Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Columns, capillary

There are numerous textbooks on and guides to chromatography, it is not intended to discuss the details of this technique in this chapter. The use of capillary columns, particularly in conjunction with headspace or purge and trap analysis, and modem bench-top mass spectrometers with large desk-top computing capacity, has produced a powerful technique for the analysis of VOCs. Typically, the separation, identification and quantification of up to about one hundred compounds in a single analysis are possible and routinely carried out. [Pg.182]

Fused silica is the most common inert and temperature stable material used in the manufacture of capillary columns. They are flexible, robust and easy to install. Many laboratories find difficulty in separating the ever increasing list of volatile compounds requiring analysis. Many phases suffer from low thermal stability, and analysis times are unacceptably long. The most common column used in the analysis of VOCs is the 624 column, available from several suppliers worldwide. [Pg.182]

This method is especially suitable for thermolabile components and for components which may form artifacts at high injector temperatures. An example of this is oxazepam. Use of a split/splitless injector leads to artifact formation, but this can be largely prevented by use of a temperature-programmable injector. [Pg.13]

In this method, the sample is injected directly onto the column. Here, the sample is not contained in the glass insert. Columns with a small inside diameter are unsuitable for this technique of sample itroduction. As on-column injection is a splitless method, only low-concentration samples can be injected. This method is suitable for polar and thermally unstable components. [Pg.13]

This special injection technique has advantages over all other methods of analyzing volatile substances with respect to separating power, sensitivity, ease of handling, and automation possibilities, and the headspace technique is therefore becoming more and more widely used. [Pg.14]

In a static headspace, the sample is in a closed static system in which conditions are at thermodynamic equilibrium. After establishment of the equilibrium, the [Pg.14]

For reproducible analysis, it is important that equilibrium between the phases should first be established. This equilibrium is influenced by the conditioning time and temperature. The removal of the sample for the GC analysis can be by means of overpressure or underpressure, although care must be taken that the temperature not only of the sample vessel but also of the metering device is kept constant. The HS vessels (vials) are closed with a PTFE (polytetrafluoroethylene) or aluminum-coated silicone septum. The septum itself is protected from the overpressure of the interior space by an aluminum cap. Before each analysis, a blank determination with an empty HS vial is carried out to detect whether any constituents of the materials of the septum or vial are being emitted. [Pg.15]


Used in virtually all organic chemistry analytical laboratories, gas chromatography has a powerful separation capacity. Using distillation as an analogy, the number of theoretical plates would vary from 100 for packed columns to 10 for 100-meter capillary columns as shown in Figure 2.1. [Pg.20]

The column is swept continuously by a carrier gas such as helium, hydrogen, nitrogen or argon. The sample is injected into the head of the column where it is vaporized and picked up by the carrier gas. In packed columns, the injected volume is on the order of a microliter, whereas in a capillary column a flow divider (split) is installed at the head of the column and only a tiny fraction of the volume injected, about one per cent, is carried into the column. The different components migrate through the length of the column by a continuous succession of equilibria between the stationary and mobile phases. The components are held up by their attraction for the stationary phase and their vaporization temperatures. [Pg.20]

Interest in this method has decreased since advances made in gas chromatography using high-resolution capillary columns (see article 3.3.3.) now enable complete identification by individual chemical component with equipment less expensive than mass spectrometry. [Pg.51]

The hydrocarbons are separated in another column and analyzed by a flame ionization detector, FID. As an example, Figure 3.13 shows the separation obtained for a propane analyzed according to the ISO 7941 standard. Note that certain separations are incomplete as in the case of ethane-ethylene. A better separation could be obtained using an alumina capillary column, for instance. [Pg.72]

The resolution of capillary columns enables the separation of all principal components of a straight-run gasoline. The most frequently used stationary phases are silicone-based, giving an order of hydrocarbon elution times close to the order of increasing boiling point. [Pg.73]

Gas chromatography is not an identification method the components must be identified after their separation by capillary column. This is done by coupling to the column a mass spectrometer by which the components can be identified with the aid of spectra libraries. However tbe analysis takes a long time (a gasoline contains aboutTwo hundred components) so it is not practical to repeat it regularly. Furthermore, analysts have developed te hpiques for identifying... [Pg.73]

One of the most important advances in column construction has been the development of open tubular, or capillary columns that contain no packing material (dp = 0). Instead, the interior wall of a capillary column is coated with a thin film of the stationary phase. The absence of packing material means that the mobile phase... [Pg.562]

Another approach to improving resolution is to use thin films of stationary phase. Capillary columns used in gas chromatography and the bonded phases commonly used in HPLC provide a significant decrease in plate height due to the reduction of the Hs term in equation 12.27. [Pg.563]

In gas chromatography (GC) the sample, which may be a gas or liquid, is injected into a stream of an inert gaseous mobile phase (often called the carrier gas). The sample is carried through a packed or capillary column where the sample s components separate based on their ability to distribute themselves between the mobile and stationary phases. A schematic diagram of a typical gas chromatograph is shown in Figure 12.16. [Pg.563]

The most common mobile phases for GC are He, Ar, and N2, which have the advantage of being chemically inert toward both the sample and the stationary phase. The choice of which carrier gas to use is often determined by the instrument s detector. With packed columns the mobile-phase velocity is usually within the range of 25-150 mF/min, whereas flow rates for capillary columns are 1-25 mF/min. Actual flow rates are determined with a flow meter placed at the column outlet. [Pg.563]

A chromatographic column provides a location for physically retaining the stationary phase. The column s construction also influences the amount of sample that can be handled, the efficiency of the separation, the number of analytes that can be easily separated, and the amount of time required for the separation. Both packed and capillary columns are used in gas chromatography. [Pg.564]

To minimize the multiple path and mass transfer contributions to plate height (equations 12.23 and 12.26), the packing material should be of as small a diameter as is practical and loaded with a thin film of stationary phase (equation 12.25). Compared with capillary columns, which are discussed in the next section, packed columns can handle larger amounts of sample. Samples of 0.1-10 )J,L are routinely analyzed with a packed column. Column efficiencies are typically several hundred to 2000 plates/m, providing columns with 3000-10,000 theoretical plates. Assuming Wiax/Wiin is approximately 50, a packed column with 10,000 theoretical plates has a peak capacity (equation 12.18) of... [Pg.564]

A capillary column that does not contain a particulate packing material. [Pg.564]

Capillary Columns Capillary, or open tubular columns are constructed from fused silica coated with a protective polymer. Columns may be up to 100 m in length with an internal diameter of approximately 150-300 )J,m (Figure 12.17). Larger bore columns of 530 )J,m, called megabore columns, also are available. [Pg.564]

Capillary columns are of two principal types. Wall-coated open tuhular columns (WCOT) contain a thin layer of stationary phase, typically 0.25 pm thick, coated on the capillary s inner wall. In support-coated open tuhular columns (SCOT), a thin layer of a solid support, such as a diatomaceous earth, coated with a liquid stationary phase is attached to the capillary s inner wall. [Pg.565]

An important problem with all liquid stationary phases is their tendency to bleed from the column. The temperature limits listed in Table 12.2 are those that minimize the loss of stationary phase. When operated above these limits, a column s useful lifetime is significantly shortened. Capillary columns with bonded or... [Pg.566]

A technique for injecting samples onto a capillary column in which only a small portion of the sample enters the column. [Pg.568]

A technique for injecting a sample onto a capillary column that allows a higher percentage of the sample to enter the column. [Pg.568]

The direct injection of thermally unstable samples onto a capillary column. [Pg.568]

Environmental Analysis One of the most important environmental applications of gas chromatography is for the analysis of numerous organic pollutants in air, water, and wastewater. The analysis of volatile organics in drinking water, for example, is accomplished by a purge and trap, followed by their separation on a capillary column with a nonpolar stationary phase. A flame ionization, electron capture, or... [Pg.571]

Time, Cost, and Equipment Analysis time can vary from several minutes for samples containing only a few constituents to more than an hour for more complex samples. Preliminary sample preparation may substantially increase the analysis time. Instrumentation for gas chromatography ranges in price from inexpensive (a few thousand dollars) to expensive (more than 50,000). The more expensive models are equipped for capillary columns and include a variety of injection options and more sophisticated detectors, such as a mass spectrometer. Packed columns typically cost 50- 200, and the cost of a capillary column is typically 200- 1000. [Pg.578]

Despite their importance, gas chromatography and liquid chromatography cannot be used to separate and analyze all types of samples. Gas chromatography, particularly when using capillary columns, provides for rapid separations with excellent resolution. Its application, however, is limited to volatile analytes or those analytes that can be made volatile by a suitable derivatization. Liquid chromatography can be used to separate a wider array of solutes however, the most commonly used detectors (UV, fluorescence, and electrochemical) do not respond as universally as the flame ionization detector commonly used in gas chromatography. [Pg.596]

The narrow bore of the capillary column and the relative thickness of the capillary s walls are important. When an electric field is applied to a capillary containing a conductive medium, such as a buffer solution, current flows through the capillary. This current leads to Joule heating, the extent of which is proportional to the capillary s radius and the magnitude of the electric field. Joule heating is a problem because it changes the buffer solution s viscosity, with the solution at the center of the... [Pg.601]

Schematic diagram showing a cross section of a capillary column for capillary electrophoresis. Schematic diagram showing a cross section of a capillary column for capillary electrophoresis.
An injection technique in capillary electrophoresis in which pressure is used to inject sample into the capillary column. [Pg.602]

A means of concentrating solutes in capillary electrophoresis after their injection onto the capillary column. [Pg.603]

A form of capillary electrophoresis in which the capillary column contains a gel enabling separations based on size. [Pg.606]

Procedure. A vitamin B complex tablet Is crushed and placed In a beaker with 20.00 mL of a 50% v/v methanol solution that Is 20 mM In sodium tetraborate and contains 100.0 ppm of o-ethoxybenzamIde. After mixing for 2 min to ensure that the B vitamins are dissolved, a 5.00-mL portion Is passed through a 0.45- xm filter to remove Insoluble binders. An approximately 4-nL sample Is loaded Into a 50- xm Internal diameter capillary column. For CZE the capillary column contains a 20 mM pH 9 sodium tetraborate/sodlum dIhydrogen phosphate buffer. For MEKC the buffer Is also 150 mM In sodium dodecylsulfate. A 40-kV/m electric field Is used to effect both the CZE and MEKC separations. [Pg.607]

Although aimed at the introductory class, this simple experiment provides a nice demonstration of the use of GG for a qualitative analysis. Students obtain chromatograms for several possible accelerants using headspace sampling and then analyze the headspace over a sealed sample of charred wood to determine the accelerant used in burning the wood. Separations are carried out using a wide-bore capillary column with a stationary phase of methyl 50% phenyl silicone and a flame ionization detector. [Pg.610]

This somewhat lengthy experiment provides a thorough introduction to the use of GG for the analysis of trace-level environmental pollutants. Sediment samples are extracted by sonicating with 3 X 100-mL portions of 1 1 acetone hexane. The extracts are then filtered and concentrated before bringing to a final volume of 10 mL. Samples are analyzed with a capillary column using a stationary phase of 5% phenylmethyl silicone, a splitless injection, and an EGD detector. [Pg.611]

The following data were obtained for four compounds separated on a 20-m capillary column. [Pg.615]

Zhou and colleagues determined the %w/w H2O in methanol by GG, using a capillary column coated with a nonpolar stationary phase and a thermal conductivity detector. A series of calibration standards gave the following results. [Pg.616]

The following data have been reported for the gas chromatographic analysis of p-xylene and methylisobutylketone (MIBK) on a capillary column. ... [Pg.617]


See other pages where Columns, capillary is mentioned: [Pg.20]    [Pg.21]    [Pg.21]    [Pg.73]    [Pg.9]    [Pg.562]    [Pg.563]    [Pg.565]    [Pg.565]    [Pg.567]    [Pg.568]    [Pg.609]    [Pg.609]    [Pg.610]    [Pg.611]    [Pg.612]   
See also in sourсe #XX -- [ Pg.9 ]

See also in sourсe #XX -- [ Pg.461 ]

See also in sourсe #XX -- [ Pg.519 ]

See also in sourсe #XX -- [ Pg.883 ]

See also in sourсe #XX -- [ Pg.188 , Pg.191 ]

See also in sourсe #XX -- [ Pg.199 ]

See also in sourсe #XX -- [ Pg.219 ]

See also in sourсe #XX -- [ Pg.122 ]

See also in sourсe #XX -- [ Pg.519 ]

See also in sourсe #XX -- [ Pg.102 , Pg.105 , Pg.108 , Pg.110 , Pg.115 ]

See also in sourсe #XX -- [ Pg.15 , Pg.16 ]

See also in sourсe #XX -- [ Pg.950 , Pg.958 ]

See also in sourсe #XX -- [ Pg.182 ]

See also in sourсe #XX -- [ Pg.183 ]

See also in sourсe #XX -- [ Pg.791 , Pg.795 , Pg.798 , Pg.800 , Pg.801 ]

See also in sourсe #XX -- [ Pg.8 , Pg.13 , Pg.14 , Pg.84 ]

See also in sourсe #XX -- [ Pg.127 , Pg.327 ]

See also in sourсe #XX -- [ Pg.124 ]

See also in sourсe #XX -- [ Pg.31 , Pg.42 ]

See also in sourсe #XX -- [ Pg.12 , Pg.39 ]

See also in sourсe #XX -- [ Pg.2 , Pg.11 , Pg.37 , Pg.54 , Pg.60 , Pg.62 , Pg.121 , Pg.122 , Pg.124 , Pg.146 ]

See also in sourсe #XX -- [ Pg.489 , Pg.495 ]

See also in sourсe #XX -- [ Pg.5 , Pg.68 ]

See also in sourсe #XX -- [ Pg.132 ]




SEARCH



Adsorption capillary columns

Aluminum clad capillary column

Aluminum-clad fused-silica capillary columns

Biogenic amines capillary columns

Bleed capillary column

Capacity capillary column performance

Capacity fused-silica capillary columns

Capillary Column Choice and Separation Optimization

Capillary Column Selection

Capillary Columns and Inlets

Capillary cation exchange column

Capillary column SFC

Capillary column chromatography

Capillary column columns)

Capillary column columns)

Capillary column composition

Capillary column coupling isotachoelectrophoresis

Capillary column gas chromatography

Capillary column gas-liquid chromatography

Capillary column guard columns

Capillary column isomeric

Capillary column manufacturers

Capillary column overview

Capillary column preparation

Capillary column regeneration

Capillary column selection factors

Capillary column selection guidelines

Capillary column separation number

Capillary column separations

Capillary column separations packed

Capillary column stabilization

Capillary column technology

Capillary column technology fused-silica

Capillary column technology glass

Capillary column technology materials

Capillary column wall coated, efficiency

Capillary columns analysis time

Capillary columns chromatograms

Capillary columns chromatographic performance

Capillary columns conditioning

Capillary columns flow rate through

Capillary columns ideal

Capillary columns inner diameter

Capillary columns installation

Capillary columns length

Capillary columns monolith silica

Capillary columns open geometry

Capillary columns particle size

Capillary columns particle-packed

Capillary columns plate numbers

Capillary columns pressure

Capillary columns retention time reproducibility

Capillary columns sample capacity

Capillary columns static coating

Capillary columns tandem

Capillary columns temperature

Capillary columns, glass

Capillary columns, high resolution capability

Capillary columns, in gas chromatography

Capillary electrochromatography open-tubular column

Capillary liquid chromatography columns

Capillary liquid columns

Capillary wall coated open tubular WCOT) columns

Capillary zone electrophoresis columns

Capillary, coatings column

Capillary-column GCMS

Chiral capillary columns

Chiraldex capillary columns

Chromatographic Performance of Capillary Columns

Cleaning procedure, capillary column

Coating fused-silica capillary columns

Column capillary electrophoresis

Column coupling capillary isotachoelectrophoresis and isotachelectrophoresis

Column coupling, capillary

Column coupling, capillary isotachophoresis

Column preparation drawing capillary columns

Column, capillary characteristics

Column, capillary dead volume

Column, capillary dispersion

Column, capillary dispersion, extra

Column, capillary efficiency

Column, capillary environment

Column, capillary equation

Column, capillary equation for

Column, capillary flow rate

Column, capillary heat capacity

Column, capillary limitations

Column, capillary longitudinal diffusion

Column, capillary minimum HETP

Column, capillary multipath

Column, capillary optimized packed columns

Column, capillary optimized preparative columns

Column, capillary optimum, open tubular

Column, capillary overload

Column, capillary particle limits

Column, capillary phase

Column, capillary physical nature

Column, capillary preparative

Column, capillary program for

Column, capillary purpose

Column, capillary radius

Column, capillary resolving power

Column, capillary stationary phase

Column, capillary theories

Column, capillary thermostat

Column, capillary velocity

Column, capillary wall thickness, limitations

Column, capillary weight, limits

Columns packed versus capillary

Columns, packed or capillaries

Commercially prepared capillary columns

Composition of a Capillary Column

Connectors, capillary columns

Deactivation capillary columns

Deactivation fused-silica capillary columns

Detector capillary columns

Disposability, capillary column

Durability, capillary columns

Equipment capillary columns

Extrusion of a Fused-Silica Capillary Column

Fast capillary columns

Fast capillary columns, sample capacity

Film thickness capillary columns

Fused silica capillary columns

Fused-silica capillary columns characteristics

Fused-silica capillary columns polyimide-clad

Fused-silica capillary columns popularity

Fused-silica capillary columns preparation

Fused-silica capillary columns stainless steel

Fused-silica capillary columns synthetic

Fused-silica capillary columns tubing

Fused-silica-lined stainless-steel capillary columns

GC capillary column

Gas-solid adsorption capillary columns

Glass capillary columns characteristics

Glass capillary columns coating

Glass capillary columns development

Glass capillary columns, wide-bore

High performance capillary electrophoresis columns

Hydrocarbons capillary column chromatography

Inert-fused silica capillary columns

Injection capillary columns

Injection systems large bore capillary columns

Injection systems small bore capillary columns

Installation preparations, capillary columns

Instrumentation capillary column

Inverse capillary column

Liquid Column in a Capillary

Materials, capillary column

Megabore capillary columns

Metal capillary columns

Metal-clad capillary columns

Monolith Capillary Columns

Monolithic capillary column technologies

Monolithic capillary columns

Monolithic imprinted polymer capillary columns

Multiplexed capillary columns

Open tubular capillary columns

Organic acids capillary columns

PDMS-coated capillary columns

Packed Capillary Column and Unified Chromatography

Packed and Capillary Columns

Packed capillary columns

Packed capillary columns chromatography

Packed capillary columns electrokinetic packing

Packed capillary columns monolithic

Packed capillary columns optimization

Packed capillary columns slurry packing

Packed fused silica capillary columns

Plate height capillary columns

Preparation of Fused-Silica Capillary Columns

Preparation of capillary columns

Programmable selectivity, with tandem capillary columns

Resolution capillary columns

Retention capillary columns

Retention time fused-silica capillary columns

Retention times during capillary column

Retention times during capillary column separation

Selection of Capillary Column for GC Analysis

Selectivity fused-silica capillary columns

Silica capillary column for

Silica capillary columns

Stainless steel capillary columns

Stationary phase fused-silica capillary columns

Temperature fused-silica capillary columns

Tubing capillary columns

Use in capillary columns

Validation, capillary columns

Vaporizer, capillary column chromatography

Velocity flow profile capillary columns

Wall-coated capillary columns

© 2024 chempedia.info