Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample volume-based

Based on the optimization criterion, SpinPro can select the most appropriate rotor. For example, suppose the investigator has a relatively large sample volume, all of which needs to be processed as soon as possible. The "minimize cumulative run time" criterion would be the appropriate choice. SpinPro would then initiate the following rotor selection procedure SpinPro determines the total sample volume based on inputs of the sample volume, the current concentration of the sample, and a correction for any pre-run dilutions of the sample. Next, consideration is made for whether tubes or bottles will be used. The program then evaluates rotors for the number of tube positions and the amount of sample per tube. At this point, SpinPro will have estimated for each rotor the number of runs required to process the sample. SpinPro then estimates the run time for each rotor to perform a single run. Based on these estimates, SpinPro selects the rotor that will give the shortest total run time when the run time is summed over the total number of runs. Similarly, the investigator can select any of the optimization criteria and initiate a variety of precise rotor selection procedures. [Pg.300]

V Vo fo Time propagation of the sample volume based on die change in diameter... [Pg.306]

Figure 4 Sample-volumes based on the modelled angular response of a Nal 3" x 3 detector fitted with a 45 degree 20 mm lead collimator, with base thickness of 25 mm and internal aperture of 25 mm. Angular response was modelled using a series of ISOCS calculations for a small particle positioned at varying angles from the detector axis. A non-active surface layer of 10 mm thickness was defined... Figure 4 Sample-volumes based on the modelled angular response of a Nal 3" x 3 detector fitted with a 45 degree 20 mm lead collimator, with base thickness of 25 mm and internal aperture of 25 mm. Angular response was modelled using a series of ISOCS calculations for a small particle positioned at varying angles from the detector axis. A non-active surface layer of 10 mm thickness was defined...
The method of volume rendering uses the whole sample volume for visualization. Therefor semitransparent representations of the samples inner structure are possible and the detection of small cracks or faults is much easier compared to the surfaces based techniques (Fig. 4 b). From its principle volume rendering is more time consuming compared to surface representation. [Pg.495]

Column Si. Size-exclusion chromatography columns are generally the largest column on a process scale. Separation is based strictly on diffusion rates of the molecules inside the gel particles. No proteins or other solutes are adsorbed or otherwise retained owing to adsorption, thus, significant dilution of the sample of volume can occur, particularly for small sample volumes. The volumetric capacity of this type of chromatography is determined by the concentration of the proteins for a given volume of the feed placed on the column. [Pg.50]

Biomolecule Separations. Advances in chemical separation techniques such as capillary zone electrophoresis (cze) and sedimentation field flow fractionation (sfff) allow for the isolation of nanogram quantities of amino acids and proteins, as weU as the characterization of large biomolecules (63—68) (see Biopolymers, analytical techniques). The two aforementioned techniques, as weU as chromatography and centrifugation, ate all based upon the differential migration of materials. Trends in the area of separations are toward the manipulation of smaller sample volumes, more rapid purification and analysis of materials, higher resolution of complex mixtures, milder conditions, and higher recovery (69). [Pg.396]

Capillary Electrophoresis. Capillary electrophoresis (ce) is an analytical technique that can achieve rapid high resolution separation of water-soluble components present in small sample volumes. The separations are generally based on the principle of electrically driven ions in solution. Selectivity can be varied by the alteration of pH, ionic strength, electrolyte composition, or by incorporation of additives. Typical examples of additives include organic solvents, surfactants (qv), and complexation agents (see Chelating agents). [Pg.246]

These reactors all work on very similar principles and will be discussed based on the example of the Berty reactor, of which more than 500 are in operation around the world. The Berty reactor shown in Figure 2.4.3 a has much empty volume and is laborious to open and close. Another version of the Berty reactor (made by Basic Technology, Inc.) is shown in Figure 2.4.3 b. This 2-inch model was developed for quick exploratory studies on small samples of catalysts. The maximum catalyst sample volume is 15... [Pg.50]

Consider the separation depicted in Figure 1. It is assumed that the pair of solutes represent the elution of the solute of interest and its nearest neighbor. Now, when the sample volume becomes extreme, the dispersion that results from column overload, to the first approximation, becomes equivalent to the sample volume itself as the sample volume now contributes to the elution of the solutes. Thus, from Figure 1, the peak separation in milliliters of mobile phase will be equivalent to the volume of sample plus half the sum of the base widths of the respective peaks. [Pg.421]

The water analysis is incomplete unless the number of coliform bacteria present is determined as well. A multiple-tube fermentation technique can be used to enumerate positive presumptive, confirmed, and fecal coliform tests. Results of the tests are expressed in terms of the most probable number (MPN). That is, the count is based on a statistical analysis of sets of tubes in a series of serial dilutions. MPN is related to a sample volume of 100 ml. Thus, an MPN of 10 means 10 coliforms per 100 ml of water. [Pg.462]

S mg), dimer (peak I) and monomer (peak 2), ovalbumin (S mg) (peak 3), and cytochrome c (3 mg) (peak 4) was loaded onto a Fractogel EMD BioSEC column (600 X 16 mm) with a bed height of 600 mm. PBS (pH 7.2) was used as the eluent at a flow rate of I ml/min the sample volume was O.S ml. (B) The same protein sample as in A was injected onto a column of identical dimensions packed with unmodified Fractogel HW 6S. Without the tentacle modification the base matrix displays only a poor resolution of the test mixture. [Pg.223]

Several experimental techniques may be used, such as acid/base titration, electrical conductivity measurement, temperature measurement, or measurement of optical properties such as refractive index, light absorption, and so on. In each case, it is necessary to specify the manner of tracer addition, the position and number of recording stations, the sample volume of the detection system, and the criteria used in locating the end-point. Each of these factors will influence the measured value of mixing time, and therefore care must be exercised in comparing results from different investigations. [Pg.299]

To determine the band dispersion that results from a significant, but moderate, sample volume overload the summation of variances can be used. However, when the sample volume becomes excessive, the band dispersion that results becomes equivalent to the sample volume itself. In figure 10, two solutes are depicted that are eluted from a column under conditions of no overload. If the dispersion from the excessive sample volume just allows the peaks to touch at the base, the peak separation in milliliters of mobile phase passed through the column will be equivalent to the sample volume (Vi) plus half the base width of both peaks. It is assumed in figure 10 that the efficiency of each peak is the same and in most cases this will be true. If there is some significant difference, an average value of the efficiencies of the two peaks can be taken. [Pg.119]

Figure 4. (a) Volume-based and (B) chlorophyll-based photosynthesis-irradiance data for phytoplankton sampled on December 6, 1987 from the experimental chambers (after flow had been stopped for 24 h). Samples were incubated for 4 h under the four treatment conditions described in the text. (Reproduced with permission from reference 23. Copyright 1990 Springer-Verlag, Berlin.)... [Pg.198]

Analysis of Corexit 9527. Corexit 9527 in natural waters can be analyzed. The method is based on the formation of a Z>w(ethylenediamine) copper(II) complex, extraction of the complex into methylisobutylketone, and atomic absorption spectroscopy [1564]. The method is suitable for a concentration range of 2 to 100 mg/liter, with a precision as low as 5% relative to standard deviation for samples in the middle- to high range. Only a small sample volume (10 ml) is required. The sensitivity may be substantially increased for trace analysis by increasing the sample volume. [Pg.306]

Calculation of ID using biological monitoring techniques requires the knowledge of the pharmacokinetics of the parent pesticide in laboratory animals. This will allow the use of the parent or its urine metabolite(s) to calculate the total amount of the parent that had been absorbed through the skin of the test subject. The amount of the residue in the urine should be corrected for any molecular weight differences between the parent and its urine metabolite(s) and also corrected for daily urine excretion volumes based on creatinine analysis of the urine samples. [Pg.1021]

Principles and Characteristics In metastable atom bombardment (MAB), a metastable atom beam, generated by a gun external to the ion volume, is used to bombard the sample. MAB, based on Penning ionisation, offers unique features for gas-phase ionisation. The energy available for ionisation and fragmentation is discrete (excitation energy of the atom, from 8 to 20 eV),... [Pg.367]

Nonetheless, near-IR is the most widely used IR technique. Less intense water absorptions permit to increase the sampling volume to compensate, to some extent, for the lower near-IR absorption coefficients and the inferior specificity of the absorption bands can for many applications be overcome by application of advanced chemometric methods. Miniaturised light sources, various sensor probes, in particular based on transmission or transflectance layouts, and detectors for this spectral range are available at competitive prices, as are (telecommunications) glass or quartz fibres. [Pg.123]

The gas pycnometer is based on the ideal gas law. A known quantity of gas (characterized by a determined temperature and pressure) is allowed to flow from a calibrated reference volume into a calibrated sample cell containing the solid. A second pressure reading is obtained, and the sample volume is calculated. [Pg.273]

Monien et al. [515] have compared results obtained in the determination of molybdenum in seawater by three methods based on inverse voltammetry, atomic absorption spectrometry, and X-ray fluorescence spectroscopy. Only the inverse voltammetric method can be applied without prior concentration of molybdenum in the sample, and a sample volume of only 10 ml is adequate. Results of determinations by all three methods on water samples from the Baltic Sea are reported, indicating their relative advantages with respect to reliability. [Pg.206]


See other pages where Sample volume-based is mentioned: [Pg.31]    [Pg.507]    [Pg.480]    [Pg.31]    [Pg.507]    [Pg.480]    [Pg.67]    [Pg.495]    [Pg.1939]    [Pg.378]    [Pg.265]    [Pg.18]    [Pg.30]    [Pg.33]    [Pg.104]    [Pg.220]    [Pg.64]    [Pg.425]    [Pg.463]    [Pg.582]    [Pg.322]    [Pg.643]    [Pg.132]    [Pg.272]    [Pg.141]    [Pg.129]    [Pg.146]    [Pg.207]    [Pg.803]    [Pg.96]    [Pg.381]    [Pg.178]    [Pg.236]    [Pg.334]   
See also in sourсe #XX -- [ Pg.15 ]




SEARCH



Injection valves volume-based sampling with

Sample volume

Sampling volume

Volume-based sampling

Volume-based sampling

© 2024 chempedia.info