Microcolumns use less solvent and, because the sample is diluted to a lesser extent, produce larger signals at the detector. These columns are made from fused silica capillaries with internal diameters of 44—200 pm and lengths of up to several meters. Microcolumns packed with 3-5-pm particles have been prepared with column efficiencies of up to 250,000 theoretical plates. [Pg.579]

We have developed the method for quantitative analysis of urinary albumin with CE. A capillary electrophoresis systems Nanophor 01 (Institute of Analytical Instmmentation, Russian Academy of Sciences, Saint-Petersburg) equipped with a UV-detector was used to determine analyte. Separation was achieved using 45 cmx30 p.m I.D. fused silica capillary column with UV-detection at 214 nm. [Pg.100]

The mechanism by which analytes are transported in a non-discriminate manner (i.e. via bulk flow) in an electrophoresis capillary is termed electroosmosis. Eigure 9.1 depicts the inside of a fused silica capillary and illustrates the source that supports electroosmotic flow. Adjacent to the negatively charged capillary wall are specifically adsorbed counterions, which make up the fairly immobile Stern layer. The excess ions just outside the Stern layer form the diffuse layer, which is mobile under the influence of an electric field. The substantial frictional forces between molecules in solution allow for the movement of the diffuse layer to pull the bulk... [Pg.198]

Figure 9.1 The hydrated inner surface of a fused silica capillary is where electi oosmotic flow originates. Figure 9.1 The hydrated inner surface of a fused silica capillary is where electi oosmotic flow originates.
Figure 12.18 LC-SFC analysis of mono- and di-laurates of poly (ethylene glycol) ( = 10) in a surfactant sample (a) normal phase HPLC trace (b) chromatogram obtained without prior fractionation (c) chromatogram of fraction 1 (FI) (d) chromatogram of fraction 2 (F2). LC conditions column (20 cm X 0.25 cm i.d.) packed with Shimpak diol mobile phase, w-hexane/methylene chloride/ethanol (75/25/1) flow rate, 4 p.L/min UV detection at 220 nm. SFC conditions fused-silica capillary column (15 m X 0.1 mm i.d.) with OV-17 (0.25 p.m film thickness) Pressure-programmed at a rate of 10 atm/min from 80 atm to 150 atm, and then at arate of 5 atm/min FID detection. Reprinted with permission from Ref. (23). Figure 12.18 LC-SFC analysis of mono- and di-laurates of poly (ethylene glycol) ( = 10) in a surfactant sample (a) normal phase HPLC trace (b) chromatogram obtained without prior fractionation (c) chromatogram of fraction 1 (FI) (d) chromatogram of fraction 2 (F2). LC conditions column (20 cm X 0.25 cm i.d.) packed with Shimpak diol mobile phase, w-hexane/methylene chloride/ethanol (75/25/1) flow rate, 4 p.L/min UV detection at 220 nm. SFC conditions fused-silica capillary column (15 m X 0.1 mm i.d.) with OV-17 (0.25 p.m film thickness) Pressure-programmed at a rate of 10 atm/min from 80 atm to 150 atm, and then at arate of 5 atm/min FID detection. Reprinted with permission from Ref. (23).
As a unichiral template which is used as a stationary phase itself, or which is bonded to a solid support (silica particles or fused silica capillaries). [Pg.185]

For routine separations, there are about a dozen useful phases for capillary columns. The best general-purpose columns are the dimethylpolysiloxane (DB-1 or equivalent) and the 5% phenyl, 95% dimethylpolysiloxane (DB-5 or equivalent). These relatively nonpolar columns are recommended because they provide adequate resolution and are less prone to bleed than the more polar phases. If a DB-1, DB-5, or equivalent capillary column does not give the necessary resolution, try a more polar phase such as DB-23, CP-Sil88, or Carbowax 20M, providing the maximum operating temperature of the column is high enough for the sample of interest. See Appendix 3 for fused silica capillary columns from various suppliers. [Pg.173]

FIGURE 3-23 Schematic of a carbon-fiber amperometric detector for capillary electrophoresis A, fused silica capillary B, eluent drop C, stainless steel plate RE, reference electrode WE, working electrode, AE, auxiliary electrode. (Reproduced with permission from reference 58.)... [Pg.89]

The gas chromatographic separation of some sulphoxide enantiomers was observed on quartz fused silica capillaries coated with the chiral silicon phase chirasil-val280. [Pg.287]

The use of a fused silica capillary column for the GC analysis of the neutral oil extract has provided the means for improving the resolution of components in a more inert system. The sultones are determined by temperature-programmed GC over CP-Sil-5 CB (methyl silicone fluid) in a 25 m x 0.2 mm fused silica capillary column using nonadecane as internal standard. A sample split ratio of 1 100 is recommended for a 3-pl injection. [Pg.448]

Chromatography Hewlett-Packard Ultra2 (25 m x 0.33 mm) cross-linked phenyl methyl silicone fused silica capillary column. Temperature program 55° (1 min.) -I- 30°C/min. to 180°C, and 4°C/min. to 320°C. [Pg.253]

Eichelberger JW, Kerns EH, Olynyk P, et al. 1983. Precision and accuracy in the determination of organics in water by fused silica capillary column gas chromatography/mass spectrometry and packed column gas chromatography/mass spectrometry. Anal Chem 55 1471-1479. [Pg.284]

Capillary electrophoresis (CE) has several unique advantages compared to HPLC, snch as higher efficiency dne to non-parabolic fronting, shorter analytical time, prodnction of no or much smaller amounts of organic solvents, and lower cost for capillary zone electrophoresis (CZE) and fused-silica capillary techniques. However, in CZE, the most popular separation mode for CE, the analytes are separated on the basis of differences in charge and molecular sizes, and therefore neutral compounds snch as carotenoids do not migrate and all co-elute with the electro-osmotic flow. [Pg.463]

Use of 10 pm LiChrosorb RP18 column and binary eluent of methanol and aqueous 0.1 M phosphate buffer (pH 4.0) according to suitable gradient elution program in less than 20-min run time with satisfactory precision sensitivity of spectrophotometric detection optimized, achieving for all additives considered detection limits ranging from 0.1 to 3.0 mg/1, below maximum permitted levels Simultaneous separation (20 min) of 14 synthetic colors using uncoated fused silica capillary column operated at 25 kV and elution with 18% acetonitrile and 82% 0.05 M sodium deoxycholate in borate-phosphate buffer (pH 7.8), recovery of all colors better than 82%... [Pg.538]

Glycosyl-linkages were determined by GC-EIMS of the partially methylated alditol acetates. RG-II samples (2 mg) were methylated using sodium methyl sulfmyl carbanion and methyl iodide in dimethyl sulfoxide [24] followed by reduction of the uronosyl groups with lithium triethylborodeuteride (Superdeuteride , Aldrich) [23,25]. Methylated and carboxyl-reduced samples were then submitted to acid hydrolysis, NaBIlt reduction and acetylation, partially methylated alditol acetates being analysed by EIMS on two fused-silica capillary columns (DB-1 and DB-225) [20]. [Pg.70]

P 32] Pyrene (20 mM), 1,4-dicyanobenzene (40 mM) and sodium cyanide (1 M) were reacted in propylene carbonate and water. A 100 pi solution of pyrene (20 mM), 1.4-dicyanobenzene (40 mM) in propylene carbonate and a 100 pi solution of sodium cyanide (1 M) in water were fed by programmable dual-syringe pumps via fused-silica capillary tubes into a micro-channel chip [29]. Both solutions were fed with equal flow velocity. A 300 W high-pressure mercury lamp was used as light source. After passing an optical filter made of a CUSO4 solution, the whole chip was irradiated after formation of a stable oil/water interface inside. The oil phase was collected at the exit. [Pg.477]

Early work relied on the use of packed columns, but all modern GC analyses are accomplished using capillary columns with their higher theoretical plate counts and resolution and improved sensitivity. Although a variety of analytical columns have been employed for the GC of triazine compounds, the columns most often used are fused-silica capillary columns coated with 5% phenyl-95% methylpolysiloxane. These nonpolar columns in conjunction with the appropriate temperature and pressure programming and pressure pulse spiking techniques provide excellent separation and sensitivity for the triazine compounds. Typically, columns of 30 m x 0.25-mm i.d. and 0.25-qm film thickness are used of which numerous versions are commercially available (e.g., DB-5, HP-5, SP-5, CP-Sil 8 CB, etc.). Of course, the column selected must be considered in conjunction with the overall design and goals of the particular study. [Pg.440]

Oxyfluorfen column, fused-silica capillary column coated with cross-linked methyl silicone (25 m x 0.3-mm i.d., 0.52- am film thickness) temperature, column 200 °C (1 min), 10°Cmin to 250 °C (5 min), inlet and detector 250 and 300 °C, respectively gas flow rates, N2 carrier gas 30mLmin , N2 makeup gas 30mLmin H2 3.5mLmin" air llOmLmin injection volume, 2 p.L. ... [Pg.457]

The residue levels of 46 pesticides, including oxyfluorfen in soil, were determined using GC/ITDMS as described in S ection 3.2.1. The conditions for GC/ITDMS were as follows column, fused-silica capillary (30 m x 0.25-mm-i.d.) with a0.25- am bonded phase ofDB-5 column temperature, 50 °C (1 min), 30 °Cmin to 130 °C, 5 °C min to 270 °C inlet and transfer temperature, 270 and 220 °C, respectively He gas with column head pressure, 12psi injection method, splitless mode. The retention time and quantitation ion of oxyfluorfen were 23.9 min and mjz 252, respectively. ... [Pg.460]

Chlornitrofen in water samples was determined by GC/MS using an SPB-1 fused-silica capillary (15 mx 0.53-mm i.d., 0.5-pm film thickness), with the ionization... [Pg.463]

Fused-silica capillary column, HP-SMS, 30 m x 0.25-mm i.d., 0.25- am film thickness, (5% phenyl)-methylpolysiloxane... [Pg.543]

Lord and Pawliszyn" developed a related technique called in-tube SPME in which analytes partition into a polymer coated on the inside of a fused-silica capillary. In automated SPME/HPLC the sample is injected directly into the SPME tube and the analyte is selectively eluted with either the mobile phase or a desorption solution of choice. A mixture of six phenylurea pesticides and eight carbamate pesticides was analyzed using this technique. Lee etal. utilized a novel technique of diazomethane gas-phase methylation post-SPE for the determination of acidic herbicides in water, and Nilsson et al. used SPME post-derivatization to extract benzyl ester herbicides. The successful analysis of volatile analytes indicates a potential for the analysis of fumigant pesticides such as formaldehyde, methyl bromide and phosphine. [Pg.732]

Capillary electrophoresis (CE) or capillary zone electrophoresis (CZE) is the technique most often employed in pesticide residue analysis. In its most basic form, free zone electrophoresis, a fused-silica capillary is filled with electrolyte (running buffer or background electrolyte). A potential is applied across the capillary and the cations... [Pg.743]

Gas chromatograph for fused-silica capillary or packed columns, equipped with a flame photometric detector (with sulfur filter), Hewlett-Packard, Carlo Erba, or equivalent... [Pg.1096]

Under these chromatographic conditions, the CS2 retention time is about 3 min on a fused-silica capillary column and about 2 min on a Teflon Chromosil 330 packed column. [Pg.1097]

The silica gel column eluates (Module Cl or C2) are injected, if necessary with the addition of an internal standard, into a gas chromatograph followed by ECD or NPD. The determinations can be performed with different gas chromatographs and fused-silica capillary columns. [Pg.1117]


See other pages where Capillaries fused-silica is mentioned: [Pg.546]    [Pg.4]    [Pg.35]    [Pg.220]    [Pg.221]    [Pg.306]    [Pg.309]    [Pg.157]    [Pg.178]    [Pg.200]    [Pg.162]    [Pg.258]    [Pg.81]    [Pg.81]    [Pg.82]    [Pg.81]    [Pg.965]    [Pg.422]    [Pg.545]    [Pg.545]    [Pg.727]    [Pg.827]   
See also in sourсe #XX -- [ Pg.268 ]

See also in sourсe #XX -- [ Pg.9 , Pg.45 , Pg.399 ]




SEARCH



Aluminum-clad fused-silica capillary

Aluminum-clad fused-silica capillary columns

Capacity fused-silica capillary columns

Capillary column technology fused-silica

Capillary electromigration techniques fused-silica capillaries

Coating fused-silica capillary columns

Deactivation fused-silica capillary columns

Extrusion of a Fused-Silica Capillary Column

Functionalized Fused Silica Capillaries

Fused silica

Fused silica capillaries modifications

Fused silica capillary columns

Fused-silica capillary columns characteristics

Fused-silica capillary columns polyimide-clad

Fused-silica capillary columns popularity

Fused-silica capillary columns preparation

Fused-silica capillary columns stainless steel

Fused-silica capillary columns synthetic

Fused-silica capillary columns tubing

Fused-silica capillary tubes

Fused-silica capillary tubes metal coated

Fused-silica capillary tubes polyimide coated

Fused-silica transfer capillary

Fused-silica-lined stainless-steel capillary

Fused-silica-lined stainless-steel capillary columns

Inert-fused silica capillary columns

Packed fused silica capillary columns

Polyimide-coated fused silica capillarie

Preparation of Fused-Silica Capillary Columns

Retention time fused-silica capillary columns

Selectivity fused-silica capillary columns

Silanol groups fused silica capillary tubes

Stationary phase fused-silica capillary columns

Temperature fused-silica capillary columns

Untreated Fused Silica Capillaries

© 2024 chempedia.info