Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detection of Ions

For ion detection in time of flight analysers, usually secondary electron multipliers are used. [Pg.92]

An important parameter for the quality of a mass analyser is its capability to separate ions with small mass differences Am. This is described by the resolution, Rs, the ratio of the mass, m, over the difference Am of an ion with mass m + Am  [Pg.92]

Generally, the higher the resolution the better is the separation. But when are two peaks considered as being separated This is a question of definition and depends on the analyser. For TOF, Am is deflned as the full width at half maximum (FWHM), i.e. the width of the peak at half its height (Fig. 4.5). With this definition, it is possible to read Rs out of a single peak. Typical resolutions obtained for TOF instruments are Rs = 15,000 (FWHM). For other mass analysers, other definitions like the 70 % valley or 50 % valley are used (Fig. 4.6). For the 50 % valley definition, two peaks are considered separated if the minimum between them (the valley) is not more than 50 % of the peak height whereas for the 10% valley, the minimum between two peaks must not be more than 10% of the peak height. [Pg.92]

Solutions containing ions give chemical reactions expected of the ions. For example, mesitoic acid, which resists esterification by other methods, is readily esterified when its sulfuric acid solution is poured into an alcohol.177 [Pg.86]

Nevertheless, chemical methods have not been used for determining ionization equilibrium constants. The analytical reaction would have to be almost instantaneous and the formation of the ions relatively slow. Also the analytical reagent must not react directly with the unionized molecule. In contrast to their disuse in studies of ionic equilibrium, fast chemical reactions of the ion have been used extensively in measuring the rate of ionization, especially in circumstances where unavoidable irreversible reactions make it impossible to study the equilibrium. The only requirement for the use of chemical methods in ionization kinetics is that the overall rate be independent of the concentration of the added reagent, i.e., that simple ionization be the slow and rate-determining step. [Pg.86]


B) The multiphoton excitation of electronic levels of atoms and molecules with visible or UV radiation generally leads to ionization. The mechanism is generally a combination of direct, Goeppert-Mayer, and quasi-resonant stepwise processes. Since ionization often requires only two or tln-ee photons, this type of multiphoton excitation is used for spectroscopic purposes in combination with mass-spectrometric detection of ions. [Pg.2131]

Detection of ions. The observation of electrical signals due to particular ionic species by a detector under conditions that minimize ambiguities from interferences. Ions can be detected by photographic or suitable electrical means. [Pg.436]

The sensitivity, accuracy, and precision of solid-sample analysis have been greatly improved by coupling LA with ICP-OES-MS. The ablated species are transported by means of a carrier gas (usually argon) into the plasma torch. Further atomization, excitation, and ionization of the ablated species in the stationary hot plasma result in a dramatic increase in the sensitivity of the detection of radiation (LA-ICP-OES) or of the detection of ions (LA-ICP-MS). [Pg.234]

Chitosan-clay bio-nanocomposites are very stable materials without significant desorption of the biopolymer when they are treated with aqueous salt solutions for long periods of time. In this way, they act as active phases of electrochemical sensors for detection of ions (Figure 1.8). The particular nanostructuration of the biopolymer in the interlayer region drives the selective uptake of monovalent versus polyvalent anions, which has been applied in electrode arrays of electronic tongues [132]. [Pg.15]

The m/z values of peptide ions are mathematically derived from the sine wave profile by the performance of a fast Fourier transform operation. Thus, the detection of ions by FTICR is distinct from results from other MS approaches because the peptide ions are detected by their oscillation near the detection plate rather than by collision with a detector. Consequently, masses are resolved only by cyclotron frequency and not in space (sector instruments) or time (TOF analyzers). The magnetic field strength measured in Tesla correlates with the performance properties of FTICR. The instruments are very powerful and provide exquisitely high mass accuracy, mass resolution, and sensitivity—desirable properties in the analysis of complex protein mixtures. FTICR instruments are especially compatible with ESI29 but may also be used with MALDI as an ionization source.30 FTICR requires sophisticated expertise. Nevertheless, this technique is increasingly employed successfully in proteomics studies. [Pg.383]

At this point we must emphasise that the spectroscopic detection of ions does not mean necessarily that those same ions are the chain-carriers in polymerization. Up to 1965 it was believed by most workers that the absorption near 420 pm and 310 pm (with e420 > e310), which develops when styrene or 1-phenylethanol are treated with acid, was due to the 1-phenylethyl ion (see reference 15 and references quoted therein). [Pg.112]

Precursor ion scan3 Detection of ions with identical fragments High Medium... [Pg.149]

Ion recognition is a subject of considerable interest because of its implications in many fields chemistry, biology, medicine (clinical biochemistry), environment, etc. In particular, selective detection of metal cations involved in biological processes (e.g., sodium, potassium, calcium, magnesium), in clinical diagnosis (e.g., lithium, potassium, aluminum) or in pollution (e.g., lead, mercury, cadmium) has received much attention. Among the various methods available for detection of ions, and more... [Pg.21]

Although the above mass spectrometric tools have mass ranges and resolving powers adequate for chemical analysis, mass spectral characterization and structural analysis of biopolymers generally demand efficient detection of ions over a wide mass range, accurate mass measurements, and high mass resolution. The FT-ICR analyzer is able to combine high resolution and MS" capabilities. ... [Pg.172]

Obviously, almost any technique to achieve the goals of ionization, separation and detection of ions in the gas phase can be applied - and actually has been applied -in mass spectrometry. This leads to a simple basic setup having all mass spectrometers in common. A mass spectrometer consists of an ion source, a mass... [Pg.3]

Note Although the range up to m/z 3000 is normally employed for the detection of ions generated by ESI, ions of much higher m/z can be formed. [23,24] Even ions at m/z 85,000 have been observed. [25]... [Pg.441]

The pump and probe pulses employed may be subjected to a variety of nonlinear optical mixing processes they may be prepared and characterized by intensity, duration, spectral band width, and polarization. They may arrive in the reaction chamber at a desired time difference, or none. The probe pulse may lead to ionizations followed by detections of ions by mass spectrometry, but many alternatives for probing and detection have been used, such as laser-induced fluorescence, photoelectron spectroscopic detection, absorption spectroscopy, and the like. [Pg.904]

Figure 1.1 Basic diagram of mass spectrometry generation of ions from compounds in the sample, introduction into the ion source, separation of these ions by their mass-to-charge ratio in the mass separator and detection of ions in the ion detector. Figure 1.1 Basic diagram of mass spectrometry generation of ions from compounds in the sample, introduction into the ion source, separation of these ions by their mass-to-charge ratio in the mass separator and detection of ions in the ion detector.
Conductimetry is used to detect and determine ionic species in solution. It plays important roles in environmental analyses and in the detection of ion chromatography. It is also used in acid-base, precipitation and chelometric titrations to detect end-points. However, the biggest advantage of conductimetry is displayed in the fundamental studies of solution chemistry. Its applications to chemical studies in non-aqueous solutions will he discussed in Chapter 7. [Pg.157]

Figure 16.22—MS detectors, a) Multiple stage electron multipliers (reproduced by permission of ETP Scientific Inc.) b) channeltron the conical shape of the cathode allows the detection of ions with slightly different trajectories c) electron multiplication within a channeltron d) entrance of a multichannel plate detector (microchanneltron). Figure 16.22—MS detectors, a) Multiple stage electron multipliers (reproduced by permission of ETP Scientific Inc.) b) channeltron the conical shape of the cathode allows the detection of ions with slightly different trajectories c) electron multiplication within a channeltron d) entrance of a multichannel plate detector (microchanneltron).
Figure 27-15 Detection of ion-cyclotron resonance. When o) = o)c, energy is absorbed by the ions and the ammeter registers a current. Figure 27-15 Detection of ion-cyclotron resonance. When o) = o)c, energy is absorbed by the ions and the ammeter registers a current.
The detection of contamination with mercury is a very serious issue due to the extreme toxicity and ease of reaction of this heavy metal. Mercury is commonly used in many everyday applications, such as thermometry, barometry or scientific apparatus. Therefore, it is important to note the development of a fluorescent IIP membrane for detection of mercury [63], This membrane was based on the combination of two fluorescent functional monomers, namely 4-vinyl pyridine and 9-vinylcarbazole. Binding of Hg2+ was proposed to be based on complexation of the metal ion with lone pairs of both pyridine and carbazole. The IIP membrane was capable of recognizing Hg2+in the linear range of concentrations from 5 x 10 7 to 1 x 10 4 M. Selectivity of the chemosensor was studied for a wide group of possible competitive ions and the chemosensor responded to all of them in the presence of a fixed amount of Hg2+ at 1 x 10 5 M. Nevertheless, the highest influence on recognition properties was observed in the presence of Cu2+ (4.87%) and Pb2+ (4.57%), which are still acceptable values for the successful detection of Hg2+. This progress stimulates further development of the MIP detection of ions. [Pg.182]

Figure 2. Schematic diagram of LAMMA 500 instrument, adapted for simultaneous detection of ions and neutrals. Figure 2. Schematic diagram of LAMMA 500 instrument, adapted for simultaneous detection of ions and neutrals.

See other pages where Detection of Ions is mentioned: [Pg.92]    [Pg.86]    [Pg.515]    [Pg.37]    [Pg.92]    [Pg.295]    [Pg.211]    [Pg.215]    [Pg.222]    [Pg.498]    [Pg.47]    [Pg.116]    [Pg.86]    [Pg.359]    [Pg.376]    [Pg.372]    [Pg.392]    [Pg.215]    [Pg.221]    [Pg.117]    [Pg.234]    [Pg.29]    [Pg.196]    [Pg.682]    [Pg.86]    [Pg.117]    [Pg.138]    [Pg.138]    [Pg.315]    [Pg.662]    [Pg.2406]    [Pg.75]   


SEARCH



Analysis by the Detection of Scattered Ions

Detection efficiency of product ions

Detection of Metal Ions

Effects of transmembrane ion flux on detection limit

Ion Detection and Resolving Power of the Orbitrap

Low-Pressure Detection of Mobility-Separated Ions

Potentiometric Porphyrinic Sensors for Detection of Ions

Principles of Ion Chromatographic Separation and Detection

The Ion Chemistry Underlying Detection of Drugs

Tools of the Trade VI. Ion Detection and Data Processing

© 2024 chempedia.info