Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Tuning the spectrometer

It is hands-on-experience time let us tune the spectrometer for optimal performance, and then take a spectrum. The cooking recipe in Table 2.3 is spelled out for an old fashioned all-hardware X-band spectrometer with 200 mW maximal output power. [Pg.25]

Turn on the cooling water (to magnet coils, magnet power supply, microwave bridge). [Pg.25]

Turn on die main electrical switch(es) (to the magnet, the console). [Pg.25]

Open die drying gas (filtered pressurized air, or any inert gas) to the waveguide just above die resonator. [Pg.25]

Optionally, turn on the regulated temperature device and select a temperature (kelvin). [Pg.26]


When tuning the spectrometer at another wavelength, the centre of the convolution function is moved to that wavelength. If we encode the convolution function relative to the set point h 0), then we obtain the following discrete values (normalized to a sum = 1) ... [Pg.532]

In principle, EPR spectrometry is well suited as a method to monitor kinetic events however, in practice, the time required to tune the spectrometer, and its intrinsically low sensitivity compared to fluorescence or light-absorption spectrometry, affect its competitiveness. Relatively slow reactions on the timescale of minutes, such as the decomposition of the DMPO-superoxide adduct and the subsequent formation of the hydroxyl radical adduct (cf. Pou et al. 1989) are readily followed, either as the first-order disappearance of the DMPO/ OOH signal... [Pg.221]

I have a minor comment concerning the polyvinyl alcohol spectrum. The one shown (Figure 5) is similar to one obtained in our laboratory in 1973. Last year. Dr. Lana Sheer obtained a polyvinyl alcohol spectrum where the methlne resonance was resolved into a triplet of triplets. By studying the effect of temperature upon the spectrum and by tuning the spectrometer carefully, one sometimes obtains better hyperflne structure. [Pg.314]

In rough-tuning, the spectrometer is put at the tune mode. A low microwave power (e.g., 25 dB, 0.63 mW) is applied to generate the model pattern , which is the microwave power reflected from the resonator as a function of the microwave frequency. Microwave frequency is adjusted to... [Pg.310]

Figure 13 shows ACN anti-Stokes transients after C-H stretch excitation (3000 cm 1) obtained by tuning the spectrometer to a particular transition... [Pg.577]

Once we have an equivalent circuit, we may manipulate the circuit equations to explore the effect on spectrometer performance of changing the values of circuit elements. It would be useful to have a variable coupling scheme in order to tune the spectrometer for optimum performance, just as in the microwave spectrometer case. Such a scheme is described in Section X. [Pg.288]

The absolute measurement of areas is not usually usefiil, because tlie sensitivity of the spectrometer depends on factors such as temperature, pulse length, amplifier settings and the exact tuning of the coil used to detect resonance. Peak intensities are also less usefiil, because linewidths vary, and because the resonance from a given chemical type of atom will often be split into a pattern called a multiplet. However, the relative overall areas of the peaks or multiplets still obey the simple rule given above, if appropriate conditions are met. Most samples have several chemically distinct types of (for example) hydrogen atoms within the molecules under study, so that a simple inspection of the number of peaks/multiplets and of their relative areas can help to identify the molecules, even in cases where no usefid infonnation is available from shifts or couplings. [Pg.1442]

In the vast majority of GC-MS applications, the chromatographic conditions employed have little or no effect on the operation of the mass spectrometer. This means that the spectrometer may be tuned for optimum performance and a number of samples containing different analytes can be analysed without operator intervention. This is not the case with LC-MS where the chromatographic conditions will invariably have a significant, compound-dependent, effect on the mass spectrometry conditions required to obtain useful analytical data. [Pg.188]

Reflectance measurements involve measurements of the intensity of light reflected from a flat specular surface of an electrode in a spectroelectrochemical cell. The incident light is polarized either parallel (p) or perpendicular s) to the plane of incidence, as shown in Fig. 27.24. A detector monitors the intensity of the reflected beam. The light is monochromatic, but the spectrometers usually can be tuned over large wavelength ranges. There are excellent reviews of reflectance by McIntyre (1973) and Plieth et al. (1992). [Pg.492]

Since the actual motion of the Mossbauer drive, as for any frequency transmission system, can show phase shifts relative to the reference signal, the ideal folding point (FP) of the raw data in terms of channel numbers may be displaced from the center at channel number (N — l)/2 (= 255.5 in the example seen earlier). The folding routine must take this into account. Phase shift and FP depend on the settings of the feedback loop in the drive control unit. Therefore, any change of the spectrometer velocity tuning requires the recording of a new calibration spectrum. [Pg.30]

Let us now consider the individual components in some detail (Poole 1996 Czoch and Francik 1989), and then go through the tuning and operating procedure of the spectrometer. [Pg.17]

FIGURE 2.8 Tuning mode patterns. The scope mode shows a dip when the frequency of the microwave fits into the resonator. This dip has maximal depth when the spectrometer is tuned (with the adjustable iris) to be reflectionless, that is, when the resonator is critically coupled to the bridge. Shown patterns are (a) off resonance, (b) slightly off resonance, (c) either under- or over-coupled, (d) critically coupled, (e) asymmetry from out-of-phase reference. [Pg.26]

Analysis of terrestrial samples demonstrated that the tuning of the ion extraction system, and consequently the amount of instrumental mass fractionation, was very sensitive to charge build-up on the sample, the position of the primary beam relative to the spectrometer optic axis, and the position of the sample relative to the extraction lens. To optimize reproducible tuning of the extraction system from sample to sample, we developed the following criteria (1) resistance of the sample (Au-coated) to ground less than 106 fi (2) alignment of the primary beam to within 10 pm using... [Pg.106]

An automatic probe tuning and matching (ATM) accessory allows one to automatically tune the NMR probe to the desired nuclei s resonant frequency and match the resistance of the probe circuit to 50 Q [7]. Traditional NMR instruments are designed so that one must perform these adjustments manually prior to data acquisition on a new sample. The advent of the ATM accessory allows the sampling of many different NMR samples without the need for human intervention. The ATM in conjunction with a sample changer enables NMR experiments to be conducted under complete automation. The sample changers are designed so that once the samples are prepared, they are placed into the instrument s sample holders. Data are then acquired under software control of both the mechanical sample delivery system as well as the electronics of the spectrometer. [Pg.277]


See other pages where Tuning the spectrometer is mentioned: [Pg.503]    [Pg.25]    [Pg.57]    [Pg.104]    [Pg.104]    [Pg.217]    [Pg.106]    [Pg.367]    [Pg.503]    [Pg.25]    [Pg.57]    [Pg.104]    [Pg.104]    [Pg.217]    [Pg.106]    [Pg.367]    [Pg.421]    [Pg.626]    [Pg.235]    [Pg.259]    [Pg.77]    [Pg.532]    [Pg.532]    [Pg.26]    [Pg.41]    [Pg.400]    [Pg.8]    [Pg.131]    [Pg.368]    [Pg.384]    [Pg.391]    [Pg.10]    [Pg.15]    [Pg.51]    [Pg.169]    [Pg.173]    [Pg.102]    [Pg.364]    [Pg.77]    [Pg.316]    [Pg.144]    [Pg.177]   


SEARCH



The Spectrometer

Tuning

Tuning the Mass Spectrometer

© 2024 chempedia.info