Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Some instruments

Other instrumental advantages include its high sensitivity and a linear mass scale to m/z 10,000 at full sensitivity. The linearity of the mass scale means that it is necessary to calibrate the spectrometer using a single or sometimes two known mass standards. Some calibration is necessary because the start of the mass scale is subject to some instrumental zero offset. The digitized accumulation of spectra provides a better signal-to-noise ratio than can be obtained from one spectrum alone. [Pg.167]

Composition The law of mass aclion is expressed as a rate in terms of chemical compositions of the participants, so ultimately the variation of composition with time must be found. The composition is determined in terms of a property that is measured by some instrument and cahbrated in terms of composition. Among the measures that have been used are titration, pressure, refractive index, density, chromatography, spectrometry, polarimetry, conduclimetry, absorbance, and magnetic resonance. In some cases the composition may vary linearly with the observed property, but in every case a calibration is needed. Before kinetic analysis is undertaken, the data are converted to composition as a function of time (C, t), or to composition and temperature as functions of time (C, T, t). In a steady CSTR the rate is observed as a function of residence time. [Pg.707]

Lateral resolution Better than 0.2 nm on some instruments... [Pg.10]

The sensor is the element of an instrument directly influenced by the measured quantity. In temperature measurement the thermal mass (capacity) of the sensor usually determines the meter s dynamics. The same applies to thermal anemometers. In IR analyzers used for concentration measurement, the volume of the flow cell and the sample flow rate are the critical factors. Some instruments, like sound-level meters, respond very fast, and follow the pressure changes up to several kHz. [Pg.1132]

A laser diffraction spectrometer can measure particles as small as 0.2 pm and up to about 1000 pm. Some instruments allow the operation of the analyzer for... [Pg.1294]

DATA BOUNDARY Pumps, valves, electrical, some instrumentation... [Pg.125]

A number of techniques have been developed since the original material was written . Some instrumentation/transducer developments also merit comment. [Pg.37]

The low-concentration eluants used to separate the sample ions on the separator column allow a substantial number of samples (typically about 50) to be analysed before the suppressor column is completely exhausted. Clearly an important practical consideration is the need to minimise the frequency of regeneration of the suppressor column and, for this reason, the specific capacity of the column is made as large as possible by using resins of moderate to high cross-linking. Some instruments contain two suppressor columns in parallel,... [Pg.199]

The Nernst equation shows that the glass electrode potential for a given pH value will be dependent upon the temperature of the solution. A pH meter, therefore, includes a biasing control so that the scale of the meter can be adjusted to correspond to the temperature of the solution under test. This may take the form of a manual control, calibrated in 0 C, and which is set to the temperature of the solution as determined with an ordinary mercury thermometer. In some instruments, arrangements are made for automatic temperature compensation by inserting a temperature probe (a resistance thermometer) into the solution, and the output from this is fed into the pH meter circuit. [Pg.566]

Because an increase in resolution causes a decrease in sensitivity, it is best to operate at the lowest resolution commensurate with good results. Some instrument data systems will allow calibration with an external reference material such as perfluorokerosene and then use of a secondary reference material for the internal mass reference. Tetraiodothiophene, vaporized using the solids probe inlet, is recommended as the secondary reference. The accurate masses are 79.9721, 127.9045, 162.9045, 206.8765, 253.8090, 293.7950, 333.7810, 460.6855, and 587.5900. For a higher mass standard, use hexaiodobenzene. Because the mass defect for these internal reference ions are so large, a resolution of 2000 is ample to separate these ions from almost any sample ions encountered in GC/MS. [Pg.375]

Annual checks of all instruments, recorders, and controls are required. Recalibration of some instruments may be required at more frequent intervals. [Pg.121]

SPM force curves are acquired by moving the tip toward the sample and recording the cantilever deflection as a function of the so-called Z position. Cantilever deflection is directly proportional to the force exerted on the sample by the tip. If the spring constant (fc) of the cantilever is known, the force can be calculated. The Z position defines the distance from the sample to the piezo, to which the base of the cantilever is attached (Figure 9.13). By convention the closest point of approach by the piezo is designated as zero on the x-axis. Note that for some instruments the piezo is attached to the sample stage and thus moves the sample up toward the tip however, this does not change the analysis. [Pg.266]

The complex situation The analysis involves some instruments (e.g., dissolution apparatus, HPLC, balance), dozens of samples, and several analysts, and could take days to perform. If nobody notices anything particular about the individual snippets of work or any of the results, no observations of probable operator error are documented. When the supervisor then inserts number after unspectacular number into the formula, all results turn out to be within specification limits, with the exception of one or two that are barely outside. But, no documentation, no operator error, just inexplicable OOS, full investigation ... [Pg.265]

Instruments. Some instruments are checked out to students only a few times each semester. These may include pH meters, small spectrophotometers, and other items. They are in storage... [Pg.12]

The purchase price of some instruments may include set-up and check-out by a manufacturer s representative. In such cases, the laboratory operator should only check for physical condition of a shipment and call the representative about its arrival. Other equipment will generally be set up at its permanent location as soon as it is unpacked and checked. Supplies will be taken to their designated storage areas. The accompanying warranty cards should be filled out and mailed as soon as possible. [Pg.106]

For some instruments, the manual may be needed with the instrument for proper operation. This is particularly true in educational laboratories. In such cases, the manufacturer vdll generally be happy to supply extra copies, or the significant parts of a manual may be copied on the office copy machine. In all cases, the original manual and instructions must be properly filed. [Pg.106]

Some instruments come with special tools for adjustment or routine maintenance. These should normally be kept with their instruments for ready access. Using them for other purposes in other locations should be discouraged. In an educational laboratory, such tools should be labeled and kept in the stockroom unless absolutely needed dining routine work. [Pg.133]

Monochromators. Replica gratings and narrow band-pass filters are used commonly, more so than quartz prisms. Computer control of the monochromator is available in some instruments, so that optimum intensity at the desired wavelength or maximum absorption by the examined substance can be obtained. [Pg.177]

What if instruments could be designed which were remoteiy self-calibrating with respect to calibration services of a National Metrology Institute (NMI) While this may sound very far-fetched , it is not difficult to speculate that future application of telemetry, transducers, intrinsic standards, detectors and computers will link some instruments directly to a NMI to obviate their need for RMs (Rasberry 1998). [Pg.282]

All MC-ICPMS instruments are equipped with a multiple Faraday collector array oriented perpendicular to the optic axis, enabling the simultaneous static or multi-static measurement of up to twelve ion beams. Most instruments use Faraday cups mounted on motorized detector carriers that can be adjusted independently to alter the mass dispersion and obtain coincident ion beams, as is the approach adopted for MC-TIMS measurement. However, some instruments instead employ a fixed collector array and zoom optics to achieve the required mass dispersion and peak coincidences (e.g., Belshaw et al. 1998). [Pg.43]

Most of the hplc instrumentation now in use is unsuitable for small bore columns. At the moment, the technique is used mainly in the applications laboratories of some instrument manufacturers (they are interested in selling it ). The method is potentially attractive in areas where sample sizes are very limited, for example in biochemical or life sciences applications, but whether or not it becomes widely accepted remains to be seen. [Pg.45]


See other pages where Some instruments is mentioned: [Pg.173]    [Pg.191]    [Pg.276]    [Pg.283]    [Pg.766]    [Pg.99]    [Pg.626]    [Pg.1137]    [Pg.1295]    [Pg.116]    [Pg.239]    [Pg.411]    [Pg.566]    [Pg.6]    [Pg.4]    [Pg.186]    [Pg.118]    [Pg.62]    [Pg.104]    [Pg.104]    [Pg.54]    [Pg.538]    [Pg.348]    [Pg.367]    [Pg.305]    [Pg.351]    [Pg.386]    [Pg.394]    [Pg.178]    [Pg.236]   


SEARCH



© 2024 chempedia.info