Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Solutions calibration

Various methods can be used to analy2e succinic acid and succinic anhydride, depending on the characteristics of the material. Methods generally used to control specifications of pure products include acidimetric titration for total acidity or purity comparison with Pt—Co standard calibrated solutions for color oxidation with potassium permanganate for unsaturated compounds subtracting from the total acidity the anhydride content measured by titration with morpholine for content of free acid in the anhydride atomic absorption or plasma spectroscopy for metals titration with AgNO or BaCl2 for chlorides and sulfates, respectively and comparison of the color of the sulfide solution of the metals with that of a solution with a known Pb content for heavy metals. [Pg.538]

The catalytic efficiency increases, under comparable conditions (pH, concentration of catalyst, etc.) in the sequence Cl < Br - S(CH3)2 < SCN < SC(NH2)2 < I . Titration with a calibrated solution of NaN02 (usually 0.05 to 0.10 m) is used for the analytical determination of aromatic amines, dissolved in aqueous H2S04 or HC1. Here nucleophilic catalysis is achieved by adding KBr. This allows a titration to be completed much faster than without that addition. [Pg.56]

In the context of the stability of the nitrosoamine intermediate in the diazotization of heteroaromatic amines relative to that in the case of aromatic amines, the reversibility of diazotization has to be considered. To the best of our knowledge the reverse reaction of a diazotization of an aromatic amine has never been observed in acidic solutions. This fact is the basis of the well-known method for the quantitative analysis of aromatic amines by titration with a calibrated solution of sodium nitrite (see Sec. 3.3). With heteroaromatic amines, however, it has been reported several times that, when using amine and sodium nitrite in the stoichiometric ratio 1 1, after completion of the reaction nitrous acid can still be detected with Kl-starch paper,... [Pg.62]

The technician needs 11 minutes to prepare every sample solution and 13 minutes to prepare a calibration solution. A vial is filled in 1 minute. [Pg.188]

Experimental A photometric method was found in the literature which seemed to suit the particular circumstances. Two cyanide stock solutions were prepared, and an electromechanical dispenser was used to precisely prepare solutions of 20, 40,. .., 240 respectively 10, 30, 50,. .., 250 fig CN /100 ml. 10 ml of each calibration solution were added to 90 ml of the color-forming reagent solution and the absorbance was measured using 1-cm cuvettes. (See Table 4.17 (left and middle panels) and data file CYANIDE.dat.)... [Pg.221]

Table 4.17. Absorption Measurements on Cyanide Calibration Solutions... Table 4.17. Absorption Measurements on Cyanide Calibration Solutions...
Figure 4.36. Cross validation between two HPLCs A stock solution containing two compounds in a fixed ratio was diluted to three different concentrations (1 10 20) and injected using both the 10 and the 20 /xl loop on both instruments. The steps observed at Amount = 100 (gray ellipses) can be explained with effective loop volumes of 9.3 and 20 pi (model 1) and 14.3 and 20 pi (model 2) instead of nominally 10 and 20 pi. This is irrelevant as both a sample and the calibration solution will be run using the same equipment configuration. The curved portion of the model 2 calibration function was fitted using Y = A /x this demonstrates the nonlinearity of the response at these high concentrations. The angle between the full and the dotted line indicates the bias that would obtain if a one-point calibration scheme were used. Figure 4.36. Cross validation between two HPLCs A stock solution containing two compounds in a fixed ratio was diluted to three different concentrations (1 10 20) and injected using both the 10 and the 20 /xl loop on both instruments. The steps observed at Amount = 100 (gray ellipses) can be explained with effective loop volumes of 9.3 and 20 pi (model 1) and 14.3 and 20 pi (model 2) instead of nominally 10 and 20 pi. This is irrelevant as both a sample and the calibration solution will be run using the same equipment configuration. The curved portion of the model 2 calibration function was fitted using Y = A /x this demonstrates the nonlinearity of the response at these high concentrations. The angle between the full and the dotted line indicates the bias that would obtain if a one-point calibration scheme were used.
Repeatability Fiveteen placebo tablets are dissolved in water and spiked with the appropriate amount of a stock solution LO and/or HI so as to obtain the same concentrations of A and B as for the calibration solutions. Aliquots of each of these solutions are injected three times, for a total of 45 results of A and 45 for B. [Pg.288]

In Fig. 4.39, results for spiked placebo and for the verum tablets are given for compound A (bold lines) and B all horizontal bars should be at 100%, and the vertical lines should be centered at the same height. The gray trendlines, particularly for the LO- and Hl-range A-values indicate a systematic difference in response between tbe calibration solutions and the spiked placebo tablets (extraction efficiency, interference, etc.). For same ranges, the verum-tablets assays either underestimate the content of A by 4—5%, or A is underdosed. For compound A the repeatability figures are as follows (%-of-nom-inal, see file Fig4 39.dat), see Table 4.36. [Pg.288]

The HPLC method for which data are given had previously been shown to be linear over a wide range of concentrations what was of interest here was whether acceptable linearity and accuracy would be obtained over a relatively narrow concentration range around the nominal concentration in the product the specification limits were 90-110% of nominal. Three concentrations were chosen and three repeat determinations were carried out at each. Two different samples were prepared at each concentration, namely an aqueous calibration solution and a spiked placebo. All samples were worked up according to the method and appropriate aliquots were injected. The area counts are given in the second, respectively the fifth column of Table 4.42. [Pg.313]

All three monomers were soluble In the chromatographic mobile phase and standard analytical techniques were used for calibration. Solutions containing known quantities of monomer were chromatographed to establish a peak area concentration relationship for the appropriate detector. The homopolymer of methylacrylate was also soluble In the mobile phase. Thus, both UV and refractometer detectors were calibrated for polymerized methylmethacrylate by chromatographing solutions of PM ... [Pg.79]

The pure substances are provided with a statement of the purity and serve as authentic reference compounds to prepare calibration solutions for measurement of trace levels of the compound in natural matrix samples. [Pg.84]

The calibration solutions, which typically contain a number of analytes at known concentrations, are useful for validating the chromatographic separation step (e.g. retention times and analyte detector response). [Pg.84]

A further distinction is made between pure solutions and matrix-based RMs. The former are available for many organic and inorganic analytes, with certified concentrations, but their role in quality control and assessment is limited. They may be used for the preparation of calibration solutions for a particular measurement but more usually these materials represent the base for traceability, through secondary... [Pg.111]

Residue study protocols typically either include quality specifications for analytical procedures or refer to a written analytical method that includes such specifications. The protocol for an LSMBS should also include analytical quality specifications, either directly or by reference to a method. Analytical specifications usually include minimum and maximum recovery of analyte from fortified control samples, minimum number of such fortifications per set of samples, minimum linearity in calibration, minimum stability of response to injection of calibration solutions, and limits of quantitation and of detection. [Pg.239]

Solution H2S can also be lowered by spontaneous oxidation catalyzed by a number of compounds in the presence of 02. Accordingly, calibration solutions should contain DTPA and should be sparged with nitrogen or argon to remove 02. [Pg.248]

Distribute the concentration values (x) of the calibrating solutions evenly over the range of interest. One or two values at the extreme ends can cause leverage of the line [15]. [Pg.89]

Unfortunately, none of the commonly used molecular probes is adequate to evaluate column-to-column variabilities [88]. The absolute prediction of retention of any compound involves the use of a rather complex equation [89,90] that necessitates the knowledge of various parameters for both the solute and the solvent [91]. The relative prediction of retention is based on the existence of a calibration line describing the linearity between log and interaction index. This second approach, although less general than the first, is simpler to use in practice, and it often gives more accurate results than the first. With a proper choice of calibration solutes, it is possible to take into account subtle mobile phase effects that cannot be included in the theoretical treatment. [Pg.541]

Potentiometric measurements with ISEs can be approached by direct potentiometry, standard addition and titrations. The determination of an ionic species by direct potentiometry is rapid and simple since it only requires pretreatment and electrode calibration. Here, the ion-selective and reference electrodes are placed in the sample solution and the change in the cell potential is plotted against the activity of the target ion. This method requires that the matrix of the calibration solutions and sample solutions be well matched so that the only changing parameter allowed is the activity of the target ion. [Pg.643]

If it is not possible to include a particular element in the calibration solutions, it is possible to perform a semiquantitative analysis. This uses the response of those elements which are in the calibration solution, but predicts the sensitivity (defined as cps/concentration) for the missing element(s) by interpolating between the sensitivities of known elements. By plotting sensitivity against mass for all the elements present in the calibration solutions (Fig. 9.7) and fitting a curve through the points, it is possible to predict the sensitivity of the instrument for any particular mass number, and hence use this sensitivity to convert cps to concentration at that mass number. As can be seen from the figure, however, this is a very crude approximation, and any data produced in this way must be treated with some caution. [Pg.206]

For solution-based analyses, it is normal to make up a set of synthetic standards from commercial calibration solutions (normally supplied as 1000 ppm stock solutions, e.g., from Aldrich, BDH, Fisons, or ROMIL). These are available to different degrees of purity, and it is necessary to use the level of purity commensurate with the sensitivity of the analytical technique to be used it is, however, better not to use the highest purity in all circumstances, since these are very expensive. Ideally each element to be determined in the sample should be calibrated against a standard solution containing that element, although interpolation is sometimes possible between adjacent elements in the periodic table, if some elements are missing. For most techniques, it is better to mix up a single standard solution... [Pg.306]


See other pages where Solutions calibration is mentioned: [Pg.507]    [Pg.287]    [Pg.26]    [Pg.189]    [Pg.189]    [Pg.283]    [Pg.288]    [Pg.304]    [Pg.379]    [Pg.84]    [Pg.85]    [Pg.85]    [Pg.598]    [Pg.629]    [Pg.1239]    [Pg.1286]    [Pg.290]    [Pg.306]    [Pg.226]    [Pg.122]    [Pg.257]    [Pg.193]    [Pg.1]    [Pg.541]    [Pg.87]    [Pg.205]    [Pg.206]   
See also in sourсe #XX -- [ Pg.127 , Pg.129 , Pg.133 , Pg.158 , Pg.238 ]

See also in sourсe #XX -- [ Pg.2 , Pg.317 ]

See also in sourсe #XX -- [ Pg.39 , Pg.40 , Pg.41 , Pg.43 , Pg.47 ]




SEARCH



Calibration Solution CRMs

Calibration problems proposed solutions

Calibration solutions for

Calibration standard solutions

Humidity, relative solutions for calibration

Matrix matched calibration solutions

Matrix-Free Calibration Solutions

NIST SRMs calibration solutions

Preparing Accurate Activity Calibration Solutions

Quantification of Analytical Data via Calibration Curves in Mass Spectrometry Using Certified Reference Materials or Defined Standard Solutions

Solution Based Calibration in LA-ICP-MS

Standard KC1 Solutions for Calibrating

Standard KC1 Solutions for Calibrating Conductivity Cells

Standard Salt Solutions for Humidity Calibration

© 2024 chempedia.info