Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Emission spectroscopies

The 77 K emission of [M2(C ANAN)2(/U-dppm)]2+ in frozen CH3CN occurs at /Lmax 598 and 638 nm for M=Pd 20 and Pt 7b respectively (Fig. 15). Since a n-n excimeric 3IL excited state is invoked for the Pd2 species, the red shift for the Pt2 analogue is indicative of an excited state arising from n-n (excimeric 3IL) and metal-metal (previously assigned as da — 7r ) interactions. The 77 K solid-state emission spectra for [M2(ClANAN)2(/t-dppm)](C104)2 [Pg.47]

At this juncture, in order to facilitate comparisons of the families of lu-minophores described in this account, a table illustrating the correlation between the metal-metal distances and photophysical properties of relevant cy-clometalated diimine Pt(II)/Pd(II) complexes is provided (Table 3). [Pg.48]

Emission spectroscopy is exclusively related to atoms whereas a number of other spectroscopic techniques deal with molecules. The fundamental fact of emission spectroscopy is very simple, wherein the atoms present in a sample undergo excitation due to the absorption of either electrical or thermal energy. Subsequently, the radiation emitted by atoms in an excited sample is studied in an elaborated manner both qualitatively and quantitatively. Therefore, emission spectroscopy is considered to be an useful analytical tool for the analysis of  [Pg.357]

In short, emission spectroscopy is considered to be the most accurate, precise and reliable means of quantitative analysis of elements as on date. If proper skill, precautions and wisdom are applied together this method may be adopted safely and conveniently to analyze approximately seventy elements from the periodic table at a concentration as low as 1 ppm. [Pg.357]

The theoretical aspects of emission spectroscopy may be categorized into the following four heads, namely  [Pg.357]

However, in emission spectroscopy the band spectra provided by molecules may be eliminated completely by giving energy to the corresponding molecules so that they may be split-up into separate atoms. [Pg.358]

Example Incandescent solids, e.g., carbon and iron give rise to continuous emission spectra when they are heated until they glow. [Pg.358]

If a molecule has emission properties such as fluorescence or phosphorescence, addition of DNA may, as in the spectrophotometric titrations, result in a batho-chromic shift of the emission maximum. More significantly, the emission intensity may change on complex formation. In the latter circumstances either an increase or a decrease of the emission intensity might be observed [28]. [Pg.179]

Dyes whose fluorescence intensity increases on binding to DNA (e.g. 3 and 4) have especially high potential as DNA marker or detector molecules. In the absence of DNA the relatively low fluorescence quantum yield of these dyes results from a radiationless deactivation of the excited state by conformational changes or acid-base reactions with the solvent. On association with DNA, however, significant suppression of the conformational flexibility and a shielding of the dye from solvent molecules within the complex occurs, leading to an increase of the emission intensity. [Pg.179]

The inverse ET mechanism has recently also been proposed for fluorescence [Pg.179]

If collected correctly, emission spectra mirror absorbance spectra and can be collected even from opaque samples. The material must not have thermal gradients, readsorption of emission, or self-adsorption. [Pg.13]

The energy of the absorbed radiation corresponds to the energy of a transition from ground to an excited state. Decay of an excited state back to the ground state may take place by a radiative or non-radiative process. The spontaneous emission of radiation from an electronically excited species is called luminescence and this term covers both fluorescence and phosphorescence. A discussion of these phenomena requires an understanding of the electronic states of multi-electron systems, and we return to emission spectra in Section 20.8. [Pg.106]

The production of excited species in flames has already been mentioned in Section 1.3. A hydrogen-oxygen flame exhibits a well defined band system in the near ultraviolet, which has been shown to originate from electronically excited hydroxyl radicals . A very wide variety of emitting species has been identified in flames we are not concerned here with the chemical implications of the results obttiined, and the reader must be referred to more specific articles which have appetu ed elsewhere . [Pg.286]

Transition probabilities are known for some of the emitting species in flames (e.g. OH , CN , CH and C2 ). A method has been developed recently by [Pg.286]

Dalby and Bennett which has given accurate probabilities for a series of transitions. The technique is described briefly onpp. 291-2. Accurate determination of concentrations may still be hindered by self-absorption of the radiation, particularly in the case of the hydroxyl radical. Penner and co-workers have overcome the difficulty by the use of a double path technique, and are able to determine the rotational temperature and concentration of hydroxyl radicals in both flame and shock-tube studies. The single and double path emissivities are compared simultaneously, the double path beam being chopped to give modulation at about 5 sec intervals. The method of correction for line widths and Doppler broadening is discussed . [Pg.287]

A further requirement for measurement of absolute concentrations of excited species in flames is that the volume from which emission is collected be known. The simplest experimental arrangement for flames at atmospheric pressures is to focus the radiation from the flame onto the entrance slit of a spectrograph. Reasonable assumptions can be made about the thickness of the emitting layer, and Ausloos and van Tiggelen have used the arrangement successfully in semi-quantitative determinations of excited OH, NH, NO and NH2 in flames emitting the bands of these species. [Pg.287]

Studies of low-pressure flames offer several advantages. In particular, the flame can be maintained flat, and the light from different parts of the reaction zone studied separately the reaction volume from which light is collected is determined with much greater accuracy for such flames. At low pressures, chemiluminescent processes are more important than thermal excitation, collisional quenching of excited species is reduced, and self-absorption is diminished. A typical investigation of the low pressure flame is that of Gaydon and Wolfhard quantitative measurements of the C2 emission were made. [Pg.287]

A Xe flashlamp-boxcar integrator system has been used for excitation, fluorescence, and time-resolved fluorescence studies of high-temperature [Pg.25]

Wetzel, B. Pettinger, and U. Wenning, Chem. Phys. Lett., 1980, 75, 173. [Pg.25]

Of possible concern to any worker observing vacuum-u.v. fluorescence signals are Wren s observations that SI-u.v. quartz, u.v. grade sapphire, Mgp2, and BaFj (commonly used materials for vacuum-u.v. optical components) show strong luminescence when excited above 200nm.  [Pg.26]

The deconvolution of multicomponent fluorescence spectra may be accomplished using a ratio method providing spectral regions may be located where the luminescence is due to one species only.  [Pg.26]

Various competitive routes are available for dissipation of absorbed radiant energy. These include both non-radiative transitions and radiative photophysical processes, such as fluorescence and phosphorescence. Energy can be transferred directly to other molecules by a process known as quenching dissipated through the vibrational motion of the molecule. Quenching depends on collisions between molecules internal transfer of energy as a result of which a molecule passes over into a lower-lying electronic state facilitates the vibrational process. [Pg.72]

Some examples of emission techniques that have been used to polymer/additive studies are  [Pg.72]

Emission spectroscopic techniques are more sensitive than absorption or reflectance spectrophotometry. Excitation by narrow-band lasers may result in the selective population of wanted levels, which emit their excitation energy as fluorescence photons. A laser-induced fluorescence (LIE) spectrum is much simpler than the emission spectrum of a gas discharge, where the superposition of fluorescence from many emitting levels is observed. [Pg.72]

Principles and Characteristics In all ranges of the infrared (NIR, mid-IR, FIR) mainly absorption of radiation by a sample is used as an analytical tool. Emission spectra are rarely recorded, even though they are powerful for problems which cannot be investigated by other methods. [Pg.72]

In order to observe emission it is necessary to populate a higher lying unoccupied quantised state. A molecule in a vibrationally excited state has a certain probability of emitting IR radiation in the presence or absence of incident electromagnetic radiation, resulting in induced and spontaneous emission, respectively. At r.t. the number of molecules in a first excited state is less than 1% of the population in the ground state, when the separation of energy levels is [Pg.72]

In this simple technique, the metal to be determined, in the form of a solution of a suitable compound, is sprayed into a flame. As in atomic absorption, when the solvent evaporates in the flame, the solid obtained is atomised and a gaseous metal ion is excited to a higher electronic level. When this drops to a lower level, a line spectrum is emitted and its intensity is measured. Flame photometers rely on the use of filters to isolate the line emitted, which is detected by a photocell and its output is measured by a calibrated galvanometer. The method is applicable to 16 metals. Reliable results are only obtainable by careful control of the experimental conditions. These depend on temperature (i.e. the type and rate of flow of the flammable gas and the oxidant which is usually air), the rate of flow of the solution to the flame as well as the compound tested and solvent used. A method used to minimise the effects of these variables is to add a known constant amount of an internal standard of a compound of a metal other than the metal to be determined but with similar excitation characteristics. Ihe ratio of the intensities of the standard and the test sample is determined. A calibration plot of the logarithm of the intensity ratio and the logarithm of the concentration of the test element is drawn. The concentration of an unknown is found by interpolation of the calibration plot. Alternatively, the standard additions method as in Sec.2.4.3 is used. In all cases, allowance should be made for any dilution effects. [Pg.29]

In these techniques, the excited atom source is not a flame but a plasma e.g. Ar plasma, a d.c. or a.c. arc or a spark. Since higher temperatures are achieved in plasmas, excited ions may be also formed as well as excited atoms. In any case, the lines are more numerous than in other techniques. The lines to be detected are selected by narrow band pass monochromator and are detected by sensitive photomultipliers. The functions of the instrument are controlled by a micro-processor and are displayed on a photographic plate, a cathode ray tube, a recorder or a printer. Qualitatively, the lines obtained from a sample are compared with tables compiled for various elements as atoms or ions. The latter are more intense when plasmas are used. The presence of 3 major lines of an element is taken as positive identification. Most elements can be identified by these methods. Solid samples can be embedded in an electrode of the source. Quantitative analysis is possible for instruments giving an intensity reading. By assigning a channel for each element, the measured intensity depends on the amount of element present. [Pg.29]

Inductively coupled plasma-atomic emission spectrometry (ICP-AES) [Pg.29]

This is a recent rapid technique particularly suitable for the determination of a number of elements simultaneously (Anal.Chem.6Xl99l)l2A). A demonstration of inductively [Pg.29]


Vibrational Spectroscopy. Infrared absorption spectra may be obtained using convention IR or FTIR instrumentation the catalyst may be present as a compressed disk, allowing transmission spectroscopy. If the surface area is high, there can be enough chemisorbed species for their spectra to be recorded. This approach is widely used to follow actual catalyzed reactions see, for example. Refs. 26 (metal oxide catalysts) and 27 (zeolitic catalysts). Diffuse reflectance infrared reflection spectroscopy (DRIFT S) may be used on films [e.g.. Ref. 28—Si02 films on Mo(llO)]. Laser Raman spectroscopy (e.g.. Refs. 29, 30) and infrared emission spectroscopy may give greater detail [31]. [Pg.689]

The above fomuilae for the absorption spectrum can be applied, with minor modifications, to other one-photon spectroscopies, for example, emission spectroscopy, photoionization spectroscopy and photodetachment spectroscopy (photoionization of a negative ion). For stimulated emission spectroscopy, the factor of fflj is simply replaced by cOg, the stimulated light frequency however, for spontaneous emission... [Pg.248]

The observation of a bend progression is particularly significant. In photoelectron spectroscopy, just as in electronic absorption or emission spectroscopy, the extent of vibrational progressions is governed by Franck-Condon factors between the initial and final states, i.e. the transition between the anion vibrational level u" and neutral level u is given by... [Pg.879]

There are two fimdamental types of spectroscopic studies absorption and emission. In absorption spectroscopy an atom or molecule in a low-lying electronic state, usually the ground state, absorbs a photon to go to a higher state. In emission spectroscopy the atom or molecule is produced in a higher electronic state by some excitation process, and emits a photon in going to a lower state. In this section we will consider the traditional instrumentation for studying the resulting spectra. They define the quantities measured and set the standard for experimental data to be considered. [Pg.1120]

Pibel C D, Sirota E, Brenner J and Dai H L 1998 Nanosecond time-resolved FTIR emission spectroscopy monitoring the energy distribution of highly vibrationally excited molecules during collisional deactivation J. Chem. Phys. 108 1297-300... [Pg.1176]

In principle, emission spectroscopy can be applied to both atoms and molecules. Molecular infrared emission, or blackbody radiation played an important role in the early development of quantum mechanics and has been used for the analysis of hot gases generated by flames and rocket exhausts. Although the availability of FT-IR instrumentation extended the application of IR emission spectroscopy to a wider array of samples, its applications remain limited. For this reason IR emission is not considered further in this text. Molecular UV/Vis emission spectroscopy is of little importance since the thermal energies needed for excitation generally result in the sample s decomposition. [Pg.434]

The focus of this section is the emission of ultraviolet and visible radiation following thermal or electrical excitation of atoms. Atomic emission spectroscopy has a long history. Qualitative applications based on the color of flames were used in the smelting of ores as early as 1550 and were more fully developed around 1830 with the observation of atomic spectra generated by flame emission and spark emission.Quantitative applications based on the atomic emission from electrical sparks were developed by Norman Lockyer (1836-1920) in the early 1870s, and quantitative applications based on flame emission were pioneered by IT. G. Lunde-gardh in 1930. Atomic emission based on emission from a plasma was introduced in 1964. [Pg.434]

Multielemental Analysis Atomic emission spectroscopy is ideally suited for multi-elemental analysis because all analytes in a sample are excited simultaneously. A scanning monochromator can be programmed to move rapidly to an analyte s desired wavelength, pausing to record its emission intensity before moving to the next analyte s wavelength. Proceeding in this fashion, it is possible to analyze three or four analytes per minute. [Pg.436]


See other pages where Emission spectroscopies is mentioned: [Pg.44]    [Pg.156]    [Pg.156]    [Pg.176]    [Pg.128]    [Pg.244]    [Pg.248]    [Pg.1120]    [Pg.1120]    [Pg.1121]    [Pg.1131]    [Pg.1132]    [Pg.1142]    [Pg.1179]    [Pg.1591]    [Pg.2507]    [Pg.226]    [Pg.373]    [Pg.395]    [Pg.434]    [Pg.435]    [Pg.446]   


SEARCH



A-Ray Photoelectron Spectroscopy (XPS) - Adsorbate-core Emission

ATOMIC emission spectroscopy 1 Technique

Absorption and Emission Spectroscopy of Carbon Nanotubes

Absorption and emission spectroscopy

Absorption and emission spectroscopy with polarized light

Acoustic emission spectroscopy

And emission spectroscopy

Applications of Arc and Spark Emission Spectroscopy

Applications of Emission Spectroscopy

Applications of Flame Emission Spectroscopy in Pharmaceutical Analysis

Applications of ICP and DCP Atomic Emission Spectroscopy

Applications of ICP, DCP, and MP Atomic Emission Spectroscopy

Applications of plasma emission spectroscopy Biomedical and environmental

Arc and spark emission spectroscopy

Arc emission spectroscopy

Argon plasma emission spectroscopy

Atomic absorption and flame emission spectroscopy

Atomic emission spectroscopy

Atomic emission spectroscopy (AES

Atomic emission spectroscopy accuracy

Atomic emission spectroscopy apparatus

Atomic emission spectroscopy equipment

Atomic emission spectroscopy evaluation

Atomic emission spectroscopy flame sources

Atomic emission spectroscopy group 1 metals

Atomic emission spectroscopy hydrogen

Atomic emission spectroscopy multielement detection

Atomic emission spectroscopy plasma sources

Atomic emission spectroscopy precision

Atomic emission spectroscopy preparing sample

Atomic emission spectroscopy procedure

Atomic emission spectroscopy quantitative applications using

Atomic emission spectroscopy representative methods

Atomic emission spectroscopy selectivity

Atomic emission spectroscopy sensitivity

Atomic emission spectroscopy sources

Atomic emission spectroscopy spectra

Atomic emission spectroscopy standardizing method

Atomic emission spectroscopy studies

Atomic fluorescence emission spectroscopy

Atomic optical emission spectroscopy

Atomic optical emission spectroscopy applications

Atomic optical emission spectroscopy calibration

Atomic optical emission spectroscopy detectors

Atomic optical emission spectroscopy excitation sources

Atomic optical emission spectroscopy instrumentation

Atomic optical emission spectroscopy interferences

Atomic plasma emission spectroscopy

Basic atomic emission spectroscopy

Box 20-1 Atomic Emission Spectroscopy on Mars

Buffer exchange-atomic emission spectroscopy

Calibration atomic emission spectroscopy

Calibration plasma emission spectroscopy

Capillary atomic emission spectroscopy

Charge emission spectroscopy

Charge-Transfer Emission Spectroscopy

Chemical interferences plasma emission spectroscopy

Chemiluminescent emission spectroscopy

Cobalt emission spectroscopy

Complexation emission spectroscopy

Conventional emission spectroscopy

Conventional emission spectroscopy techniques

Coupled Plasma Atomic Emission Spectroscopy

Coupled plasma optical emission spectroscopy

Coupled plasma optical emission spectroscopy ICP-OES)

Defect level spectroscopy - thermal emission energies

Direct Couple Optical Emission Spectroscopy

Direct current plasma emission spectroscopy

Dispersed emission spectroscopy

Effect of Nuclear Decay Mossbauer Emission Spectroscopy

Electromagnetic spectrum atomic emission spectroscopy

Electron (Emission) Spectroscopies (EES)

Electron impact emission spectroscopy

Emission IR spectroscopy

Emission electronic spectroscopy

Emission spectroscopy analyzer

Emission spectroscopy biology

Emission spectroscopy excited state

Emission spectroscopy group 1 metals

Emission spectroscopy hydrogen

Emission spectroscopy laser-induced fluorescence

Emission spectroscopy metal complexes

Emission spectroscopy observation

Emission spectroscopy of dissociating molecules

Emission spectroscopy oxygen reactions

Emission spectroscopy transitions

Emission spectroscopy, accuracy

Emission spectroscopy, accuracy detection limits

Emission spectroscopy, accuracy precision

Emission spectroscopy, analytical method

Emission spectroscopy, determination

Emission spectroscopy, determination anions

Emission spectroscopy, lead

Emission spectroscopy, limitations

Emission spectroscopy, metal analysis

Emission spectroscopy, schematic

Emission spectroscopy. Inductively coupled

Energy levels emission/absorption spectroscopy

Environmental applications plasma emission spectroscopy

Excitation emission matrix spectroscopy

Excitation plasma emission spectroscopy

Excitation-emission fluorescence spectroscopy

FTIR emission spectroscopy

Field-emission spectroscopy

Flame emission spectroscopy

Flame emission spectroscopy (FES

Flame emission spectroscopy, water analysis

Flames atomic emission spectroscopy

Fluorescence emission spectroscopy

Fluorescence emission spectroscopy of plasma

Fourier Transform Infrared Emission Spectroscopy (FT-IRES)

Gas chromatography atomic emission spectroscopy

Glow atomic emission spectroscopy

Glow discharge optical emission spectroscopy

Glow discharge optical emission spectroscopy GD-OES)

Glow discharge optical emission spectroscopy GDOES)

High-performance liquid chromatography-inductively coupled plasma atomic emission spectroscopy

Hollow cathodes, atomic emission spectroscopy

Hydride generation-direct current plasma emission spectroscopy

ICP Atomic Emission Spectroscopy

ICP emission spectroscopy

ICP-AES atomic emission spectroscopy

ICP-optical emission spectroscopy

Impurity concentration measurements emission spectroscopy

In atomic emission spectroscopy

Induced coupled plasma atomic emission spectroscopy

Inductively atomic emission spectroscopy

Inductively coupled plasma atomic emission spectroscopy

Inductively coupled plasma atomic emission spectroscopy, ICP-AES

Inductively coupled plasma emission spectroscopy

Inductively coupled plasma optical emission spectroscopy ICP-OES)

Inductively coupled plasma optical emission spectroscopy instrumentation

Inductively coupled plasma with atomic emission spectroscopy

Inductively coupled plasma-optical emission spectroscopy

Infrared emission spectroscopy applications

Infrared spectroscopy emission measurements

Instrumentation atomic emission spectroscopy

Instrumentation for Emission Spectroscopy

Interferences in Arc and Spark Emission Spectroscopy

Interferences plasma emission spectroscopy

Internal standards plasma emission spectroscopy

Laser sources, atomic emission spectroscopy

Liquid samples plasma emission spectroscopy

Mass spectrometry optical emission spectroscopy

Mdssbauer emission spectroscopy

Metals flame atomic emission spectroscopy

Microwave-induced plasma atomic emission spectroscopy

Molecular emission spectroscopy

Mossbauer emission spectroscopy

Mossbauer emission spectroscopy catalysts

Nebulizers atomic emission spectroscopy

Optical emission spectroscopy

Optical emission spectroscopy (OES

Optical emission spectroscopy applications

Optical emission spectroscopy atomic fluorescence spectrometry

Optical emission spectroscopy excitation process

Optical properties, spectroscopy emission

Organic solutions, atomic emission spectroscopy

Photoelectron emission spectroscopy

Photoelectron spectroscopy electron emission from core

Photoelectron spectroscopy electron emission from valence

Picosecond emission spectroscopy

Plasma emission spectroscopy

Plasma emission spectroscopy applications

Plasma emission spectroscopy chemical speciation

Plasma emission spectroscopy excitation sources

Plasma emission spectroscopy hyphenated techniques

Plasma emission spectroscopy instrumentation

Plasma emission spectroscopy sample introduction systems

Plasma emission spectroscopy, inductively

Polymers plasma emission spectroscopy

Powders, atomic emission spectroscopy

Process acoustic emission spectroscopy

Product emission spectroscopy

Proton-induced x-ray emission spectroscopy

Qualitative atomic emission spectroscopy

Quantitative atomic emission spectroscopy

Secondary electron emission spectroscopy

Single molecule emission spectroscopy

Single photon emission computed tomography spectroscopy

Soft x-ray emission spectroscopy ,

Solid atomic emission spectroscopy

Solid samples plasma emission spectroscopy

Solid-State Emission Spectroscopy

Sources in atomic emission spectroscopy

Spark emission spectroscopy

Spark source emission spectroscopy

Spectral interferences plasma emission spectroscopy

Spectroscopy Auger electron emission

Spectroscopy dispersed emission spectrum

Spectroscopy infrared emission

Spectroscopy laser-induced emission

Spectroscopy stimulated emission pumping

Steady-State Emission and Transient Absorption Spectroscopy

Steady-state emission spectroscopy

Studies of Cyclodextrin Inclusion Complexes by Electronic (UV-Vis Absorption and Emission) Spectroscopy

Time-Gated Excitation-Emission Matrix Spectroscopy

Time-resolved emission spectroscopy

Time-resolved spectroscopies spontaneous emission

Trace elements optical emission spectroscopy

Transferred plasmas, atomic emission spectroscopy

Transient Emission Spectroscopy

Transient absorption spectroscopy stimulated emission

Transient infrared emission spectroscopy

Transient infrared emission spectroscopy TIRES)

Trapping atomic emission spectroscopy

Water flame emission spectroscopy

X-ray Emission and (Photo)Electron Spectroscopies

X-ray absorption and emission spectroscopies

X-ray emission spectroscopy

X-ray emission spectroscopy applications

X-ray photo-emission spectroscopy

© 2024 chempedia.info