Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample calibration-check

Performance test samples (also called quality control samples or blind samples) are a quality control measure to help eliminate bias introduced by the analyst knowing the concentration of the calibration check sample. These samples of known composition are provided to the analyst as unknowns. Results are then compared with the known values, usually by a quality assurance manager. [Pg.81]

For quality assessment of an analytical process, a control chart could show the relative deviation of measured values of calibration check samples or quality control samples from their known values. Another control chart could display the precision of replicate analyses of unknowns or standards as a function of time. [Pg.81]

Calibration-check sample Chemometrics Cluster analysis Control chart Correlation coefficient Dependent variable Duplicates EDA... [Pg.82]

Calibration check sample A sample prepared independently of the calibration curve and used to detect problems with the curve. [Pg.617]

Continuing calibration for a Series Method is performed using calibration check compounds. Surrogate compounds are added to the matrix before sample preparation to evaluate recovery levels. To check GC retention times, internal standards are added to a sample after its preparation for analysis. [Pg.418]

In general, it appears that the Micromerograph, provided that frequent calibration checks are performed, is a good, reproducible instrument for size measurement. The operator time involved is less than with most other methods, and the calcns are not complicated. As in all sedimentation methods, only when the sample particles are spherical does the Stokes diameter that is measured become a measure of absolute particle size. Microscopic examination should be used to check on particle shape and the effect of deagglomeration... [Pg.520]

Display Calibration Check Graph) The calibration points as obtained under (Accept) above remain as is, but a renewed measurement yes, i of these same samples as unknowns is simulated vertical lines indicate the CL(X) that would be determined for these X = fiyes, mean)- The variability so observed mimics the within-calibration repeatability. Use the button [New Check] to repeat the simulation. [Pg.380]

It remains possible to check the correctness of the end-point detection by calibration on samples of known composition under the same measurement conditions a similar procedure consists in the differential titrations introduced by Pinkhof and Treadwell, who used a reference electrode, identical with the indicator electrode, but dipped it into a solution buffered to the end-point potential value67. [Pg.111]

Step 5 Off-line method or analyzer development and validation This step is simply standard analytical chemistry method development. For an analyzer that is to be used off-line, the method development work is generally done in an R D or analytical lab and then the analyzer is moved to where it will be used (QA/ QC lab, at-line manufacturing lab, etc.). For an analyzer that is to be used on-line, it may be possible to calibrate the analyzer off-line in a lab, or in situ in a lab reactor or a semiworks unit, and then move the analyzer to its on-line process location. Often, however, the on-line analyzer will need to be calibrated (or recalibrated) once it is in place (see Step 7). Off-line method development and validation generally includes method development and optimization, identification of appropriate check samples, method validation, and written documentation. Again, the form of the documentation (often called the method or the procedure ) is company-specific, but it typically includes principles behind the method, equipment needed, safety precautions, procedure steps, and validation results (method accuracy, precision, etc.). It is also useful to document here which approaches did not work, for the benefit of future workers. [Pg.496]

It has already been mentioned that control of a continuous process includes control and monitoring of raw materials used in the process. This is a very successful area of application for NIR because of the minimal sample preparation required, the robustness of the instrumentation and its ability to measure both chemical and physical properties [117]. Libraries of raw material can be compiled, which can then be shared between manufacturing locations given adequate calibration checks between measuring equipment [118]. [Pg.258]

When dealing with large numbers of samples and replicates, we perform periodic calibration checks to make sure that our instrument continues to work properly and the calibra-... [Pg.80]

Together, raw data and results from calibration checks, spike recoveries, quality control samples, and blanks are used to gauge accuracy. Analytical performance on replicate samples and replicate portions of the same sample measures precision. Fortification also helps ensure that qualitative identification of analyte is correct. If you spike the unknown in Figure 0-5 with extra caffeine and the area of a chromatographic peak not thought to be caffeine increases, then you have misidentified the caffeine peak. [Pg.81]

Specifications How good do the numbers have to be Write specifications Pick methods to meet specifications Consider sampling, precision, accuracy, selectivity, sensitivity, detection limit, robustness, rate of false results Employ blanks, fortification, calibration checks, quality control samples, and control charts to monitor performance Write and follow standard operating procedures... [Pg.82]

Documentation is critical for assessment. Standard protocols provide directions for what must be documented and how the documentation is to be done, including how to record information in notebooks. For labs that rely on manuals of standard practices, it is imperative that tasks done to comply with the manuals be monitored and recorded. Control charts (Box 5-1) can be used to monitor performance on blanks, calibration checks, and spiked samples to see if results are stable over time or to compare the work of different employees. Control charts can also monitor sensitivity or selectivity, especially if a laboratory encounters a wide variety of matrixes. [Pg.82]

Government agencies such as the U.S. Environmental Protection Agency set requirements for quality assurance for their own labs and for certification of other labs. Published standard methods specify precision, accuracy, numbers of blanks, replicates, and calibration checks. To monitor drinking water, regulations state how often and how many samples are to be taken. Documentation is necessary to demonstrate that all requirements have been met. Table 5-1 summarizes the quality assurance process. [Pg.82]

What is the difference between a calibration check and a performance test sample ... [Pg.93]

Calibration is time consuming when performed correctly. It may require 1 or 2 days to perform all the necessary steps (i.e., prepare stocks, filter, measure absorbance, check purity, dilute, mix, and inject calibrants). Once the stock solutions and mixed calibration solutions have been prepared, a calibration check can be performed in -4 hr. Sample preparation, depending on the matrix, may require a few minutes or a few hours. If an autosampler is unavailable for overnight injection the extracts are typically stable overnight, refrigerated at - 20° to 4°C. It is prudent to maintain the autosampler tray temperature from 4° to 15°C to reduce sample degradation. HPLC analysis of the extracted sample requires 20 to 60 min. Typically one technician can extract 12 to 24 samples per day to be analyzed overnight or the next day. [Pg.873]

Each application allows the use of eight independent channels that depending on the procedure can be used to perform individual or duplicate analysis. For instance, in the OTA application the default setting of the software permit to dedicate two channels for the blank measurement and two channels for each of the three calibrators or samples whereas in the OPs protocol the user can utilize single channel or multiple channels. In each application, a check biosensor option is available to test the correct functioning and positioning of the sensors. [Pg.700]

The preparation of a standard calibration curve is required for many colorimetric and gas chromatography analyses. A fresh calibration check standard at any selected concentration should be prepared daily and analyzed prior to sample analysis. If the response for the check standard falls outside of 15% standard deviation for the same concentration in the standard calibration curve, then a new calibration curve should be prepared. [Pg.22]

Routine GC analysis for environmental samples involve running one of the calibration check standards before sample analysis to determine if the area or height response is constant (i.e., within 15% standard deviation of the response factor or calibration factor, and to check if there is a shift in the retention times of the analytes peaks. The latter can occur to a significant degree due to any variation in conditions, such as temperature or the flow rate of the carrier gas. Therefore, an internal standard should be used if possible in order to determine the retention time shift or to compensate for any change in the peak response. If an analyte is detected in the sample, its presence must be ascertained and then confirmed as follows ... [Pg.40]

Three calibration blank standards should be analyzed to establish a representative blank level, after which the calibration standards are analyzed. After calibration, the quality control standard should be analyzed to verify the calibration. The sample introduction system is flushed with rinse blank, and the blank solution is analyzed to check for carry-over and the blank level. If the blank level is acceptable, the samples can be analyzed. If the blank values are too high, the flushing of the sample introduction system and analysis of the blank solution should be repeated until an acceptable blank level is reached. The calibration blank value, which is the same as the absolute value of the instrument response, must be lower than the method s detection limit. [Pg.407]

Analytical measurements should be made with properly tested and documented procedures. These procedures should utilise controls and calibration steps to minimise random and systematic errors. There are basically two types of controls (a) those used to determine whether or not an analytical procedure is in statistical control, and (b) those used to determine whether or not an analyte of interest is present in a studied population but not in a similar control population. The purpose of calibration is to minimise bias in the measurement process. Calibration or standardisation critically depends upon the quality of the chemicals in the standard solutions and the care exercised in their preparation. Another important factor is the stability of these standards once they are prepared. Calibration check standards should be freshly prepared frequently, depending on their stability (Keith, 1991). No data should be reported beyond the range of calibration of the methodology. Appropriate quality control samples and experiments must be included to verify that interferences are not present with the analytes of interest, or, if they are, that they be removed or accommodated. [Pg.260]

A horizontal audit examines one aspect of the analysis. For example, a horizontal audit might check the balance used to weigh samples for analysis. Is the balance calibration checked as required by the quality system Has it been regularly serviced Have the weights used to perform the balance calibration themselves been calibrated ... [Pg.59]

Calibration checks are nsnally carried out on analyzers by injecting known samples into the sample conditioning systems. This mnst be determined for each type of analyzer by reference to the mannfacrnrer s handbook or by consnltation with the instrument vendor, and it mnst be agreed on with the cnstomer. Complex analyzer systems usually require specialist personnel from the analyzer mannfacmrer to assist in precalibration and commissioning and are generally ontside the scope of the installation contractor s responsibility and experience. [Pg.564]

Standard addition is a useful calibration method when the sample matrix cannot be reproduced well enough to prepare standard solutions for calibration. The sample itself is used to prepare calibration standards by addition of known amounts of the analyte to three or more aliquots of the sample. The amount of added standard should cover a range of about 10-100% of the sample concentration. When the instrument response is plotted against the concentration added, the sample concentration is determined by calculating the absolute value of the x intercept. See Fig. 2. This technique is often used in conjunction with internal standards, where the instrument response is then the ratio of the analyte to internal standard response. This is a very good reference method to cross-check a primary method. It is rarely used for routine analysis of multiple samples because each test article requires multiple sample preparations and analyses. [Pg.99]

The filter was leached with 5 mL of Na0H-Na2C03 buffer solution samples were heated to near boiling point for 45 min, and diluted with H2O. Detection was by ICP-MS, using a commercial solution of Cr as calibrant (checked against NIST SRM 3112). [Pg.159]


See other pages where Sample calibration-check is mentioned: [Pg.76]    [Pg.76]    [Pg.251]    [Pg.583]    [Pg.603]    [Pg.8]    [Pg.240]    [Pg.249]    [Pg.81]    [Pg.91]    [Pg.282]    [Pg.14]    [Pg.21]    [Pg.72]    [Pg.109]    [Pg.256]    [Pg.241]    [Pg.138]    [Pg.100]   
See also in sourсe #XX -- [ Pg.75 , Pg.78 ]




SEARCH



CHECK

Calibration check

Calibration samples

Check sample

Checking

Checking calibration

© 2024 chempedia.info