Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Silicas columns

For most samples liquid-solid chromatography does not offer any special advantages over liquid-liquid chromatography (LLC). One exception is for the analysis of isomers, where LLC excels. Figure 12.32 shows a typical LSC separation of two amphetamines on a silica column using an 80 20 mixture of methylene chloride and methanol containing 1% NH4OH as a mobile phase. Nonpolar stationary phases, such as charcoal-based absorbents, also may be used. [Pg.590]

The analysis of cigarette smoke for 16 different polyaromatic hydrocarbons is described in this experiment. Separations are carried out using a polymeric bonded silica column with a mobile phase of 50% v/v water, 40% v/v acetonitrile, and 10% v/v tetrahydrofuran. A notable feature of this experiment is the evaluation of two means of detection. The ability to improve sensitivity by selecting the optimum excitation and emission wavelengths when using a fluorescence detector is demonstrated. A comparison of fluorescence detection with absorbance detection shows that better detection limits are obtained when using fluorescence. [Pg.613]

Coenzyme Q4 (Ubiquinone-4, 2,3-dimethoxy-5-methyl-6-[3,7,ll,15-tetrametbyl-hexadeca-2/,6/,10/,14-tetraenyl]-[l,4]benzoquinone [4370-62-l]M 454.7, m 30 , 33-45 , A (275nm) 185. A red oil purified by TLC chromatography on Si02 and eluted with Et20-hexane. Purity can be checked by HPLC (silica column using 7% Et20-hexane). It has A- ax 270 nm (e 14,800) in pet ether. [NMR and MS Naruta J Org Chem 45 4097 1980 cf Morton Biochemical Spectroscopy (Adam Hilger, London, 1975) p 491]. It has also been dissolved in MeOH/EtOH (1 1 v/v) and kept at 5 until crystals appear [Lester and Crane Biochim Biophys Acta 32 497 1958]. [Pg.522]

Silica gel, per se, is not so frequently used in LC as the reversed phases or the bonded phases, because silica separates substances largely by polar interactions with the silanol groups on the silica surface. In contrast, the reversed and bonded phases separate material largely by interactions with the dispersive components of the solute. As the dispersive character of substances, in general, vary more subtly than does their polar character, the reversed and bonded phases are usually preferred. In addition, silica has a significant solubility in many solvents, particularly aqueous solvents and, thus, silica columns can be less stable than those packed with bonded phases. The analytical procedure can be a little more complex and costly with silica gel columns as, in general, a wider variety of more expensive solvents are required. Reversed and bonded phases utilize blended solvents such as hexane/ethanol, methanol/water or acetonitrile/water mixtures as the mobile phase and, consequently, are considerably more economical. Nevertheless, silica gel has certain areas of application for which it is particularly useful and is very effective for separating polarizable substances such as the polynuclear aromatic hydrocarbons and substances... [Pg.93]

The authors repeated the experiment with two, more strongly retained, solutes m-dimethoxy benzene and benzyl acetate. These solutes were found to elute at (k ) values of 10.5 and 27.0 respectively on a silica column operated with the same mobile phase. The results obtained are shown as similar curves in Figure 13. The m dimethoxy benzene, which eluted at a (k ) of 10.5, also failed to displace any ethyl acetate from the silica gel even when more than 0.5 g of solute resided on the silica surface. Consequently, the m-dimethoxy benzene must have also interacted with the surface by a sorption process. [Pg.104]

The initial work was carried out using a silica column 25 cm long and 9 mm I.D. packed with Partisil 10 silica (actual mean particle diameter 8.5 pm) thermostatted at... [Pg.317]

Toyopearl HW-50S resin has been used to help isolate the ubiquitin-histone conjugate mH2A from the unicellular ciliated protozoan Tetrahymena pyriformis. Figure 4.49 shows the separation of mH2A from the histone, H2A. The sole difference between these two components is a small polypeptide, ubiquitin (approximately 8500 Da). The mH2A fraction was then further purified by HPLC on a Tosoh ODS-silica column (52). One of the many benefits... [Pg.154]

Let us consider the separation of polymethylmethacrylate (PMMA) on a nonmodified silica column as an example. In THE (medium polar eluent) the PMMA eludes in size exclusion mode because the dipoles of the methylmethacrylate (MMA) are masked by the dipoles of the THE. Using the nonpolar toluene as the eluent on the same column, the separation is governed by adsorption because the dipoles of the carbonyl group in the PMMA will interact with the dipoles on the surface of the stationary phase. The separation of PMMA in the critical mode of adsorption can be achieved by selecting an appropriate THF/toluene mixture as the eluent. In this case all PMMA samples... [Pg.274]

The use of bonded, silica column supports has also become a useful way to characterize cationic, water-soluble polymers. CATSEC SEC columns from Micra Scientific contain a silica support with a polymerized polyamine-bonded phase. This imparts a cationic surface charge on the packing that can be... [Pg.573]

Figure 1.2 Chromatogram of coal-tar oil obtained by using the following conditions column, Waters Spherisorb PAH 5 mm in 250 p.m id X 30 cm fused silica column oven temperature, 100°C UV detector wavelength to 254 nm mobile phase, 100 to 300 bar CO2 and 0.10 to 1.00 p.L min methanol over 30 minutes. Figure 1.2 Chromatogram of coal-tar oil obtained by using the following conditions column, Waters Spherisorb PAH 5 mm in 250 p.m id X 30 cm fused silica column oven temperature, 100°C UV detector wavelength to 254 nm mobile phase, 100 to 300 bar CO2 and 0.10 to 1.00 p.L min methanol over 30 minutes.
Figure 12.8 Mia ocolumn size exclusion chromatogram of a styrene-aaylonitrile copolymer sample fractions ti ansfeired to the pyrolysis system are indicated 1-6. Conditions fused-silica column (50 cm X 250 p.m i.d.) packed with Zorbax PSM-1000 (7p.m 4f) eluent, THF flow rate, 2.0 p.L/min detector, Jasco Uvidec V at 220 nm injection size, 20 nL. Reprinted from Analytical Chemistry, 61, H. J. Cortes et al, Multidimensional chromatography using on-line microcolumn liquid chromatography and pyrolysis gas chromatography for polymer characterization , pp. 961 -965, copyright 1989, with peimission from the American Chemical Society. Figure 12.8 Mia ocolumn size exclusion chromatogram of a styrene-aaylonitrile copolymer sample fractions ti ansfeired to the pyrolysis system are indicated 1-6. Conditions fused-silica column (50 cm X 250 p.m i.d.) packed with Zorbax PSM-1000 (7p.m 4f) eluent, THF flow rate, 2.0 p.L/min detector, Jasco Uvidec V at 220 nm injection size, 20 nL. Reprinted from Analytical Chemistry, 61, H. J. Cortes et al, Multidimensional chromatography using on-line microcolumn liquid chromatography and pyrolysis gas chromatography for polymer characterization , pp. 961 -965, copyright 1989, with peimission from the American Chemical Society.
Figure 12.10 Microcolumn SEC-LC analysis of an acrylonitrile-butadiene-styrene (ABS) teipolymer sample (a) SEC ti ace (b) EC ti ace. SEC conditions fused-silica column (30 cm X 250 mm i.d.) packed with PL-GEL (50 A pore size, 5 mm particle diameter) eluent, THE at a flow rate of 2.0 mL/min injection size, 200 nL UV detection at 254 nm x represents the polymer additive fraction (6 p-L) tr ansferred to EC system. EC conditions NovaPak CIS Column (15 cm X 4.6 mm i.d.) eluent, acetonitrile-water (60 40) to (95 5) in 15 min gradient flow rate of 1.5 mL/min detection at 214 nm. Peaks identification is follows 1, styrene-acrylonitrile 2, styrene 3, benzylbutyl phthalate 4, nonylphenol isomers 5, Vanox 2246 6, Topanol 7, unknown 8, Tinuvin 328 9, Irganox 1076 10, unknown. Reprinted with permission from Ref. (14). Figure 12.10 Microcolumn SEC-LC analysis of an acrylonitrile-butadiene-styrene (ABS) teipolymer sample (a) SEC ti ace (b) EC ti ace. SEC conditions fused-silica column (30 cm X 250 mm i.d.) packed with PL-GEL (50 A pore size, 5 mm particle diameter) eluent, THE at a flow rate of 2.0 mL/min injection size, 200 nL UV detection at 254 nm x represents the polymer additive fraction (6 p-L) tr ansferred to EC system. EC conditions NovaPak CIS Column (15 cm X 4.6 mm i.d.) eluent, acetonitrile-water (60 40) to (95 5) in 15 min gradient flow rate of 1.5 mL/min detection at 214 nm. Peaks identification is follows 1, styrene-acrylonitrile 2, styrene 3, benzylbutyl phthalate 4, nonylphenol isomers 5, Vanox 2246 6, Topanol 7, unknown 8, Tinuvin 328 9, Irganox 1076 10, unknown. Reprinted with permission from Ref. (14).
The preseparation utilized a 5 pim cyano column (250 cm X 4.6 mm i.d.) and a 5 p.m silica column (250 cm X 4.6 mm i.d.) in series, followed by GC analysis on an SE-54 column (25 m X 0.2 mm i.d., 0.33 p.m film thickness). The SFC system separated the aviation sample into two peaks, including saturates and single-ring aromatics as the first peak, and two-ring aromatic fractions as the second peak. These fractions were selectively cut and then transferred to the GC unit for further analysis. (Figure 12.20). [Pg.327]

Zebiihr et al. (29) developed an automated system for determining PAHs, PCBs and PCDD/Fs by using an aminopropyl silica column coupled to a porous graphitic carbon column. This method gives five fractions, i.e. aliphatic and monoaromatic hydrocarbons, polycyclic aromatic hydrocarbons, PCBs with two or more ortho-chlorines, mono-ort/io PCBs, and non-ortho PCBs and PCDD/Fs. This method employed five switching valves and was successfully used with extracts of sediments, biological samples and electrostatic filter precipitates. [Pg.343]

Figure 14.14 Schematic diagram of the SFC olefins analyser system Cl, high-surface-area silica column C2, silver-loaded silica column VI and V2, six-poit valves FID, flame-ionisation detector UV, ulcaviolet monitor detector. Figure 14.14 Schematic diagram of the SFC olefins analyser system Cl, high-surface-area silica column C2, silver-loaded silica column VI and V2, six-poit valves FID, flame-ionisation detector UV, ulcaviolet monitor detector.
A) 1-(2-Amino-5-chlorophenyll-1-(2-fluorophenyll-2-a2a-but-1-en-4-ol A mixture of 40 g of 2-methylimidazole hydrochloride and of 90 g of 2-amino-5-chloro-2 -fluoro-benzophenone in 240 ml of ethanolamine is heated at 135 for 2 hours. After cooling, the reaction mixture is poured into an aqueous sodium bicarbonate solution. The mixture is extracted with ether, the organic phase is washed repeatedly with water and is dried over sodium sulfate, and the solvent is evaporated to dryness. The residual oil is chromatographed on a silica column, elution being carried out with a 50/50 mixture of cyclohexane and ethyl acetate. [Pg.882]

From the reaction of 5-0-benzoyl-l,2-0-isopropylidene-o -D-en/t/iro-pentofuranos-3-ulose (prepared in 80% yield by oxidation of 5-0-benzoyl-l,2-0-isopropylidene- -D-xylofuranose (35,36) with ruthenium tetroxide) with an excess of diazomethane in methanol-ether, two main products (m.p. 44°-45°C. and 76°-77°C.), both epoxides, could be isolated by chromatography of the product on a silica column. An... [Pg.154]

C and the residual solid was dried in vacuo over P205 overnight and then heated in toluene (20 mL) under reflux until TLC showed that the reaction was complete (15 min). The sodium p-toluenesulfinate produced was filtered off and the filtrate was washed with H20 (2 x 50 mL), dried and evaporated to leave the product yield 0.40 g (78%) oil bp 80 C/10 Torr. HPLC analysis (silica column at 0 T7) showed that the tautomers 7a and 8a were present in the ratio 5.9 1. [Pg.348]

HPLC on a Cl 8-silica column, with 27% (v/v) acetonitrile (pH 2.6 with H3PO4). The active fraction eluted is concentrated, and the panal in the residue is extracted with ethyl acetate. After evaporating ethyl acetate, panal is redissolved in 30% methanol, and stored at -30°C. [Pg.278]

Protein mixtures were well resolved on poly(aspartic acid)-silica columns using 0.05 mol/1 phosphate buffer, pH 6.0 and a gradient of sodium chloride from 0 to 0.6 mol/1. The columns displayed a high capacity and selectivity. Figure 3 shows the separation of several standard proteins with isoelectric points ranging from 4.7 to over 11. Peaks are sharp and show minimal tailing. The poly(aspartic acid) coating was quite stable the columns lasted for hundreds of hours of use without decrease in efficiency and capacity. [Pg.151]

I Most of the GC conditions given in this book are for 0.25-mm ID columns, but 0.32- or 0.53-mm ID columns also can be used. The wide bore fused silica columns are found to be more inert, probably because of the greater film thicknesses. A splitter arrangement with a jet separator is used with 0.53-mm ID columns. This arrangement shown in Figure 11.1 has the advantage of simultaneous flame ionization quantitation. [Pg.362]

The submitters determined the crystalline hydroxy esters to be >99.9% diastereomerically pure by supercritical fluid chromatography (EMdiol silica column and a Chiralcel (+) OD-(H) column (Chiral... [Pg.95]

Naphthalenedisulfonate-acetonitrile as the only mobile phase with a silica column coated with a crosslinked aminofluorocarbon polymer has proven to be an effective combination for the separation of aliphatic anionic surfactants. Indirect conductivity and photometric detection modes are used to monitor these analytes. The retention of these surfactants is found to depend on both the ionic strength and the organic solvent content of the mobile phase. The mechanism of retention is considered to be a combination of both reverse phase and ion exchange processes. Selective separation of both alkanesulfonates and... [Pg.168]

Gillespie AM, Walters SM. 1986. HPLC silica column fractionation of pesticides and PCB from butterfat. J Liq Chromatogr 9 2111-2142. [Pg.209]

Liver Addition of water to sample followed by homogenization extraction with benzene, clean-up on silica column and HPLC GC/ECD No data No data Demeter and Heyndrickx 1979... [Pg.252]

The catalyst testing was carried out in a gas phase downflow stainless steel tubular reactor with on-line gas analysis using a Model 5890 Hewlett-Packard gas chromatograph (GC) equipped with heated in-line automated Valeo sampling valves and a CP-sD 5 or CP-sil 13 capillary WCOT colunm. GC/MS analyses of condensable products, especially with respect to O-isotopic distribution, was also carried out using a CP-sil 13 capillary column. For analysis of chiral compounds, a Chirasil-CD capillary fused silica column was employed. [Pg.602]


See other pages where Silicas columns is mentioned: [Pg.55]    [Pg.522]    [Pg.4]    [Pg.84]    [Pg.221]    [Pg.221]    [Pg.240]    [Pg.263]    [Pg.306]    [Pg.314]    [Pg.324]    [Pg.329]    [Pg.391]    [Pg.394]    [Pg.98]    [Pg.171]    [Pg.272]    [Pg.292]    [Pg.1179]    [Pg.1179]    [Pg.318]    [Pg.445]    [Pg.242]    [Pg.456]    [Pg.410]   
See also in sourсe #XX -- [ Pg.250 ]




SEARCH



Acclaim Mixed-Mode silica columns

Alkyl-silica columns

Aluminum-clad fused-silica capillary columns

Box 25-1 Monolithic Silica Columns

Capacity fused-silica capillary columns

Capillary column technology fused-silica

Capillary columns monolith silica

Coating fused-silica capillary columns

Column chromatography silica-gel

Column of silica gel

Column packings silica

Column packings silica morphology

Column silica-based

Columns silica gel

Cyclodextrin-silica columns, displacement

Deactivation fused-silica capillary columns

Extrusion of a Fused-Silica Capillary Column

Fused silica capillary columns

Fused silica column micro HPLC

Fused silica column resolution enhancement

Fused silica columns

Fused-silica capillary columns characteristics

Fused-silica capillary columns polyimide-clad

Fused-silica capillary columns popularity

Fused-silica capillary columns preparation

Fused-silica capillary columns stainless steel

Fused-silica capillary columns synthetic

Fused-silica capillary columns tubing

Fused-silica packed columns

Fused-silica-lined stainless-steel capillary columns

Hybrid organo-silica columns

Inert-fused silica capillary columns

Microparticulate silica column

Monolithic porous silica columns

Monolithic porous silica columns porosity

Monolithic silica columns

Nonporous silica column

Normal-phase liquid chromatography silica column

Open column silica

Open tubular column fused silica

Organo-silica hybrid monolithic columns for

Packed fused silica capillary columns

Porosity monolith silica columns

Preparation of Fused-Silica Capillary Columns

Protective coatings, fused-silica columns

Retention time fused-silica capillary columns

Reverse phase chromatography silica based columns

Reversed-phased silica gel column chromatography cleanup

Selectivity fused-silica capillary columns

Separation efficiency porous silica columns

Separator Acclaim Mixed-Mode silica columns

Silica capillary column for

Silica capillary columns

Silica column chromatography

Silica column chromatography general methods

Silica columns chromatographic separations

Silica columns samples

Silica gel column chromatograph

Silica gel flash column chromatography

Silica saturator columns

Silica supports column packing structure

Silica-Based Mixed-Mode Columns

Silica-based SEC column

Silica-based stationary phases monolithic columns

Stationary phase fused-silica capillary columns

Structural Properties of Monolithic Silica Columns

Temperature fused-silica capillary columns

Tubing fused-silica columns

Water-column responses silica

© 2024 chempedia.info