Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Detector overload

The low MW power levels conuuonly employed in TREPR spectroscopy do not require any precautions to avoid detector overload and, therefore, the fiill time development of the transient magnetization is obtained undiminished by any MW detection deadtime. (3) Standard CW EPR equipment can be used for TREPR requiring only moderate efforts to adapt the MW detection part of the spectrometer for the observation of the transient response to a pulsed light excitation with high time resolution. (4) TREPR spectroscopy proved to be a suitable teclmique for observing a variety of spin coherence phenomena, such as transient nutations [16], quantum beats [17] and nuclear modulations [18], that have been usefi.il to interpret EPR data on light-mduced spm-correlated radical pairs. [Pg.1566]

Linearity of response versus absolute amount injected must be confirmed for each different sample type and for each different set of chromatographic operating conditions. This linearity cannot be assumed. Nonlinearity may result from column overload, detector overload, or adsorption problems. [Pg.202]

The understanding of the effects of sample concentration (sample mass) in field-flow fractionation (FFF) has being obtained gradually with the improvement of the sensitivity (detection limit) of high-performance liquid chromatography (HPLC) detectors. Overloading, which was used in earlier publications, emphasizes that there is an upper limit of sample amount (or concentration) below which sample retention will not be dependent on sample mass injected into the FFF channels [1]. Recent studies show that such limits may not exist for thermal FFF (may be true for all the FFF techniques in polymer separation), although some of the most sensitive detectors on the market were used [2]. [Pg.423]

Detector overloaded. Outside linear dynamic range. Use a low volume detector cell. [Pg.1656]

Detector overload is indicated when the signal can no longer be recorded as a whole but appears as a rectangle, even at the highest attenuation. This gives the false impression of column overload. As already mentioned in Section 21.2, an insensitive detector is recommended in preparative work in the case of a UV detector this means a small optical path length. [Pg.328]

In both cases the data acquisition in the TCSPC channels is synchronised with the scanning by clock pulses from the scan controller. It must, however, be taken into account that the length of the lines of the scan varies since the return points of the scan are controlled by the detector overload signals. Therefore, the scan software must store the positions of the return points and the number of pixels between. These positions are used later to adjust the lines horizontally. [Pg.103]

Fig. 7.41 Components for detector overload protection HFAC-26 preamplifier, DCC-lOO detector controller card, and shutter assembly for R3809U MCP. Becker Hickl, Berlin... Fig. 7.41 Components for detector overload protection HFAC-26 preamplifier, DCC-lOO detector controller card, and shutter assembly for R3809U MCP. Becker Hickl, Berlin...
If the peaks that exhibit detector overload do not need to be quantitated, then we can just ignore this overload phenomenon. Otherwise we should inject a smaller amount or use a detector (or detector setting) with a larger linear dynamic range. [Pg.185]

For the highly volatile components, commonly present in essential oils, the most adequate transfer technique is partially concurrent eluent evaporation [137]. In the latter technique, proposed by Grob, a retention gap is installed, followed by a few meters of precolumn and the analytical capillary GC column, both with identical stationary phase, for the separation of the LC fractionated components. A vapor exit is placed between the precolumn and the analytical column, allowing partial evaporation of the solvent. Hence, column and detector overloading are avoided. This transfer technique can be applied to the analysis of GC components with a boiling point of at least 50°C higher than the solvent. [Pg.221]

The column saturation capacity is calculated from two chromatograms. One of these is made at analytical loads, whilst the other is run at a preparative loading. It is important that the load is chosen such that the peak due to the component of interest is not deformed by detector overload and that a reasonably large change in capacity factor is seen. It is recommended that a relative change of more than 10 or 15% in k is used. [Pg.64]


See other pages where Detector overload is mentioned: [Pg.305]    [Pg.6]    [Pg.449]    [Pg.203]    [Pg.325]    [Pg.141]    [Pg.172]    [Pg.302]    [Pg.303]    [Pg.178]    [Pg.185]    [Pg.207]    [Pg.386]    [Pg.289]    [Pg.55]    [Pg.140]   
See also in sourсe #XX -- [ Pg.328 ]

See also in sourсe #XX -- [ Pg.292 ]




SEARCH



Detector Control and Overload Protection

Detector overload protection

Overload

Overloading

Overloading overload

© 2024 chempedia.info