Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Splitless methods

In this method, the sample is injected directly onto the column. Here, the sample is not contained in the glass insert. Columns with a small inside diameter are unsuitable for this technique of sample itroduction. As on-column injection is a splitless method, only low-concentration samples can be injected. This method is suitable for polar and thermally unstable components. [Pg.13]

This type of injection allows the introduction of much higher sample volumes than those in split or splitless procedures. The volume depends on the length of the pre-column it can reach several dozens of milliliters, unlike the 0.1 to 2.0 (xL generally injected in split and splitless methods. [Pg.7]

Conditions apparatus, Hewlett-Packard HP5890 equipped with an HP5972 mass-selective ion detector (quadruple) column, PTE-5 (30 m x 0.25-mm i.d.) with 0.25- am film thickness column temperature, 50 °C (1 min), increased at 20 °C min to 150 °C(5 min) and then at 4 °Cmin to 280 °C (30 min) inlet and detector (GC/MS transfer line) temperature, 250 and 280 °C, respectively gas flow rate, He carrier gas ImLmin" injection method, splitless mode solvent delay, 3 min electron ionization voltage, 70eV scan rate, 1.5 scanss scanned-mass range, m/z 50-550. The retention times of benfluralin, pendimethalin and trifluralin are 15.2, 25.1 and... [Pg.393]

Chlornitrofen and nitrofen conditions for GC/MS column, cross-linked methyl silicone capillary (12 m x 0.22-mm i.d., 0.33- am film thickness) column temperature, 60 °C (1 min), 18 °C min to 265 °C inlet, transfer line and ion source temperature, 260, 200 and 200 °C, respectively He gas column head pressure, 7.5 psi injection method, splitless mode solvent delay, 3 min electron ionization voltage, 70 eV scan rate, 0.62 s per scan cycle scanned mass range, m/z 100-400. The retention times for chlornitrofen and nitrofen were 11.8 and 11.3 min, respectively. The main ions of the mass spectrum of chlornitrofen were at m/z 317, 319 and 236. Nitrofen presented a fragmentation pattern with the main ions at m/z 283, 202 and 285. ... [Pg.457]

The residue levels of 46 pesticides, including oxyfluorfen in soil, were determined using GC/ITDMS as described in S ection 3.2.1. The conditions for GC/ITDMS were as follows column, fused-silica capillary (30 m x 0.25-mm-i.d.) with a0.25- am bonded phase ofDB-5 column temperature, 50 °C (1 min), 30 °Cmin to 130 °C, 5 °C min to 270 °C inlet and transfer temperature, 270 and 220 °C, respectively He gas with column head pressure, 12psi injection method, splitless mode. The retention time and quantitation ion of oxyfluorfen were 23.9 min and mjz 252, respectively. ... [Pg.460]

In the above-mentioned method by Hodgeson el acifluorfen was determined by GC/ECD after the extraction with a 47-mm PS-DVB disk and derivatization with diazomethane. The conditions for GC/ECD were as follows column, DB-5 fused silica (30 m X 0.32-mm i.d., 0.25-p.m film thickness) He carrier gas velocity, 25 cm s (210°C), detector makeup gas, methane-argon (5 95), 30mLmin column temperature, 50 °C (5 min), 10°Cmin to 210 °C (5 min) and to 230 °C (10 min) injection port and detector temperature, 220 and 300 °C, respectively injection method, splitless mode. The recovery of acifluorfen from purified water, dechlorinated tap water and high humic content surface water fortified at 0.5-2.0 ug was 59-150% and the LOD was 25 ngL Acifluorfen after derivatization with various chlorofor-mates was also determined by GC/MS using an SE-54 column (25 m x 0.20-mm i.d., 0.32- um Aim thickness), and the average recovery of acifluorfen fortified between 0.05 and 0.2 ugL was 78%. ... [Pg.464]

Major advantages of LVI methods are higher sensitivity (compare the 100-1000 iL volume in LVI to the maximum injection volume of about 1 iL in conventional splitless or on-column injection), elimination of sample preparation steps (such as solvent evaporation) and use in hyphenated techniques (e.g. SPE-GC, LC-GC, GC-MS), which gives opportunities for greater automation, faster sample throughput, better data quality, improved quantitation, lower cost per analysis and fewer samples re-analysed. At-column is a very good reference technique for rapid LVI. Characteristics of LVI methods are summarised in Tables 4.19 and 4.20. Han-kemeier [100] has discussed automated sample preparation and LVI for GC with spectrometric detection. [Pg.191]

One method that prevents overloading of narrow-bore capillary columns is split injection. The inlets are usually bimodal split/splitless inlets, and either mode can be selected for a given analytical method. In the split mode, the sample is rapidly vaporized in the inlet and a portion is introduced into the column in a narrow band with carrier gas, while the rest of the sample is vented (Dybowski and Kaiser, 2002). The amount introduced can vary for each method and is chosen as a ratio, e.g., 1 100. Easily vaporized compounds may preferentially vent, leading to the introduction of a nonrepresentative sample to the column (Watson, 1999). When the splitless mode is chosen, the entire sample is introduced into the column and the vent is opened after a predetermined period of time, to flush the excess solvent from the injector (Dybowski and Kaiser, 2002). [Pg.3]

Chromatographic Conditions. GC/MS-MS analyses were performed on a Varian 3800 gas chromatograph (Varian Chromatography Systems, Walnut Creek, CA) equipped with a 1079 split/splitless injector and a ion trap spectrometer (Varian Saturn 2000, Varian Chromatography Systems) with a waveboard for MS-MS analysis. The system was operated by Saturn GC/MS Workstation v5.4 software. The MS-MS detection method was adapted from reference. PCBs were separated on a 25 m length x 0.32 mm i.d., CPSil-8 column coated with a 0.25-pm film. The GC oven temperature program was as follows 90 °C hold 2 min, ramp 30 °C/min to 170 °C, hold for 10 min, rate 3 °C/ min to 250 °C, rate 20 °C/min to a final temperature of 280 °C, and hold for 5 min. Helium was employed as the carrier gas, with a constant column flow of 1.0 mL/min. [Pg.94]

The mode of injection in GC-based methods can affect the recoveries of diazinon. In a study of the determination of organophosphorus pesticides in milk and butterfat, it was found that the recoveries of diazinon from butterfat, calculated relative to organic solutions of standard compounds, were 125% and 84% for splitless and hot on-column injections, respectively (Emey et al. 1993). Recoveries from milk were not dependent on the mode of injection. It was concluded that the sample matrix served to increase diazinon transfer to the GC column by reducing thermal stress imposed on the analytes and by blocking active sites within the injector. Therefore, on-column injection should be used in order to prevent bias when organic solutions of standard compounds are used for quantitation if this is not possible, the matrix must be present at low concentrations or the calibration standards must be prepared in residue-free samples to avoid unknown bias. [Pg.174]

Glycerin is used in Nasonex primarily as a humectant. For its quantification, both capillary gas chromatography method and HPLC methods may be selected. The GC is equipped with a flame-ionization defector, a 0.53 mm x 30 m fused silica analytical column coated with 3.0-p,mG43 stationary phase, and a 0.53 mm x 5 m silica guard column deactivated with phenylmethyl siloxane. The carrier gas was helium with a linear velocity of about 35 cm/s. The injection port and detector temperature was maintained at 240 and 260°C, respectively. The injection mode is splitless. The column temperature is programmed to be maintained at about 40°C for 20 min, then to increase to 250°C at a rate of 10°C/min and to hold at 250°C for 15 min. [Pg.88]

When working with capillary columns, the splitless mode is used for very dilute samples. In this mode, the injection is made very slowly, leaving valve no. 2 in the closed position (Fig. 2.5) for approximately 0.5 to 1 min. This allows vaporisation of the compounds and solvent in the first decimetre of the column by a complex mechanism of dissolution in the stationary phase, which is saturated with solvent. Compound discrimination is very weak using this method. The proper use of this injection mode, which demands some experience, requires a temperature program that starts with a colder temperature so that the solvent can precede the analytes in the column. This mode is typically used for trace analyses. The opening of valve no. 2 eliminates, from the injector, compounds which are less volatile and that can interfere with the analyses. [Pg.27]

Splitless or direct sampling procedures have been extensively studied. These methods involved trapping of the high-boiling materials in a small cooled zone followed by rapid desorption and introduction to the column. The column itself has been used at a low temperature to allow the solvent to pass while the high-boiling material was trapped on the front end. Once the solvent has passed the column is quickly heated to the desired temperature. [Pg.315]

The core technology used in the analysis of aroma chemicals is gas chromatography (GC) therefore, foods must be sampled so they can be introduced on to a GC column. For liquid samples it is possible to inject them into split, splitless, or on-column injectors directly. This is the preferred method for the analysis of synthetic aromas, essential oils, and aroma standards however, solid or dilute liquid samples need to be extracted, distilled, or gas-phase generated in order to obtain useful results. This unit begins with simple direct analysis of a synthetic flavor (see Basic Protocol 1) followed by the analysis of a dilute liquid sample by solvent extraction (see Basic Protocol 2). It ends with a protocol for determining retention indices (see Support Protocol). [Pg.993]


See other pages where Splitless methods is mentioned: [Pg.200]    [Pg.58]    [Pg.217]    [Pg.325]    [Pg.186]    [Pg.200]    [Pg.58]    [Pg.217]    [Pg.325]    [Pg.186]    [Pg.26]    [Pg.393]    [Pg.393]    [Pg.127]    [Pg.129]    [Pg.131]    [Pg.323]    [Pg.190]    [Pg.191]    [Pg.464]    [Pg.468]    [Pg.32]    [Pg.374]    [Pg.356]    [Pg.64]    [Pg.157]    [Pg.88]    [Pg.92]    [Pg.252]    [Pg.27]    [Pg.338]    [Pg.541]    [Pg.208]    [Pg.211]    [Pg.212]    [Pg.42]    [Pg.23]    [Pg.261]    [Pg.47]   
See also in sourсe #XX -- [ Pg.58 ]




SEARCH



© 2024 chempedia.info