Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

The Spectrometer

Permanent magnets generate a very stable magnetic field, but they should be well thermally insulated, and they are very heavy. They have been the most commonly used instruments until mid 1970s. Electromagnets, are lighter. [Pg.220]

Although standard IR spectrometers are used for studying the amide bands, FTIR spectrometers are more accurate and reliable. FT-IR spectrophotometers are based upon the Michelson interferometer. A typical instrument (Fig. 7.1) comprises an optical bench housing the interferometer, sample, infrared source and detector, coupled to a computer, which controls the spectral scanning, analysis and data processing (for review see Griffiths, 1980). [Pg.210]

The Michelson interferometer consists simply of two mutually perpendicular plane mirrors one of which is fixed and the other able to move at 90° to its plane. A semi-reflecting film or beamsplitter  [Pg.210]

This chapter is a guided tour of the standard EPR spectrometer. The goal is not to give a rigorous description of the underlying physics, but to develop a feel for basic parts and principles sufficient to make you an independent, intelligent operator of any X-band machine. [Pg.9]

Now that the principles of NMR spectroscopy have been introduced, we will see how NMR spectra of the two most common nuclei—hydrogen and carbon-13—are obtained. The principles described for carbon-13 are applicable to many other spin- /2 nuclei, such as nitrogen-15, fluorine-19, silicon-29, and phosphorus-31. Topics to be discussed include the components of a typical NMR spectrometer, preparation of a sample, signal optimization techniques, spectral acquisition, selection of processing parameters, spectral presentation, and calibration of the spectrometer. [Pg.31]

Modern spectrometers typically have a dedicated acquisition controller/processor, plus a main computer. The latter is usually UNIX based with as much RAM and as much permanent data storage as possible. Ideally, one or more workstations are linked to the host computer to permit data processing and spectral plotting away from the spectrometer. In addition, a recorder to display the signals and a printer to list spectral parameters, chemical shifts, coupling constants, and integral values (all described later) are usually located next to the main computer and monitor. [Pg.32]

If a particular sample is in short supply and solubility is not a problem, microtubes (120-150 p-1) and submicrotubes (25-30 xl) can be used. With their receiver coils placed very close to the small, but concentrated, samples, these microprobes are excellent at scavenging their signals. Conversely, the use of wider diameter tubes, such as 10 or 15 mm, is appropriate for (i) a relatively large amount of sample that can be readily put in solution, (ii) a relatively small quantity of sample that cannot be adequately dissolved, and (iii) the experimental examination of low-sensitivity nuclei for which microprobes have not been developed. For commonly studied nuclei, microtubes should be considered when one is sample limited, while large-diameter tubes should be employed when one is solubility limited. [Pg.32]


X-ray spectrometer An apparatus used in the X-ray study of crystals in which a fine beam of monochromatic X-rays impinges at a measured angle on the face of a crystal mounted in its path, and in which the intensity of the X-rays diffracted in various directions by the crystal is measured with an ionization chamber mounted on an arm of the spectrometer table, or is recorded photographically. [Pg.429]

Light sources can either be broadband, such as a Globar, a Nemst glower, an incandescent wire or mercury arc lamp or they can be tunable, such as a laser or optical parametric oscillator (OPO). In the fomier case, a monocln-omator is needed to achieve spectral resolution. In the case of a tunable light source, the spectral resolution is detemiined by the linewidth of the source itself In either case, the spectral coverage of the light source imposes limits on the vibrational frequencies that can be measured. Of course, limitations on the dispersing element and detector also affect the overall spectral response of the spectrometer. [Pg.1162]

The absolute measurement of areas is not usually usefiil, because tlie sensitivity of the spectrometer depends on factors such as temperature, pulse length, amplifier settings and the exact tuning of the coil used to detect resonance. Peak intensities are also less usefiil, because linewidths vary, and because the resonance from a given chemical type of atom will often be split into a pattern called a multiplet. However, the relative overall areas of the peaks or multiplets still obey the simple rule given above, if appropriate conditions are met. Most samples have several chemically distinct types of (for example) hydrogen atoms within the molecules under study, so that a simple inspection of the number of peaks/multiplets and of their relative areas can help to identify the molecules, even in cases where no usefid infonnation is available from shifts or couplings. [Pg.1442]

The principal dilTerence from liquid-state NMR is that the interactions which are averaged by molecular motion on the NMR timescale in liquids lead, because of their anisotropic nature, to much wider lines in solids. Extra infonnation is, in principle, available but is often masked by the lower resolution. Thus, many of the teclmiques developed for liquid-state NMR are not currently feasible in the solid state. Furthemiore, the increased linewidth and the methods used to achieve high resolution put more demands on the spectrometer. Nevertheless, the field of solid-state NMR is advancing rapidly, with a steady stream of new experiments forthcoming. [Pg.1466]

Another problem in many NMR spectrometers is that the start of the FID is corrupted due to various instrumental deadtimes that lead to intensity problems in the spectrum. The spectrometer deadtime is made up of a number of sources that can be apportioned to either the probe or the electronics. The loss of the initial part of the FID is manifest in a spectrum as a rolling baseline and the preferential loss of broad components of... [Pg.1471]

M continually decreases under the influence of spin-spin relaxation which destroys the initial phase coherence of the spin motion within they z-plane. In solid-state TREPR, where large inliomogeneous EPR linewidths due to anisotropic magnetic interactions persist, the long-time behaviour of the spectrometer output, S(t), is given by... [Pg.1566]

The low MW power levels conuuonly employed in TREPR spectroscopy do not require any precautions to avoid detector overload and, therefore, the fiill time development of the transient magnetization is obtained undiminished by any MW detection deadtime. (3) Standard CW EPR equipment can be used for TREPR requiring only moderate efforts to adapt the MW detection part of the spectrometer for the observation of the transient response to a pulsed light excitation with high time resolution. (4) TREPR spectroscopy proved to be a suitable teclmique for observing a variety of spin coherence phenomena, such as transient nutations [16], quantum beats [17] and nuclear modulations [18], that have been usefi.il to interpret EPR data on light-mduced spm-correlated radical pairs. [Pg.1566]

Each vibrational peak within an electronic transition can also display rotational structure (depending on the spacing of the rotational lines, the resolution of the spectrometer, and the presence or absence of substantial line broadening effects such as... [Pg.415]

Finally, values of sx are directly proportional to transmittance for indeterminate errors due to fluctuations in source intensity and for uncertainty in positioning the sample cell within the spectrometer. The latter is of particular importance since the optical properties of any sample cell are not uniform. As a result, repositioning the sample cell may lead to a change in the intensity of transmitted radiation. As shown by curve C in Figure 10.35, the effect of this source of indeterminate error is only important at low absorbances. This source of indeterminate errors is usually the limiting factor for high-quality UV/Vis spectrophotometers when the absorbance is relatively small. [Pg.411]

In many applications in mass spectrometry (MS), the sample to be analyzed is present as a solution in a solvent, such as methanol or acetonitrile, or an aqueous one, as with body fluids. The solution may be an effluent from a liquid chromatography (LC) column. In any case, a solution flows into the front end of a mass spectrometer, but before it can provide a mass spectrum, the bulk of the solvent must be removed without losing the sample (solute). If the solvent is not removed, then its vaporization as it enters the ion source would produce a large increase in pressure and stop the spectrometer from working. At the same time that the solvent is removed, the dissolved sample must be retained so that its mass spectrum can be measured. There are several means of effecting this differentiation between carrier solvent and the solute of interest, and thermospray is just one of them. Plasmaspray is a variant of thermospray in which the basic method of solvent removal is the same, but the number of ions obtained is enhanced (see below). [Pg.71]

Other instrumental advantages include its high sensitivity and a linear mass scale to m/z 10,000 at full sensitivity. The linearity of the mass scale means that it is necessary to calibrate the spectrometer using a single or sometimes two known mass standards. Some calibration is necessary because the start of the mass scale is subject to some instrumental zero offset. The digitized accumulation of spectra provides a better signal-to-noise ratio than can be obtained from one spectrum alone. [Pg.167]

Example of a mass spectrum showing the peaks (or lines) corresponding to ions measured at various m/z values by the spectrometer the heights of the peaks relate to the abundance of the ions. [Pg.226]

Of course, some substances are sufficiently volatile that a heated inlet line can be used to get them into a mass spectrometer. Even here, there are practical problems. Suppose a liquid or solid is sufficiently volatile, that heating it to 50°C is enough to get the vapor into the mass spectrometer through a heated inlet line. If the mass spectrometer analyzer is at 30°C, there is a significant possibility that some of the sample will condense onto the inner walls of the spectrometer and slowly vaporize from there. If the vacuum pumps cannot remove this vapor quickly, then the mass... [Pg.278]

When a mass spectrum has been acquired by the spectrometer/computer system, it is already in digital form as m/z values versus peak heights (ion abundances), and it is a simple matter for the computer to compare each spectrum in the library with that of the unknown until it finds a match. The shortened search is carried out first, and the computer reports the best fits or matches between the unknown and spectra in the library. A search of even 60,000 to 70,000 spectra takes only a few seconds, particularly if transputers are used, thus saving the operator a great deal of time. Even a partial match can be valuable because, although the required structure may not have been found in the library, it is more than likely that some of the library compounds will have stractural pieces that can be recognized from a partial fit and so provide information on at least part of the structure of the unknown. [Pg.323]

Before ions actually pass into the spectrometer analyzer, there is usually a second drying stage to remove any final solvent. [Pg.391]

The spectrometer provides a mass spectrum of the ions, some of which come from anything dissolved in the solution or matrix (solute ions) and some from the matrix solvent itself. [Pg.394]

Components of a mixture emerging from a liquid chromatographic column are dissolved in the eluting solvent, and this solution is the one directed across the target, as described above. Thus, as the components reach the target, they produce ions. These ions are recorded by the spectrometer as an ion current. [Pg.394]

A computer attached to a mass spectrometer is used both to acquire data and to control the operation of the spectrometer. Powerful transputer systems can be used to ensure that both modes of operation can be carried out almost simultaneously. [Pg.421]

Flowever, unless is unusually large or the spectrometer is of fairly high resolution, the doubling of levels is relatively small and can be neglected. [Pg.178]

In order for the primary photoelectron, which is bound to the surface atom with binding energy to be detected ia xps, the electron must have sufficient kinetic energy to overcome, ia addition to E the overall attractive potential of the spectrometer described by its work function, Thus, the measured kinetic energy of this photoelectron, Ej is given by... [Pg.275]

If the energy of the iacident x-rays and the spectrometer work function are known, the measured kinetic energy can be used to determine the binding energy E from... [Pg.275]

Interference effects, which arise because of the extraordinary uniformity of thickness of the film over the spectrometer sample beam, superimposed on the absorption of incident light by parylene films, can be observed. Experimentally, a sinusoidal undulation of the baseline of the spectmm is seen, particularly in the spectral regions where there is Htde absorption by the sample. These so-called "interference fringe" excursions can amount to some... [Pg.438]

The helium leak detector is a common laboratory device for locating minute leaks in vacuum systems and other gas-tight devices. It is attached to the vacuum system under test a helium stream is played on the suspected leak and any leakage gas is passed into a mass spectrometer focused for the helium-4 peak. The lack of nearby mass peaks simplifies the spectrometer design the low atmospheric background of helium yields high sensitivity helium s inertness ensures safety and its high diffusivity and low adsorption make for fast response. [Pg.15]

The hnearity between M and makes the concept of absorbance so usehil that measurements made by sampling methods other than transmission are usually converted to a scale proportional to absorbance. The linearity between M and i is maintained only if the resolution of the spectrometer is adequate to eliminate contributions from wavelengths not absorbed by the species being measured. In addition, the apparent value of a is very dependent on resolution because a is 2l strong function of wavelength (30,31). [Pg.197]


See other pages where The Spectrometer is mentioned: [Pg.53]    [Pg.64]    [Pg.314]    [Pg.1307]    [Pg.1462]    [Pg.1472]    [Pg.1475]    [Pg.1560]    [Pg.1561]    [Pg.1852]    [Pg.535]    [Pg.77]    [Pg.173]    [Pg.185]    [Pg.225]    [Pg.226]    [Pg.257]    [Pg.322]    [Pg.323]    [Pg.66]    [Pg.295]    [Pg.279]    [Pg.283]    [Pg.195]    [Pg.199]    [Pg.199]    [Pg.200]   


SEARCH



Adjustment of the spectrometer

Advanced Topic The Fourier Transform Spectrometer

An overview of the NMR spectrometer

Application of field-portable x-ray fluorescence spectrometers in mineral exploration, with examples from the Abitibi Greenstone Belt

Basic Principles of the Quadrupole Mass Spectrometer

Classical Detectors Parallel to the Mass Spectrometer

Decomposition in the mass spectrometer

Detecting the Signal Fourier Transform NMR Spectrometers

Experimental Technique The Mass Spectrometer

Fragmentation in the Time-of-Flight Mass Spectrometer

Instrumentation - The Mass Spectrometer

Ionization in the mass spectrometer

Loading sample strontium solution on a filament for measurement in the thermal ionization mass spectrometer (TIMS)

NanoSIMS The Mass Spectrometer

Near Infrared Camera Multiobject Spectrometer on the Hubble Space Telescope

Neutral Molecules Eliminated in the Mass Spectrometer

Resolution of the spectrometer

Resolving power of the spectrometer

Results from the Yale cold-ion spectrometer

Spectrometers with the Paschen-Runge Mount

The Components of a MMW Cavity Spectrometer for Quantitative Measurements

The Cornu Prism Spectrometer

The Czerny-Turner Spectrometer

The ESR spectrometer

The Eagle Spectrometer

The Ebert Spectrometer

The Fourier Transform Spectrometer

The Fourier transform infrared (FT-IR) spectrometer and sample preparation

The Infrared Spectrometer

The Magnetic Sector Mass Spectrometer

The Mass Spectrometer Overview

The Miniaturized Spectrometer MIMOS II

The Mossbauer Spectrometer

The NMR Spectrometer

The Nuclear Magnetic Resonance Spectrometer

The Paschen-Runge Spectrometer

The Photoelectron Spectrometer

The Plasma Desorption Mass Spectrometer

The Quadrupole Mass Spectrometer

The Rowland Spectrometer

The SNR of imaging spectrometers

The Seya-Namioka Spectrometer

The Spectrometer Slit

The Spin Echo Spectrometer

The Time-of-Flight Mass Spectrometer

The Time-of-Flight Spectrometer

The Two-Mirror, Crossed-Beam, Plane Grating Spectrometer

The Wadsworth Spectrometer

The addition of a mass spectrometer

The isotope mass spectrometer

The mass spectrometer

Tuning the Mass Spectrometer

Tuning the spectrometer

© 2024 chempedia.info