Analysis method standardization

Single-operator characteristics are determined by analyzing a sample whose concentration of analyte is known to the analyst. The second step in verifying a method is the blind analysis of standard samples where the analyte s concentration remains unknown to the analyst. The standard sample is analyzed several times, and the average concentration of the analyte is determined. This value should be within three, and preferably two standard deviations (as determined from the single-operator characteristics) of the analyte s known concentration. [Pg.683]

The most useful methods for quality assessment are those that are coordinated by the laboratory and that provide the analyst with immediate feedback about the system s state of statistical control. Internal methods of quality assessment included in this section are the analysis of duplicate samples, the analysis of blanks, the analysis of standard samples, and spike recoveries. [Pg.708]

Analysis of Standards The analysis of a standard containing a known concentration of analyte also can be used to monitor a system s state of statistical control. Ideally, a standard reference material (SRM) should be used, provided that the matrix of the SRM is similar to that of the samples being analyzed. A variety of appropriate SRMs are available from the National Institute of Standards and Technology (NIST). If a suitable SRM is not available, then an independently prepared synthetic sample can be used if it is prepared from reagents of known purity. At a minimum, a standardization of the method is verified by periodically analyzing one of the calibration standards. In all cases, the analyte s experimentally determined concentration in the standard must fall within predetermined limits if the system is to be considered under statistical control. [Pg.710]

Some solid materials are very intractable to analysis by standard methods and cannot be easily vaporized or dissolved in common solvents. Glass, bone, dried paint, and archaeological samples are common examples. These materials would now be examined by laser ablation, a technique that produces an aerosol of particulate matter. The laser can be used in its defocused mode for surface profiling or in its focused mode for depth profiling. Interestingly, lasers can be used to vaporize even thermally labile materials through use of the matrix-assisted laser desorption ionization (MALDI) method variant. [Pg.280]

Analytical Methods. Molybdenum contents in ore concentrates and technical oxide are most accurately deterrnined gravimetricaHy by precipitating lead molybdate. Molybdenum content is usually not determined on pure compounds or metal. Instead, spectrographic methods are used to measure impurity elements that must be controlled. Carbon and oxygen in metal products are measured by standard gas analysis methods. [Pg.463]

Although the most sensitive line for cadmium in the arc or spark spectmm is at 228.8 nm, the line at 326.1 nm is more convenient to use for spectroscopic detection. The limit of detection at this wavelength amounts to 0.001% cadmium with ordinary techniques and 0.00001% using specialized methods. Determination in concentrations up to 10% is accompHshed by solubilization of the sample followed by atomic absorption measurement. The range can be extended to still higher cadmium levels provided that a relative error of 0.5% is acceptable. Another quantitative analysis method is by titration at pH 10 with a standard solution of ethylenediarninetetraacetic acid (EDTA) and Eriochrome Black T indicator. Zinc interferes and therefore must first be removed. [Pg.388]

Instrumental Quantitative Analysis. Methods such as x-ray spectroscopy, oaes, and naa do not necessarily require pretreatment of samples to soluble forms. Only reUable and verified standards are needed. Other instmmental methods that can be used to determine a wide range of chromium concentrations are atomic absorption spectroscopy (aas), flame photometry, icap-aes, and direct current plasma—atomic emission spectroscopy (dcp-aes). These methods caimot distinguish the oxidation states of chromium, and speciation at trace levels usually requires a previous wet-chemical separation. However, the instmmental methods are preferred over (3)-diphenylcarbazide for trace chromium concentrations, because of the difficulty of oxidizing very small quantities of Cr(III). [Pg.141]

Specifications for soHd i7j -l,4-polyisoprenes are shown in Table 5 and include analyses for volatile matter, extractables, ash, and Mooney viscosity at 100°C. Standard method ASTM D1416 includes chemical analysis methods for volatile matter, extractables, and total ash, while ASTM D1646 includes Mooney viscosity (82). The Monsanto rheometer data for vulcanizates prepared by a standard recipe may also be specified. This formulation for vulcanizate (ASTM D3403) is mixed in a Banbury mixer in two passes with all but sulfur and accelerator added in first pass ... [Pg.6]

The analytical characteristics of the soi ption-spectroscopic and soi ption-atomic-absoi ption methods developed have been checked up by the analysis of standard solutions, natural, artesian, mineral waters and food salt. [Pg.292]

With regard to the line 1, the development eould be noted for the following methods those, where peroxide derivatives of earboxylie aeids are used methods of multiple-wavelength speetrophotometrie analysis methods of quantitative aeeounting of a priori information methods of liquid ehromatography (a Unified adsorption eenter model and a Mobile phase effeetive eoneentration eoneeption an applieation of mieellar ehromatography standardization of TLC-plates). [Pg.339]

Whilst solving some ecological problems of metals micro quantity determination in food products and water physicochemical and physical methods of analysis are employed. Standard mixture models (CO) are necessary for their implementation. The most interesting COs are the ones suitable for graduation and accuracy control in several analysis methods. Therefore the formation of poly functional COs is one of the most contemporary problems of modern analytical chemistry. The organic metal complexes are the most prospective class of CO-based initial substances where P-diketonates are the most appealing. [Pg.405]

The method of Cu determination in which ligand is monometal Tb-L complex was used in analysis of standard samples of Zn and Ni alloys. Experimental detection limit for copper is 0.01 pg/mL. [Pg.453]

The failure determining stresses are also often loeated in loeal regions of the eomponent and are not easily represented by standard stress analysis methods (Sehatz et al., 1974). Loads in two or more axes generally provide the greatest stresses, and should be resolved into prineipal stresses (Ireson et al., 1996). In statie failure theory, the error ean be represented by a eoeffieient of variation, and has been proposed as C =0.02. This margin of error inereases with dynamie models and for statie finite element analysis, the eoeffieient of variation is eited as Q = 0.05 (Smith, 1995 Ullman, 1992). [Pg.192]

Measurement of specific chemical or biological contaminants can be very expensive. Before expending time and money to obtain measurements of indoor air pollutants, you must decide how the results will be used (e.g., comparison to standards or guidelines, comparison to levels in complaint-free areas) what substances(s) should be measured where to take, samples when to take samples what sampling and analysis method to use so that the results provide useful information. [Pg.225]

Tiible 1 Detection limits, working ranges and method standard deviation V for quantitative analysis of -blockers. [Pg.19]

Determination of appropriate measuring and analysis methods Assessment of standards and acceptable tolerances Interpretation of test results... [Pg.564]

Standards should be analyzed contemporaneously for both determinative and confirmatory procedures. The method developer needs to describe fully the preparation of all the standards and the calibration procedure to be used, such as calibration prior to sample analysis, interspersed standards, or bracketing standards (confirmatory only). [Pg.84]

In the first phase, the performance of the instrumentation used for the method is demonstrated. Based on the analysis of standards, results from the participating laboratory should meet the system suitability requirements of the method. Successful completion of this phase will qualify the analyst, his or her equipment, and the laboratory for the trial. Failure in the first phase does not usually cause a method to fail the trial. However, it can slow the process. When a procedure fails during the first phase of a trial, the sponsor may need to write a cautionary note in the SOP discussing recommended or inadequate types of instruments. To correct the problem, the participating laboratory analyst can substitute equipment that gives adequate performance alternatively, the sponsor must find a different laboratory to participate in the trial. [Pg.91]

Sample preparation, injection, calibration, and data collection, must be automated for process analysis. Methods used for flow injection analysis (FLA) are also useful for reliable sampling for process LC systems.1 Dynamic dilution is a technique that is used extensively in FIA.13 In this technique, sample from a loop or slot of a valve is diluted as it is transferred to a HPLC injection valve for analysis. As the diluted sample plug passes through the HPLC valve it is switched and the sample is injected onto the HPLC column for separation. The sample transfer time typically is determined with a refractive index detector and valve switching, which can be controlled by an integrator or computer. The transfer time is very reproducible. Calibration is typically done by external standardization using normalization by response factor. Internal standardization has also been used. To detect upsets or for process optimization, absolute numbers are not always needed. An alternative to... [Pg.76]

Musumarra et al. [43] identified miconazole and other drugs by principal components analysis of standardized thin-layer chromatographic data in four eluent systems. The eluents, ethylacetate methanol 30% ammonium hydroxide (85 10 15), cyclohexane-toluene-diethylamine (65 25 10), ethylacetate chloroform (50 50), and acetone with the plates dipped in potassium hydroxide solution, provided a two-component model that accounts for 73% of the total variance. The scores plot allowed the restriction of the range of inquiry to a few candidates. This result is of great practical significance in analytical toxicology, especially when account is taken of the cost, the time, the analytical instrumentation and the simplicity of the calculations required by the method. [Pg.44]

The ASTM (American Society for Testing and Materials) has published a Standard Practice for General Techniques for Qualitative Analysis (Method E 1252-88). The method describes techniques useful for qualitative evaluation of liquids, solids, and gases using the spectral measurement region of 4000 to 50 cm 1 (above 2500 nm) [1, 2],... [Pg.497]


See other pages where Analysis method standardization is mentioned: [Pg.47]    [Pg.1149]    [Pg.47]    [Pg.1149]    [Pg.538]    [Pg.683]    [Pg.699]    [Pg.813]    [Pg.201]    [Pg.485]    [Pg.36]    [Pg.424]    [Pg.605]    [Pg.1028]    [Pg.855]    [Pg.855]    [Pg.57]    [Pg.51]    [Pg.157]    [Pg.202]    [Pg.466]    [Pg.484]    [Pg.675]    [Pg.210]    [Pg.372]    [Pg.453]    [Pg.81]    [Pg.347]    [Pg.124]    [Pg.96]   
See also in sourсe #XX -- [ Pg.6 ]




SEARCH



Analysis of Old Lubricating Oil for Total Metal Content Using a Slurry Method with Internal Standard

Analysis strategies standard methods

Method standardization

Quantitative Analysis Using Internal Standard Method

Quantitative Analysis Using Standard Addition Method

Quantitative analysis external standard method

Quantitative analysis internal standard method

Quantitative analysis standard addition method

Standard addition method, application chemical analysis

Standard method

Standard methods of solvent analysis

Standardizing method

Thermal analysis standard test methods