SRM 869a Column Selectivity Test Mixture for Liquid Chromatography [44] is composed of three shape-constrained PAHs (phenanthro[3,4-c]phenanthrene, PhPh l,2 3,4 5,6 7,8-tetrabenzonaphthalene, TBN and benzo[a]pyrene, BaP) and is routinely employed to evaluate the shape selectivity of stationary phases. The retention differences between the nonplanar TBN and planar BaP solutes (expressed as a selectivity factor axEN/BaP = provide a numerical assessment of... [Pg.240]

Each SynChropak column is tested chromatographically to assure that it has been packed according to specifications. For SynChropak GPC columns, a mixture of a high molecular weight DNA and glycyltyrosine, a dipeptide, is used to evaluate internal volume and efficiency. The mobile phase used for the test is 0.1 M potassium phosphate, pH 7, and the flow rate is 0.5 ml/min for 4.6-mm i.d. columns. Minimum plate count values and operational flow rates are listed in Table 10.4 for 4.6-mm i.d. columns of all supports and the various diameters of the SynChropak GPC 100 columns. [Pg.314]

Figure 15.11 (a) Total ion clnomatogram of a Grob test mixture obtained on an Rtx-1701 column, and (b) re-injection of the entire clnomatogram on to an Rtx-5 column. Peak identification is as follows a, 2,3-butanediol b, decane c, undecane d, 1-octanol e, nonanal f, 2,6-dimethylphenol g, 2-ethylhexanoic acid h, 2,6-dimethylaniline i, decanoic acid methyl ester ], dicyclohexylamine k, undecanoic acid, methyl ester 1, dodecanoic acid, methyl ester. Adapted from Journal of High Resolution Chromatography, 21, M. J. Tomlinson and C. L. Wilkins, Evaluation of a semi-automated multidimensional gas chromatography-infrared-mass specti ometry system for initant analysis , pp. 347-354, 1998, with permission from Wiley-VCH. [Pg.424]

Only one research group has illustrated the reversed phase retention behavior of solutes on zirconia-silica modified surfaces. In the study by Melo et al. on PMOS gamma-irradiated modified surfaces, the reversed phase retention behavior of a test mixture containing acetone, benzonitrile, benzene, toluene, and naphthalene was evaluated. The authors illustrated good resolution of the five-component text mixture as shown in Fig. 7. Retention, however, decreased after the stationary phase was washed with 5000 column volumes of base (pH 10). Despite the base washing, uniform peak shape was maintained, with only a slight reduction in resolution as discussed in Alkaline Stability above. [Pg.1747]

Test mixtures for routine quality evaluation of normal- and reversed-phase columns... [Pg.402]

The formula refers to a reflux ratio v = oo, i.e., for a liquid-vapour ratio FID = 1. Beck [244] evaluated 115 experiments on the system water-air and derived new equations for packing sizes from 8 to 80 mm and column diameters from 150 to 1200 mm. Savkovi5-Stevanovi6, Simonovic and Popovic [244 a] examined the efficiency of a laboratory column (diameter, 25 mm separating length, 1.2 m) by determining the HETP values for various pressure drops and vapour velocities. The packing units were of 0.35 nmi steel wire with a diameter of 4 mm. The system benzene/carbon tetrachloride (30 mol% carbon tetrachloride) was used as test mixture. [Pg.182]

The performance of a capillary column can be evaluated with a test mixture whose components and resulting peak shapes serve as monitors of column efficiency and diagnostic probes for adverse adsorptive effects and the acid/base character of a column. These mixtures are used by column manufactures in the quality control of their columns and are likewise recommended for the chromatographic laboratory. [Pg.126]

TABLE 3.8 Test Mixtures Used to Evaluate GC Columns... [Pg.118]

Discussion—Under these conditions no vapor can reach the head and the heat to the distillation flask must be reduced to establish normal operations again. The flood point is normally determined during the efficiency evaluation of a column using the n-heptane-methylcyclohexane test mixture (Annex Al). [Pg.456]

In addition to the visual observations of the dynamic responses, a quantitative measure is needed to provide a better comparison. With such an objective, lAE values were evaluated for each closed-loop response. The PUL option shows the lowest lAE value of 5.607 x 10 , while the value for the Petlyuk column turns out to be 2.35 x 10. Therefore, the results of the test indicate that, for the SISO control of the heaviest component of the ternary mixture, the PUL option provides the best dynamic behavior and improves the performance of the Petlyuk column. Such result is consistent with the prediction provided by the SVD analysis. [Pg.65]

The addition of zirconium to activated carbon may substantially increase the removal of arsenic from water (Daus, Wennrich and Weiss, 2004 Schmidt et al., 2008). (Daus, Wennrich and Weiss, 2004) used batch and column tests to evaluate the ability of five materials (activated carbon, zirconium-loaded activated carbon, zerovalent iron, granulated Fe(III) hydroxide, and a commercial product, Absorptionsmittel 3 ) to sorb As(III) and As(V) from water. The GAC had grain sizes between 1.0 and 1.5 mm. The material was primarily chosen as a comparison with the zirconium-loaded sample. The zirconium-loaded activated carbon contained 28 mg zirconium g 1 activated carbon and was produced by shaking activated carbon in a solution of zirconyl nitrate (Zr0(N03)2). The zerovalent iron (Fe(0)) primarily had particle sizes of 1.2-1.7 mm. Absorptionsmittel 3 is a mixture of calcite, brucite, fluorite, and iron hydroxides. The granular iron hydroxides consisted of mostly amorphous Fe(III) hydroxide coatings on sand grains (particle sizes of 3-4mm) (Daus, Wennrich and Weiss, 2004, 2950). [Pg.386]


See other pages where Test mixtures, column evaluation is mentioned: [Pg.422]    [Pg.603]    [Pg.697]    [Pg.168]    [Pg.172]    [Pg.173]    [Pg.240]    [Pg.244]    [Pg.487]    [Pg.72]    [Pg.422]    [Pg.170]    [Pg.373]    [Pg.1665]    [Pg.683]    [Pg.1625]    [Pg.950]    [Pg.1621]    [Pg.199]    [Pg.151]    [Pg.155]    [Pg.402]    [Pg.408]    [Pg.218]    [Pg.276]    [Pg.1116]    [Pg.1593]    [Pg.127]    [Pg.973]    [Pg.117]    [Pg.456]    [Pg.283]    [Pg.323]    [Pg.172]    [Pg.646]    [Pg.184]    [Pg.279]   