Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analysis sample

Before sample preparation, surrogate compounds must be added to the matrix. These are used to evaluate the efficiency of recovery of sample for any analytical method. Surrogate standards are often brominated, fluorinated, or isotopically labeled compounds that are not expected to be present in environmental media. If the surrogates are detected by GC/MS within the specified range, it is [Pg.299]

Comparison of the mass spectrum from a target compound (top), with the three best fits from the library of standard spectra (lower three traces). The closeness of fit of the mass spectra and the chromatographic retention time lead to a positive identification of 2, 6-dimethylheptane. [Pg.300]

Internal standards at a known concentration are added to the sample after its preparation but prior to analysis to check for GC retention-time accuracy and response stability. If the internal standard responses are in error by more than a factor of two, the analysis must be stopped and the initial calibration repeated. Only if all the criteria have been met can sample analysis begin. [Pg.300]

A considerable amount of time is necessary to reach the point at which sample analyses can commence, and it is essential that the stability and reliability of the mass spectrometer be high to ensure maximum sample throughput during the limited time available between calibration checta. [Pg.300]

Chromatographic peak areas are calculated automatically by the data system by reference to the response obtained from certain specified, compound-dependent ions. From the peak areas of the target compounds, quantification is achieved by comparison with the internal standards, which are present in known concentration. The laboratory responsible for the analysis must report the target compounds and all tentatively identified (nontarget) compounds. Standard EPA forms must be completed and submitted. A laboratory is said to be in compliance when it has satisfied all aspects of its CLP contract. [Pg.301]

Sample analysis requires that sufficient test data have been obtained for a sample, such that all internal and external requirements are met. It also covers the accuracy and applicability of test data. Related specific functions include manual and automated result entry, validating against method specification, checking of detection limits, and instrument management. [Pg.299]

Weigh out an appropriate weight of the weU-homogenised sample into a 150 mL beaker and add 10.0 mL of ISE solution. Heat for 1/2 hour on a steam bath. Add 100 mL of water and cool to 5 C. Skim or filter and warm to room temperatvire. Place the electrodes in the solution, wait to stabihse (usually 1 minute) and read the mV value. Use the cahbration chart to determine the Cl level and to calculate the chloride in the sample. Clean the electrode with a soft cloth or absorbent paper. The ISA solution also precipitates the interfering proteins (albuminoids). Manufacturers of some pH/potentiometers supply detachable scales which allow cahbration and analysis to be reported in % NaCl. [Pg.284]

Specification of the Problem. In the foUowing, the cardinalities 2, k are treated in detail. Hence appropriate edges in ovalene and drcumovalenes are to be counted, representing positions of naphthalene holes in the circular coronoids 2, k). [Pg.212]

Direct Counting. Divide A - ircumovalene, which belong to the symmetry I 2, into segments as shown below. [Pg.212]

The increments in the number of edges for each k are collected in the below table. In total one obtains the increments in the number of nonisomorphic isomers, say AI. [Pg.213]

The absolute values, obtained from successive additions of the numbers in the above table, are given below. [Pg.213]

Here the I numbers give the final answer the numbers of nonisomorphic isomers of 2, A coronoids, represented by the numbers of appropriate edges in A -circumovalene. [Pg.214]


An additional advantage derived from plotting the residuals is that it can aid in detecting a bad data point. If one of the points noticeably deviates from the trend line, it is probably due to a mistake in sampling, analysis, or reporting. The best action would be to repeat the measurement. However, this is often impractical. The alternative is to reject the datum if its occurrence is so improbable that it would not reasonably be expected to occur in the given set of experiments. [Pg.107]

Gardone, M. J. Detection and Determination of Error in Analytical Methodology. Part 11. Gorrection for Gorrigible Systematic Error in the Gourse of Real Sample Analysis, /. Assoc. Off. Anal. Chem. 1983, 66, 1283-1294. [Pg.134]

A spike recovery for the analysis of chloride in well water was performed by adding 5.00 mb of a 25,000-ppm solution of Ck to a 500-mL volumetric flask and diluting to volume with the sample. Analysis of the sample and the spiked sample resulted in chloride concentrations of 183 ppm and 409 ppm, respectively. Determine the percent recovery of the spike. [Pg.711]

S ample Analysis. The fohowing procedure is used for sample analysis (/) after sample connections are made and the sample tubing purged for 30 minutes, the tubing is connected to the sampling manifold which serves to distribute sample gas to each of the analy2ers, and should analy2ers have specific pressure... [Pg.90]

Microwaves have successfully been used for rewarming of blood for medical appHcations (157). Another successful appHcation, not commetciali2ed as of this writing, is the use of microwave heating for rapid tissue fixation (158,159). This procedure appears to reduce the time for tissue sample analysis... [Pg.345]

Like the refining of the PGMs, the analysis is compHcated by the chemical similarity of the metals. The techniques used depend on the elements present and their concentration in the sample. For some low grade samples, analysis is preceded by a concentration stage using fire assay with collection into a lead or nickel sulfide button. The individual metals can then be determined. [Pg.171]

Statistical control of an analysis or instmment is best demonstrated by SQC of a standard sample analysis. The preferred approach to demonstrate statistical control is to use a reference sample of the subject material that has been carefully analyzed or, alternatively, to use a purchased reference standard. Either material must be stored so that it remains unchanged, eg, sealed in ampuls or septum capped bottles. Periodically a sample can then be reanalyzed by the technique used for routine analysis. These results are plotted in a control chart. Any change in the stabihty of the test in question results in a lack of... [Pg.367]

Sample test data are either manually entered into the system or captured from analytical instmments coimected to the LIMS. The system performs any necessary calculations and compares the result to the appropriate specification stored in the database. If the comparison indicates the material is in conformance, the system can automatically provide an approval. Otherwise, the LIMS can alert lab supervision to the nonconforming sample analysis. [Pg.368]

The field of steroid analysis includes identification of steroids in biological samples, analysis of pharmaceutical formulations, and elucidation of steroid stmctures. Many different analytical methods, such as ultraviolet (uv) spectroscopy, infrared (ir) spectroscopy, nuclear magnetic resonance (nmr) spectroscopy, x-ray crystallography, and mass spectroscopy, are used for steroid analysis. The constant development of these analytical techniques has stimulated the advancement of steroid analysis. [Pg.448]

Trihalomethanes in Drinking Water (Sampling Analysis, Monitoring and Compliance), U.S. Envkonmental Piotection Agency, EPA/570/9-83-002, Washington, D.C., 1983. [Pg.492]

Suppose we have two methods of preparing some product and we wish to see which treatment is best. When there are only two treatments, then the sampling analysis discussed in the section Two-Population Test of Hypothesis for Means can be used to deduce if the means of the two treatments differ significantly. When there are more treatments, the analysis is more detailed. Suppose the experimental results are arranged as shown in the table several measurements for each treatment. The goal is to see if the treatments differ significantly from each other that is, whether their means are different when the samples have the same variance. The hypothesis is that the treatments are all the same, and the null hypothesis is that they are different. The statistical validity of the hypothesis is determined by an analysis of variance. [Pg.506]

Method or samphng, location, size and number of samples, method of sample analysis, and fraction of the batch removed for samphng all contribute to how well the samphng study reflects the actual conditions. [Pg.1763]

On-Hne Procedures The growing trend toward automation in industry has resiilted in many studies of rapid procedures for generating size information so that feedback loops can be instituted as an integral part of a process. Many of these techniques are modifications of more traditional methods. The problems associated with on-line methods include allocation and preparation of a representative sample analysis of the sample evaluation of the results. The interface between the measuring apparatus and the process has the potential of high complexity, and consequently, high costs [Leschonsld, Paiticle Cha racterization, 1, 1 (July 1984)]. [Pg.1828]

Even within a single sample analysis, it is hkely that some of the reported concentrations are known with greater accuracy than others. L oratory personnel will know which concentrations can be rehed upon and which should be questioned. The plant-performance analyst should know at this stage which of the concentrations are of greatest importance and direct the discussion to those components. [Pg.2558]

Manual sampling/ analysis is part of control scheme resulting in a large lag time in determining when to proceed and what to do (heat/cool, etc.). [Pg.122]

Ensure that safety is not compromised by using manual sampling/analysis as part of the control scheme, e.g., increasing hold times... [Pg.122]

In the analysis of trace elements or thin films on substrate using electrons, however, one finds that the MDL, may be increased by choosing Eq such that Uis just greater than 1. The reason for this is that the k factor, which is the ratio of the intensity from the sample to that from the standard, increases as Uapproaches 1 for thin films. Thus, by maximizing the k factor, the sensitivity is increased. For bulk sample analysis, however, the k factor will usually be a maximum ax. U- 2.5. [Pg.129]

The sensitivity, accuracy, and precision of solid-sample analysis have been greatly improved by coupling LA with ICP-OES-MS. The ablated species are transported by means of a carrier gas (usually argon) into the plasma torch. Further atomization, excitation, and ionization of the ablated species in the stationary hot plasma result in a dramatic increase in the sensitivity of the detection of radiation (LA-ICP-OES) or of the detection of ions (LA-ICP-MS). [Pg.234]

A bulk sample is the last choice and the least desirable. It should be submitted "for laboratory use only" if there is a possibility of contamination by other matter. The type of bulk sample submitted to the laboratory should be cross-referenced to the appropriate air samples. A reported bulk sample analysis for quartz (or cristobalite) will be semi-quantitative in nature because (1) The XRD analysis procedure requires a thin layer deposition for an accurate analysis. (2) The error for bulk samples analyzed by XRD is unknown because the particle size of nonrespirable bulk samples varies from sample to sample. [Pg.253]

FIGURE 5-2. Sample Analysis of Selected PSM Gaps and Priorities... [Pg.103]


See other pages where Analysis sample is mentioned: [Pg.1829]    [Pg.296]    [Pg.299]    [Pg.300]    [Pg.698]    [Pg.384]    [Pg.21]    [Pg.108]    [Pg.548]    [Pg.546]    [Pg.398]    [Pg.143]    [Pg.243]    [Pg.169]    [Pg.338]    [Pg.119]    [Pg.122]    [Pg.43]    [Pg.547]    [Pg.10]    [Pg.491]    [Pg.498]    [Pg.231]    [Pg.235]    [Pg.246]    [Pg.115]   
See also in sourсe #XX -- [ Pg.296 ]

See also in sourсe #XX -- [ Pg.158 ]

See also in sourсe #XX -- [ Pg.145 ]

See also in sourсe #XX -- [ Pg.296 ]

See also in sourсe #XX -- [ Pg.8 , Pg.52 ]

See also in sourсe #XX -- [ Pg.180 ]




SEARCH



© 2024 chempedia.info