Flame photometry interferences

Polarography has also been applied to the determination of potassium in seawater [535]. The sample (1 ml) is heated to 70 °C and treated with 0.1 M sodium tetraphenylborate (1 ml). The precipitated potassium tetraphenylborate is filtered off, washed with 1% acetic acid, and dissolved in 5 ml acetone. This solution is treated with 3 ml 0.1 M thallium nitrate and 1.25 ml 2M sodium hydroxide, and the precipitate of thallium tetraphenylborate is filtered off. The filtrate is made up to 25 ml, and after de-aeration with nitrogen, unconsumed thallium is determined polarographically. There is no interference from 60 mg sodium, 0.2 mg calcium or magnesium, 20 pg barium, or 2.5 pg strontium. Standard eviations at concentrations of 375, 750, and 1125 pg potassium per ml were 26.4, 26.9, and 30.5, respectively. Results agreed with those obtained by flame photometry. [Pg.210]

Atomic absorption takes advantage of the fact that most of the atoms remain in the ground state, and are capable of absorbing radiation of the appropriate wavelength corresponding to Ah. Whereas a hot flame is preferred for flame photometry, a cooler flame is preferred for atomic absorption, except in cases where chemical interference may occur. [Pg.81]

Instrumentation. Flame Characteristics. Flame Processes. Emission Spectra. Quantitative Measurements and Interferences. Applications of Flame Photometry and Flame Atomic Emission Spectrometry. [Pg.9]

The basic instrumentation used for spectrometric measurements has already been described in the previous chapter (p. 277). Methods of excitation, monochromators and detectors used in atomic emission and absorption techniques are included in Table 8.1. Sources of radiation physically separated from the sample are required for atomic absorption, atomic fluorescence and X-ray fluorescence spectrometry (cf. molecular absorption spectrometry), whereas in flame photometry, arc/spark and plasma emission techniques, the sample is excited directly by thermal means. Diffraction gratings or prism monochromators are used for dispersion in all the techniques including X-ray fluorescence where a single crystal of appropriate lattice dimensions acts as a grating. Atomic fluorescence spectra are sufficiently simple to allow the use of an interference filter in many instances. Photomultiplier detectors are used in every technique except X-ray fluorescence where proportional counting or scintillation devices are employed. Photographic recording of a complete spectrum facilitates qualitative analysis by optical emission spectrometry, but is now rarely used. [Pg.288]

Atomic absorption spectrometry is one of the most widely used techniques for the determination of metals at trace levels in solution. Its popularity as compared with that of flame emission is due to its relative freedom from interferences by inter-element effects and its relative insensitivity to variations in flame temperature. Only for the routine determination of alkali and alkaline earth metals, is flame photometry usually preferred. Over sixty elements can be determined in almost any matrix by atomic absorption. Examples include heavy metals in body fluids, polluted waters, foodstuffs, soft drinks and beer, the analysis of metallurgical and geochemical samples and the determination of many metals in soils, crude oils, petroleum products and plastics. Detection limits generally lie in the range 100-0.1 ppb (Table 8.4) but these can be improved by chemical pre-concentration procedures involving solvent extraction or ion exchange. [Pg.333]

The results obtained with ISEs have been compared several times with those of other methods. When the determination of calcium using the Orion SS-20 analyser was tested, it was found that the results in heparinized whole blood and serum were sufficiently precise and subject to negligible interference from K and Mg ([82]), but that it is necessary to correct for the sodium error, as the ionic strength is adjusted with a sodium salt [82], and that a systematic error appears in the presence of colloids and cells due to complexa-tion and variations in the liquid-junction potential [76]. Determination of sodium and potassium with ISEs is comparable with flame photometric estimation [39, 113, 116] or is even more precise [165], but the values obtained with ISEs in serum are somewhat higher than those from flame photometry and most others methods [3, 25, 27, 113, 116]. This phenomenon is called pseudohyponatremia. It is caused by the fact that the samples are not diluted in ISE measurement, whereas in other methods dilution occurs before and during the measurement. On dilution, part of the water in serum is replaced by lipids and partially soluble serum proteins in samples with abnormally increased level of lipids and/or proteins. [Pg.132]

In environmental analysis, flame photometry is most widely used for the determination of potassium, which emits at 766.5 nm. It is also often used for the determination of sodium at 589.0 nm, although spectral interference problems (see Chapter 3) then may be encountered in the presence of excess calcium because of emission from calcium-containing polyatomic species. Molecular species are more likely to be found in cooler flames than in hotter flames. Some instruments use single, interchangeable filters, while others have three or more filters, for example for the determinations of potassium, sodium and lithium,... [Pg.24]

Sodium is still often determined by flame photometry, measuring the emission intensity of the doublet at around 589 nm, but care is necessary to make sure that excess calcium does not cause spectral interference (from molecular emission). This is unlikely to be a problem if AES is used, with a narrow spectral band-pass, and the intensity of emission at 589.0 nm from an air-acetylene flame is measured. However, at low determinant concentrations it is then advisable to add 2-5 mg ml 1 potassium or caesium as an ionization buffer. This is even more true if a nitrous oxide-acetylene flame is used for FES, although its use is rarely justified in environmental analyses because the additional sensitivity gained is rarely necessary. [Pg.89]

Each spectroscopic method has a characteristic application. For example, flame photometry is still applicable to the direct determination of Ca and Sr, and to the determination of Li, Rb, Cs and Ba after preconcentration with ion-exchange resin. Fluorimetry provides better sensitivities for Al, Be, Ga and U, although it suffers from severe interference effects. Emission spectrometry, X-ray fluorescence spectrometry and neutron activation analysis allow multielement analysis of solid samples with pretty good sensitivity and precision, and have commonly been applied to the analysis of marine organisms and sediments. Recently, inductively-coupled plasma (ICP)... [Pg.95]

Flame photometry (see also p. 168) is almost exclusively used for the determination of alkali metals because of their low excitation potential (e.g. sodium 5.14eV and potassium 4.34 eV). This simplifies the instrumentation required and allows a cooler flame (air-propane, air-butane or air-natural gas) to be used in conjunction with a simpler spectrometer (interference filter). The use of an interference filter allows a large excess of light to be viewed by the detector. Thus, the expensive photomultiplier tube is not required and a cheaper detector can be used, e.g. a photodiode or photoemissive detector. The sample is introduced using a pneumatic nebulizer as described for FAAS (p. 172). Flame photometry is therefore a simple, robust and inexpensive technique for the determination of potassium (766.5 nm) or sodium (589.0nm) in clinical or environmental samples. The technique suffers from the same type of interferences as in FAAS. The operation of a flame photometer is described in Box 26.2. [Pg.175]

Herrmann and Lang (H3) studied various atomizers and recorded best results with a laboratory-built high pressure vaporizer. No ionization interference was seen in an air-propane flame and calibration curves were straight from 1 to 10 mg sodium per liter. Determinations were performed on serum diluted 1 20-1 200 and results agreed well with those concurrently obtained by emission flame photometry. [Pg.39]

The development of fast and accurate procedures for the determination of calcium in biological materials represents one of the important early achievements of atomic absorption spectroscopy. The diflBculties encountered with calcium in emission flame photometry are well known (Dll, L6, S6, SIO), but spectral interferences and extreme dependency on flame temperature, serious obstacles in emission, are either nonexistent or of lower importance in absorption. Chemical interferences, however. [Pg.41]

F6. Foster, W. H., and Hume, D. N., Mutual cation interference effects in flame photometry. Anal. Chem. 31, 2033-2036 (1959). [Pg.57]

F8. Fukushima, S., Mechanism and elimination of interferences in flame photometry. Mikrochim. Acta, pp. 596-618 (1959). [Pg.57]

P4. Porter, P., and Wyld, G., Elimination of interferences in flame photometry. Anal. Chem. 27, 732-736 (1955). [Pg.60]

Spector, J., Mutual interferences in elimination of calcium interference in flame photometry. And. Chem. 27, 1452-1455 (1955). [Pg.61]

The method of preparing the sample for assay depends upon the concentration of the element, the sensitivity of the spectrochemical procedure, and any interference phenomena arising from other constituents of the sample. In flame photometry, the sample is usually an aqueous solution, but in other methods it may be necessary to dry the sample on some form of support. With trace elements or micro samples, great care is needed to avoid contamination during sample collection and preparation. Frequently, elements are bound to protein or other organic matrices. As spectrochemical methods measure total element, sample preparation must be used to separate the metal into its various fractions. [Pg.311]

Filter monochromators are now used almost only for flame photometry. They make use of interference filters, which may have a fairly low spectral bandpass (less than a few nm). However, it is also possible to use such filters for dynamic measurements of line and background intensities, and for transient signals, as occur in gas chromatography. The use of oscillating filters has been described, where the wavelength bandpass is slightly shifted by inclining them towards the radiation beam [65]. [Pg.70]


See other pages where Flame photometry interferences is mentioned: [Pg.591]    [Pg.155]    [Pg.118]    [Pg.147]    [Pg.232]    [Pg.441]    [Pg.358]    [Pg.370]    [Pg.551]    [Pg.106]    [Pg.39]    [Pg.48]    [Pg.363]    [Pg.287]    [Pg.515]    [Pg.23]    [Pg.26]    [Pg.27]    [Pg.27]    [Pg.30]    [Pg.55]    [Pg.298]    [Pg.314]    [Pg.317]    [Pg.1762]    [Pg.986]    [Pg.199]   