Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample dispersion

In the above-mentioned studies, Snyder and Adler used a theoretical model which they contrasted with experimental results. They evaluated the dispersion by assuming the transient signal to be Gaussian (see Fig. 5.3b). By expressing the bandwith as a standard deviation (8o and 4o for signals with and without a plateau, respectively), the dispersion can be related to the analyser and sample variables through [Pg.134]

The maximum sampling rate (number of samples processed per hour) achievable with no signal overlap is given by [Pg.135]

Obviously, these sampling rates can be safely augmented (by 20-30 ) as a slight overlap between the ends of two signals is acceptable provided that it does not affect their steady-state portions. [Pg.135]


When a sample is injected into the carrier stream it has the rectangular flow profile (of width w) shown in Figure 13.17a. As the sample is carried through the mixing and reaction zone, the width of the flow profile increases as the sample disperses into the carrier stream. Dispersion results from two processes convection due to the flow of the carrier stream and diffusion due to a concentration gradient between the sample and the carrier stream. Convection of the sample occurs by laminar flow, in which the linear velocity of the sample at the tube s walls is zero, while the sample at the center of the tube moves with a linear velocity twice that of the carrier stream. The result is the parabolic flow profile shown in Figure 13.7b. Convection is the primary means of dispersion in the first 100 ms following the sample s injection. [Pg.650]

The sum expressed by equation (21) lends itself to a digital calculation and can be employed in an appropriate computer program to calculate actual peak profiles. In doing so, however, as (v) is measured in plate volumes and sample volumes are usually given in milliliters, they must be converted to plate volumes to be used with equation (21). To demonstrate the effect of a finite charge and the use of equation (21), the peak profiles resulting from a sample dispersed over the twenty-one consecutive plates of a column are shown in Figure 16. [Pg.192]

In this present study, we basically showed dependence of the number of reduced cobalt metal surface atoms on dispersion of cobalt oxides along with the presence of rutile phase in titania. Both XRD and SEM/EDX results (not shown) revealed good distribution of cobalt oxides over the titania support. However, it can not differentiate all samples containing various ratios of rutile/anatase phase. Thus, in order to determine the dispersion of cobalt oxide species on titania, a more powerful technique such as TEM was applied with all samples. The TEM micrographs for all samples are shown in Figure 1. The dark spots represented cobalt oxides species present after calcination of samples dispersing on titania consisting various... [Pg.286]

A very important concept in understanding the theoretical background of FIA is the dispersion of the sample. Dispersion is expressed as the dispersion coefficient, D, which is defined as follows ... [Pg.326]

Figure 4 Sample dispersion (A) and signal profiles (B) at different distances within the tube (2-5) after injection of the sample into the stream (1), assuming that the detector is able to follow the sample zone as it flows. Figure 4 Sample dispersion (A) and signal profiles (B) at different distances within the tube (2-5) after injection of the sample into the stream (1), assuming that the detector is able to follow the sample zone as it flows.
In matrix solid-phase dispersion (MSPD) the sample is mixed with a suitable powdered solid-phase until a homogeneous dry, free flowing powder is obtained with the sample dispersed over the entire material. A wide variety of solid-phase materials can be used, but for the non-ionic surfactants usually a reversed-phase C18 type of sorbent is applied. The mixture is subsequently (usually dry) packed into a glass column. Next, the analytes of interest are eluted with a suitable solvent or solvent mixture. The competition between reversed-phase hydrophobic chains in the dispersed solid-phase and the solvents results in separation of lipids from analytes. Separation of analytes and interfering substances can also be achieved if polarity differences are present. The MSPD technique has been proven to be successful for a variety of matrices and a wide range of compounds [43], thanks to its sequential extraction matrices analysed include fish tissues [44,45] as well as other diverse materials [46,47]. [Pg.464]

A precision injection device is required to minimize sample dispersion and keep the sample volume and length of sample zone reproducible. This is normally a rotary valve similar to that used for injection in HPLC. Exact timing from sample injection to detection is critical because of rapidly occurring reactions which are monitored before they reach completion. This demands a constant flow rate with low amplitude pulsing, normally achieved by a peristaltic... [Pg.223]

Samples of both fulvic and humic acids were suspended in methanol and methylated with diazomethane. Both H and spectra of the free acids were obtained, at 299.94 MHz and 75.42 MHz respectively, on a Varian XL-300 spectrometer having a Nicolet TT-100 PET accessory. Spectra were obtained in D2O, in a 12-mm tube, with deuterated TSP (sodium 3-(trimethylsilyl)propionate-, , 3,3- 4) added as internal reference. GC/MS of methylated acids was conducted on a Hewlett-Packard Model No 5995 GC/MS/DA system equipped with a fused silica capillary column (12 m x. 020 mm ID, Hewlett Packard) internally coated with crosslinked methylene silicone. Infrared spectra were obtained with solid samples dispersed in KBr pellets, by using a Beckman IR-33 spectrophotometer. The various absorption peaks in IR and NMR were interpreted conventionally (9-10). [Pg.385]

At infrared wavelengths extinction by the MgO particles of Fig. 11.2, including those with radius 1 jam, which can be made by grinding, is dominated by absorption. This is why the KBr pellet technique is commonly used for infrared absorption spectroscopy of powders. A small amount of the sample dispersed in KBr powder is pressed into a pellet, the transmission spectrum of which is readily obtained. Because extinction is dominated by absorption, this transmission spectrum should follow the undulations of the intrinsic absorption spectrum—but not always. Comparison of Figs. 10.1 and 11.2 reveals an interesting discrepancy calculated peak extinction occurs at 0.075 eV, whereas absorption in bulk MgO peaks at the transverse optic mode frequency, which is about 0.05 eV. This is a large discrepancy in light of the precision of modern infrared spectroscopy and could cause serious error if the extinction peak were assumed to lie at the position of a bulk absorption band. This is the first instance we have encountered where the properties of small particles deviate appreciably from those of the bulk solid. It is the result of surface mode excitation, which is such a dominant effect in small particles of some solids that we have devoted Chapter 12 to its fuller discussion. [Pg.292]

Diffuse Reflection. Using a set of flat and elliptical mirrors, this device can measure a sufficient amount of light diffused by a sample dispersed in KBr powder (Fig. 10.20). By comparing the diffused reflection obtained with neat KBr, a result resembling the transmission spectrum is obtained. Kubelka-Munk s correction can be used to improve the spectrum. [Pg.180]

For solid samples dispersed in a KBr disc which have a thickness that cannot be precisely measured, an internal standard is used (e.g. calcium carbonate, naphthalene, sodium nitrite). This reference is added in equal quantity to all standards and to the sample. [Pg.183]

Figure 1 shows the FT-1R spectra of samples dispersed in KBr. All the spectra display a strong band at 960 cm 1. This band has been assigned to Si-O-Ti bonds [14] or to Si-OH groups [15, 16]. It is usually taken as the evidence for isomorphous substitution of Si by Ti, but it cannot be used to determine quantitatively the content of titanium into the framework of mesoporous materials [17]. In addition, the broad pattern between 3700 and 3000 cm 1, originated from hydrogen-bonded surface OH groups as well as from adsorbed H20 [18], decreases dramatically in the silylated samples. [Pg.183]

Limited sample clean-up could overload the analytical column, and residual matrix components can accumulate on the column after multiple injections. The residual matrix components can also solidify and deposit over a period of time in the LC-MS ionization source or vacuum interface, resulting in a decrease in ion transfer efficiency. The decrease in instrumentation performance (i. e., signal intensity) can be monitored by the signals of system-suitability samples dispersed within an analytical batch. The practice of replacing the pre-column in every run and scrubbing the analytical column periodically with a cleaning mobile phase will help to maintain instrument performance. [Pg.165]

In SFA, the bubbles prevent contamination between successive samples, reduce sample dispersion, and facilitate mixing of sample and reagents in a way that enables physical and chemical equilibrium to be attained before the sample reaches the detector. The main elements of these analyzers are as follows ... [Pg.281]

Gallagher, P.A., S. Murray, X. Wei, et al. 2002. An evaluation of sample dispersion media used accelerated solvent extraction for the extraction and recovery of arsenicals from LFB and DORM-2. J. Anal. At. Spectrom. 17 581-586. [Pg.365]

Sample dispersion is another cause of the long lag time in flow injection techniques where an aqueous carrier fluid is used [63,64]. Dispersion is caused by axial mixing of the sample with the carrier stream. This increases the sample volume, resulting in longer residence time in the membrane. Dilution reduces the concentration gradient across the membrane, which is the driving force for diffusion. The overall effects are broadened sample band and slow permeation. [Pg.220]

Gas injection membrane extraction (GIME) of aqueous samples has been developed to address the issues of boundary layer effects and sample dispersion [66]. This is shown in Figure 4.20. An aqueous sample from the loop... [Pg.220]

Photon correlation spectroscopy (PCS), also referred to as dynamic light scattering, is a technique that is used to measure particles in the size range of 1 -0.001 p,m. Unlike particle sizing by laser diffraction, the sample, dispersed in a diluent, is not circulated, stirred, or sonicated during the measurement. The technique is dependent upon a stable suspension of particles that are in constant random motion due to collisions with molecules of the suspending liquid. [Pg.316]

At the irradiation of the mesoporous Ti02 samples, dispersed in a water-alcohol mixture (typically containing 5 vol. % of H20 and 95 vol. % of C2H5OH) without addition of any metal cations, we did not observe any molecular hydrogen evolution (Tables 2, 3). Suspension of the mesoporous titania becomes grey-blue in... [Pg.590]

To measure the emulsifying activity, each 200-mg sample dispersed in 0.01M phosphate, pH 7 (20 mL), was homogenized with 20 mL corn oil at 10,000 rpm for 3 min, and the resulting emulsion was centrifuged at 500 rpm for 5 min to separate a cream layer. The emulsifying activity was represented by the ratio of cream layer volume/total volume. [Pg.207]


See other pages where Sample dispersion is mentioned: [Pg.516]    [Pg.196]    [Pg.453]    [Pg.319]    [Pg.738]    [Pg.390]    [Pg.171]    [Pg.53]    [Pg.321]    [Pg.326]    [Pg.332]    [Pg.256]    [Pg.210]    [Pg.103]    [Pg.169]    [Pg.311]    [Pg.181]    [Pg.53]    [Pg.321]    [Pg.326]    [Pg.332]    [Pg.853]    [Pg.68]    [Pg.23]    [Pg.155]    [Pg.97]    [Pg.54]    [Pg.210]    [Pg.357]   
See also in sourсe #XX -- [ Pg.418 , Pg.420 , Pg.429 , Pg.433 ]




SEARCH



Dispersed samples

Dispersed samples

Dispersing Powder Samples for Size Characterization Studies

Dispersion and Diffusion of Sample Plug in a Carrier Stream

Dispersion from sample volume

Dispersion in sample valves

Dispersion sample volume influence

Flow-injection analysis sample dispersion

Peak Width, Dispersion Factor, and Sampling Frequency

Relationship between the dispersion coefficient and injected sample volume

Sample Volume and Maximum Extra Column Dispersion

Sample dispersion from

Sample preparation matrix solid-phase dispersion

Sample valves, dispersion

Sample volume, effect on dispersion equation for

Sampling dispersion instrumentation

Sampling dispersion parameters

© 2024 chempedia.info