Fused silica open tubular columns

Standards with a poor UV chromophore (4-methyl-2-pentanone, 1-chlorododecane, and stearic acid) were analyzed on a Perkin-Elmer 3920 gas chromatograph with a flame ionization detector and a 30-m SE-54 wall-coated open tubular (WCOT) fused-silica capillary column (J W Scientific). The injector temperature was 200 °C the detector interface temperature was 280 °C. The carrier gas was He at 16.5 lb/in.2, and the makeup gas was nitrogen at a flow of 40 cm3/min. Splitless l-/zL injections of the 25,000 1 concentrates were made by starting with an oven temperature of 45 °C and the oven door open after 2.75 min, the oven door was closed and the temperature was programmed at 4 °C/min to 280 °C, which was held for 8 min. The syringe was kept in the injection port for 15 s after injection. [Pg.547]

Gas chromatography (GC) and gas chromatography-mass spectrometry (GC-MS) analyses GC analyses are performed on a WCOT (wall-coated open tubular) polydimethylsiloxane fused silica capillary column (1 fim film thickness, 50m x 0 32mm ID ) Shorter columns for example 25m, or columns with thinner films, for example 0 2/im, may also be used successfully Injection of the silylated sample (2/d) is performed with a moving needle-type injector, other types of capillary GC injection systems may be used after optimization of the concentration and amount of sample to be injected With a 50-m capillary... [Pg.337]

The extracted degradation products were identified and quantified by a ThermoFinnigan GCQ (San Jose, CA, USA). The column used a wall-coated open tubular (WCOT) fused silica CP-WAX 58 (FFAP)-CB column from Varian (25 m x 0.32 mm id, od 0.2 pm). Helium of scientific grade purity from AGA (Stockholm, Sweden) was used as carrier gas at the constant velocity of 40 cm/s. The initial oven temperature was 40 "C, which was held for 1 minute. The oven was heated to 250 C for 15 minutes. Electron impact mode (El) detection was used with an electron energy of 70 eV. The mass-range scanned was 35-400 m/z and the ion source and transfer line temperatures were 180 and 250 C, respectively. The injector temperature was set to 250 C. A sample (1 pi) was injected in splitless injection mode and two blanks were run between each sample by injecting clean methanol (0.15 M HCl). [Pg.215]

Capillary Columns Capillary, or open tubular columns are constructed from fused silica coated with a protective polymer. Columns may be up to 100 m in length with an internal diameter of approximately 150-300 )J,m (Figure 12.17). Larger bore columns of 530 )J,m, called megabore columns, also are available. [Pg.564]

Restek Rtx-IMS fused-silica open-tubular column (FSOT) with integral guard column, 30 m X 0.25-mm i.d. and 0.25-p.m film thickness Eppendorf fixed-volume pipets, 0.50-mL Eppendorf fixed-volume pipefs, 1.0-mL Eppendorf pipel lips, 1.0-mL Amber-glass boftles wifti Teflon-lined caps, 4-oz... [Pg.370]

Adsorption or catalytic decomposition of labile substances by the syringe needle can be a problem for some compounds using hot vaporizing injectors [25]. For open tubular columns deactivated fused silica syringe needles and cold on-column Injection techniques are used to minimize this problem. Alternatively, syringes fitted with a needle shroud for cold-needle injection can be used [26]. [Pg.125]

Figure 2.15 Separation of a mixture of saturated and unsaturated hydrocarbons on a 50 m x 0.32 mm fused silica porous-layer open tubular column coated with alumina modified wl potassium chloride. The separation was performed by temperature programming from 70 to 200°C at 3 C/mln. (Reproduced with permission fr Figure 2.15 Separation of a mixture of saturated and unsaturated hydrocarbons on a 50 m x 0.32 mm fused silica porous-layer open tubular column coated with alumina modified wl potassium chloride. The separation was performed by temperature programming from 70 to 200°C at 3 C/mln. (Reproduced with permission fr<ni ref 430. Copyright Preston Publications, Inc.)...
Virtually all current research in SFC utilizes either small bore packed columns with particles of 5-10 micrometers in diameter optimized for use in liquid chromatography or narrow bore, fused silica open tubular columns with Immobilized phases similar to those used in gas chromatography. In the latter case columns of saaller internal diameter, 10-100 micrometers, shorter lengths (generally less than 20 m with 1-10 m being the most common length), and more firmly crosslinked stationary phases are used by coaparison with standard columns for gas chromatography. In all... [Pg.819]

An open-tubular column is a capillary bonded with a wall-supported stationary phase that can be a coated polymer, bonded molecular monolayer, or a synthesized porous layer network. The inner diameters of open-tubular CEC columns should be less than 25 pm that is less than the inner diameters of packed columns. The surface area of fused silica tubing is much less than that of porous packing materials. As a result, the phase ratio and, hence, the sample capacity for open-tubular columns are much less than those for packed columns. The small sample capacity makes it difficult to detect trace analytes. [Pg.451]

Packed columns are conventional stainless steel columns that contain small deactivated particles to which the stationary phases adheres. Capillary columns are open tubular columns of narrow internal diameter made of fused silica, with... [Pg.677]

The vast majority of analyses use long, narrow open tubular columns (Figure 24-2) made of fused silica (Si02) and coated with polyimide (a plastic capable of withstanding 350°C) for support and protection from atmospheric moisture.3 As discussed in Section 23-5, open tubular columns offer higher resolution, shorter analysis time, and greater sensitivity than packed columns, hut have less sample capacity. [Pg.529]

Figure 24-3 (a) Porous carbon stationary phase (2 p,m thick) on inside wall of fused silica open tubular column. (fc>) Chromatogram of vapors from the headspace of a beer can. obtained with 0.25-mm-diameter x 30-m-long porous carbon column operated at 30°C for 2 min and then ramped up to 160°C at 20°/min. [Courtesy Alltech Associates, State College, PA.]... [Pg.530]

Gas Chromatograph -- A Varian 6000 equipped with two constant-current/pulsed-frequency electron capture detectors, a 30-m x 0.53-mm ID DB-5 fused-silica open-tubular column (1.5-/xm film thickness), and a 30-m x 0.53-mm ID DB-1701 fused-silica open-tubular column (1.0-/im film thickness), both connected to a press-fit Y-shaped fused-silica inlet splitter (Restek Corporation, Bellefonte, Pennsylvania), was used to analyze for the nitroaromatic compounds. The columns were temperature-programmed from 120°C (1.0-min hold) to 200°C (1-min hold) at 3°C/min, then to 250°C (4-min hold) at 8°C/min injector temperature 250°C detector temperature 320°C helium carrier gas 6 mL/min nitrogen makeup gas 20 mL/min. [Pg.187]

High resolution fused silica capillary columns were used to examine, and in some cases to compare, flavor and aroma essences obtained by 1) direct headspace injection, 2) simultaneous steam distillation extraction, 3) standard Soxhlet extraction using dichloromethane, and 4) high pressure Soxhlet extraction using liquid CO2 Developments in smaller- and larger-diameter fused silica open tubular columns, and columns coated with a bonded polyethylene glycol stationary phase were also explored. [Pg.95]

Figure 9. Results of the direct substitution of a large-bore fused silica open tubular column. Solutes (in order of elution) 4-chlorophenol, dodecane, 1-decylamine, 1-undecanol,... Figure 9. Results of the direct substitution of a large-bore fused silica open tubular column. Solutes (in order of elution) 4-chlorophenol, dodecane, 1-decylamine, 1-undecanol,...

See other pages where Fused silica open tubular columns is mentioned: [Pg.142]    [Pg.801]    [Pg.935]    [Pg.142]    [Pg.801]    [Pg.935]    [Pg.351]    [Pg.329]    [Pg.37]    [Pg.72]    [Pg.308]    [Pg.326]    [Pg.406]    [Pg.420]    [Pg.464]    [Pg.696]    [Pg.819]    [Pg.827]    [Pg.997]    [Pg.1001]    [Pg.392]    [Pg.451]    [Pg.551]    [Pg.187]    [Pg.329]    [Pg.291]    [Pg.389]    [Pg.255]    [Pg.124]    [Pg.127]    [Pg.691]    [Pg.127]    [Pg.384]    [Pg.256]    [Pg.702]    [Pg.724]   