Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sampling volume

Using the theorem that the sufficiency condition for mathematical correctness in 3D-reconstruction is fulfilled if all planes intersecting the object have to intersect the source-trajectory at least in one point [8], it is possible to generalise Feldkamp s method. Using projection data measured after changing the sotuce-trajectory from circular to spiral focus orbit it is possible to reconstruct the sample volume in a better way with the Wang algorithm [9]. [Pg.494]

The method of volume rendering uses the whole sample volume for visualization. Therefor semitransparent representations of the samples inner structure are possible and the detection of small cracks or faults is much easier compared to the surfaces based techniques (Fig. 4 b). From its principle volume rendering is more time consuming compared to surface representation. [Pg.495]

Given that R, can be arbitrarily chosen as anywhere in the sample volume of an isotropic liquid in the absence... [Pg.560]

Titrations conducted with microliter or picoliter sample volumes require a smaller absolute amount of analyte. For example, diffusional titrations have been successfully conducted on as little as 29 femtomoles (10 mol) of nitric acid. Nevertheless, the analyte must still be present in the sample at a major or minor level for the titration to be performed accurately and precisely. [Pg.312]

Analyte Sample Volume Sample (liL) Concentration Range Sampling Frequency (h-1)... [Pg.656]

Duarte and colleagues used a factorial design to optimize a flow injection analysis method for determining penicillin potentiometricallyd Three factors were studied—reactor length, carrier flow rate, and sample volume, with the high and low values summarized in the following table. [Pg.702]

Fig. 5. Anion-exchange separation of insulin and insulin A- and B-chains, over diethylaminoethyl (DEAF) in a 10.9 x 200 mm column having a volume of 18.7 mL. Sample volume is 0.5 mL and protein concentration ia 16.7 mAf Tris buffer at pH 7.3 is 1 mg/mL for each component ia the presence of EDTA. Eluent (also 16.7 mAf Tris buffer, pH 7.3) flow rate is 1.27 ml,/min, and protein detection is by uv absorbance at 280 nm. The straight line depicts the salt... Fig. 5. Anion-exchange separation of insulin and insulin A- and B-chains, over diethylaminoethyl (DEAF) in a 10.9 x 200 mm column having a volume of 18.7 mL. Sample volume is 0.5 mL and protein concentration ia 16.7 mAf Tris buffer at pH 7.3 is 1 mg/mL for each component ia the presence of EDTA. Eluent (also 16.7 mAf Tris buffer, pH 7.3) flow rate is 1.27 ml,/min, and protein detection is by uv absorbance at 280 nm. The straight line depicts the salt...
Column Si. Size-exclusion chromatography columns are generally the largest column on a process scale. Separation is based strictly on diffusion rates of the molecules inside the gel particles. No proteins or other solutes are adsorbed or otherwise retained owing to adsorption, thus, significant dilution of the sample of volume can occur, particularly for small sample volumes. The volumetric capacity of this type of chromatography is determined by the concentration of the proteins for a given volume of the feed placed on the column. [Pg.50]

Fig. 8. Representation of measurement of elution volume, as a function of sample volume (a) <2% of bed volume, (b) >2% and (c) >2% and giving a plateau region which has the same concentration as the iajected sample A represents the inflection poiat. See text. Fig. 8. Representation of measurement of elution volume, as a function of sample volume (a) <2% of bed volume, (b) >2% and (c) >2% and giving a plateau region which has the same concentration as the iajected sample A represents the inflection poiat. See text.
Biomolecule Separations. Advances in chemical separation techniques such as capillary zone electrophoresis (cze) and sedimentation field flow fractionation (sfff) allow for the isolation of nanogram quantities of amino acids and proteins, as weU as the characterization of large biomolecules (63—68) (see Biopolymers, analytical techniques). The two aforementioned techniques, as weU as chromatography and centrifugation, ate all based upon the differential migration of materials. Trends in the area of separations are toward the manipulation of smaller sample volumes, more rapid purification and analysis of materials, higher resolution of complex mixtures, milder conditions, and higher recovery (69). [Pg.396]

The quantity of sample required comprises two parts the volume and the statistical sample size. The sample volume is selected to permit completion of all required analytical procedures. The sample size is the necessary number of samples taken from a stream to characterize the lot. Sound statistical practices are not always feasible either physically or economically in industry because of cost or accessibiUty. In most sampling procedures, samples are taken at different levels and locations to form a composite sample. If some prior estimate of the population mean, and population standard deviation. O, are known or may be estimated, then the difference between that mean and the mean, x, in a sample of n items is given by the following ... [Pg.298]

Air—electric samplers can be installed directly in the pipe wall. One type of Hquid sampler is operated by a solenoid valve that activates an air cylinder. A shaft is moved in and out of the pipe by this cylinder and samples are expeUed into a container below the sampler. Sample volumes of from 2—30 mL are possible. [Pg.303]

Radiometry. Radiometry is the measurement of radiant electromagnetic energy (17,18,134), considered herein to be the direct detection and spectroscopic analysis of ambient thermal emission, as distinguished from techniques in which the sample is actively probed. At any temperature above absolute zero, some molecules are in thermally populated excited levels, and transitions from these to the ground state radiate energy at characteristic frequencies. Erom Wien s displacement law, T = 2898 //m-K, the emission maximum at 300 K is near 10 fim in the mid-ir. This radiation occurs at just the energies of molecular rovibrational transitions, so thermal emission carries much the same information as an ir absorption spectmm. Detection of the emissions of remote thermal sources is the ultimate passive and noninvasive technique, requiring not even an optical probe of the sampled volume. [Pg.315]

Mumber Density and Volume Flux. The deterrnination of number density and volume dux requires accurate information on the sample volume cross-sectional area, droplet size and velocity, as well as the number of droplets passing through the sample volume at any given instant of time. Depending on the instmmentation, the sample volume may vary with the optical components and droplet sizes. The number density represents the number of droplets contained in a specified volume of space at a given instant. It can be expressed as follows, where u is the mean droplet velocity, t the sample time, andM the representative cross-sectional area at the sampling location. [Pg.331]

Phase Doppler particle analyzers are essentially single-particle counters because they measure one particle at a time within a small sampling volume. This volume must be kept small to minimize the probabiUty of having more than one droplet in the volume at any given instant. This probabiUty increases as the concentration of droplets becomes greater, and there is more risk of measurement errors. [Pg.334]

Capillary Electrophoresis. Capillary electrophoresis (ce) is an analytical technique that can achieve rapid high resolution separation of water-soluble components present in small sample volumes. The separations are generally based on the principle of electrically driven ions in solution. Selectivity can be varied by the alteration of pH, ionic strength, electrolyte composition, or by incorporation of additives. Typical examples of additives include organic solvents, surfactants (qv), and complexation agents (see Chelating agents). [Pg.246]

Because the cells can intermpt the optical path in random orientations, individual scattering intensities are not proportional to cell volume. However, because thousands of cells of each type pass through the flow cell, the effects of orientation can be averaged To a first approximation HCT and platelet crit (PCT), the percentage of blood sample volume occupied by platelets, is proportional to the sums of the scattering intensities of the ted cells and platelets, respectively. MCV can be computed from HCT and RBC, whereas MPV can be computed from PCT and PLT. The accuracy of MCV deterrnined by this method is tied to the RBC accuracy, as is the case for the manual MCV method. Ortho Instmments Corporation s ELT-8 uses these counting and sizing methods. [Pg.403]

Immobilized Enzymes. The immobilized enzyme electrode is the most common immobilized biopolymer sensor, consisting of a thin layer of enzyme immobilized on the surface of an electrochemical sensor as shown in Figure 6. The enzyme catalyzes a reaction that converts the target substrate into a product that is detected electrochemicaHy. The advantages of immobilized enzyme electrodes include minimal pretreatment of the sample matrix, small sample volume, and the recovery of the enzyme for repeated use (49). Several reviews and books have been pubHshed on immobilized enzyme electrodes (50—52). [Pg.102]

A few special high pressure pistons with sintered diamond working faces have been made for laboratory experiments. Although the sample volume is very small, pressures of 50 GPa (500 kbar) at temperatures of up to 500°C have been reached with such an apparatus (39). [Pg.567]

Radiation-Density Gauges Gamma radiation may be used to measure the density of material inside a pipe or process vessel. The equipment is basically the same as for level measurement, except that here the pipe or vessel must be filled over the effective, irradiated sample volume. The source is mounted on one side of the pipe or vessel and the detector on the other side with appropriate safety radiation shielding surrounding the installation. Cesium 137 is used as the radi-... [Pg.764]

There will be many times when the quantity of sample is limited. While it is best to use the 92.9 cm" (0.1 ft") area leaf in order to minimize edge effects and improve accuracy, when the sample volume is limited it is much better to have several data points with a smaller leaf than onl one or two using the larger leaf. Data from leaves as small as 23.2 cm" (0.025 ft") are reasonably accurate and can be used to scale up to commercially sized units. However, it is usually prudent to employ a more conservative scale-up factor. [Pg.1696]

Factors to be considered in maldng the selection of chromatography processing steps are cost, sample volume, protein concentration and sample viscosity, degree of purity of protein product, presence of nucleic acids, pyrogens, and proteolytic enzymes. Ease with which different types of adsorbents can be washed free from adsorbed contaminants and denatured proteins must also be considered. [Pg.2064]

A one-hoiir sampling period is generally required for both methods. Sampling periods are specified by the applicable standard e.g., standards applicable to triple-superphosphate plants require sampling of one hour or more. The standard may also specify a minimum sample volume that will dictate the minimum length of the sampling period. [Pg.2203]

A simple electrochemical flow-through cell with powder carbon as cathodic material was used and optimized. The influence of the generation current, concentration of the catholyte, carrier stream, flow rate of the sample and interferences by other metals on the generation of hydrogen arsenide were studied. This system requires only a small sample volume and is very easily automatized. The electrochemical HG technique combined with AAS is a well-established method for achieving the required high sensitivity and low detection limits. [Pg.135]

The lower limit of detection is 1 p.g/dm for sodium hexadecyl sulfonate, 2.5 p.g/dm for sodium dodecyl benzenesulfonate and 10 p.g/diW for sodium dodecyl sulfonate with sample volume of O.ldiW. The method proposed is highly sensitive, simple, rapid and guarantees environmental safety of analysis. [Pg.267]

Experimental part was provided by device Model Knauer-Compact with UV-detector (b=3 mm) at 250 nm and column Spherisorb-ODS-2 (250x4,6 mm). Sample volume was 1-2 p.1 injected by Reodyne 7725. Concentration range was 0.4-0.5 mg/ml for solutions of studied substances in DMSO. The organic modificator concentration range was 75-85 % w for methanol and 40-60 % w for acetonitrile in eluent (flow rate -1 ml/min). [Pg.286]


See other pages where Sampling volume is mentioned: [Pg.67]    [Pg.112]    [Pg.50]    [Pg.495]    [Pg.1509]    [Pg.1558]    [Pg.1939]    [Pg.1958]    [Pg.24]    [Pg.51]    [Pg.651]    [Pg.702]    [Pg.51]    [Pg.51]    [Pg.54]    [Pg.209]    [Pg.401]    [Pg.315]    [Pg.316]    [Pg.334]    [Pg.396]    [Pg.400]    [Pg.1760]    [Pg.2199]    [Pg.2203]   
See also in sourсe #XX -- [ Pg.28 ]

See also in sourсe #XX -- [ Pg.28 ]

See also in sourсe #XX -- [ Pg.28 ]




SEARCH



Boyle s law The volume of a given sample

Capillary electrophoresis sample volume/concentration

Charles s law The volume of a given sample

Chromatography sample volume injected

Clinical chemistry sample volume

Column maximum sample volume

Columns sample volume

Comparison of Fido with Canines Using High-Volume Sampling Methods (REST)

Dispersion from sample volume

Dispersion sample volume influence

Due to Sample Volume

Electrochemical detection sample volume study

Fixed volume sampling

Flow injection analysis sample volume

Flow injection analysis sample volume influence

Flow injection analysis sample volume study

Glass fiber filter sampling volume

High volume filtration sampling

High-sensitivity Determination Relationship of Sample Volume to Peak Height

Influence of the injected sample volume

Injection valves volume-based sampling with

Large-volume samples

Large-volume samples analysis

Large-volume samples concentration methods

Liquid chromatography sample volume

Maximum Sample Injection Volume for a Specific Column

Maximum sample volume

Maximum sample volume, relation

Overload Due to Excess Sample Volume

Parallel Liquid Handling of Low-Volume Samples

Peak Height and Sample Volume

Pharmaceutical analysis required sample volumes

Preparative chromatography sample volume

Rapid large-volume liquid sampling

Relationship between the dispersion coefficient and injected sample volume

Safe sampling volume

Sample Plates for Low-Volume High-Throughput Screening

Sample Volume Injected

Sample Volume and Maximum Extra Column Dispersion

Sample dilution variable volume pipettes

Sample injection volume

Sample preparation volume

Sample storage, volume and pretreatment

Sample valve injectors, injection volume

Sample volume

Sample volume

Sample volume and

Sample volume maximum allowed

Sample volume measurement

Sample volume overload

Sample volume, effect on dispersion equation for

Sample volume, influence

Sample volume, preparative

Sample volume, significance

Sample volume-based

Sample, load, effect volume, maximum

Sampling large-volume

Sampling sample volume

Sampling volume flow rate

Small volume sampling and

Sorption sample volume

The Effect of Sample Volume

The Maximum Sample Volume

The sample volume problem

Ultratrace Analysis of Long-lived Radionuclides in Very Small Sample Volumes

Volume change, tissue sample

Volume of sample

Volume of the sample

Volume tissue sample

Volume, Cell maximum sample

Volume-based sampling

© 2024 chempedia.info