Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standard absorbance

Thermogravimetry is a valuable technique for the assessment of the purity of materials. Analytical reagents, especially those used in titrimetric analysis as primary standards, e.g. sodium carbonate, sodium tetraborate, and potassium hydrogenphthalate, have been examined. Many primary standards absorb appreciable amounts of water when exposed to moist atmospheres. TG data can show the extent of this absorption and hence the most suitable drying temperature for a given reagent may be determined. [Pg.432]

Cahbration spectra must be measured at defined temperamres (ambient temperature for a-iron) because of the influence of second-order Doppler shift (see Sect. 4.2.1) for the standard absorber. After folding, the experimental spectrum should be simulated with Lorentzian lines to obtain the exact line positions in units of channel numbers which for calibration can be related to the hteramre values of the hyperfine splitting. As shown in Fig. 3.4, the velocity increment per channel, Ostep, is then obtained from the equation Ustep = D,(mm s )/D,(channel numbers). Different... [Pg.31]

The calibration procedure is sufficiently accurate for Doppler velocities in the regime 0 to 10 mm s beyond this range, laser calibration is more suitable. Calibration with ot-iron, as described, can also be used for Mossbauer measurements with other isotopes, e.g., Ni, Au, and Ir, for which suitable standard absorbers are not available (provided that the Doppler velocity range of interest is not significantly greater than 10 mm s ). This, of course, requires that the spectrometer is temporarily equipped with a Co source and an a-iron absorber. [Pg.32]

The hematite with adsorbed Co-57 or Sb-119 along with the solution was subjected to emission Mossbauer measurement at 24 1°C with the experimental setup shown in Figure 2. The absorber, Fe-57-enriched potassium ferrocyanide (0.5 mg Fe-57/cm2) or barium stannate (0.9 mg Sn-119/cm2), was driven by a Hanger 700-series Mossbauer spectrometer connected to a Tracor-Northern TN-7200 multi-channel analyzer. The Mosssbauer gamma-rays of Co-57 and Sb-119 were detected respectively with a Kr(+3% carbon dioxide)-filled proportional counter and with a 2 mm-thick Nal(Tl) scintillation counter through 65 pm-thick Pd critical absorber for Sn K X-rays. The integral errors in the relative velocity were estimated to be of the order of 0.05 mm/s by repeated calibration measurements using standard absorbers. [Pg.406]

JJor chemists interested in modem theories of chemical bonding, the most useful data obtainable by the Mossbauer technique are the magnitude and sign of the electric quadrupole field gradient tensor and the magnitude of the shift, 8, (which we prefer to call the chemical isomeric. Cl, shift), of the center of the Mossbauer spectrum relative to some standard absorber. Although a considerable amount of chemical and structural information is potentially available from quadrupole data on iron compounds, relatively little use has been made of such data in the literature, and we will not discuss this parameter here. We will instead restrict ourselves to two main points review of the explanations put forth to explain Cl shift data in iron compounds, and a survey of some of the correlations and generalizations which have been found. [Pg.86]

We close with a plea to investigators to report shift data with respect to a standard absorber, either clean iron foil or sodium nitroprusside or both. The logic behind this request is obvious since anyone who has tried to intercompare results between different papers, knows how frustrating it can be when no common substances were studied. We also suggest that inherently narrow line sources be used—i.e., Cu, Pt, Pd, or Cr. Stainless steel sources are usually so broad that they can often mask important features of the spectrum. Of the host matrices mentioned we prefer Cu since it doesn t give rise to any interfering x-rays and when prepared properly doesn t show any broadening at low temperatures. [Pg.103]

Note Tests have to be performed for each acceptance criterion. If several acceptance criteria have to be met to get overall acceptance (e.g., system suitability test relative standard deviation of standard absorbance and blank sample with no absorption), the overall functionality has to be checked. [Pg.287]

The blank absorbance was 0.038 at 562 nm in a 1.000-cm cell. A serum sample had an absorbance of 0.129. After the blank was subtracted from each standard absorbance, the points in Figure 18-8 were obtained. The least-squares line through the standard points is... [Pg.387]

Figure 6. Elution profile obtained on examination of the gel permeation behavior of 14C acetylated S. aureus Factor III. Cytochrome-c and fluorescein labeled E. coli HPr were included in the applied sample as internal standards. Absorbance at 410 nm (cytochrome-c), fluorescein fluorescence (HPr), and radioactivity (Factor III) are plotted vs. fraction number. Inset shows the elution profiles for the external (blue dextran) and internal (fiuoroglycine) column markers. Data in this figure are from unpublished work (20). Figure 6. Elution profile obtained on examination of the gel permeation behavior of 14C acetylated S. aureus Factor III. Cytochrome-c and fluorescein labeled E. coli HPr were included in the applied sample as internal standards. Absorbance at 410 nm (cytochrome-c), fluorescein fluorescence (HPr), and radioactivity (Factor III) are plotted vs. fraction number. Inset shows the elution profiles for the external (blue dextran) and internal (fiuoroglycine) column markers. Data in this figure are from unpublished work (20).
Adsorbent activity function, proportional to adsorbent surface energy a = I for standard" absorbent Parameters defined by Hammers et al. (S5, 36) which measure importance of delocalization due to intramolecular effects (/3) or to site competion (y)... [Pg.221]

The method-development laboratory is olfen set with a task of identifying unknown constituents in a complex sample. The identity or quantity of analytes is not well defined at this stage. Complicated sample-preparation steps and multiple HPLC analyses may be required to ascertain the nature of a sample. The diode array detector simplifies this process. In the case that more than 15 different peaks have been separated in a analysis of a multicomponent sample of known origin more than a full day s work might be required just to establish the retention times of the unknowns with the use of a standard absorbance detector. A number of long-run HPLC analyses would have to be made with pure standards so that retention time correlations could be determined. This necessity is accomplished by the DAD s advantage of acquiring data in both the time and spectral domain. [Pg.1114]

Fluorescence. The use of molecular fluorescence spectroscopy for the quantitation of enzyme reaction products has resulted in detection limits that are several orders of magnitude lower than those achieved by standard absorbance methods. At low analyte concentrations, fluorescence emission intensity is directly proportional to concentration, and its value depends on both the molar absorptivity of the analyte at the excitation wavelength, and the fluorescence quantum yield of the analyte, under the assay conditions. [Pg.49]

Absorbance of sample - Absorbance of sample blank Absorbance of standard - Absorbance of standard... [Pg.114]

In the solid state, the drugs are radio-resistant until the Standard Absorbed Dose (SAD) (25 kGy) but some changes in the organoleptic properties may be reported at this dose by decreasing the doses, some discoloration and/or smell may be reduced. [Pg.155]

Characterization of the electronic configuration of the resonant ions. The oxidation and spin state of the iron ion determine the isomer shift (IS) which depends on the charge density at the nucleus. It is the shift of the centre of gravity of the spectrum, generally quoted with respect to a standard absorber of iron metal. Typical values are 0.2 to 0 5 mm/sec for high spin Fe3+, 0.8 to 1.5 mm/sec for high spin Fe2 and -0.2 to 0.3 mm/sec for low spin Fe2 (9). [Pg.172]

The sample was a 0.0127 mm thick iron foil, of 99.99% purity, cut into disks (2.5 cm diameter). Room temperature Mossbauer spectra were taken in the initial state, after treatment at 800°C for 30 minutes in 1 atm. argon saturated with water at room temperature, and finally after reaction in 1 atm. acetylene at 750°C for 0.5 minute. Velocities were calibrated using the known Mossbauer parameters of metallic iron at 25°C, and zero velocity is with respect to this standard absorber. [Pg.12]

Table 3.13 Initial, standard, absorbance spectra and data regenerated by TTFA analysis of the mixtures in Table 3.11... Table 3.13 Initial, standard, absorbance spectra and data regenerated by TTFA analysis of the mixtures in Table 3.11...
The standardized absorbance value of each measured spectrum (i = l,...,n) at each spectral wavelength (j = l,...,p) is calculated by using... [Pg.1036]


See other pages where Standard absorbance is mentioned: [Pg.23]    [Pg.440]    [Pg.134]    [Pg.31]    [Pg.404]    [Pg.19]    [Pg.9]    [Pg.87]    [Pg.440]    [Pg.5]    [Pg.23]    [Pg.308]    [Pg.37]    [Pg.296]    [Pg.141]    [Pg.480]    [Pg.23]    [Pg.513]    [Pg.383]    [Pg.440]    [Pg.324]    [Pg.479]    [Pg.533]    [Pg.358]    [Pg.198]    [Pg.299]    [Pg.296]    [Pg.109]    [Pg.1433]    [Pg.323]    [Pg.341]   
See also in sourсe #XX -- [ Pg.294 ]




SEARCH



Absorbance standard addition

Absorbancy, standard values

© 2024 chempedia.info