Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column test

Each column is shipped with a PSS column quality certificate, a column connector, and the test kit, which was used at the PSS quality inspection laboratory to test the column resolution. This allows the user to reproduce the column tests without problems. It should be noted, however, that column test results not only depend on the performance of the column alone, but also on the instrument employed for the test. Figure 9.8 shows an example of a PSS column quality certificate. [Pg.284]

FIGURE 9.9 Definition of column test parameters using a SEC chromatogram of a polymer standard mixture (PSS Polystyrene ReadyCal) and a low molar mass compound (BHT). [Pg.286]

GL 1] [R 1] [R 3] [P le] The falling film micro reactor has a better selectivity-conversion performance than the two micro bubble columns tested (Figure 5.18) (3, 38]. The micro bubble column with narrow channels has a better behavior at large conversion than the version with wide channels. The behavior of the falling film micro reactor and the micro bubble column with narrow channels is characterized by a nearly constant selectivity with increasing conversion, while the bubble column with wide channels shows notably decreasing selectivity with conversion (similar to the laboratory bubble column). [Pg.603]

Figure 4.10 Typical routine column test chromatogram for a 30 cm X 4.6 mm I. D. column pacXed with an octadecylsiloxane bonded silica packing of lO micrometers particle diameter. The test mixture consisted of resorcinol (0.55 mg/ml), acetophenone (0.025 mg/ml), naphthalene (0.20 mg/ml) and anthracene (0.01 mg/ml) in acetonitrile, 10 microliters injected. The separation was performed isocratically at 23 C with acetonitrile-water (55 45) as the mobile phase at a flow rate of 1.5 ml/min. Detection was by UV at 254 nm (0.1 AUFS). Figure 4.10 Typical routine column test chromatogram for a 30 cm X 4.6 mm I. D. column pacXed with an octadecylsiloxane bonded silica packing of lO micrometers particle diameter. The test mixture consisted of resorcinol (0.55 mg/ml), acetophenone (0.025 mg/ml), naphthalene (0.20 mg/ml) and anthracene (0.01 mg/ml) in acetonitrile, 10 microliters injected. The separation was performed isocratically at 23 C with acetonitrile-water (55 45) as the mobile phase at a flow rate of 1.5 ml/min. Detection was by UV at 254 nm (0.1 AUFS).
Figure 2 Chromatogram from conventional SEC column compared to high-speed SEC column tested on an identical instrument with polystyrene standards in tetrahydrofuran (THF). (See Color Plate Section at the end of this book.)... Figure 2 Chromatogram from conventional SEC column compared to high-speed SEC column tested on an identical instrument with polystyrene standards in tetrahydrofuran (THF). (See Color Plate Section at the end of this book.)...
Sediment desorption t,/2 = 38.5 d from sediment under conditions mimicking marine disposal (Zhang et al. 2000). Soil t,/2 = 295-2448 h, based on aerobic soil column test data (Kincannon Lin 1985 quoted, Howard et al. 1991) ... [Pg.694]

Absorptionsmittel 3, zero-valent iron, granular iron hydroxide) were investigated for their capability to remove arsenic from water. Both arsenite and arsenate were investigated and batch and column tests were carried out. [Pg.26]

Column tests were carried out to investigate the behavior of the materials in a dynamic system. Figure 2 shows the results of these experiments for granular iron hydroxide, Absorptionsmittel 3 and Zr loaded activated carbon. The results of Fe° and activated carbon are not shown in the graph because they had initial high outlet concentrations. [Pg.29]

The retention indices, measured on the alkyl aryl ketone scale, of a set of column test compounds (toluene, nitrobenzene, p-cresol, 2-phenyl ethanol, and IV-methylaniline) were used to determine the changes in selectivity of a series of ternary eluents prepared from methanol/0.02M phosphate buffer pH 7 (60 40), acetonitrile/0.02 M phosphate buffer pH 7 (50 50) and tetrahydrofuran/0.02 M phosphate buffer pH 7 (25 65). The analyses were carried out on a Spherisorb ODS reversed-phase column. The selectivity changes were often nonlinear between the binary composition [83]. [Pg.538]

In order to determine the applicability of retention indices, based on the alkyl arylketone scale, as the basis of a reproducible method of reporting retentions, the separation of 10 barbiturates and a set of column test compounds were examined on an octadecylsilyl bonded silica (ODS-Hypersil) column with methanol-buffer of pH 8.5 as eluent [100]. The effects on the capacity factors and retention indices, on changing the eluent composition, pH, ionic strengthened temperature, showed that the retention indices of the barbiturates were much less susceptible to minor changes in the eluent than the capacity factors. [Pg.543]

Although the treatability of a particular groundwater by carbon and the relative capacity of different types of carbon for treatment may be estimated from adsorption isotherms, carbon performance and design criteria are best determined by pilot column tests. Design-related information that can be obtained from pilot tests includes ... [Pg.249]

Pd = the dry mass density (mass of dry solids divided by the total volume of the soil/sediment specimen used in a leaching-column test) of the test specimen, and... [Pg.198]

Several workers [ 1,29,66,67,104,146 -149] indicated that studying pollutants and/or SWM leachate migration profiles resulting from transport of pollutants with a test soil requires that replicate samples be subjected to leaching-column tests, where various pore volumes of the same solution are applied. [Pg.200]

The diffusion coefficient Dx can be expressed as a function of the concentration of the specific constituent in the pore fluid [158]. Figure 8 shows four relationships between the non-dimensional pollutant concentration and Dn (the molecular diffusion coefficient). These show the strong influence from both in characterizing the shape of the relative pollutant concentration profile. Leaching-column test information shows that case (4) in Fig. 8 is the more likely profile (i.e., Dn= ae bc). Using this function in Eq. (90), the final governing equation for (X)+ will be ... [Pg.210]

To determine the leaching of chemical constituents from SWMs/COMs under conditions of constant surface renewal, columns (2.5 cm in diameter, 25 cm long) filled with SWMs/COMs were leached with distilled water at three different flow rates. The column tests were used to simulate leaching of highway materials under conditions of subsurface percolation of rainwater. Effluent samples from the column were taken with time for up to 80 h. The filtered solutions were measured for TOC and/or individual compound concentrations, and for toxicity. [Pg.222]

Column Test for Re versed-phase Liquid Chromatography... [Pg.40]

Reversed-phase liquid chromatography shape-recognition processes are distinctly limited to describe the enhanced separation of geometric isomers or structurally related compounds that result primarily from the differences between molecular shapes rather than from additional interactions within the stationary-phase and/or silica support. For example, residual silanol activity of the base silica on nonend-capped polymeric Cis phases was found to enhance the separation of the polar carotenoids lutein and zeaxanthin [29]. In contrast, the separations of both the nonpolar carotenoid probes (a- and P-carotene and lycopene) and the SRM 869 column test mixture on endcapped and nonendcapped polymeric Cig phases exhibited no appreciable difference in retention. The nonpolar probes are subject to shape-selective interactions with the alkyl component of the stationary-phase (irrespective of endcapping), whereas the polar carotenoids containing hydroxyl moieties are subject to an additional level of retentive interactions via H-bonding with the surface silanols. Therefore, a direct comparison between the retention behavior of nonpolar and polar carotenoid solutes of similar shape and size that vary by the addition of polar substituents (e.g., dl-trans P-carotene vs. dll-trans P-cryptoxanthin) may not always be appropriate in the context of shape selectivity. [Pg.244]

Le Mapihan, K., Vial, J., and Jardy, A. (2004). Reversed-phase liquid chromatography column testing robustness study of the test.. Chromatogr. A 1061, 149—158. [Pg.221]

Soil Column Tests. In the sand penetration test, a minimal amount of water was used. No consideration was given to the hydrostatic pressure which would occur in nature from a body of surface water. A new soil infiltration test was developed to take this into consideration. This test used a maximum amount of water (200 mL) on a minimum amount of treated soil (10 g) and was restricted only by the dimensions of the laboratory equipment. Our aim was to prepare an hydrophobe for soil which would support water over an extended period of time. Whereas water passed through soil treated with hydrophilic compounds within 8 hr, 2 weeks or more were required for penetration through an hydrophobe-treated soil. In the latter case the water level dropped 6 mm or less each day, showing that the cationic surfactant greatly hindered, but did not completely restrict the passage of water. The tests were usually terminated after 2 weeks, due to the large number of samples to be tested. [Pg.218]

Based on aerobic soil column test data, the reported half-life of acenaphthene in soil ranged from 295 h to 102 d (Kincannon and Lin, 1985). [Pg.48]

Trstenjak and Perdih (1999) reported that the fungus Phanerochaete chrysosporium removed the majority of PCB-1242 from a saturated water solution at room temperature up to a volumetric ratio of one to several thousand. In static column tests containing pelletized Phanerochaete chrysosporium, the percentage of PCB-1242 not removed ranged from 2.4 to 12% at pellets to contaminated water ratios of 1 5,000 and 1 500,000, respectively. [Pg.902]

Soil sorption distribution coefficients (Kd) were determined from centrifuge column tests using kaolinite as the absorbent (Celorie et al, 1989). Values for Kd ranged from 0.010 to 0.054 L/g. [Pg.952]

Toxin Separations. A number of columns have been evaluated for suitability using the HPLC method. Of all columns tested to date, the Hamilton PRP-1 column (polystyrene divinylbenzene resin) has proven to be the most useful. Toxin retention on this column is controlled by 1) methanol concentration, 2) mobile phase ionic strength, 3) chain length of the ion-pair reagent, and M) mobile phase pH. [Pg.202]


See other pages where Column test is mentioned: [Pg.55]    [Pg.173]    [Pg.1395]    [Pg.302]    [Pg.306]    [Pg.430]    [Pg.433]    [Pg.43]    [Pg.86]    [Pg.674]    [Pg.697]    [Pg.697]    [Pg.699]    [Pg.237]    [Pg.693]    [Pg.155]    [Pg.97]    [Pg.205]    [Pg.133]    [Pg.200]    [Pg.340]    [Pg.343]    [Pg.112]    [Pg.10]    [Pg.23]   
See also in sourсe #XX -- [ Pg.141 , Pg.146 ]

See also in sourсe #XX -- [ Pg.130 ]




SEARCH



Beam-to-column joints test set-up for determining the moment

Buckling test columns

Collapse test columns

Column Selectivity Testing

Column Testing (Stability and Selectivity)

Column durability tests

Column evaluation Grob test

Column evaluation routine tests

Column evaluation test mixtures

Column leaching test

Column oven, test

Column stability testing

Column testing

Column with test compounds

Electropulse column testing

HPLC Column Tests

Method development column testing during

Open tubular column testing

Other Column Tests

Rapid small-scale column tests

Rapid small-scale column tests RSSCTs)

Simple tests for HPLC columns

Small column adsorption capacity test

Test Method for Distillation of Crude Petroleum (15-Theoretical Plate Column)

Tests for columns

Tests for the Comparison of Columns and Their Expressiveness

Water column testing

© 2024 chempedia.info