectrometers are considered first. As NIR spectra depend sensitively on temperature, thermostats to stabilize the temperature of samples, mainly liquid samples in an optical cell, are usually provided. An integrating sphere is necessary for measurements of diffuse reflection. Attachments for measuring transmission spectra from samples outside the body of the spectrometer are used for samples that cannot be put into the sample compartment of the spectrometer, or are used with an automatic sampler. Ports for the optical fibers are of especial practical importance, because optical fibers of many kinds are used to measure NIR spectra of samples in various shapes and states, including high pressure and high temperatures an appropriate optical fiber should be used for the sample to be studied. [Pg.263]

Range of temperature (thermostat fluid sihcon liquid) essentially the range of temperature is determined by the boiling point resp. freezing point of the thermostat liquid... [Pg.23]


See other pages where Temperature Thermostats is mentioned: [Pg.518]    [Pg.100]    [Pg.155]    [Pg.518]    [Pg.113]    [Pg.143]    [Pg.214]    [Pg.113]    [Pg.261]    [Pg.325]    [Pg.143]    [Pg.165]    [Pg.568]    [Pg.382]    [Pg.48]    [Pg.809]    [Pg.543]    [Pg.39]    [Pg.543]    [Pg.305]    [Pg.289]    [Pg.189]    [Pg.88]    [Pg.153]    [Pg.671]    [Pg.221]    [Pg.59]   


SEARCH



Temperature control thermostat blocks

Thermostatting

© 2024 chempedia.info