Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical Preparative

These separations are run at 1.0-2.0 mL/min on the same 4.5-mm x 25-cm column used for our analytical runs. Normally, in analysis we shoot from picogram to nanogram quantities. Most separations maintain their resolution until we reach an injection quantity of about 1 fig. The valleys between peaks begin to rise indicating some overlapping. [Pg.138]

If we increase our first peak k to 8-10, we can increase the interpeak gap allowing us to load to about 10,ug of compound/injection. As we increase the amount of sample, we need to go to lower detector sensitivity. We can increase flow rate to 2.0 mL/min, but we will lose some resolution by doing so. Generally, we have no problem increasing sample concentration and keeping the same injection loop size. If necessary, we can increase to the next size larger loop without affecting resolution. [Pg.138]


A standard solution containing a known amount of analyte, prepared separately from samples containing the analyte. [Pg.109]

Although isotope-dilution analysis can be very accurate, a number of precautions need to be taken. Some of these are obvious ones that any analytical procedure demands. For example, analyte preparation for both spiked and unspiked sample must be as nearly identical as possible the spike also must be intimately mixed with the sample before analysis so there is no differential effect on the subsequent isotope ration measurements. The last requirement sometimes requires special chemical treatment to ensure that the spike element and the sample element are in the same chemical state before analysis. However, once procedures have been set in place, the highly sensitive isotope-dilution analysis gives excellent precision and accuracy for the estimation of several elements at the same time or just one element. [Pg.366]

CO2 extraction has been prevalent for the isolation of essential oils and other natural lipophilic pigments like carotenoids. Hot water and superheated water extraction methods are used for analytical preparation of polar pigments. The technique is commonly referred to as subcritical water extraction because the practitioners of this approach come from SEE backgrounds. [Pg.305]

From the stock solutions of the individual native analytes, prepare a mixed fortification solution at 50 pgmL in methanol. Take an aliquot of this solution and dilute with aqueous methanol (1 1, v/v) to obtain a 0.25 p.gmL solution that will be used for... [Pg.1322]

The introduction of high-resolution, high-efficiency /-ray detectors composed of lithium-drifted germanium crystals has revolutionised /-measurement techniques. Thus, /-spectrometry allows the rapid measurement of relatively low-activity samples without complex analytical preparations. A technique described by Michel et al. [25] uses Ge(Li) /-ray detectors for the simultaneous measurements of 228radium and 226radium in natural waters. This method simplifies the analytical procedures and reduces the labour while improving the precision, accuracy, and detection limits. [Pg.347]

A control is a standard solution of the analyte prepared independently, often by other laboratory personnel, for the purpose of cross-checking the analyst s work. If the concentration found for such a solution agrees with the concentration it is known to have (within acceptable limits based on statistics), then this increases the confidence a laboratory has in the answers found for the real samples. If, however, the answer found differs significantly from the concentration it is known to have, then this signals a problem that would not have otherwise been detected. The analyst then knows to scrutinize his or her work for the purpose of discovering an error. [Pg.164]

Many of the parameters above have a direct influence on the efficiency of CLEC systems. Hence, they must be considered if the full analytical preparative potential of CLEC is to be utilized. [Pg.216]

Speed, load, and resolution are the three trade-off considerations that must be balanced to optimize the three levels of preparative runs (Fig. 11.1). Analytical preparative is concerned with isolation of up to microgram quantities of material and with obtaining enough material for spectrometric analysis the most important factors are speed and resolution. [Pg.137]

The laboratory quality control program has several components documentation of standard operating procedures for all analytical methods, periodic determination of method detection levels for the analytes, preparation of standard calibration curves and daily check of calibration standards, analysis of reagent blank, instrument performance check, determination of precision and accuracy of analysis, and preparation of control charts. Determination of precision and accuracy of analysis and method detection limits are described under separate subheadings in the following sections. The other components of the quality control plan are briefly discussed below. [Pg.22]

Figure 6.17 The schematic of the parallel analytical/preparative LC/MS system devised by Zeng and Kassel. (Reprinted with permission from Zeng and Kassel, 1998. Copyright 1998 American Chemical Society.)... Figure 6.17 The schematic of the parallel analytical/preparative LC/MS system devised by Zeng and Kassel. (Reprinted with permission from Zeng and Kassel, 1998. Copyright 1998 American Chemical Society.)...
Zeng, L. Burton, L. Yung, K. Shushan, B. Kassel, D. B. 1998b. Automated analytical/preparative high-performance liquid chromatography-mass spectrometry system for the rapid characterization and purification of compound libraries. /. Chromatogr. A, 794,3-13. [Pg.234]

The activities of some isotopes, in particular °Sr- °Y, can also be detected by liquid-crystal spectrometry with the use of the Cherenkov phenomenon [10, 11]. The Cherenkov effect is used to determine beta isotopes emitting particles whose iiniax IS above 500 keV [12]. The main advantage of beta activity determination by the Cherenkov effect is the use of analytical preparation used for another chemical analysis (e.g. calculation of recovery). Moreover, the addition of low energy beta or alpha radiation does not disturb the measurement, thereby lowering the cost of analysis. The weakness of this method is the decreased recovery registration and the decline in information about the realistic appearance of the beta spectrum [13]. The determination of beta isotopes in environmental samples is very difficult and requires their chemical isolation. The type of sample and the time of chemical analysis determine the choice of analytical method. Also, the time between contamination and sample collection is important procedures used for samples recently contaminated are different to those used for old samples in which the decay of short-lived radionuclides has aheady taken place [1, 5]. [Pg.435]

Given a sample to analyse for a known or suggested compound or group of compounds, how is a HPLC procedure, whether analytical, preparative or a combination, established The following lists suggest some of the preliminary observations the analyst should make. [Pg.208]

Accuracy in MS depends on calibration of the mlz axis of the spectra. Calibration is typically performed using separate samples or solutions of the calibrant studied under identical instrumental conditions as the analyte. Calibrants are selected based on the ionization technique and the mlz range required for the analyte typical calibrants used for soft ionization methods with biomolecules include bovine trypsin fragments, bovine insulin (5734 Da), cytochrome c (12,361 Da), horse myoglobin (16,951 Da), lysozyme (14,317 Da), and propylene glycols. Separate solutions of the analyte prepared in the absence and presence of a calibrant species may also be used. Table 15.6 shows typical calibration peaks found with bovine trypsin autolysis fragments.15... [Pg.308]

Discontinuous rectification is characterised by non-recurring feed and a separation column. To generate a reflux, an overhead condenser is inserted and a cold trap cools the obtained distillate. For analytical-preparative purposes in the laboratory, an apparatus with annulus columns has proved to be successful. This apparatus has a vacuum of up to 0.01 mbar, an output of 1-50 ml/h and low hold-up volume. Larger rectification units have tray columns, sieve plates and packing material with a vacuum of up to 0.1 mbar and outputs of several litres per hour. [Pg.90]

For analytical LC, standard packed columns (4-8 mm I.D.), capillary packed colunms (50-100 pm I.D.), and microbore packed colunms (0.5-1.5 mm I.D.) are used widely. Colunm sizes depend on the application, e.g., analytical, preparative, or commercial separations. Special configurations also exist, including membrane chromatography modules (stacks and hollow fibers) that offer lower pressure drops and easier scale-up than packed beds. [Pg.486]

The development of the first HPLC system with MW-triggered fraction collection was described by Zeng et al. [50] at CombiChem. This system was a dual analytical-preparative instrument with parallel-column format, termed parallel Analyl/PrepLCMS." developed by the modification of commercially available instrumentation. This system had software-controlled valves that applied sample to each path from a single autosampler and had the capacity to purify and analyze more than 100 samples per day. Initial analytical LC-MS data acquired by the system allowed the identification of samples that require... [Pg.195]

Fast LC-MS methods have been used to assess library quantity and purity, as well as to triage purification of compounds. Zeng et al. [51] developed one of the first fully automated analytical/preparative LC-MS systems for the characterization and purification of compound libraries derived by parallel synthesis. The system incorporated fast, reverse-phase LC/ESI-MS analysis (5-10 minutes). Post-data-acquisition purity assessment of compound ti-braries was performed automatically with software control. Compounds that were below a threshold level of purity were automatically purified with HPLC. The real-time purity assessment eliminated the need for postpurification analysis or pooling of fractions collected. [Pg.202]

The MS instrumentation is the most expensive part of the LC-MS system, hence efforts to improve the throughput of the LC-MS analysis often involve the use of parallel multiple columns that feed into a single mass spectrometer. Zeng and Kassel [99] developed an automated parallel analytical/preparative LC-MS workstation to increase the throughput for the characterization and purification of combinatorial libraries. The system incorporates two columns operated in parallel for both LC-MS analytical and preparative LC-MS purifications. A multiple-sprayer ESI interface was designed to support flows from multiple columns. The system is under complete software control and delivers the crude samples to the two HPLC columns from a single autosampler. The authors demonstrated characterization of more than 200 compounds per instrument per day, and purification of more than 200 compounds per instrument per night. De Biasi et al. [100] described a four-channel multiplexed... [Pg.205]

Wu, J.T. Zeng, H. Deng, Y. Unger, S.E. Automated Analytical/Preparative High-performance Liquid Chromatography-Mass Spectrometry System for the Rapid Characterization and Purification of Compound Libraries, Rapid Commun. Mass Spectrom. 15, 1113-1119 (2001). [Pg.222]

Wiechert U, Hoefs J (1995) An excimer laser-based micro analytical preparation technique for in situ oxygen isotope analysis of silicate and oxide minerals. Geochim Cosmochim Acta 59 4093-4101 Wright K, Freer R, Catlow CRA (1995) Oxygen diffusion in grossular and some geological implications. Am Mineral 80 1020-1025... [Pg.412]

Witherup, K.M. Look, S.A. Stasko, M.W. Ghiorzi, T.J. Muschik, G.M. Cragg, G.M. Taxus spp. needles contain amounts of taxol comparable to the bark of Taxus brevifolia analysis and isolation. J.Nat.Prod., 1990, 53, 1249-1255 [plants analytical preparative]... [Pg.1084]

There are two common ways to operate SFE, in online mode or offline mode. In the online mode, the outlet of the SFE instrument is directly hnked to an analytical instrument. Direct coupling with different chromatographic techniques allows simultaneous analyte preparation, separation and determination. The main drawback of the online extraction method is the limited sample size. In the offline mode, the extracted analytes are trapped and later, the extracted analytes can be analyzed by means of different chromatographic techniques. The offline collection system is chosen according to the extracted analyte characteristics. The use of online methods reduces possible errors and sources of contamination. Studies on trapping methods have been performed to optimize the corresponding parameters. ... [Pg.63]


See other pages where Analytical Preparative is mentioned: [Pg.772]    [Pg.440]    [Pg.930]    [Pg.19]    [Pg.559]    [Pg.90]    [Pg.138]    [Pg.103]    [Pg.254]    [Pg.266]    [Pg.295]    [Pg.364]    [Pg.550]    [Pg.249]    [Pg.323]    [Pg.108]    [Pg.204]    [Pg.192]    [Pg.198]    [Pg.178]   


SEARCH



Analytical Uses of Ultrasound Prior to Sample Preparation

Analytical and Preparative Separation

Analytical aspects sample preparation

Analytical electron microscopy specimen preparation

Analytical methods herbal preparations

Analytical preparative chromatography

Analytical preparative-scale batch chromatographs

Analytical procedures preparation of reagents

Microarrays preparing analytes

Preparation of Special Analytical Reagents

Preparation of analytical samples

Preparative and analytical methods

Preparative chromatography analytical HPLC

Recommended procedures for preparation of analytical samples

Rough preparation of the analytical sample

Sample Preparation Techniques in Analytical Chemistry, Edited by Somenath Mitra

Sample preparation analyte/matrix

Sample preparation analytical equipment

Sampling and preparation of analytical samples

Sodium Azide, Plant Analytical Procedures Analysis of Technical SA Prepared from

© 2024 chempedia.info