Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mode of operation,

Most Mossbauer spectroscopy experiments are conducted either in the transmission mode, in which a source of well-defined characteristics is used to examine the spectral properties of an unknown absorber, or in the emission mode, in which the source of radiation becomes the sample under investigation and a known or standard absorber is employed to determine the transition energy differences. In both cases, any of a number of y-ray detectors is utilized to record the amount of radiation transmitted through the absorber. [Pg.406]

An alternate way of acquiring Mossbauer spectra consists in measuring events associated with the deexcitation of nuclei in the sample following the absorption of y-rays. Such a process involves the emission of y-rays. X-rays, or electrons, which are most often detected in a backscattering geometry. The [Pg.406]

TABLE 1. Probability of Relaxation Products from Excited Fe (1=1) [Pg.407]

TABLE II. Backscattering Detection in Mossbauer Spectroscopy Measurements [Pg.408]

Another mechanism by which the measured Fe charged state may be different from that of the original parent Co involves an actual electron transfer reaction in which other host lattice species serve as either donor or acceptor sites. The difficulties in the interpretation of emission Mossbauer spectra introduced by chemical and after effects could be diminished to some extent by conducting detailed studies of the behavior of known solids. [Pg.408]

The electron microscope can be operated to form on the fluorescent screen either (i) a diffraction pattern of the specimen or (ii) one of several types of image, depending on which beams are allowed to pass through the aperture in the back focal plane of the objective lens. The geometry of the electron diffraction pattern will be discussed in detail in Chapter 3. [Pg.50]

The most commonly employed imaging mode is bright field (BF), which uses only the central beam, usually called the transmitted beam. Image contrast is observed if, for whatever reason, the diffracted beams originating from different parts of the specimen vary in intensity. If a crystalline specimen is so oriented with respect to the incident beam that many strong diffracted beams are excited, then an extremely complex BF image [Pg.50]

The detailed interpretation of electron microscope images produced using any of the operating modes discussed in this chapter requires as complete an understanding as possible of the diffraction process. The next two chapters develop and explain as simply as possible the current theories of electron diffraction by crystals in order to provide a basis for the interpretation of images of crystal defects (such as dislocations, stacking faults, and twins) and of lattice images. [Pg.51]

At present the ESA Coulochem is the only commercially available coulometric detector. Several different types of cell are now available, most of which contain two analytical electrodes in series. The standard cell (Model 5010) contains equal sized PGEs, but in the high sensitivity cell (Model 5011) the surface area of the second electrode is reduced with the aim of producing a better S/N ratio. Despite the argument that the increased surface area available in PGE systems only increases the noise in proportion to the increased signal, such systems do appear to give an enhanced S/N ratio in certain applications, as compared to glassy carbon electrodes in thin-layer or wall-jet assemblies, and may offer more flexibility in routine use. Even better S/N ratios are claimed for the Model 5014 cell, which has been developed specifically for use with microdialysis samples. [Pg.37]

Coulochem cells with dual PGEs can be operated in three modes, often referred to as screen , difference and redox .In screen mode the potential of the first electrode (Ei) in the analytical cell is set, typically, 0.2-0.3 V below that of [Pg.38]

A further way of reducing interference is to use difference mode. The Coulochem has three outputs those for the two electrodes, Ei and E2, and the difference (Ej minus E2). If the electrodes are set to similar potentials, such that the analyte of interest is oxidised at the first electrode with little or no signal at the second, then the difference will contain a signal due to the analyte. If impurities produce similar [Pg.39]

A specialised form of redox mode is when the conditioning electrode is used to produce species that are then detected at the analytical electrode(s). The conditioning cell was developed primarily for catecholamine analysis and, like the analytical cells, is designed to minimise peak broadening. The guard cell, on the other hand, is [Pg.40]


In the standard method, the metal enclosure (called the air chamber) used to hold the hydrocarbon vapors is immersed in water before the test, then drained but not dried. This mode of operation, often designated as the wet bomb" is stipulated for all materials that are exclusively petroleum. But if the fuels contain alcohols or other organic products soluble in water, the apparatus must be dried in order that the vapors are not absorbed by the water on the walls. This technique is called the dry bomb" it results in RVP values higher by about 100 mbar for some oxygenated motor fuels. When examining the numerical results, it is thus important to know the technique employed. In any case, the dry bomb method is preferred. [Pg.189]

The UK government enquiry into the Piper Alpha disaster in the North Sea in 1988 has had a significant impact on working practices and equipment and has helped to improve offshore safety around the world. One result has been the development of a Safety Management System (SMS) which is a method of integrating work practices, and is a form of quality management system. Major oil companies have each developed their own specific SMS, to suit local environments and modes of operation, but the SMS typically addresses the following areas (recommended by the Cullen Enquiry into the Piper Alpha disaster) ... [Pg.68]

The operations group will develop general operating and maintenance objectives for the facilities which will address product quality, costs, safety and environmental issues. At a more detailed level, the mode of operations and maintenance tor a particular project will be specified in the field development plan. Both specifications will be discussed in this section, which will focus on the input of the production operations and maintenance departments to a field development plan. The management of the field during the producing period is discussed in Section 14.0. [Pg.278]

The production operations and maintenance group will develop a set of operating and maintenance objectives for the project. This will be a guideline when specifying the mode of operation and maintenance of the equipment items and systems, and will incorporate elements of... [Pg.278]

One of the primary objectives of production operations is to deliver product at the required rate and quality. Therefore the product quality specification and any agreed contract terms will drive the activities of the production operations department, and will be a starting point for determining the preferred mode of operation. The specifications, such as delivery of stabilised crude with a BS W of less than 0.5%, and a salinity of 70 g/m, ... [Pg.279]

The service provided by maintenance engineering was traditionally that of repairing equipment items when they failed. This is no longer the case, and a maintenance department is now pro-active rather than reactive in its approach. Maintenance of equipment items will be an important consideration in the FDP, because the mode and cost of maintaining equipment plays an important part in the facilities design and in the mode of operation. [Pg.286]

AFM measures the spatial distribution of the forces between an ultrafme tip and the sample. This distribution of these forces is also highly correlated with the atomic structure. STM is able to image many semiconductor and metal surfaces with atomic resolution. AFM is necessary for insulating materials, however, as electron conduction is required for STM in order to achieve tiumelling. Note that there are many modes of operation for these instruments, and many variations in use. In addition, there are other types of scaiming probe microscopies under development. [Pg.310]

As noted earlier, most electron diffraction studies are perfonned in a mode of operation of a transmission electron microscope. The electrons are emitted themiionically from a hot cathode and accelerated by the electric field of a conventional electron gun. Because of the very strong interactions between electrons and matter, significant diffracted intensities can also be observed from the molecules of a gas. Again, the source of electrons is a conventional electron gun. [Pg.1379]

The original method employed was to scan eitiier the frequency of the exciting oscillator or to scan the applied magnetic field until resonant absorption occiined. Flowever, compared to simultaneous excitation of a wide range of frequencies by a short RF pulse, the scanned approach is a very time-inefficient way of recording the spectrum. Flence, with the advent of computers that could be dedicated to spectrometers and efficient Fourier transfomi (FT) algoritluns, pulsed FT NMR became the nomial mode of operation. [Pg.1470]

Figure Bl.19.1. Principle of operation of a scanning tiimrelling microscope. The x- andj -piezodrives scan the tip across the surface. In one possible mode of operation, the current from the tip is fed into a feedback loop that controls the voltage to die z-piezo, to maintam constant current. The Ime labelled z-displacement shows the tip reacting both to morphological and chemical (i.e. electronic) inliomogeneities. (Taken from [213].)... Figure Bl.19.1. Principle of operation of a scanning tiimrelling microscope. The x- andj -piezodrives scan the tip across the surface. In one possible mode of operation, the current from the tip is fed into a feedback loop that controls the voltage to die z-piezo, to maintam constant current. The Ime labelled z-displacement shows the tip reacting both to morphological and chemical (i.e. electronic) inliomogeneities. (Taken from [213].)...
Figure Bl.19.2. The two modes of operation for scanning tunnelling microscopes (a) constant current and (b) constant height. (Taken from [214], figure 1.)... Figure Bl.19.2. The two modes of operation for scanning tunnelling microscopes (a) constant current and (b) constant height. (Taken from [214], figure 1.)...
Laser type Lasing wavelength >. Efficiency r (%) Mode of operation Typical output power... [Pg.2862]

A regenerative fuel cell system can also be a single electrochemical cell in which both the oxidation of fuels (i.e., production of electric power) and reduction of CO2 (to obtain fuels) can be carried out by simply reversing the mode of operation. [Pg.220]

Many designs of nebulizer are commonly used in ICP/MS, but their construction and mode of operation can be collated into a small number of groups pneumatic, ultrasonic, thermospray, APCI, and electrospray. These different types are discussed in the following sections, which are followed by further sections on spray and desolvation chambers. [Pg.139]

Quadmpole mass spectrometers (mass filters) allow ions at each m/z value to pass through the analyzer sequentially. For example, ions at m/z 100, 101, and 102 are allowed to pass one after the other through the quadmpole assembly so that first m/z 100 is transmitted, then m/z 101, then m/z 102, and so on. Therefore, the ion collector at the end of the quadmpole unit needs to cover only one point or focus in space and can be placed immediately behind the analyzer (Figure 30.1). A complete mass spectram is recorded over a period of time (temporally), which is set by the voltages on the quadmpole analyzer. In this mode of operation, the ions are said to be scanned sequentially. The resolution of m/z values is dependent solely on the analyzer and not on the detector. The single-point collector is discussed in detail in Chapter 28. [Pg.211]

To differentiate tteir functions and modes of operation, the array collector of spatially dispersed m/z values is still called an array collector for historical reasons, but the other multipoint detector of a temporally dispersed range of m/z values is called a microchannel plate (typically used in time-of-flight instruments). [Pg.410]

A computer attached to a mass spectrometer is used both to acquire data and to control the operation of the spectrometer. Powerful transputer systems can be used to ensure that both modes of operation can be carried out almost simultaneously. [Pg.421]

The constant height mode of operation results in a faster measurement. In this analysis, the tip height is maintained at a constant level above the surface and differences in tunneling current ate measured as the tip is scaimed across the surface. This approach is not as sensitive to surface irregularities as the constant current mode, but it does work well for relatively smooth surfaces. [Pg.273]

In most electron spectroscopic analyses, the kinetic energies of the electrons entering the analyzer are retarded to either a constant energy or by a constant factor. These approaches lead to two modes of operation the constant analyzer energy (CAE) mode and the constant retard ratio (CRR) mode. [Pg.283]

The CRR mode involves retarding the electron kinetic energies to a constant ratio of H /H where H is the energy passed by the analyzer. Thus, the energies are retarded by a constant factor. Spectra acquired in this mode ate less easy to quantify, but small peaks at low kinetic energies ate readily detected. This mode of operation results in spectra of constant relative resolution throughout. The relative resolution is improved in this mode by a factor of E. ... [Pg.283]

These modes of operation ate used in conjunction with the two most popular energy analyzers, the cylindrical mirror analyzer (CMA) and the concentric hemispherical analyzer (CHA). The most common form of the CMA used today is the double-pass version diagramed in Eigute 21. This device consists of two perfectly coaxial cylinders of radii r and r. The outer cylinder is held at a potential of (— ) and the inner cylinder is held at ground. The... [Pg.283]

In aes, the resolution is largely independent of the characteristics of the analy2er or source and is dictated by the natural linewidth of the Auger line (usually several eV). Therefore, in using a CMA for aes, the analyst is more concerned with transmission (and hence, sensitivity) than with resolution. In contrast to xps, the optimhation of variables is achieved for aes in the CRR mode of operation. The large transmission of the CMA relative to the CHA make it the more desirable analy2er for aes. [Pg.284]

Fig. 17. The two basic modes of operation for an adsorption process (a) cycHc batch system (b) continuous countercurrent system with adsorbent... Fig. 17. The two basic modes of operation for an adsorption process (a) cycHc batch system (b) continuous countercurrent system with adsorbent...
Addition of Inert Filter Aids. FUtet aids ate rigid, porous, and highly permeable powders added to feed suspensions to extend the appheabUity of surface filtration. Very dilute or very fine and slimy suspensions ate too difficult to filter by cake filtration due to fast pressure build-up and medium blinding addition of filter aids can alleviate such problems. Filter aids can be used in either or both of two modes of operation, ie, to form a precoat which then acts as a filter medium on a coarse support material called a septum, or to be mixed with the feed suspension as body feed to increase the permeabihty of the resulting cake. [Pg.389]

Two modes of operation are available. In the simpler mode, a fixed current is appHed to the wire. In the other mode, the sensor temperature (hence is maintained at a fixed value above that of the gas by means of a feedback amplifier. In both cases, velocity is measured by the voltage drop. [Pg.110]

Because of its small size and portabiHty, the hot-wire anemometer is ideally suited to measure gas velocities either continuously or on a troubleshooting basis in systems where excess pressure drop cannot be tolerated. Furnaces, smokestacks, electrostatic precipitators, and air ducts are typical areas of appHcation. Its fast response to velocity or temperature fluctuations in the surrounding gas makes it particularly useful in studying the turbulence characteristics and rapidity of mixing in gas streams. The constant current mode of operation has a wide frequency response and relatively lower noise level, provided a sufficiently small wire can be used. Where a more mgged wire is required, the constant temperature mode is employed because of its insensitivity to sensor heat capacity. In Hquids, hot-film sensors are employed instead of wires. The sensor consists of a thin metallic film mounted on the surface of a thermally and electrically insulated probe. [Pg.110]


See other pages where Mode of operation, is mentioned: [Pg.280]    [Pg.280]    [Pg.281]    [Pg.1348]    [Pg.1687]    [Pg.1941]    [Pg.2083]    [Pg.2861]    [Pg.172]    [Pg.112]    [Pg.156]    [Pg.158]    [Pg.160]    [Pg.160]    [Pg.164]    [Pg.211]    [Pg.77]    [Pg.273]    [Pg.274]    [Pg.284]    [Pg.295]    [Pg.165]    [Pg.230]    [Pg.529]   
See also in sourсe #XX -- [ Pg.234 ]

See also in sourсe #XX -- [ Pg.66 , Pg.67 , Pg.68 , Pg.70 ]




SEARCH



Batch mode of operation

Choice of operation mode

Continuous mode of operation

Cycling Mode of Operation

Determination of Operating Conditions in Batch Mode

Discontinuous-mode of operation

Fed-batch mode of operation

Feedback mode of SECM operation

Illustrative Examples Using the Zero Effluent Mode of Operation

Mode of operation for

Modes of Fermentor Operation

Modes of Operation for LC-MS-NMR

Modes of SECM Operation

New Modes of Operation

New Operating Modes of Production

OCM Reactors and Modes of Operation

Operating Modes of Calorimeters

Operating mode

Operating modes of the nuclear power unit

Operation Modes of STM

Operation mode

Operation modes mode

Range of single-mode operation

Reactor configurations and modes of operation

Selection of Most Suitable Mode for MS Operation

Selection of the operation mode

Single mode operation, of laser

System Modes of operation

Tasks and Modes of Operation

Various Operation Modes of AFM

© 2024 chempedia.info