Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Absorption pumps

The experimental configuration of the pump-probe experiment is similar to Ref. [5]. A home built non-collinear optical parametric amplifier (nc-OPA) was used as a pump, providing Fourier-transform-limited 30 fs pulses, which could be spectrally tuned between 480-560 nm. In all experiments white-light generated in a sapphire crystal using part of the fundamental laser (800 nm), was used as probe light. In the pump-probe experiments the pump was tuned to the S2 0-0 band for carotenoids with n>l 1. In the case of M9, it was not possible to tune the nc-OPA to its 0-0 transition, and hence another nc-OPA tuned to 900 nm was frequency doubled and used for excitation. In addition to conventional transient absorption pump-probe measurements, we introduce pump-deplete-probe spectroscopy, which is sensitive to the function of an absorbing state within the deactivation network. In this technique, we... [Pg.454]

All these studies with femtosecond pulses on the primary photochemical processes of rhodopsin were done by means of transient absorption (pump probe) spectroscopy [10]. However, absorption spectroscopy may not be the best way to probe the excited-state dynamics of rhodopsin, because other spectral features, such as ground-state depletion and product absorption, are possibly superimposed on the excited-state spectral features (absorption and stimulated emission) in the obtained data. Each spectral feature may even vary in the femtosecond time domain, which provides further difficulty in analyzing the data. In contrast, fluorescence spectroscopy focuses only on the excited-state processes, so that the excited-state dynamics can be observed more directly. [Pg.62]

Figure C3.5.4. Ensemble-averaged loss of energy from vibrationally excited I2 created by photodissoeiation and subsequent recombination in solid Kr, from [M ]. The inset shows calculated transient absorption (pump-probe) signals for irmer turning points at 3.5, 3.4 or 3.3 A. [Pg.3041]

Chilled water can be generated using heat pumps. They can be either absorption or mechanical chillers. Absorption chillers, which use waste heat (as a source of thermal energy) rather than mechanical/shaft energy for operation, can be economical. Heat source for an absorption chiller can be LLP steam (at 1.5 barg) or a hot process stream. Lithium bromide (LiBr) absorption pumps are frequently used due to their lower cost and application range down to the freezing point of water. [Pg.326]

There are two main types of heat pumps vapor-compression pumps and absorption pumps. Vapor-compression heat pumps are the more common of the two. [Pg.946]

Capturing and keeping the gas molecules (adsorption pump, absorption or reaction pump) - e.g. cryopump, sorption pump, ion pump, getter pump, absorption pump, getter pump. [Pg.110]

The resonance vector analysis has been used to explore all of the questions raised above on the fate of the polyad numbers in larger molecules, the most thoroughly investigated case so far probably being C2FI2- This molecule has been very extensively probed by absorption as well as stimulated emission pumping and dispersed fluorescence teclmiques [, 53, 70 and 71], the experimental spectra have been analysed in... [Pg.73]

Pollard W T, Lee S-Y and Mathies R A 1990 Wavepacket theory of dynamic absorption spectra in femtosecond pump-probe experiments J. Chem. Phys. 92 4012... [Pg.280]

At still shorter time scales other techniques can be used to detenuiue excited-state lifetimes, but perhaps not as precisely. Streak cameras can be used to measure faster changes in light intensity. Probably the most iisellil teclmiques are pump-probe methods where one intense laser pulse is used to excite a sample and a weaker pulse, delayed by a known amount of time, is used to probe changes in absorption or other properties caused by the excitation. At short time scales the delay is readily adjusted by varying the path length travelled by the beams, letting the speed of light set the delay. [Pg.1124]

An interferometric method was first used by Porter and Topp [1, 92] to perfonn a time-resolved absorption experiment with a -switched ruby laser in the 1960s. The nonlinear crystal in the autocorrelation apparatus shown in figure B2.T2 is replaced by an absorbing sample, and then tlie transmission of the variably delayed pulse of light is measured as a fiinction of the delay This approach is known today as a pump-probe experiment the first pulse to arrive at the sample transfers (pumps) molecules to an excited energy level and the delayed pulse probes the population (and, possibly, the coherence) so prepared as a fiinction of time. [Pg.1979]

Pump-probe absorption experiments on the femtosecond time scale generally fall into two effective types, depending on the duration and spectral width of the pump pulse. If tlie pump spectrum is significantly narrower in width than the electronic absorption line shape, transient hole-burning spectroscopy [101. 102. 103. 104. 105. 106. 107. 108. 109. 110. 111. 112 and 113] can be perfomied. The second type of experiment, dynamic absorption spectroscopy [57, 114. 115. 116. 117. 118. 119. 120. 121 and 122], can be perfomied if the pump and probe pulses are short compared to tlie period of the vibrational modes that are coupled to the electronic transition. [Pg.1979]

Figure B2.1.7 Transient hole-burned speetra obtained at room temperature with a tetrapyrrole-eontaining light-harvesting protein subunit, the a subunit of C-phyeoeyanin. Top fluoreseenee and absorption speetra of the sample superimposed with die speetnuu of the 80 fs pump pulses used in the experiment, whieh were obtained from an amplified CPM dye laser operating at 620 mn. Bottom absorption-diflferenee speetra obtained at a series of probe time delays. Figure B2.1.7 Transient hole-burned speetra obtained at room temperature with a tetrapyrrole-eontaining light-harvesting protein subunit, the a subunit of C-phyeoeyanin. Top fluoreseenee and absorption speetra of the sample superimposed with die speetnuu of the 80 fs pump pulses used in the experiment, whieh were obtained from an amplified CPM dye laser operating at 620 mn. Bottom absorption-diflferenee speetra obtained at a series of probe time delays.
So far we have exclusively discussed time-resolved absorption spectroscopy with visible femtosecond pulses. It has become recently feasible to perfomi time-resolved spectroscopy with femtosecond IR pulses. Flochstrasser and co-workers [M, 150. 151. 152. 153. 154. 155. 156 and 157] have worked out methods to employ IR pulses to monitor chemical reactions following electronic excitation by visible pump pulses these methods were applied in work on the light-initiated charge-transfer reactions that occur in the photosynthetic reaction centre [156. 157] and on the excited-state isomerization of tlie retinal pigment in bacteriorhodopsin [155]. Walker and co-workers [158] have recently used femtosecond IR spectroscopy to study vibrational dynamics associated with intramolecular charge transfer these studies are complementary to those perfomied by Barbara and co-workers [159. 160], in which ground-state RISRS wavepackets were monitored using a dynamic-absorption technique with visible pulses. [Pg.1982]

The detector D monitors the absorption of the probe beam as a function of the delay between the pulses given by xHc, where c is the speed of light and v is the difference between the optical path travelled by the probe and by the pump pulse. Adapted from [110],... [Pg.2127]

In a 1 litre round-bottomed flask provided with an efficient double surface condenser, place 40 g. (39 ml.) of aniline, 50 g. (40 ml.) of carbon sulphide CAUTION inflammable) (1), and 50 g. (63-5 ml.) of absolute ethyl alcohol (2). Set up the apparatus in the fume cupboard or attach an absorption device to the top of the condenser (see Fig. 11, 8, 1) to absorb the hydrogen sulphide which is evolved. Heat upon an electrically-heated water bath or upon a steam bath for 8 hours or until the contents of the flask sohdify. When the reaction is complete, arrange the condenser for downward distillation (Fig. 11, 13, 3), and remove the excess of carbon disulphide and alcohol (CA UTION inflammable there must be no flame near the receiver). Shake the residue in the flask with excess of dilute hydrochloric acid (1 10) to remove any aniline present, filter at the pump, wash with water, and drain well. Dry in the steam oven. The yield of crude product, which is quite satisfactory for the preparation of phenyl iao-thiocyanute (Section IV.95), is 40-45 g. Recrystalhse the crude thiocarbanihde by dissolving it, under reflux, in boiling rectified spirit (filter through a hot water funnel if the solution is not clear), and add hot water until the solution just becomes cloudy and allow to cool. Pure sj/m.-diphenylthiourea separates in colourless needles, m.p, 154°,... [Pg.642]

Equip a 1 litre bolt-head flask with dropi)ing fuuncl and a double surface reflux condenser to the top of the latter attach a device (e.g.. Fig. II, 8, 1. c) for the absorption of the hydrogen bromide evolved. Place 100 g. (108 ml.) of dry iso-valeric acid (Section 111,80) and 12 g. of pmified red phosphorus (Section 11,50,5) in the flask. Add 255 g. (82 ml.) of dry bromine (Section 11,49,5) slowly through the dropping funnel at such a rate that little or no bromine is lost with the hydrogen bromide evolved the addition occupies 2-3 hours. Warm the reaction mixture on a water bath until the evolution of hydrogen bromide is complete and the colour of the bromine has disappeared. Pour off the liquid reaction product into a Claisen flask and distil mider the reduced pressure of a water pump. Collect the a-bromo-wo-valeryl bromide at 117-122°/25-30 mm. The yield is 150 g. [Pg.999]

Depending on the method of pumping, the population of may be achieved by — Sq or S2 — Sq absorption processes, labelled 1 and 2 in Figure 9.18, or both. Following either process collisional relaxation to the lower vibrational levels of is rapid by process 3 or 4 for example the vibrational-rotational relaxation of process 3 takes of the order of 10 ps. Following relaxation the distribution among the levels of is that corresponding to thermal equilibrium, that is, there is a Boltzmann population (Equation 2.11). [Pg.360]

In order to prevent this occurring a pulsed method of pumping is used with a repetition rate low enough to allow time for Tj — Sq relaxation. For CW operation either Tj must be sufflciently short or another dye has to be used for which T2 — Ti absorption does not overlap with the fluorescence. [Pg.361]


See other pages where Absorption pumps is mentioned: [Pg.145]    [Pg.6]    [Pg.136]    [Pg.145]    [Pg.6]    [Pg.136]    [Pg.263]    [Pg.264]    [Pg.1591]    [Pg.1968]    [Pg.1971]    [Pg.1979]    [Pg.1980]    [Pg.1981]    [Pg.1982]    [Pg.1990]    [Pg.2116]    [Pg.2126]    [Pg.3027]    [Pg.244]    [Pg.430]    [Pg.738]    [Pg.791]    [Pg.840]    [Pg.961]    [Pg.1006]    [Pg.132]    [Pg.337]    [Pg.340]    [Pg.390]    [Pg.43]   
See also in sourсe #XX -- [ Pg.50 , Pg.144 ]




SEARCH



Absorption cycle heat pump

Absorption heat pump

Pump absorption coefficient

Pump probe absorption spectroscopy

Pump probe transient absorption

Pump probe transient absorption microscopy

Pump transient absorption scheme

Pump-probe absorption

Pump-probe electronic absorption spectroscopy

Transient absorption spectroscopy pump-probe measurement

Ultrafast dynamics pump-probe absorption

Vibrational relaxation pump-probe absorption

© 2024 chempedia.info