Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration with Standards

In order to use Beer s law to determine the concentration of analyte in an unknown, it is necessary to establish the relationship between absorbance at a given wavelength and the concentration of the analyte. Solutions containing known concentrations of analyte are called standard solutions or, more simply, standards. For some types of analyses, the standards may be in the form of solids [Pg.76]

Standards must be prepared accurately from high-purity materials so that the concentration of analyte is known as accurately as possible. A series of standards covering an appropriate concentration range is prepared. The standards should include one solution with no added analyte the concentration of analyte in this standard is zero. This solution is called the reagent blank and accounts for absorbance due to impurities in the solvent and other reagents used to prepare the samples. It also accounts for the instrumental baseline. The absorbance of the reagent blank and each standard is [Pg.76]

Concentration of n-Hexadecane (mg/100 mL Solution) Absorbance at 3.41 im Corrected Absorbance [Pg.77]

Performing a linear regression on the data in Table 2.4 provides us with the exact Beer s law relationship for this method A = 0.0250t - 0.001, where x is the concentration of n-hexadecane in mg/100 mL. From the equation for the calibration curve, the concentration can be determined for any measured absorbance. For example, an unknown sample of contaminated soil is prepared according to US EPA Method 8440, and the absorbance of the sample solution is measured. The measured absorbance is 0.302, so the corrected absorbance would be 0.302 - 0.002 = 0.300. From our calibration curve, we can see visually that this corresponds to a concentration of 12.0 mg n-hexadecane/100 mL. The exact concentration can be calculated from the linear regression equation and is found to be 11.96 mg/100 mL or 12.0 mg/100 mL rounded to three significant figures. [Pg.78]

Suppose that in addition to the standards listed in Table 2.4, we also prepared n-hexadecane standards containing 60.0 mg n-hexadecane/100 mL solution and 100.0 mg n-hexadecane/100 mL solution. If we measure the absorbances for these standards, we obtain (he data shown in Table 2.5. The additional points, shown as open circles, are plotted along with the original standards in Hgure 2.12. Clearly, we now see a deviation from Beer s law at these high concentrations. The points no longer fit a straight fine the measured absorbances are lower than they should be if Beer s law was followed [Pg.78]


In Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES), a gaseous, solid (as fine particles), or liquid (as an aerosol) sample is directed into the center of a gaseous plasma. The sample is vaporized, atomized, and partially ionized in the plasma. Atoms and ions are excited and emit light at characteristic wavelengths in the ultraviolet or visible region of the spectrum. The emission line intensities are proportional to the concentration of each element in the sample. A grating spectrometer is used for either simultaneous or sequential multielement analysis. The concentration of each element is determined from measured intensities via calibration with standards. [Pg.48]

Previous studies of the interaction of energetic particles with suri ces have made it clear that under nearly all conditions the majority of atoms or molecules removed from a surface are neutral, rather than charged. This means that the chained component can have large relative fluctuations (orders of magnitude) depending on the local chemical matrix. Calibration with standards for surfaces is difficult and for interfaces is nearly impossible. Therefore, for quantification ease, the majority neutral component of the departing flux must be sampled, and this requires some type of ionization above the sample, often referred to as post-ionization. SALI uses effi-... [Pg.561]

Resistivity. Control of the resistivity of the mud and mud filtrate while drilling may be desirable to permit better evaluation of formation characteristics from electric logs. The determination of resistivity is essentially the measurement of the resistance to electrical current flow through a known sample configuration. Measured resistance is converted to resistivity by use of a cell constant. The cell constant is fixed by the configuration of the sample in the cell and is determined by calibration with standard solutions of known resistivity. The resistivity is expressed in ohm-meters. [Pg.656]

In sugar refinery control operations, pH electrodes should not be, as unfortunately they sometimes are, calibrated with standard buffer solutions, then placed on stream in sugar liquors, and assumed to read equivalent pH. Implicit in this operation is the equivalence of pH electrode response in dilute aqueous buffers at 24°C and in high Brix sugar solutions at elevated temperatures. Such equivalence does not exist. [Pg.464]

Chromatorapf c methods For gel filtration of polysaccharide fraction PI, a Sephacryl S-300 chromatographyc column (1,1 X 46,7 cm) was calibrated with standard dextrans (molecular mass range 266, 72, 40, and 17 KDa Sigma Chemicals), and the void volume determined with blue dextran. Polysaccharide sample (0.5 mL 2 mg/mL) was applied and eluted with 50 mM NaOH, fractions 1 mL being collected and carbohydrate absorbance (phenol-H2S04) being monitored. [Pg.551]

Gel Filtration. The lyophilized protein was redissolved in 50 mM phosphate buffer, pH 7.4 0.15 m NaCl 0.013 % sodium azide and loaded on a Superdex 75HR1030 column equilibrated with the same buffer. Elution was downward flow (0.15 ml/min) and 0.25 ml fi actions were collected. Fractions with pectin lyase activity were combined, dialyzed against distilled water and used in the next step. To estimate the molecular mass of PNL, the column was calibrated with standard proteins (Sigma MW-GF-70 Albumin, 66,000 Da Carbonic Anhidrase, 29,00 Cytochrome, 12,400 and Aprotinin, 6,500). The proteins were eluted in the conditions described above and their volumes (F ) were calculated fi om the peak maximum of the absorbance at 280 nm. The partition coefficient was obtained fi om the relationship where F, represents the bed volmne of column and F the void volume (which was calculated using blue dextran. Sigma). The molecular mass was determined using a standard curve of vs the logarithm of the molecular masses of the standards [28, 29]... [Pg.750]

Calibration range Not specified usually > 4 Not specified but at least one calibration in each sequence Linearity not required any vahd calibration is accepted Calibration with standards in matrix strongly recommended... [Pg.126]

In summary, official German analytical methods for pesticide residues are always validated in several laboratories. These inter-laboratory studies avoid the acceptance of methods which cannot readily be reproduced in further laboratories and they do improve the ruggedness of analytical procedures applied. The recently introduced calibration with standards in matrix improves the trueness of the reported recovery data. Other aspects of validation (sample processing, analyte stability, extraction efficiency) are not considered. [Pg.128]

Partial) dialysis in flow analysis. The sample solution flows along one side of the membrane, while the analyser solution passing (often in counter-current) on the other side takes up the diffused components from the sample. A dynamic equilibrium is reached (under steady-state conditions) in the leaving analyser solution, which is then analysed and from the result of which the analyte content can be derived via calibration with standard solutions treated in exactly the same way. This is a common procedure, e.g., in Technicon AutoAnalyzers, and has also been applied in haemoanalysis by Ammann et al.154 as described above. [Pg.370]

For PyMS to be used for (1) routine identification of microorganisms and (2) in combination with ANNs for quantitative microbiological applications, new spectra must be comparable with those previously collected and held in a data base.127 Recent work within our laboratory has demonstrated that this problem may be overcome by the use of ANNs to correct for instrumental drift. By calibrating with standards common to both data sets, ANN models created using previously collected data gave accurate estimates of determi-nand concentrations, or bacterial identities, from newly acquired spectra.127 In this approach calibration samples were included in each of the two runs, and ANNs were set up in which the inputs were the 150 new calibration masses while the outputs were the 150 old calibration masses. These associative nets could then by used to transform data acquired on that one day to data acquired at an earlier data. For the first time PyMS was used to acquire spectra that were comparable with those previously collected and held in a database. In a further study this neural network transformation procedure was extended to allow comparison between spectra, previously collected on one machine, with spectra later collected on a different machine 129 thus calibration transfer by ANNs was affected. Wilkes and colleagues130 have also used this strategy to compensate for differences in culture conditions to construct robust microbial mass spectral databases. [Pg.333]

Calibration with standards allows accurate determination of the molecular mass of the product itself, as well as any impurities. Batch-to-batch variation can also be assessed by comparison of chromatograms from different product runs. [Pg.184]

Automatic instruments are usually reflection-type instruments that measure the deflection of a beam of light as it passes from one medium, into the sample and then is reflected back to a detector. The angle at which the light beam exits the medium is related to the refractive index of the sample. Automated instruments are calibrated with standard substances of precisely known refractive index prior to use. [Pg.66]

Number-average molar masses were determined using a vapor pressure osmometer (VPO) (Hitachi 117 Molecular Weight Apparatus) at 54.8 0.1°C in toluene (Fisher Scientific, certified A.C.S.) which was distilled from freshly crushed CaH2. The VPO apparatus was calibrated with pentaerythritol tetrastearate (Pressure Chemical). Gel permeation chromatographic (GPC) analyses were performed in tetrahydrofuran by HPLC (Perkin-Elmer 601 HPLC) using six y-Styragel columns (106, 105, 10l, 103, 500, and 100 A) after calibration with standard polystyrene samples. [Pg.140]

However, salinity values are easily obtained with a salinometer (which measures electrical conductivity and is appropriately calibrated with standard solutions and adjusted to account for T effects). The salinity of seawater increases if the loss of H2O (evaporation, formation of ice) exceeds the atmospheric input (rain plus rivers), and diminishes near deltas and lagoons. Salinity and temperature concur antithetically to define the density of seawater. The surface temperature of the sea reflects primarily the latitude and season of sampling. The vertical thermal profile defines three zones surface (10-100 m), where T is practically constant thermoclinal (100-1000 m), where T diminishes regularly with depth and abyssal... [Pg.601]

After calibration with standard of betulinic acid, the monthly extracts from leaves of Eugenia florida were analyzed. Those extracts were analyzed in triplicate and the average areas... [Pg.188]

Spectroscopic techniques require calibration with standards of known analyte concentration. Atomic spectrometry is sufficiently specific for a simple solution of a salt of the analyte in dilute acid to be used, although it is a wise precaution to buffer the standards with any salt which occurs in large concentration in the sample solution, e.g. 500 pg ml-i or above. Calibration curves can be obtained by plotting absorbance (for AAS), emission signal (for AES), fluorescence signal (for AFS) or ion count rate (for MS) as the dependent variable against concentration as the independent variable. Often the calibration curve will bend towards the concentration axis at higher concentrations, as shown in Fig. [Pg.6]

Other Rotational Viscometers. Some rotational viscometers employ a disk as the inner member or bob, eg, the Brookfield and Mooney viscometers others use paddles (a geometry of the Stormer). These nonstandard geometries are difficult to analyze, particularly for an infinite bath, as is usually employed with the Brookfield and the Stormer. The Brookfield disk has been analyzed for Newtonian and non-Newtonian fluids and shear rate corrections have been developed (22). Other nonstandard geometries are best handled by determining instrument constants by calibration with standard fluids. [Pg.187]

The speed at which a sphere rolls down a cylindrical tube filled with a fluid or down an angled plate covered with a film of the fluid also gives a measure of viscosity. For the cylindrical tube geometry, equation 35, a generalized form of the Stokes equation is used for any given instrument, where v is the translational velocity of the rolling sphere and k is the instrument constant determined by calibration with standard fluids. [Pg.190]

In typical experiments, 4-/uL aliquots of concentrates were injected. Recoveries were determined from the areas of peaks, and each determination was calibrated with standards prepared from the same stock solutions used to... [Pg.495]

For calibrating with standards, use the 3 mg/ml standard protein solution to prepare dilutions of 20, 50, 100, 250, 500, 1000, 2000, and 3000 pg/ml in the same solvent as used to prepare the sample protein. Prepare a blank consisting of solvent alone. [Pg.115]

Standards. There is no standard or calibration with standards, so these issues are moot. [Pg.1237]

For most ISS/GSSs used in clinical analysis, these conditions are satisfied. In such cases, the galvanic cell can be calibrated easily with a two-point calibration with standards of different but known activities and traceable concentrations of the components. [Pg.13]

The relationship between the electromotive force and pH is defined as a straight line. To establish the slope and the intercept of a meter system at a given temperature, the meter must be calibrated with standard solutions prior to use. A typical field meter kit includes three calibration standards, which are buffer solutions with known pH values (usually pH 4.01, 7.00, and 10.01 at 25°C). By immersing the probe into the buffer solution with pH 7, we establish the intercept (also called offset or zero) of the probe. If the reading is different from 7.0 at this point, we must adjust it with a control knob labelled Offset or Zero. The buffer solutions with the pH values of 4.01 and 10.01 allow us to verify and adjust the slope (span) of the calibration line. [Pg.170]

Calibration with standards that are traceable to NIST-certified reference materials... [Pg.252]

The ion chromatograph is calibrated with standard solutions containing known concentrations of the target ions. Calibration curves are constructed from which the concentration of each ion in the unknown sample is determined. It is strongly recommended to match the calibration solutions with the sample matrix. Five calibration solutions and one zero standard (blank, normally water) are needed to generate a suitable calibration curve. The range to be used will depend on the concentration range for the different samples. [Pg.404]


See other pages where Calibration with Standards is mentioned: [Pg.247]    [Pg.361]    [Pg.5]    [Pg.742]    [Pg.367]    [Pg.143]    [Pg.185]    [Pg.445]    [Pg.1253]    [Pg.303]    [Pg.526]    [Pg.415]    [Pg.1253]    [Pg.181]    [Pg.190]    [Pg.384]    [Pg.477]    [Pg.80]    [Pg.303]    [Pg.526]    [Pg.59]    [Pg.248]    [Pg.154]   


SEARCH



Standardization calibration standards

© 2024 chempedia.info