Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Standards atomic emission lines

Atomic emission spectroscopy can be employed, generally with an inductively coupled plasma for thermal excitation. The sample is introduced into the plasma as a mist of ultrafine droplets, and the monochromator and detector are set to measure the intensity of an atomic emission line characteristic of the element. This technique is powerful, general, sensitive, linear, and able to measure over 70 elements, and, as a result, is widely used. Response is typically linear over four orders of magnitude in concentration with relative standard deviations of 1 to 3%. In low-salt aqueous solutions, detection limits range from 10 to 1000 nanomolar without preconcentration. Significant problems with saline samples remain, but use of Babington nebulizers alleviates these problems somewhat. [Pg.60]

Elements were monitored using the following atomic emission lines As 189.042 nm, Mg 280.270 nm, Na 589.592 nm, P 213.618 nm. Prior to determining the concentration of each sample, the instrument was flushed with dilute nitric acid and then 1 mL of the analyte. The concentration of the remaining 3 mL of sample was determined three times sequentially. The average of the three measurements was used in the final analysis. However, spurious measurements among the three measurements can be an indication of a problem. The instrument was minimally programmed to measure a reference standard every 15 samples. [Pg.387]

Standardizing the Method Equation 10.34 shows that emission intensity is proportional to the population of the excited state, N, from which the emission line originates. If the emission source is in thermal equilibrium, then the excited state population is proportional to the total population of analyte atoms, N, through the Boltzmann distribution (equation 10.35). [Pg.438]

When possible, quantitative analyses are best conducted using external standards. Emission intensity, however, is affected significantly by many parameters, including the temperature of the excitation source and the efficiency of atomization. An increase in temperature of 10 K, for example, results in a 4% change in the fraction of Na atoms present in the 3p excited state. The method of internal standards can be used when variations in source parameters are difficult to control. In this case an internal standard is selected that has an emission line close to that of the analyte to compensate for changes in the temperature of the excitation source. In addition, the internal standard should be subject to the same chemical interferences to compensate for changes in atomization efficiency. To accurately compensate for these errors, the analyte and internal standard emission lines must be monitored simultaneously. The method of standard additions also can be used. [Pg.438]

In Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES), a gaseous, solid (as fine particles), or liquid (as an aerosol) sample is directed into the center of a gaseous plasma. The sample is vaporized, atomized, and partially ionized in the plasma. Atoms and ions are excited and emit light at characteristic wavelengths in the ultraviolet or visible region of the spectrum. The emission line intensities are proportional to the concentration of each element in the sample. A grating spectrometer is used for either simultaneous or sequential multielement analysis. The concentration of each element is determined from measured intensities via calibration with standards. [Pg.48]

Spark sources are especially important for metal analysis. To date, medium-voltage sparks (0.5-1 kV) often at high frequencies (1 kHz and more), are used under an argon atmosphere. Spark analyses can be performed in less than 30 s. For accurate analyses, extensive sets of calibration samples must be used, and mathematical procedures may be helpful so as to perform corrections for matrix interferences. In arc and spark emission spectrometry, the spectral lines used are situated in the UV (180-380nm), VIS (380-550nm) and VUV (<180 nm) regions. Atomic emission spectrometry with spark excitation is a standard method for production and product control in the metal industry. [Pg.615]

Table 8.7). Thus, intensity and concentration are directly proportional. However, the intensity of a spectral line is very sensitive to changes in flame temperature because such changes can have a pronounced effect on the small proportion of atoms occupying excited levels compared to those in the ground state (p. 274). Quantitative measurements are made by reference to a previously prepared calibration curve or by the method of standard addition. In either case, the conditions for measurement must be carefully optimized with reference to the choice of emission line, flame temperature, concentration range of samples and linearity of response. Relative precision is of the order of 1-4%. Flame emission measurements are susceptible to interferences from numerous sources which may enhance or depress line intensities. [Pg.318]

Atomic absorption spectroscopy is highly specific and there are very few cases of interference due to the similar emission lines from different elements. General interference effects, such as anionic and matrix effects, are very similar to those described under flame emission photometry and generally result in reduced absorbance values being recorded. Similarly, the use of high temperature flames may result in reduced absorbance values due to ionization effects. However, ionization of a test element can often be minimized by incorporating an excess of an ionizable metal, e.g. potassium or caesium, in both the standards and samples. This will suppress the ionization of the test element and in effect increase the number of test atoms in the flame. [Pg.84]

The standard astm test method (D-1149-64) for rubber damage includes a test chamber (volume, 0.11-0.14 m ) through which ozonized air flows at a rate greater than 0.6 m/s. Because the residence time of the ozonized air in the test chamber is about 1 s, the ozone may be expected to reach the material in about 0.1 s. A somewhat similar test procedure (aatcc test method 109-1972 ansi L14, 174-1973) is used in testing colorfastness. The ozone generator is usually (but not necessarily) a mercury-vapor resonance lamp with emission lines at 184.9 and 253.7 nm. The 184.9-nm line is absorbed, and two ground-state oxygen atoms are produced ... [Pg.644]

It is also possible to use an internal standard to correct for sample transport effects, instrumental drift and short-term noise, if a simultaneous multi-element detector is used. Simultaneous detection is necessary because the analyte and internal standard signals must be in-phase for effective correction. If a sequential instrument is used there will be a time lag between acquisition of the analyte signal and the internal standard signal, during which time short-term fluctuations in the signals will render the correction inaccurate, and could even lead to a degradation in precision. The element used as the internal standard should have similar chemical behaviour as the analyte of interest and the emission line should have similar excitation energy and should be the same species, i.e. ion or atom line, as the analyte emission line. [Pg.105]

The procedure for the method of standard additions was described in Section 1.4.1. The following data were obtained for the determination of copper in a contaminated stream by inductively coupled plasma atomic emission using the 324.754 nm copper line. [Pg.162]

There are many atomic emission lamps which give very precise line spectra. These are little used in photochemical applications, but are useful as wavelength calibration standards. A small selection of available wavelengths is listed in Table 7.1. [Pg.218]

There is also a standard test method for determination of major and minor elements in coal ash by inductively coupled plasma (ICP)-atomic emission spectrometry (ASTM D-6349). In the test method, the sample to be analyzed is ashed under standard conditions and ignited to constant weight. The ash is fused with a fluxing agent followed by dissolution of the melt in dilute acid solution. Alternatively, the ash is digested in a mixture of hydrofluoric, nitric, and hydrochloric acids. The solution is analyzed by (ICP)-atomic emission spectrometry for the elements. The basis of the method is the measurement of atomic emissions. Aqueous solutions of the samples are nebulized, and a portion of the aerosol that is produced is transported to the plasma torch, where excitation and emission occurs. Characteristic line emission spectra are produced by a radio-frequency inductively coupled plasma. A grating monochromator system is used to separate the emission lines, and the intensities of the lines are monitored by photomultiplier tube or photodiode array detection. The photocurrents from the detector... [Pg.104]

Official methods have been published for the determination of nitric-perchloric acid-soluble lead [111] and ammonium pyrrolidine dithiocarbamate-extractable lead [ 112] in soil. Atomic absorption spectrometric evaluations of the digest or extract is conducted at the 217 nm emission line from a lead hollow cathode lamp. Rigin and Rigina [122] determined lead in soil by flameless atomic fluorescence using electrolytic preconcentration. The limit of detection is 15 pg lead and the standard deviation is not greater than 0.04. [Pg.42]

The tail of the plasma formed at the tip of the torch is the spectroscopic source, where the analyte atoms and their ions are thermally ionized and produce emission spectra. The spectra of various elements are detected either sequentially or simultaneously. The optical system of a sequential instrument consists of a single grating spectrometer with a scanning monochromator that provides the sequential detection of the emission spectra lines. Simultaneous optical systems use multichannel detectors and diode arrays that allow the monitoring of multiple emission lines. Sequential instruments have a greater wavelength selection, while simultaneous ones have a better sample throughput. The intensities of each element s characteristic spectral lines, which are proportional to the number of element s atoms, are recorded, and the concentrations are calculated with reference to a calibration standard. [Pg.231]

Procedure Concomitantly determine the absorbances of the Standard Solutions and the Test Solution at the potassium emission line of 766.5 nm, using a suitable atomic absorption spectrophotometer equipped with a potassium hollow-cathode... [Pg.40]

Procedure Concomitantly determine the absorbance of the Blank, the Standard Preparation, and the Test Preparation at the lead emission line at 283.3 nm, with a suitable atomic absorption spectrophotometer equipped with a lead hollow-cathode lamp and an air-acetylene flame, using 4-methyl-2-pentanone to set the instrument to zero. In a suitable analysis, the absorbance of the Blank is not greater than 20% of the difference between the absorbance of the Standard Preparation and the absorbance of the Blank. The absorbance of the Test Preparation does not exceed that of the Standard Preparation. [Pg.171]


See other pages where Standards atomic emission lines is mentioned: [Pg.46]    [Pg.487]    [Pg.263]    [Pg.134]    [Pg.212]    [Pg.457]    [Pg.514]    [Pg.579]    [Pg.134]    [Pg.212]    [Pg.62]    [Pg.109]    [Pg.265]    [Pg.227]    [Pg.717]    [Pg.381]    [Pg.531]    [Pg.986]    [Pg.331]    [Pg.37]    [Pg.42]    [Pg.48]    [Pg.59]    [Pg.96]    [Pg.17]    [Pg.263]    [Pg.381]    [Pg.126]    [Pg.331]    [Pg.143]    [Pg.161]   
See also in sourсe #XX -- [ Pg.252 ]




SEARCH



Atom line

Atomic emission

Atomic lines

Atomic standard

Emission standards

Standard line

© 2024 chempedia.info