Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Slope control

A pH electrode is normally standardized using two buffers one near a pH of 7 and one that is more acidic or basic depending on the sample s expected pH. The pH electrode is immersed in the first buffer, and the standardize or calibrate control is adjusted until the meter reads the correct pH. The electrode is placed in the second buffer, and the slope or temperature control is adjusted to the-buffer s pH. Some pH meters are equipped with a temperature compensation feature, allowing the pH meter to correct the measured pH for any change in temperature. In this case a thermistor is placed in the sample and connected to the pH meter. The temperature control is set to the solution s temperature, and the pH meter is calibrated using the calibrate and slope controls. If a change in the sample s temperature is indicated by the thermistor, the pH meter adjusts the slope of the calibration based on an assumed Nerstian response of 2.303RT/F. [Pg.492]

Remove the electrode assembly (and the thermometer probe if used), rinse in distilled water, and place into a small beaker containing a little of the second buffer solution. If the meter reading does not agree exactly with the known pH, adjust the Slope control until the required reading is obtained. [Pg.567]

At low frequencies the loss modulus is linear in frequency and the storage modulus is quadratic for both models. As the frequency exceeds the reciprocal of the relaxation time ii the Rouse model approaches a square root dependence on frequency. The Zimm model varies as the 2/3rd power in frequency. At high frequencies there is some experimental evidence that suggests the storage modulus reaches a plateau value. The loss modulus has a linear dependence on frequency with a slope controlled by the solvent viscosity. Hearst and Tschoegl32 have both illustrated how a parameter h can be introduced into a bead spring... [Pg.189]

Set sensitivity or slope control to 100% position (ignore this step for pH meters that do not have the 100% indicated). [Pg.1115]

Adjust the sensitivity or slope control dial so that the display reads pH 4.00. [Pg.1115]

Calibration Select two standard buffers to bracket, if possible, the anticipated pH of the unknown substances. These commercially available standards and the sample should be at the same temperature, within two degrees. Set the temperature compensator of the pH meter to the temperature of the samples and standards. Follow the manufacturer s instructions for setting temperature compensation and for adjusting the output during calibration. Rinse the electrodes with distilled or deionized water, and blot them dry with clean, absorbent laboratory tissue. Place the electrode(s) in the first standard buffer solution, and adjust the standardization control so that the pH reading matches the stated pH of the standard buffer. Repeat this procedure with fresh portions of the first buffer solution until two successive readings are within 0.02 pH units with no further adjustment. Rinse the electrodes, blot them dry, and place them in a portion of the second standard buffer solution. Following the manufacturer s instructions, adjust the slope control (not the standardization control) until the output displays the pH of the second standard buffer. [Pg.845]

Fig. 7.3 The relationship between electrical potential and pH. The solid line shows the response of a calibrated electrode while the other plots are for instruments requiring calibration 1 has the correct slope but incorrect isopotential point (calibration control adjustment is needed) 2 has the correct isopotential point but incorrect slope (slope control adjustment is needed). Fig. 7.3 The relationship between electrical potential and pH. The solid line shows the response of a calibrated electrode while the other plots are for instruments requiring calibration 1 has the correct slope but incorrect isopotential point (calibration control adjustment is needed) 2 has the correct isopotential point but incorrect slope (slope control adjustment is needed).
The delay in alkaline "breakthrough can be explained by the theory of chrcmatograjtiy. According to the theory, the reversible adsorptiai isotherm slope controls the advance rate... [Pg.243]

If all test result values are above the LOQ, the distribution is normal, and the variance is constant, the regression control chart method or the slope control chart method may be applied to the data. [Pg.269]

Automations and automatic level and slope control device... [Pg.412]

The tractor unit also carries all the automations for material handling and the automatic level (grade) and slope control device. [Pg.412]

Apart from the levelling control system, there is also the slope control system that is integrated with the levelling control system in modern pavers. The slope control system is particularly useful when laying base asphalt layers, thick asphalt layers or asphalt layers in general, on curved sections. [Pg.413]

If the extrapolations of curves 1 and III are not co-linear, the slope control of the instrument is adjusted until this condition is satisfied and steps 3 5 are repeated. Once the above conditions have been satisfied, the slope, the horizontal and vertical axis sensitivities, the position of the zero point, the gas flow rate, the level of coolant, the orientation of the sample holder lid and the position of the recorder pen (if a chart recorder is being used) should be kept at those values for the duration of the experiment. (6) A 10 30 mg amoung of a standard sapphire sample is weighed with a precision of 0.01 mg and placed in the second sample vessel (previously weighed). The sapphire sample is inserted into the sample holder and steps 3 to 5 are repeated to obtain curve IV. (7) The sapphire is removed from the sample vessel and replaced with the sample of known mass 0.01 mg. The sample is inserted into the sample holder and steps 3 to 5 are repeated to obtain curve V. The sample mass should be approximately 10 mg. The three curves II, IV and V should be coincident at Z and. If not, the... [Pg.98]

The electrodes are normally stored in distilled or deionized water and should be rinsed with deionized water before use. The electrodes can be dried by draining the excess water off onto a tissue. The electrode must be standardized by immersion in buffer solutions whose pH are well separated. The procedure is to immerse the electrodes in the buffer solution for a period of 1 min with continuous stirring and observe the value which is then adjusted to the buffer value by the pH or calibration control. The electrode is then rinsed and dried and immersed in the second buffer solution and allowed to equilibrate for 1 min with stirring and the second buffer value is read. If this differs from that which is expected then the slope control of the pH meter should be altered. This whole procedure is then repeated until the values obtained in the buffer solutions at each end of the pH range agree with those of the known buffer solution. [Pg.2385]

The characteristics of sedimentary. The over-lying slope proluvium and the loess comprise the materials of the landslide. They covered on the original underlying bedrock slopes, and aU the bedrock layers are steeper and incline to the outer slope. Control by the gravity, the upper proluvium and the loess layers with downward trend along the contact surfaces between up soils and the underlying bedrock. [Pg.861]

Adjust the slope controller of the apparatus in order that line I meets line III on the same line, as shown in Figure 6.1. [Pg.155]

In the last decade, several techniques have been developed in order to improve the resolution that can be obtained with optical lithography. This review describes a number of these techniques. Some techniques make use of a photosensitive layer on top of the photoresist, which results in enhancement of the contrast of the aerial image (Sects. 3.1.1 and 3.1.2). Other techniques use the possibility of slope control during development (Sects. 3.1.3 and 3.1.4). Also discussed are some techniques involving anisotropic plasma etching. [Pg.91]

This chapter describes how a pH meter functions and how its various controls affect the pH reading. The types of readout and the effects of temperature compensation and standardization controls are discussed. Answers are given to questions such as when to use a slope control or how a zero control relates to the slope adjustment. In conclusion, the various types of meters and their performance specifications are discussed. [Pg.17]

The purpose of the slope and zero (isopotential) controls is to provide a pH readout that closely follows the electrode response, thus increasing the accuracy of the measurement. If a high degree of accuracy is not required, typically greater than 0.05 pH unit or more, the slope control should be turned off or set at 100% and the zero control set to display the typical isopotential point at pH 7 and no second slope/span adjustment made. [Pg.27]

The first step of this test, to short the glass and reference inputs, is done using a terminal connector and shorting strap. The terminal connector allows a pin jack to make connection with the glass input. The shorting strap with pin jacks at either end is connected between the terminal connector and the reference jack (see Figure 7.1). The meter is placed in the pH function, the temperature control at 30°C, and the slope control at 100% or off. If the meter has a zero control, it should be set to provide pH 7.00 on the display when the meter is in the stand-by mode. The readout is then activated to conduct the test. [Pg.140]

Then, set the display to read pH 6.00 with the standardization control. When the temperature compensating control is rotated to 0°C, the display should read about 5.9. When it is rotated to lOO C, it should read about 6.2. Reset the temperature control to 30°C and rotate the slope control. At 80% slope, the reading should be lower than pH 5.8. [Pg.140]

Set the controls at the starting point temperature control, SC C, slope control, 100%, zero control, pH 7. [Pg.141]

On the basis of the equivalent circuit adopted, the slope controlled by the iron dissolution, /Fe/Fe oxide formation, /pe/FeOx hydrogen evolution, /h /h2 is given by the expression... [Pg.224]


See other pages where Slope control is mentioned: [Pg.566]    [Pg.469]    [Pg.914]    [Pg.18]    [Pg.59]    [Pg.788]    [Pg.95]    [Pg.59]    [Pg.393]    [Pg.268]    [Pg.416]    [Pg.463]    [Pg.465]    [Pg.18]    [Pg.27]    [Pg.29]    [Pg.108]    [Pg.165]    [Pg.553]    [Pg.553]   
See also in sourсe #XX -- [ Pg.27 ]

See also in sourсe #XX -- [ Pg.126 ]




SEARCH



Slope

Sloping

© 2024 chempedia.info