Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Response of the blank

Another criterion for linearity is that the y-intercept of the calibration curve (after the response of the blank has been subtracted from each standard) should be close to 0. An acceptable degree of closeness to 0 might be 2% of the response for the target value of analyte. For the assay of impurities, which are present at concentrations lower than that of the major component, an acceptable value of R2 might be 0.98 for the range 0.1 to 2 wt% and the y-intercept should be 10% of the response for the 2 wt% standard. [Pg.84]

Solvent peak areas are compared to the calibration curves to determine the concentration of each solvent in the sample. The response of the blank is also converted to concentration and subtracted from the sample concentration. The blank includes the background of carbon disulfide as well as the background related to the charcoal tube. With the concentration as ul of solvent/ml of carbon disulfide, conversion to PPM of solvent/volume of air sampled is determined by the following equation ... [Pg.207]

Fig. 4 Calibration graph showing the LOD. This graph illustrates the lUPAC method for determining the LOD. i B=response of the blank b = standard deviation of multiple determinations of the blank responses. View this art in color at WWW. dekker. com.)... Fig. 4 Calibration graph showing the LOD. This graph illustrates the lUPAC method for determining the LOD. i B=response of the blank b = standard deviation of multiple determinations of the blank responses. View this art in color at WWW. dekker. com.)...
Fig. 6. (continued) incubated with the probe-modified electrode (B2), hybridization takes place between the complementary strands resulting in the formation of a hybrid-modified electrode (B3), Hoechst 33258, is incubated with the hybrid-modified-electrode (B4), since Hoechst 33258 intercalates into the double-stranded hybrids on the surface, a high electrochemical current response is obtained (B5). The peak current intensity at the peak potential (—0.6V) is recorded relative to the background response of the blank buffer solution (dashed grey line). [Pg.108]

Pesticide and insecticide standards were tested as possible interferences against each other. Standard 100 ppm solutions were prepared by the dissolution and/or dilution of the samples in deionized water when possible. The pesticides/insecticides with limited solubility in water were prepared using 50 50 water/methanol or 50 50 water/acetone. The pH of each of the solutions was adjusted to 9.5 using IM sodium hydroxide. Spectra for each analyte against the sensors were taken at regular intervals for 30 minutes, and then conqiared with the response from the imprinted analyte. The sensors were rinsed with deionized water between each analysis. The response of the blank sensor to each of the pesticides was also evaluated. The sensors were also evaluated using a variety of water sources including tap water, seawater (Ocean City, MD), pond water (Baltimore, MD), and deionized water. [Pg.68]

Equations (1.16) and (1.17) show that Af and AD are related not only to the inherent properties of the quartz crystal but also to the solvent viscosity and density. Therefore, the shifts in A/and AD induced by the polymer behavior at the resonator surface can be extracted by taking the background response of the blank resonator as a Ref. [19]. [Pg.6]

On emergency shutdowns or any outage with less than five days planning time, planning should prepare a rough estimate (RE) for the job and issue a blank work plan with the appropriate work orders. Planning will not order miscellaneous parts, materials, tools, etc. this will be the responsibility of the assigned supervisor and area planner. [Pg.832]

To determine die diffusion current, it is necessary to subtract the residual current. This can be achieved by extrapolating the residual current prior to the wave or by recording die response of the deaerated supporting electrolyte (blank) solution. Addition of a standard or a calibration curve are often used for quantitation. Polarograms to be compared for this purpose must be recorded in the same way. [Pg.63]

The current-time response of the system during Sn(Il) addition presents the same features as the mass intensity-time curve. For comparison the i-t curve for a blank experiment (only adsorbed methanol being present, no tin addition) is also shown in Fig. 4.6a. The observed response is not simply the sum of the individual responses of Sn(II) (Fig. 4.2b) and adsorbed methanol (see dashed curve in Fig. 4.6a), to the applied potential step. [Pg.164]

In a bioanalytical method, analyses of blank samples (plasma, urine, or other matrix) should be obtained from at least six sources. Each blank sample should be tested for the possible interference of endogenous substances, metabolites, or degradation products. The response of the peaks interfering at the retention time of the analyte should be less than 20% of the response of a lower quantitation limit standard, and should be less than 5% of the response of the internal standard that was used [18, 19]. For dissolution studies, the dissolution media or excipients should not give a peak or spot that has an identical Rt or Rf value with the analyte [20]. [Pg.248]

The DL and QL for chromatographic analytical methods can be defined in terms of the signal-to-noise ratio, with values of 2 1-3 1 defining the DL and a value of 10 1 defining the QL. Alternatively, in terms of the ratio of the standard deviation of the blank response, the residual standard deviation of the calibration line, or the standard deviation of intercept (s) and slope (5) can be used [40, 42], where ... [Pg.255]

After a carrier protein has been activated with sulfo-SMCC, it is often useful to measure the degree of maleimide incorporation prior to coupling an expensive hapten. Ellman s reagent may be used in an indirect method to assess the level of maleimide activity of sulfo-SMCC-activated proteins and other carriers. First, a sulfhydryl-containing compound such as 2-mercaptoethanol or cysteine is reacted in excess with the activated protein. The amount of unreacted sulfhydryls remaining in solution is then determined using the Ellman s reaction (Chapter 1, Section 4.1). Comparison of the response of the sample to a blank reaction using... [Pg.768]

A comparison of the blank corrected values before and after conjugation should give an indication of the percent of peptide coupled. To be more quantitative, a standard curve must be run to focus in on the linear response range of the peptide-Ellman s reaction. Using cysteine as a representative sulfhydryl compound (similar in Ellman s response to a peptide having one free sulfhydryl), it is possible to obtain very accurate determinations of the amount which coupled to the activated carrier. Figure 19.20, discussed previously in this section, shows the results of this type of assay. [Pg.773]

Petty et al. (1998, 2000) used a vitellogenin (VGT) assay to assess the endocrine disrupting potential of contaminants in purified SPMD extracts. VGT is an egg yolk phosphoprotein precursor that is synthesized in the liver of female teleosts in response to estrogen from the ovary (Bailey, 1957). A wide variety of environmental contaminants have been shown to have estrogenic activity (Colborn et al., 1993). Equal portions of purified extracts from SPMDs, exposed in the Missouri River after the flood of 1993 and from the IWWTP at the Nogales Wash deployment were individually injected into immature rainbow trout (Oncorhynchus mykiss) as described in Section 6.4. The SPMD extracts contained elevated levels of complex mixtures of contaminants, including PAHs and pesticides. The fish injected with these sample extracts exhibited VGT induction, while no induction was observed in fish injected with any of the blank sample extracts. [Pg.131]

Before stepping through the several dimensions, it is worthwhile to examine the general analytical model which applies and, through that, consider the implications of the necessary assumptions in practical applications. To begin, let us express the observed response (y) and its error (e) in terms of the blank (B) and concentrations of all contributing analytes (xj). [Pg.52]

Depending on method type or intent, the LOD or LOQ may need to be determined. ICH guidelines describe several approaches and allow alternative approaches, if scientifically justifiable. Suggested approaches include calculation based on signal-to-noise ratio (typically set at 3 1 for LOD and 10 1 for LOQ) standard deviation of the response and slope standard deviation of the response of a blank, or a calibration curve. [Pg.389]

The exponential nature of the fall-curve is responsible for sample interaction. Provided that the between-sample time is made long enough, or blanks are inserted between samples, the effect is negHgible, but if the sampHng rate is increased then the response to any given sample will be influenced by the tail of the response of the preceding one. The effect can be particularly severe when a concentrated sample precedes a dilute one the peak due to the latter may appear as a shoulder on the tail on the one due to the former, or, in severe cases, it may be entirely hidden. [Pg.51]

Figure 12.3 Colorimetric detection of influenza by sialoside bilayer assembly (2% sialoside lipid 2 and 98% matrix lipid 1). (a) The colorimetric response of the film, supported on a glass microscope slide, is readily visible to the naked eye for quahtative evaluation of the presence of virus. The film on the left (blue) has been exposed to a blank solution of PBS. The film on the right (red) has been exposed to 100 hemagglutinin units (HAU) of vims (CR = 77%). (b) The... Figure 12.3 Colorimetric detection of influenza by sialoside bilayer assembly (2% sialoside lipid 2 and 98% matrix lipid 1). (a) The colorimetric response of the film, supported on a glass microscope slide, is readily visible to the naked eye for quahtative evaluation of the presence of virus. The film on the left (blue) has been exposed to a blank solution of PBS. The film on the right (red) has been exposed to 100 hemagglutinin units (HAU) of vims (CR = 77%). (b) The...
A straightforward and widely accepted approach is to deem that an instrument response greater than the blank signal plus three times the standard deviation of the blank signal indicates the presence of the analyte. This is consistent with the approach shown in figure 8.4 if it is assumed that the standard deviation of a measurement result at the LOD is the same as that of a blank measurement. Suppose there is a linear calibration relation... [Pg.239]

Carryover. Small amounts of analyte may get carried over from the previous injection and contaminate the next sample to be injected [10]. The carryover will affect the accurate quantitation of the subsequent sample. The problem is more serious when a dilute sample is injected after a concentrated sample. To avoid cross-contamination from the preceding sample injection, all the parts in the injector that come into contact with the sample (the injection loop, the injection needle, and the needle seat) have to be cleaned effectively after the injection. The carryover can be evaluated by injecting a blank after a sample that contains a high concentration of analyte. The response of the analyte found in the blank sample expressed as a percentage of the response of the concentrated sample can be used to determine the level of carryover. Caffeine can be used for the system carryover test for assessing the performance of an injector and serves as a common standard for comparing the performance of different injectors. [Pg.178]

Furthermore, many bioassay procedures can be affected synergis-tically or antagonistically by artifacts. In current practice, chemists use various solvents to clean the resin as completely as possible, run a resin blank, and chromatographically analyze the blank sample. Rarely are the artifacts identified. This procedure does not help the biologist who requires a blank that does not show a positive response in the bioassay test of the blank. Biologists usually clean the resin as the chemists do and complete a blank for the bioassay. Artifacts should be identified to assure that the bioassay results are in response to the compounds under study and not the artifacts in combination with the sample. [Pg.271]

A calibration curve shows the response of an analytical method to known quantities of analyte.8 Table 4-7 gives real data from a protein analysis that produces a colored product. A spectrophotometer measures the absorbance of light, which is proportional to the quantity of protein analyzed. Solutions containing known concentrations of analyte are called standard solutions. Solutions containing all the reagents and solvents used in the analysis, but no deliberately added analyte, are called blank solutions. Blanks measure the response of the analytical procedure to impurities or interfering species in the reagents. [Pg.69]

Step 2 Subtract the average absorbance (0.0993) of the blank samples from each measured absorbance to obtain corrected absorbance. The blank measures the response of the procedure when no protein is present. [Pg.70]

A calibration curve shows the response of a chemical analysis to known quantities (standard solutions) of analyte. When there is a linear response, the corrected analytical signal (= signal from sample — signal from blank) is proportional to the quantity of analyte. Blank solutions are prepared from the same reagents and solvents used to prepare standards and unknowns, but blanks have no intentionally added analyte. The blank tells us the response of the procedure to impurities or interfering species in the reagents. The blank value is subtracted from measured values of standards prior... [Pg.73]

A method blank is a sample containing all components except analyte, and it is taken through all steps of the analytical procedure. We subtract the response of the method blank from the response of a real sample prior to calculating the quantity of analyte in the sample. A reagent blank is similar to a method blank, but it has not been subjected to all sample preparation procedures. The method blank is a more complete estimate of the blank contribution to the analytical response. [Pg.80]


See other pages where Response of the blank is mentioned: [Pg.275]    [Pg.103]    [Pg.108]    [Pg.2024]    [Pg.275]    [Pg.103]    [Pg.108]    [Pg.2024]    [Pg.390]    [Pg.393]    [Pg.334]    [Pg.720]    [Pg.211]    [Pg.87]    [Pg.643]    [Pg.129]    [Pg.179]    [Pg.204]    [Pg.208]    [Pg.362]    [Pg.769]    [Pg.132]    [Pg.754]    [Pg.758]    [Pg.168]   
See also in sourсe #XX -- [ Pg.239 , Pg.240 ]




SEARCH



Blank

Blank, blanking

Blanking

Response blank

© 2024 chempedia.info