Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Slope compensation

A pH electrode is normally standardized using two buffers one near a pH of 7 and one that is more acidic or basic depending on the sample s expected pH. The pH electrode is immersed in the first buffer, and the standardize or calibrate control is adjusted until the meter reads the correct pH. The electrode is placed in the second buffer, and the slope or temperature control is adjusted to the-buffer s pH. Some pH meters are equipped with a temperature compensation feature, allowing the pH meter to correct the measured pH for any change in temperature. In this case a thermistor is placed in the sample and connected to the pH meter. The temperature control is set to the solution s temperature, and the pH meter is calibrated using the calibrate and slope controls. If a change in the sample s temperature is indicated by the thermistor, the pH meter adjusts the slope of the calibration based on an assumed Nerstian response of 2.303RT/F. [Pg.492]

In addition, most devices provide operator control of settings for temperature and/or response slope, isopotential point, zero or standardization, and function (pH, mV, or monovalent—bivalent cation—anion). Microprocessors are incorporated in advanced-design meters to faciHtate caHbration, calculation of measurement parameters, and automatic temperature compensation. Furthermore, pH meters are provided with output connectors for continuous readout via a strip-chart recorder and often with binary-coded decimal output for computer interconnections or connection to a printer. Although the accuracy of the measurement is not increased by the use of a recorder, the readabiHty of the displayed pH (on analogue models) can be expanded, and recording provides a permanent record and also information on response and equiHbrium times during measurement (5). [Pg.467]

A cocurrent evaporator train with its controls is illustrated in Fig. 8-54. The control system applies equally well to countercurrent or mixed-feed evaporators, the princip difference being the tuning of the dynamic compensator/(t), which must be done in the field to minimize the short-term effects of changes in feed flow on product quality. Solid concentration in the product is usually measured as density feedback trim is applied by the AC adjusting slope m of the density function, which is the only term related to x. This recahbrates the system whenever x must move to a new set point. [Pg.750]

Single-Effect Evaporators The heat requirements of a singleeffect continuous evaporator can be calculated by the usual methods of stoichiometry. If enthalpy data or specific heat and heat-of-solution data are not available, the heat requirement can be estimated as the sum of the heat needed to raise the feed from feed to product temperature and the heat required to evaporate the water. The latent heat of water is taken at the vapor-head pressure instead of at the product temperature in order to compensate partiaUv for any heat of solution. If sufficient vapor-pressure data are available for the solution, methods are available to calculate the true latent heat from the slope of the Diihriugliue [Othmer, Ind. Eng. Chem., 32, 841 (1940)]. [Pg.1145]

At night when radiational cooling occurs on slopes, the cool dense air near the surface descends along the slope (Fig. 17-19). This is the downslope wind. To compensate for this descending air, air farther from the slope that is cooled very little is warmer relative to the descending air and rises, frequently resulting in a closed circular path. Where the downslope winds... [Pg.263]

First one assumes that the final closed loop compensation network will have a continuous -20dB/decade slope. To achieve a 15 kHz cross-over frequency, the amplifier must add gain to the input signal and push-up the gain curve of the Bode plot. [Pg.104]

In order to improve the slope compensation circuitry that depends upon a resistor to ground all the time, I will split this resistor between the secondary winding of the current transformer and after the rectifiers. I will double the value of the two resistors (150 ohms each), so that when the diodes are conducting, the net value is the same. [Pg.127]

Every current-mode control application that exceeds 50 percent duty cycle must have slope compensation on the current ramp waveform. Otherwise an instability will occur whenever the duty cycle exceeds 50 percent. This is typically done by summing into the current waveform some of the oscillator ramp waveform. This will increase the slope of the current waveform and therefore trip the current sense comparator earlier. A common problem is the inadvertent loading of the oscillator, so I will use a PNP emitter-follower to buffer the oscillator. The circuit configuration can be seen in Figure 3-74. [Pg.127]

The design of the slope compensation circuit is almost fairly qualitative and may eventually need to be adjusted at the breadboard stage. To estimate how much additional ramp voltage is needed to keep the power supply stable, one performs the following equation. Aj is the gain or step-down influences of the transformers between the output and the current sense pin. [Pg.127]

Applications of Rp techniques have been reported by King et al. in a study of the corrosion behavior of iron pipes in environments containing SRB. In a similar study, Kasahara and Kajiyama" used Rp measurements with compensation of the ohmic drop and reported results for active and inactive SRB. Nivens et al. calculated the corrosion current density from experimental Rp data and Tafel slopes for 304 stainless steel exposed to a seawater medium containing the non-SRB Vibrio mtriegens. [Pg.211]

In order to go further into the experimental check we constructed Arrhenius plots of the fluorescence quantum yield of BMPC in a few solvents (methanol, ethanol, propanol, hexanol and methylene chloride), all of which showed good linearity. The activation energies and A/kp ratios, calculated from the slopes and intercepts of those plots, are collected in Table 1. The smooth increase of both parameters in the alcohol series is mainly associated with the increase of solvent viscosity. On the other hand, decrease of the solvent dielectric constant from 32.7 (methanol) to 8.9 (dichloromethane) causes a small but significant increase of the activation energy also, this increase is probably somewhat compensated by the decrease of the viscous-flow... [Pg.393]

The pH (or pI) term of the Nemst equation contains the electrode slope factor as a linear temperature relationship. This means that a pH determination requires the instantaneous input, either manual or automatic, of the prevailing temperature value into the potentiometer. In the manual procedure the temperature compensation knob is previously set on the actual value. In the automatic procedure the adjustment is permanently achieved in direct connection with a temperature probe immersed in the solution close to the indicator electrode the probe usually consists of a Pt or Ni resistance thermometer or a thermistor normally based on an NTC resistor. An interesting development in 1980 was the Orion Model 611 pH meter, in which the pH electrode itself is used to sense the solution temperature (see below). [Pg.94]

Boss Ng You must have been dreaming all right. Because I think you have gone and ruined the slope compensation You should never, never, never fool around with that capacitor in this current mode control IC. If you were just a little smarter, you would have tried paralleling a larger resistor across the sense resistor to adjust its value, but you had to go and do just that ... [Pg.25]

Fixed ramp added to switch current sense signal (crude but popular method of introducing slope compensation)... [Pg.26]

It is clear from the Nemst equation that the temperature of the solution affects the response slope (2.303A7//0 of the calibration curve. The electrode voltage changes linearly in relationship to changes in temperature at a given pH therefore, the pH of any solution is a function of its temperature. For example, the electrode response slope increases from 59.2mV/pH at 25°C to 61.5 mV/pH at a body temperature of 37°C. For modem pH sensing systems, a temperature probe is normally combined with the pH electrode. The pH meter with an automatic temperature compensation (ATC) function automatically corrects the pH value based on the temperature of the solution detected with the temperature probe. [Pg.290]

The standard curve will in some instances not follow a straight course or will have a wrong slope. If the slope is too steep, i.e., if the double concentration gives more than double a value of optical density, one will suspect the presence of an inhibitor or toxic factor, relatively more effective at lower concentrations, or of a stimulating factor relatively more effective at higher concentrations. In the case of strongly colored samples, optical compensation can be achieved in the reading of the... [Pg.190]

Figure 9 demonstrates this compensation effect by the linear relationship between AS and AH. This indicates that both activation parameters depend equally on a and that the isokinetic temperature, i.e. the slope of the line, amounts to 256°K. Thus, at -17°C the rate would become independent of a, whereas it increases with a at higher temperatures. [Pg.20]

Figure 6.29 Schematic representation of the adjustment which is made to compensate for the sloping baseline in stripping voltammetry, itself due to charging of the electric double-layer at the WE. Figure 6.29 Schematic representation of the adjustment which is made to compensate for the sloping baseline in stripping voltammetry, itself due to charging of the electric double-layer at the WE.

See other pages where Slope compensation is mentioned: [Pg.195]    [Pg.62]    [Pg.79]    [Pg.325]    [Pg.127]    [Pg.151]    [Pg.9]    [Pg.204]    [Pg.191]    [Pg.145]    [Pg.463]    [Pg.47]    [Pg.92]    [Pg.93]    [Pg.94]    [Pg.146]    [Pg.206]    [Pg.215]    [Pg.276]    [Pg.293]    [Pg.176]    [Pg.295]    [Pg.314]    [Pg.147]    [Pg.33]    [Pg.244]    [Pg.778]   


SEARCH



Slope

Sloping

© 2024 chempedia.info