Ion genera lion can be achieved in a number of ways electron impact (Eh ionization, chemical ionization (CI). fas I atom bombardment (FAB), matrix assisted taser desorption ionization (MAI.DI), eleclrospray ionization (ESI) and atmospheric pressure chemical ionization (APC I are the most common methods,... [Pg.149]

These direct ion sources exist under two types liquid-phase ion sources and solid-state ion sources. In liquid-phase ion sources the analyte is in solution. This solution is introduced, by nebulization, as droplets into the source where ions are produced at atmospheric pressure and focused into the mass spectrometer through some vacuum pumping stages. Electrospray, atmospheric pressure chemical ionization and atmospheric pressure photoionization sources correspond to this type. In solid-state ion sources, the analyte is in an involatile deposit. It is obtained by various preparation methods which frequently involve the introduction of a matrix that can be either a solid or a viscous fluid. This deposit is then irradiated by energetic particles or photons that desorb ions near the surface of the deposit. These ions can be extracted by an electric field and focused towards the analyser. Matrix-assisted laser desorption, secondary ion mass spectrometry, plasma desorption and field desorption sources all use this strategy to produce ions. Fast atom bombardment uses an involatile liquid matrix. [Pg.15]

Earlier methods of ionization applied to carotenoids, including electron impact (El), chemical ionization (Cl), a particle beam interface with El or Cl, and continuous-flow fast atom bombardment (CF-FAB), have been comprehensively reviewed elsewhere (van Breemen, 1996, 1997 Pajkovic and van Breemen, 2005). These techniques have generally been replaced by softer ionization techniques like electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI), and more recently atmospheric pressure photoionization (APPI). It should be noted that ESI, APCI, and APPI can be used as ionization methods with a direct infusion of an analyte in solution (i.e. not interfaced with an HPLC system), or as the interface between the HPEC and the MS. In contrast, matrix-assisted laser desorption ionization (MALDI) cannot be used directly with HPEC. [Pg.127]

One of the most significant developments in mass spectrometry in the recent years is the introduction of a new class of ionization methods where samples in either solid or liquid state can be directly ionized in their native environment under ambient conditions (rather than inside a mass spectrometer) without any sample preparation. This new class of ionization methods is often referred to as ambient ionization methods [1,2], Because these methods generally ionize analytes on the surface or near the surface of the samples at atmospheric pressure, they have also been called atmospheric pressure surface sampling/ionization methods or direct/open air ionization methods [3], Since the first reports on ambient ionization with desorption electrospray ionization (DESI) [4] and direct analysis in real time (DART) [5], numerous reports have been published on the applications of these new ionization methods as well as the introduction of many related ambient ionization methods such as desorption atmospheric pressure chemical ionization (DAPCI) [6], atmospheric solid analysis probe (ASAP) [7], and electrospray-assisted laser desorption/ionization (ELDI) [8], Recently, two reviews of the various established and emerging ambient ionization methods have been published [2,3],... [Pg.377]

Nearly all known ionization methods of mass spectrometry (including electron impact, laser desorption and fast atom bombardment) were already successfully applied to lipids. However, many ionization techniques are not very suitable for the analysis of complex PL mixtures as they provide considerable amounts of fragment ions. Therefore, only three soft-ionization methods play nowadays a major role in lipid analysis. Beside atmospheric pressure chemical ionization (APCI) (Byrdwell 2001), electrospray ionization (ESI) (Pulfer and Murphy... [Pg.543]

Fayad PB, Prevost M, Sauve S, Laser diode thermal desorption/atmospheric pressure chemical ionization tandem mass spectrometry analysis of selected steroid hormones in wastewater Method optimization and application. Anal. Chem. 2010 82(2) 639-645. [Pg.225]

Thome FA, Heavner DL, Ingebrethsen BJ, Eudy LW, Green CR (1986) Environmental tobacco smoke monitoring with an atmospheric pressure chemical ionization mass spectrometer/mass spectrometer coupled to a test chamber. Proc 79th Annual Meet Air Pollution Control Assoc. Air Pollution Control Assoc, Pittsburgh, paper 86-37.6 Thompson CV, Jenkins RA, Higgins CE (1989) A thermal desorption method for the determination of nicotine in indoor environments. Environ Sci Technol 23 429-435 Thomson BA, Davidson WR, Lovett AM (1980) Applications of a versatile technique for trace analysis atmospheric pressure negative chemical ionization. Environ Health Perspect 36 77-84... [Pg.190]

Atmospheric pressure ionization (API). The need to analyze polar componnds and the necessity to interface LC with MS led to the development of techniqnes where the ionization occurs at atmospheric pressure outside the vacuum chamber, and the resulting ions are transferred directly into the mass analyzer. Electrospray ionization (ESI) is the most successful of the API methods because of the range of molecular masses to which it can be applied, from small molecules to proteins. Other API methods include atmospheric pressure chemical ionization (APCI) and atmospheric pressure photo-ionization (APPI), and also the recently developed surface ionization methods such as desorption electrospray ionization (DESI) and direct analysis in real time (DART) (see below and Sections 2.2.2 and 2.2.3). [Pg.16]

Ions can be formed within the vacuum chamber of the mass spectrometer or outside the instrument at atmospheric pressure. Examples of in vacuo ionization are electron ionization (El), chemical ionization (Cl), and matrix-assisted laser desorption/ion-ization (MALDI). Ionization techniques carried out outside the vacuum system are collectively termed atmospheric pressure ionization (API). The most important API methods are electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI). Among other API techniques are atmospheric pressure photo-ionization... [Pg.45]

Several other ionization methods have been developed based on DESI, including desorption atmospheric pressure chemical ionization (DAPCI), desorption atmospheric pressure photo-ionization (DAPPI), laser ablation electrospray ionization (LAESI), and extractive electrospray ionization (EESI). Each technique uses variations of the solvent, how the charged beam is formed, and how the beam is nsed to facilitate the prodnction of analyte ions. Because these are surface methods (except EESI), they are incompatible with LC. [Pg.65]

Methods Ambient ionization methods, of which there are now over 20, e.g., desorption electrospray ionization (DESI), desorption atmospheric pressure chemical ionization (DAPC), desorption atmospheric pressme photo-ionization (DAPPI), and direct analysis in real time (DART), are now joined by paper spray, a method where ESI is initiated at the pointed tip of a piece of filter paper. A drop of blood ( 15 pi) is dried on the paper, and then the paper is moistened with 25 pi of a solvent suited to both the extraction of the analytes from the blood and the ESI process (e.g., 90% methanol 10% water with either 100 ppm acetic acid or 200 ppm sodium acetate). When the paper is exposed to high voltage (3-5 kV) while held close ( 5 mm) to the entrance of the mass analyzer, a spray (similar to electrospray) is induced at the tip of the paper as capillary action carries extracted compounds through the paper (Figure 4.5). The spray is maintained for 30-90 s at a flow rate comparable to that used in nano-electrospray. [Pg.216]

El and Cl methods can be used if the compound to be studied is sufficiently volatile and stable to be vaporized intact. However, only 20% of the organics found in surface water are volatile enough to be amenable to GC-EI-MS or GC-CI-MS. Today, there are a variety of other ionization techniques available electrospray ionization (ESI), atmospheric pressure chemical ionization, matrix-assisted laser desorption ionization, and fast atom bombardment. Each of these has its advantages and disadvantages. A simple guideline to the most likely optimum ionization technique for a given class of substance is given in Table 1. [Pg.2798]

The successful on-line interfacing of several ion sources has made them dominant players in quantitative analyses using mass spectrometry. These include electron ionization (El) and chemical ionization (Cl) both coupled to GC, and the atmospheric pressure ionization (API) methods of atmospheric pressure chemical ionization (APCI) atmospheric pressure photoionization (APPI), and electrospray ionization (ESI) coupled to LC. In addition, matrix assisted laser desorption ionization (MALDI) is seeing increased application in off-line LC/MS applications. [Pg.238]

In terms of the hardware, TRMS methods described in this book use most common types of ion sources and analyzers. Electrospray ionization (ESI), electron ionization (El), atmospheric pressure chemical ionization (APCI), or photoionization systems, and their modified versions, are all widely used in TRMS measurements. The newly developed atmospheric pressure ionization schemes such as desorption electrospray ionization (DESI) and Venturi easy ambient sonic-spray ionization (V-EASI) have already found applications in this area. Mass analyzers constitute the biggest and the most costly part of MS hardware. Few laboratories can afford purchasing different types of mass spectrometers for use in diverse applications. Therefore, the choice of mass spectrometer for TRMS is not always dictated by the optimum specifications of the instrument but its availability. Fortunately, many real-time measurements can be conducted using different mass analyzers equipped with atmospheric pressure inlets - with better or worse results. For example, triple quadrupole mass spectrometers excel at quantitative capabilities however, in many cases, popular ion trap (IT)-MS instruments can be used instead. On the other hand, applications of TRMS in fundamental studies often require a particular type of instrument (e.g., Fourier transform ion cyclotron resonance mass spectrometer for photodissociation studies on trapped ions). [Pg.338]

The first method for liquid chromatography-mass spectrometry analysis of TATP by APCI-MS was reported by Widmer et al. [59]. The atmospheric pressure chemical ionization (APCI) was operated in the positive ion mode and resulted in a TATP LOD of 100 pg/pL. A lower LOD, 3.3ng, has been reported for a method that couples HPLC with an APCI-MS/MS full-scan method [52]. The LOD was reduced to 0.8 ng when SRM was employed with quantitative analysis on ions m/z 223, 132, 91, and 74 [52]. Detection limits in the low nanogram range has been reported for TATP analyzed by desorption atmospheric pressure chemical ionization (DAPCI), a technology similar to DESI [27]. [Pg.383]

Chapter 6, titled Selection of Ionization Methods of Analytes in the TLC-MS Techniques provides an overview of mass spectrometric techniques that can be coupled with TLC and act as specific detectors in this hyphenated approach. The mass spectrometric techniques discussed in this chapter are secondary mass spectrometry (SIMS), liquid secondary ion mass spectrometry (LSIMS), fast atom bombardment (FAB), matrix-assisted laser desorption/ionization (MALDI), atmospheric pressure matrix-assisted laser desorption/ionization (AP-MALDI), electrospray ionization (ESI), desorption electrospray ionization (DESI), electrospry-assisted laser desorption/ionization (ELDI), easy ambient sonic spray ionization (EASI), direct analysis in real time (DART), laser-induced acoustic desorption/electrospray ionization (LIAD/ESI), plasma-assisted multiwavelength laser desorption/ionization (PAMLDI), atmospheric-pressure chemical ionization (APCI), and dielectric barrier discharge ionization (DBDI). For the sake of illustration, the authors introduce practical examples of implementing TLC separations with detection carried out by means of individual mass spectrometric techniques for the systematically arranged compounds belonging to different chemical classes. [Pg.9]

Most of the mass spectrometry analyses are conducted under vacuum environment. However, ambient mass spectrometry is a rapidly growing field that provides fast and direct analysis of solid sample surfaces or liquid samples introduced on a suitable surface (Alberici et al. 2010 Weston 2010 Huang et al. 2010 Chen et al. 2010). For that, different ambient ionization MS methods, such as atmospheric pressure desorption/ionization on porous silicon (AP-DIOS) (Huikko et al. 2003), desorption electrospray ionization (DESI) (Takats et al. 2004), direct analysis in real time (DART) (Cody et al. 2005), desorption atmospheric pressure chemical ionization (DAPCI) (Takats et al. 2005), and desorption atmospheric pressure photoionization (DAPPI) (Haapala et al. 2007), have been successfully used in the direct analysis of compounds fi"om various samples, such as body fluids (Cody et al. 2005 Chen et al. 2006), finiits, plant leaves (Luosujarvi et al. 2010), milk (Yang et al. 2009), banknotes (Cody et al. 2005), textiles (Cody et al. 2005 Chen et al. 2007), and pharmaceutical formulations (Ifa et al. 2009 Gheen et al. 2010), just to mention a few, without any sample pretreatment. [Pg.768]

Ambient MS is another advance in the field. It allows the analysis of samples with little or no sample preparation. Following the introduction of desorption electrospray ionization (DESI) [108,109], direct analysis in real time (DART) [110], and desorption atmospheric pressure chemical ionization (DAPCI) [111, 112], a number of ambient ionization methods have been introduced. They include electrospray-assisted laser desorption/ionization (ELDI) [113], matrix-assisted laser desorption electrospray ionization (MALDESI) [114], atmospheric solids analysis probe (ASAP) [115], jet desorption ionization (JeDI) [116], desorption sonic spray ionization (DeSSI) [117], field-induced droplet ionization (FIDI) [118], desorption atmospheric pressure photoionization (DAPPI) [119], plasma-assisted desorption ionization (PADI) [120], dielectric barrier discharge ionization (DBDI) [121], and the liquid microjunction surface sampling probe method (LMJ-SSP) [122], etc. All these techniques have shown that ambient MS can be used as a rapid tool to provide efficient desorption and ionization and hence to allow mass spectrometric characterization of target compounds. [Pg.41]

Desorption electrospray ionization (DESI) [1] was introduced at the end of 2004, and direct analysis in real time (DART) [2] soon after in 2005. The apparent potential of both DESI and DART in high-throughput applications soon led to the development of some derivatives with the intention to broaden the field of applications or to adapt the underlying methodology to specific analytical needs. Now, the repertoire of methods includes variations of the DESI theme such as desorption sonic spray ionization (DeSSI) [3], later renamed easy sonic spray ionization (EASI) [4] or extractive electrospray ionization (EESI) [5,6]. Then, there are the DESI analogs of APCI and APPI, i.e., desorption atmospheric-pressure chemical ionization (DAPCI) [7,8] and desorption atmospheric pressure photoionization (DAPPI) [9]. [Pg.621]

Triple quadrupole MS instruments have been the most common ones in studies involving lipid analysis, butnovel hybrid (quadrupoletime-of-flight, etc.)instruments are rapidly gaining popularity due to their ability for multiple precursor ion scans simultaneously. Besides ESI, atmospheric pressure chemical ionization (APCI), atmospheric pressure photoionization (APPI), and matrix-assisted laser desorption ionization (MALDI) have been employed in analysis of lipids. However, these methods seem to have an advantage over ESI only in special cases. For instance, APPI and APCI allow analysis of sterols without derivatization, which is needed for ESI. [Pg.225]

The type of ion source used in an SID experiment depends on the projectile ion under study. For instance, if volatile, small organic compounds are of interest then an electron ionization (El) and/or chemical ionization (Cl) source will be sufficient. However, if one wishes to study biological compounds then a spray or desorption ionization method will be required, such as electrospray (ESI), atmospheric pressure chemical ionization (APCI), desorption chemical ionization (DCI), or matrix-assisted laser desorption ionization (MALDI). [Pg.1116]


See other pages where Atmospheric pressure chemical desorption methods is mentioned: [Pg.468]    [Pg.544]    [Pg.48]    [Pg.54]    [Pg.60]    [Pg.329]    [Pg.694]    [Pg.875]    [Pg.959]    [Pg.373]    [Pg.283]    [Pg.282]    [Pg.277]    [Pg.382]    [Pg.611]    [Pg.13]    [Pg.55]    [Pg.67]    [Pg.325]    [Pg.1463]    [Pg.86]    [Pg.241]    [Pg.297]    [Pg.40]    [Pg.199]    [Pg.338]    [Pg.4560]    [Pg.264]    [Pg.319]    [Pg.115]   