Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Values measured

Unfortunately, many commonly used methods for parameter estimation give only estimates for the parameters and no measures of their uncertainty. This is usually accomplished by calculation of the dependent variable at each experimental point, summation of the squared differences between the calculated and measured values, and adjustment of parameters to minimize this sum. Such methods routinely ignore errors in the measured independent variables. For example, in vapor-liquid equilibrium data reduction, errors in the liquid-phase mole fraction and temperature measurements are often assumed to be absent. The total pressure is calculated as a function of the estimated parameters, the measured temperature, and the measured liquid-phase mole fraction. [Pg.97]

Table 1 gives the measured data, estimates of the true values corresponding to the measurements, and deviations of the measured values from model predictions. Figure 1 shows the phase diagram corresponding to these parameters, together with the measured data. [Pg.100]

In the maximum-likelihood method used here, the "true" value of each measured variable is also found in the course of parameter estimation. The differences between these "true" values and the corresponding experimentally measured values are the residuals (also called deviations). When there are many data points, the residuals can be analyzed by standard statistical methods (Draper and Smith, 1966). If, however, there are only a few data points, examination of the residuals for trends, when plotted versus other system variables, may provide valuable information. Often these plots can indicate at a glance excessive experimental error, systematic error, or "lack of fit." Data points which are obviously bad can also be readily detected. If the model is suitable and if there are no systematic errors, such a plot shows the residuals randomly distributed with zero means. This behavior is shown in Figure 3 for the ethyl-acetate-n-propanol data of Murti and Van Winkle (1958), fitted with the van Laar equation. [Pg.105]

INITIAL ESTIMATES OF TRUE VALUES OF THE VARIABLES ARE SET EQUAL TO THE MEASURED VALUES. [Pg.241]

APRAV OF TRUE VALUES CORRESPCNDING TC THE MEASURED VALUES IN THE XM ARRAY... [Pg.252]

VECTOR OF TRUE VALUES CORRESPCNDING TO THE MEASURED VALUES IN TH ZM VECTOR... [Pg.252]

MEASURED VALUES, CALCULATED VALUES, AND RESIDUALS (MEAS-CALC)... [Pg.276]

To verify the modelling of the data eolleetion process, calculations of SAT 4, in the entrance window of the XRII was compared to measurements of RNR p oj in stored data as function of tube potential. The images object was a steel cylinder 5-mm) with a glass rod 1-mm) as defect. X-ray spectra were filtered with 0.6-mm copper. Tube current and exposure time were varied so that the signal beside the object. So, was kept constant for all tube potentials. Figure 8 shows measured and simulated SNR oproj, where both point out 100 kV as the tube potential that gives a maximum. Due to overestimation of the noise in calculations the maximum in the simulated values are normalised to the maximum in the measured values. Once the model was verified it was used to calculate optimal choice of filter materials and tube potentials, see figure 9. [Pg.212]

Fig. 3.2 optimum excitation frequency for cracks in varing depths. The dotted line is for a sheet inducer and the solid one is for a Double-D coil. The indicated points represent the measured values. [Pg.258]

Fig. 3.3 Dependence of the signal strength A B, at the surface (z=0) on the depth of the cracks. The dots represent the measured values (double-D coil). Fig. 3.3 Dependence of the signal strength A B, at the surface (z=0) on the depth of the cracks. The dots represent the measured values (double-D coil).
For the field calculation it more convenient to use a tx(B) curve than the normal ix(H) curve because the calculated vector potential A is derived from the flux density B. This ii(B) curve however can be calculated easily from the measured values. [Pg.314]

The system compares its own values for good and a bad magnetic particle suspensions with the actually measured values. The system values are stored in the form of reference curves for a fresh MP-suspension used on the machine and the used MP-suspension which has to be changed. [Pg.629]

The resistance due to a circular junction is given by / = /2ak, where a is the radius of the junction and k is specific conductivity of the metal. For the case of two steel plates, the measured resistance is 5 x 10" Q for a load of 50 kg the yield pressure of steel is 60 kg/mm, and the specific resistance is 5x 10 Q/cm. Calculate the number of junctions, assuming that it is their combined resistance that is giving the measured value. [Pg.458]

The fifth postulate and its corollary are extremely important concepts. Unlike classical mechanics, where everything can in principle be known with precision, one can generally talk only about the probabilities associated with each member of a set of possible outcomes in quantum mechanics. By making a measurement of the quantity A, all that can be said with certainty is that one of the eigenvalues of /4 will be observed, and its probability can be calculated precisely. However, if it happens that the wavefiinction corresponds to one of the eigenfunctions of the operator A, then and only then is the outcome of the experiment certain the measured value of A will be the corresponding eigenvalue. [Pg.11]

The enthalpy, entropy and free energy changes for an isothennal reaction near 0 K caimot be measured directly because of the impossibility of carrying out the reaction reversibly in a reasonable time. One can, however, by a suitable combination of measured values, calculate them indirectly. In particular, if the value of... [Pg.369]

Caution Restrain ts may sign ificantly affect the property yon are observing. When using restrain is, ch cck th eir effects on measured values. [Pg.81]

Thus, if measured values of the left hand side of equation (6.4) are plotted against the mean pressure (pj + p2)/2, the theory predicts a straight line with slope B /pu RT and and intercept D /RT at (p +p )/2 = 0, and experiments of Carman [35] confirm that this is the case. In contrast,... [Pg.54]

If, on the other hand, one were able to measure values when r is equal to some specified bond length (this is only a hypothetical example there is no known way to perform such a measurement), then the probability would equal ... [Pg.46]

The system of atomic units was developed to simplify mathematical equations by setting many fundamental constants equal to 1. This is a means for theorists to save on pencil lead and thus possible errors. It also reduces the amount of computer time necessary to perform chemical computations, which can be considerable. The third advantage is that any changes in the measured values of physical constants do not affect the theoretical results. Some theorists work entirely in atomic units, but many researchers convert the theoretical results into more familiar unit systems. Table 2.1 gives some conversion factors for atomic units. [Pg.9]

Note for highly substituted allenes containing silicon, gernanium or tin we observed deviations of the calculated chemical shifts from the measured values, and one should therefore treat these particular cases with special care. [Pg.256]

Table III-16 shows the estimations of AH for di- and trimethyl-thiazoles based on measured values for thiazole and methylthiazoles (150). Table III-16 shows the estimations of AH for di- and trimethyl-thiazoles based on measured values for thiazole and methylthiazoles (150).
These results show that the measured values of H° are close to those calculated from the thiazole value and the increments. That compounds substituted ortho-ortho to nitrogen (2,4-dimethyI and 2,4,5-trimethyl-thiazole) also obey this rule shows that the methyl groups do not interact sterically (150). The same conclusion had been reached in the case of the picolines and lutidines (151). [Pg.356]

In all other cases the quantity / calculated from the specific surface is a mean diameter. Unless there is some definite and detailed evidence as to particle shape, the simplest such diameter to aim at is the mean diameter obtained by substituting the measured value of A in Equation (1.79)... [Pg.35]

From standard Student tables, the value for t l/fT= 1.049 at the 95% confidence level. Thus, mean value = 56.3 13.0 (95% confidence level) and one could be confident that 95% of measured values would fall in the range 69.3 - 43.3. This is a large range and is not very precise. [Pg.362]

In theory, this process can be reversed in that any measured mass leads to an elemental composition. For example, a measured value of 17 would imply the composition, NH3. [Pg.416]

From the ratio of activities and measured values of n, the average number of initiator fragments per polymer can be determined. Carry out a similar argument for the ratio of activities for polystyrene and evaluate the average number of initiator fragments per molecule for each polymer from the following datat ... [Pg.416]

Use the values determined in Example 7.6 for the vinylidene chloride (M )-isobutylene (M2) systemf to calculate for various values of fi according to the terminal mechanism. Prepare a plot of the results. On the same graph, plot the following experimentally measured values of fi and ... [Pg.499]

Since capillary tubing is involved in osmotic experiments, there are several points pertaining to this feature that should be noted. First, tubes that are carefully matched in diameter should be used so that no correction for surface tension effects need be considered. Next it should be appreciated that an equilibrium osmotic pressure can develop in a capillary tube with a minimum flow of solvent, and therefore the measured value of II applies to the solution as prepared. The pressure, of course, is independent of the cross-sectional area of the liquid column, but if too much solvent transfer were involved, then the effects of dilution would also have to be considered. Now let us examine the practical units that are used to express the concentration of solutions in these experiments. [Pg.550]

Figure 10 shows a comparison of Scofield s calculated values with experimentally measured values (24) which, in addition to (, are dependent on spectrometer transmission function. The overall agreement between the calculated and experimental values is quite good. The far rightjy-axis in Figure 10 indicates the experimentally accessible surface sensitivities in monolayers (ML) as a function of atomic number. For most elements, sensitivities on the order of 1% of a ML are achievable. [Pg.276]


See other pages where Values measured is mentioned: [Pg.68]    [Pg.99]    [Pg.28]    [Pg.218]    [Pg.221]    [Pg.159]    [Pg.240]    [Pg.251]    [Pg.759]    [Pg.1146]    [Pg.1338]    [Pg.1875]    [Pg.492]    [Pg.58]    [Pg.82]    [Pg.84]    [Pg.917]    [Pg.362]    [Pg.363]    [Pg.53]    [Pg.276]   
See also in sourсe #XX -- [ Pg.6 , Pg.10 ]

See also in sourсe #XX -- [ Pg.14 , Pg.292 ]

See also in sourсe #XX -- [ Pg.14 , Pg.292 ]

See also in sourсe #XX -- [ Pg.29 ]




SEARCH



Average values of measurements

Central value, measures

Color measurements tristimulus values

Comparison of measured values

Correlation and regression - relationships between measured values

Credit analysis and relative value measurement

Density values, measurement

Differences between calculated and measured values

Economic value added measures

Evaluation of linear dependent measured values

Expectation values position measurements

FIGURE 6.10 Empirical p-box corresponding to a data set with measurement error including 4 nondetect values

Health care economics value measurements

Ideal measured value

Inputs measured values

Lipophilicity values, HPLC measurements

Market prices, economic value measures

Measured digitized value

Measured quantity value

Measurements average values

Measurements of f values for high Rydberg members

Measuring coupling constants proton chemical shift values

Model with Measured Values for Dissolved Iron

Model with Measured Values of DOC

Outputs measured values

P-Value measurement

Performance measures for optimal and baseline values of inventory parameters

Peroxide value , measurement

Peroxide value , measurement stability measurements

Polymers measurement values

Positive-operator-valued measurement

Postulate of the Averaged Measured Values

Rf values measurement

Stable isotope measurement 8-value

Texture measurements yield value

The Postulate Relating Measured Values to Eigenvalues

The Value of Accurate Mass Measurement

Thiocyanate peroxide value measurement

Toxicity measurements threshold limit values

Yield value, measurement

© 2024 chempedia.info