Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Instruments ammeter

Seebeck used antimony and copper wires and found the current to be affected by the measuring instrument (ammeter). But, he also found that the voltage generated (EMF) was directly proportional to the difference in temperature of the two junctions. Peltier, in 1834, then demonstrated that if a current was induced in the circuit of 7.1.3., it generated heat at the junctions. In other words, the SEEBECK EFFECT was found to be reversible. Further work led to the development of the thermocouple, which today remains the primary method for measurement of temperature. Nowadays, we know that the SEEBECK EFFECT arises because of a difference in the electronic band structure of the two metals at the junction. This is illustrated as follows ... [Pg.359]

Control system. Current source (batteries), ammeter, voltmeter, instrument panel, and so on. [Pg.2430]

The height (we may consider it from the bottom line) of the indicating instruments such as a voltmeter and an ammeter, which the operator may have to read often, is also recommended to be no higher than 2000 mm or less than 300 mm from ground level. [Pg.374]

These are employed for the measurement of power circuit currents through an ammeter, kW, kWh or KVAr and power factor meter, or similar instruments requiring a current measurement. They must have a specified accuracy class as in lEC 60044-1 and the secondary current substantially proportional to the primary within a working range of about 5-120% of its primary rated current. They... [Pg.475]

I When the system voltage is linear (an ideal condition that would seldom exist) but the load is non-linear The current will be distorted and become non-sinusoidal. The actual current /, (r.m.s.) (equation (23.2)) will become higher than could be measured by an ammeter or any other measuring instrument, at the fundamental frequency. Figure 23.13 illustrates the difference between the apparent current, measured by an instrument, and the actual current, where / = active component of the current... [Pg.744]

This instrument was developed from the hot-wire ammeter, some examples of which can still be found. In the modem equivalent, the current to be measured (or a known proportion of it) flows through a small element that heats a thermocouple, so producing a rms voltage at its terminals, which is a function of the current. This voltage then supplies a current to a permanent magnet, moving-coil movement. [Pg.238]

The Kelvin double bridge is a more sophisticated variant used for the measurement of very low resistance such as ammeter shunts or short lengths of cable. This is also operated on D.C. In industrial terms the digital D.C. low-resistance instruments are more convenient although somewhat less accurate. [Pg.245]

By the use of suitable shunts, the basic moving-coil movement can be adapted to measure m almost unlimited range of currents. Figure 10.46 illustrates a direct-indicating instrument with shunt, to measure current up to 5 A d.c. To ensure that the resistance of the circuit is not materially altered by the insertion of an ammeter, it is usual to install either a shunt or the meter itself (usually a moving-coil meter with internal shunt) permanently in the circuit. Ammeter shunts are normally of the four-terminal type, to avoid contact resistance errors, i.e. two current terminals and two potential terminals, as shown, in fig. J0.46. [Pg.249]

In making measurements of current flowing within a structure, it is extremely important that additional resistance, as for example a shunt, is not introduced into the circuit, as otherwise erroneous results will be obtained. One method is to use a tong test meter. Such instruments are, however, not particularly accurate, especially at low currents, and are obviously jmpracticablein thecaseof, say, a 750 mm diameter pipeline. A far moreaccurate method and onethat can beapplied to ail structures, isthe zero-resistance ammeter or, as it is sometimes called, the zero-current ammeter method. The basic circuit of such an instrument is shown in Fig. 10.47. [Pg.249]

The zero-resistance ammeter is seldom employed for routine testing. This instrument requires careful handling to avoid damage, in particular to the galvanometer. Usually two permanent test leads are installed at a set distance apart, and by the initial use of a zero-resistance ammeter a calibration chart of potential between the two leads and current in the structure is drawn up. Thus when routine testing is made, it is only necessary to measure the... [Pg.250]

As their name suggests, these instruments are capable of carrying out a variety of measurements, e.g. structure/electrolyte potentials, current, resistivity and voltage. Most instruments of this type contain two meters in one case, one being a low-resistance millivolt/voltmeter and milliamp/ammeter, and the second a high-resistance voltmeter. [Pg.255]

In more sophisticated instruments, the modern tendency is to replace the micro-ammeter by a digital read-out, and there is an increasing trend to use visual display units to show the results. Such instruments are controlled by microprocessors which may either show sequentially the successive operations which must be performed to measure the absorbance of a solution at a fixed wavelength or to observe the absorption spectrum of a sample alternatively the whole procedure may be automated. Such instruments will display the absorption spectrum on the VDU screen, and by linking to a printer, a permanent record is produced. [Pg.666]

The hot-wire anemometer is very accurate even for very low rates of flow. It is one of the most convenient instruments for the measurement of the flow of gases at low velocities accurate readings are obtained for velocities down to about 0.03 m/s. If the ammeter has a high natural frequency, pulsating flows can be measured. Platinum wire is commonly used. [Pg.265]

By this method the weighed dry product is dissolved in methanol and titrated with the Karl Fischer solution until the color changes from brown to yellow. The visual observation can be replaced by an ammeter, which shows an steep increase in current, when the titration is terminated (dead-stop-titration). The samples can be two to four times smaller than for the gravimetric method. To avoid the visual observation completely, iodine can be produced by electrolyzation and the water content is calculated by Coulomb s law. Such an apparatus (e. g. Fig. 1.97.1 and 1.97.2) is available commercially. The smallest amount of water to be detected by such instruments is 10 pg. Wekx and De Kleijn [1.84) showed, how the Karl Fischer method can be used directly in the vial with the dried product. The Karl... [Pg.111]

Electric Instruments, Many and various kinds of electric instruments are used in industrial processes and in analytical and research, laboratories, including those dealing with expls and ammunition. Among such instruments may be mentioned ammeters, voltmeters, galvanometers, thermocouples, ohm meters, wattmeters, frequency meters, oscillographs, computers, etc... [Pg.677]

Since the values of Rb R2, and R3 are known values, the only unkown is Rx. The value of Rx can be calulated for the bridge during an ammeter zero current condition. Knowing this resistance value provides a baseline point for calibration of the instrument attached to the bridge circuit. The unknown resistance, Rx, is given by Equation 1-2. [Pg.29]

Its successor, used for all early instruments designed to measure electrical currents, and still used in ammeters and voltmeters today, is the d Arsonval144 galvanometer (Fig. 10.28) developed by d Arsonal and Deprez145 in the 1880s, which consists of a permanent magnet B0/ within... [Pg.640]

Electrochemical calorimetry — is the application of calorimetry to thermally characterize electrochemical systems. It includes several methods to investigate, for instances, thermal effects in batteries and to determine the -> molar electrochemical Peltier heat. Instrumentation for electrochemical calorimetric studies includes a calorimeter to establish the relationship between the amount of heat released or absorbed with other electrochemical variables, while an electrochemical reaction is taking place. Electrochemical calorimeters are usually tailor-made for a specific electrochemical system and must be well suited for a wide range of operation temperatures and the evaluation of the heat generation rate of the process. Electrochemical calorimeter components include a power supply, a device to control charge and discharge processes, ammeter and voltmeter to measure the current and voltage, as well as a computerized data acquisition system [i]. In situ calorimetry also has been developed for voltammetry of immobilized particles [ii,iii]. [Pg.186]

The direct reading type of instrument, although possibly less accurate than potentiometric is also used exclusively in modem soil laboratories. The e.m.f of the glass electrode-calomel electrode cell is applied across a resistance, and the resulting current after amplification is passed through an ammeter causing deflection of the pointer across a scale marked in pH rmits. These instruments are available to operate on mains A.C. current. In most pH meters temperature control knob is provided to adjust at temperature of the test solution. [Pg.5]


See other pages where Instruments ammeter is mentioned: [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.339]    [Pg.1238]    [Pg.1019]    [Pg.727]    [Pg.44]    [Pg.295]    [Pg.505]    [Pg.641]    [Pg.369]    [Pg.26]    [Pg.437]    [Pg.1061]    [Pg.706]    [Pg.3]    [Pg.427]    [Pg.427]    [Pg.447]    [Pg.450]   
See also in sourсe #XX -- [ Pg.300 ]




SEARCH



Ammeters

© 2024 chempedia.info