Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Ion cyclotron

The most widely used type of trap for the study of ion-molecule reactivity is the ion-cyclotron-resonance (ICR) [99] mass spectrometer and its successor, the Fourier-transfomi mass spectrometer (FTMS) [100, 101]. Figure A3.5.8 shows the cubic trapping cell used in many FTMS instmments [101]. Ions are created in or injected into a cubic cell in a vacuum of 10 Pa or lower. A magnetic field, B, confines the motion in the x-y... [Pg.810]

The chapter is divided into sections, one for each general class of mass spectrometer magnetic sector, quadnipole, time-of-flight and ion cyclotron resonance. The experiments perfonned by each are quite often unique and so have been discussed separately under each heading. [Pg.1329]

B1.7.6 FOURIER TRANSFORM ION CYCLOTRON RESONANCE MASS SPECTROMETERS... [Pg.1354]

Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s). Figure Bl.7.18. (a) Schematic diagram of the trapping cell in an ion cyclotron resonance mass spectrometer excitation plates (E) detector plates (D) trapping plates (T). (b) The magnetron motion due to tire crossing of the magnetic and electric trapping fields is superimposed on the circular cyclotron motion aj taken up by the ions in the magnetic field. Excitation of the cyclotron frequency results in an image current being detected by the detector electrodes which can be Fourier transfonned into a secular frequency related to the m/z ratio of the trapped ion(s).
Mclver R T 1970 A trapped ion analyzer cell for ion cyclotron resonance spectroscopy Rev. Sc/. Instrum. 41 555-8... [Pg.1360]

Vartanian V H, Anderson J S and Laude D A 1995 Advances in trapped ion cells for Fourier transform ion cyclotron resonance mass spectrometry Mass Spec. Rev. 41 1-19... [Pg.1360]

Comisarow M B and Marshall A G 1996 Early development of Fourier transform ion cyclotron resonance (FT-ICR) spectroscopy J. Mass Spectrom. 31 581-5... [Pg.1360]

Grover R, Decouzon M, Maria P-C and Gal J-F 1996 Reliability of Fourier transform-ion cyclotron resonance determinations of rate constants for ion/molecule reactions Eur. Mass Spectrom. 2 213-23... [Pg.1360]

Fisher J J and McMahon T B 1990 Determination of rate constants for low pressure association reactions by Fourier transform-ion cyclotron resonance Int. J. Mass Spectrom. Ion. Proc 100 707-17... [Pg.1360]

The reactivity of size-selected transition-metal cluster ions has been studied witli various types of mass spectrometric teclmiques [1 ]. Fourier-transfonn ion cyclotron resonance (FT-ICR) is a particularly powerful teclmique in which a cluster ion can be stored and cooled before experimentation. Thus, multiple reaction steps can be followed in FT-ICR, in addition to its high sensitivity and mass resolution. Many chemical reaction studies of transition-metal clusters witli simple reactants and hydrocarbons have been carried out using FT-ICR [49, 58]. [Pg.2394]

Other types of mass spectrometer may use point, array, or both types of collector. The time-of-flight (TOF) instrument uses a special multichannel plate collector an ion trap can record ion arrivals either sequentially in time or all at once a Fourier-transform ion cyclotron resonance (FTICR) instrument can record ion arrivals in either time or frequency domains which are interconvertible (by the Fourier-transform technique). [Pg.201]

In modem mass spectrometry, ion collectors (detectors) are generally based on the electron multiplier and can be separated into two classes those that detect the arrival of all ions sequentially at a point (a single-point ion collector) and those that detect the arrival of all ions simultaneously (an array or multipoint collector). This chapter compares the uses of single- and multipoint ion collectors. For more detailed discussions of their construction and operation, see Chapter 28, Point Ion Collectors (Detectors), and Chapter 29, Array Collectors (Detectors). In some forms of mass spectrometry, other methods of ion detection can be used, as with ion cyclotron instmments, but these are not considered here. [Pg.211]

Other types of mass spectrometer can use point, array, or both types of ion detection. Ion trap mass spectrometers can detect ions sequentially or simultaneously and in some cases, as with ion cyclotron resonance (ICR), may not use a formal electron multiplier type of ion collector at all the ions can be detected by their different electric field frequencies in flight. [Pg.212]

Other techniques for mass measurement are available, but they are not as popular as those outlined above. These other methods include mass measurements on a standard substance to calibrate the instrument. The standard is then withdrawn, and the unknown is let into the instrument to obtain a new spectrum that is compared with that of the standard. It is assumed that there are no instrumental variations during this changeover. Generally, this technique is less reliable than when the standard and unknown are in the instrument together. Fourier-transform techniques are used with ion cyclotron mass spectrometers and give excellent mass accuracy at lower mass but not at higher. [Pg.274]

Commercial mass analyzers are based almost entirely on quadrupoles, magnetic sectors (with or without an added electric sector for high-resolution work), and time-of-flight (TOE) configurations or a combination of these. There are also ion traps and ion cyclotron resonance instruments. These are discussed as single use and combined (hybrid) use. [Pg.280]

An added consideration is that the TOF instruments are easily and quickly calibrated. As the mass range increases again (m/z 5,000-50,000), magnetic-sector instruments (with added electric sector) and ion cyclotron resonance instruments are very effective, but their prices tend to match the increases in resolving powers. At the top end of these ranges, masses of several million have been analyzed by using Fourier-transform ion cyclotron resonance (FTICR) instruments, but such measurements tend to be isolated rather than targets that can be achieved in everyday use. [Pg.281]

A simple mass spectrometer of low resolution (many quadrupoles, magnetic sectors, time-of-flight) cannot easily be used for accurate mass measurement and, usually, a double-focusing magnetic/electric-sector or Fourier-transform ion cyclotron resonance instrument is needed. [Pg.416]

Ion cyclotron resonance analyzer. A device to determine the mass-to-charge ratio (m/z) of an ion in the presence of a magnetic field by measuring its cyclotron frequency. [Pg.429]

Mass spectrometer configuration. Multianalyzer instruments should be named for the analyzers in the sequence in which they are traversed by the ion beam, where B is a magnetic analyzer, E is an electrostatic analyzer, Q is a quadrupole analyzer, TOP is a time-of-flight analyzer, and ICR is an ion cyclotron resonance analyzer. For example BE mass spectrometer (reversed-geometry double-focusing instrument), BQ mass spectrometer (hybrid sector and quadrupole instrument), EBQ (double-focusing instrument followed by a quadrupole). [Pg.430]

FTICR. Fourier-transform ion cyclotron resonance GC/IRMS. gas chromatography isotope ratio mass spectrometry... [Pg.445]

ICP/MS. inductively coupled plasma and mass spectrometry used as a combined technique ICR. ion cyclotron resonance (spectroscopy)... [Pg.445]

Asamoto, B. and Dunbar, R.C., Analytical Applications of Fourier Transform Ion Cyclotron Resonance Spectroscopy, VCH, New York, 1991. [Pg.449]

Lehman, T.A. and Bursey, M.M., Ion Cyclotron Resonance Spectroscopy, Wiley, New York, 1976. [Pg.451]

The result of the Back-to-Basics series is an accumulation of some 50 separate but interrelated expositions of mass spectrometric principles and apparatus. Some areas of mass spectrometry, such as ion cyclotron resonance and ion trap instruments, have not been covered except for passing references. This decision has not been due to any bias by the authors or Micromass but simply reflects the large amount of writing that had to be done and the needs of the greatest proportion of users. [Pg.478]

A particularly useful property of the PX monomer is its enthalpy of formation. Conventional means of obtaining this value, such as through its heat of combustion, are, of course, excluded by its reactivity. An experimental attempt was made to obtain this measure of chemical reactivity with the help of ion cyclotron resonance a value of 209 17 kJ/mol (50 4 kcal/mol) was obtained (10). Unfortunately, the technique suffers from lack of resolution in addition to experimental imprecision. It is perhaps better to rely on molecular orbital calculations for the formation enthalpy. Using a semiempirical molecular orbital technique, which is tuned to give good values for heat of formation on experimentally accessible compounds, the heat of formation of /5-xylylene has been computed to be 234.8 kj/mol (56.1 kcal/mol) (11). [Pg.429]

The existence of the XeCHg [34176-86-8] cation has been estabtished ia the gas phase. The Xe—C bond energy of the XeCHg cation has been estimated to be 180 A 33 kJ/mol (112) and more recently, 231 A 10 kJ/mol (113) by ion cyclotron resonance. The compound Xe(CF3)2 [72599-34-9] is reported to be a waxy white sotid having a half-life of ca 30 min at room temperature (114). The synthesis iavolved the addition of XeF2 to a tritiuoromethyl plasma, but the characterization of this compound is limited and has not been iadependently confirmed. [Pg.24]

The most common modes of operation for ms/ms systems include daughter scan, parent ion scan, neutral loss scan, and selected reaction monitoring. The mode chosen depends on the information required. Stmctural identification is generally obtained using daughter or parent ion scan. The mass analyzers commonly used in tandem systems include quadmpole, magnetic-sector, electric-sector, time-of-flight, and ion cyclotron resonance. Some instmments add a third analyzer such as the triple quadmpole ms (27). [Pg.405]

Photoelectron spectroscopic studies show that the first ionization potential (lone pair electrons) for cyclic amines falls in the order aziridine (9.85 eV) > azetidine (9.04) > pyrrolidine (8.77) >piperidine (8.64), reflecting a decrease in lone pair 5-character in the series. This correlates well with the relative vapour phase basicities determined by ion cyclotron resonance, but not with basicity in aqueous solution, where azetidine (p/iTa 11.29) appears more basic than pyrrolidine (11.27) or piperidine (11.22). Clearly, solvation effects influence basicity (74JA288). [Pg.239]

ITiscusaioa of (be techniques for gas-(>haau equilibrium meaauiuncaita can be found in T. A. Lehman and M. M. Buisey, Ion Cyclotron Resonance ectrometry, Wiky-Interscience, New yfaik, 1976 M. T. Bowers, ed., Gas Phase Ion Chemistry, Vols. I and 2, Aeadetnic Press, New Ypik, 1979. [Pg.244]

By measurements in an ion cyclotron resonance spectrometer, the proton affinity (PA) of free radicals can be measured. [Pg.741]


See other pages where Ion cyclotron is mentioned: [Pg.810]    [Pg.1355]    [Pg.2390]    [Pg.85]    [Pg.89]    [Pg.189]    [Pg.195]    [Pg.281]    [Pg.282]    [Pg.284]    [Pg.25]    [Pg.542]    [Pg.23]    [Pg.135]    [Pg.4]    [Pg.40]    [Pg.70]    [Pg.887]   


SEARCH



Cyclotron

ESI-fourier transform ion cyclotron resonance

ESI—Fourier-transform ion cyclotron

Electron and ion cyclotron resonance studies

Electron cyclotron resonance ion sources

Electron cyclotron resonance ion trap

Electrospray Fourier transform-ion cyclotron

Electrospray ionization Fourier transform ion cyclotron resonance

Electrospray ionization-fourier transform ion cyclotron

Electrospray ionization-fourier transform ion cyclotron resonance-mass spectrometry

FT-ICR (Fourier transform-ion cyclotron

FTICR (Fourier transform ion cyclotron

Fourier Transform Ion Cyclotron Resonance Spectroscopy

Fourier Transform-ion cyclotron resonance FT-ICR) mass spectrometry

Fourier transform ion cyclotron

Fourier transform ion cyclotron mass

Fourier transform ion cyclotron mass spectrometer

Fourier transform ion cyclotron resonance FTICR)

Fourier transform ion cyclotron resonance FTICR) mass spectrometers

Fourier transform ion cyclotron resonance FTICR) mass spectrometry

Fourier transform ion cyclotron resonance MS

Fourier transform ion cyclotron resonance mass

Fourier transform ion cyclotron resonance mass spectra

Fourier transform ion cyclotron resonance mass spectrometer

Fourier transform ion cyclotron resonance spectrometer

Fourier transform ion cyclotron spectrometry

Fourier transform-ion cyclotron resonance FT-ICR)

Fourier transform-ion cyclotron resonance spectrometry

Fourier-transform ion cyclotron mass spectroscopy

Fourier-transform ion cyclotron resonance

Fourier-transform ion cyclotron resonance instruments

Fourier-transform ion-cyclotron resonance mass spectrometry

Ion cyclotron experiments

Ion cyclotron resonance

Ion cyclotron resonance (ICR

Ion cyclotron resonance ICR) spectrometer

Ion cyclotron resonance analysers (ICRMS)

Ion cyclotron resonance analyzer

Ion cyclotron resonance cells

Ion cyclotron resonance experiment

Ion cyclotron resonance instrument

Ion cyclotron resonance mass

Ion cyclotron resonance mass analyzer

Ion cyclotron resonance mass spectra

Ion cyclotron resonance mass spectrometr

Ion cyclotron resonance mass spectrometry

Ion cyclotron resonance mass spectrometry ICR-MS)

Ion cyclotron resonance mass spectroscopy

Ion cyclotron resonance spectrometry

Ion cyclotron resonance spectroscopy

Ion cyclotron resonance spectroscopy, and

Ion cyclotron resonance studies

Ion cyclotron resonance technique

Ion cyclotron resonance trap

Ion cyclotron resonance, Fourier

Ion-cyclotron motion

Laser ablation Fourier transform ion cyclotron resonance

Magnetic and Electrostatic Field Ion Cyclotron Resonance (ICR) Analyzers

Mass Analysis in Fourier Transform Ion Cyclotron Resonance Instruments

Mass Spectrometry and Ion Cyclotron Resonance Studies

Mass spectrometry Fourier transform ion cyclotron

Mass spectrometry Fourier transformed ion cyclotron resonance

Metastable ion cyclotron resonance

Principle of Ion Cyclotron Resonance

Pulsed ion cyclotron resonance

Spectrometers, ion-cyclotron resonance

Trapped ion cyclotron resonance

© 2024 chempedia.info