Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Background response

Cq ), where is the blended impurity concentration of impurity a Cq, the background impurity level and the multiplication constant. Possible sources of background response include instmment noise, sample system outgassing, or interference from other impurity response signals. Proper setup, purging, and operation of the instmment should reduce background levels weU below ippb. [Pg.90]

The FDA requests that the method exhibit sufficient sensitivity to measure accurately the residue of interest after fortification of the control matrix at half the tolerance concentration. Minimally, the detector response at the tolerance should be at least 10 times the average background response. [Pg.85]

The practical limit of determination for these herbicides is between 0.01 and 0.05pg gp1 depending on the background response from the soil extract. [Pg.266]

Different organic and inorganic buffers, such as ammonium acetate, ammonium formate, HEPES, Gly-Gly, and triethanolamine, were selected to study the response of biotin and fluorescein-biotin in MS and compared to phosphate buffer. Biotin and fluorescein-biotin were dissolved in the carrier solution compositions of buffer (10 mM pH 7.5)/methanol (50 50, v/v) at concentrations of 10 ng pl k Both infusion and 20 pl-loop injection experiments were performed with detection by MS in full-scan and SIM mode. Main optimization criteria are the maximum response of biotin and fluorescein-biotin with lowest interference of the carrier solution. HEPES, Gly-Gly, and triethanolamine give very high background response, which significantly hampers the detection of biotin and fluorescein-... [Pg.201]

Another technique that has been used in CE format for chiral drug-protein interactions is the Hummel-Dreyer method (52). In this technique, the solute is dissolved in the run buffer at varying concentrations, creating a high detector background response. After equilibration of the system, mixtures of the ligand and protein in various ratios are injected into this system as a sample. [Pg.194]

The slope S may be estimated from the calibration curve of the analyte. The value of may be estimated by (1) calculating the standard deviation of the responses obtained from the measurement of the analytical background response of an appropriate number of blank samples or (2) calculating the residual standard deviation of the regression line from the calibration curve using samples containing the analyte in the range of the QL. [Pg.734]

Current as a function of time is the system response as well as the monitored response in chronoamperometry. A typical double-potential-step chronoamperogram is shown by the solid line in Figure 3.3B. (The dashed line shows the background response to the excitation signal for a solution containing supporting electrolyte only. This current decays rapidly when the electrode has been charged to the applied potential.) The potential step initiates an instantaneous current as a result of the reduction of O to R. The current then drops as the electrolysis proceeds. [Pg.56]

Run the first stripping cycle so as to obtain the SWV peak of the background response. [Pg.1014]

Run the analysis and wait until a steady-state signal of the background response of the buffer is obtained. [Pg.1080]

After construction of the biosensor (thick- or multi-layer), run a DPV scan in pH 4.5 0.1 M acetate buffer to obtain the background response. [Pg.1158]

Determination Determining the limit of quantitation of an analytical method may vary depending on whether it is an instrumental or a noninstrumental procedure. For instrumental procedures, a common approach is to measure the magnitude of analytical background response by analyzing a number of blank samples and calculating the standard deviation of this response Multiplying the standard deviation by a factor, usually 10, provides an estimate of the limit of quantitation. This limit is subsequently validated by the analysis of a suitable number of samples known to be close to or at the limit of quantitation. [Pg.1022]

Since the rank of Y can be estimated by factor-analytical technique with consideration of experimental noise, the number of unexpected interferents, M, can be obtained easily by subtracting N from the rank of Y. The information on the number of interferents is crucial in this situation, this makes the distinction between matrix calibration and vector calibration. Assuming the bilinear structure of the response, one can factor-decompose the overall background responses of M interferents into the product of two matrices... [Pg.74]

The mathematical dose-response models described in the preceding sections have assumed responses of the subjects to be due solely to the applied stimuli. However, many toxicity experiments and observational studies show clear evidence that responses can occur even at a zero dose. Thus, any mathematical dose-response function should properly allow for this natural, or background, responsiveness. [Pg.66]


See other pages where Background response is mentioned: [Pg.111]    [Pg.602]    [Pg.648]    [Pg.298]    [Pg.302]    [Pg.115]    [Pg.462]    [Pg.202]    [Pg.46]    [Pg.40]    [Pg.758]    [Pg.107]    [Pg.246]    [Pg.247]    [Pg.77]    [Pg.32]    [Pg.46]    [Pg.127]    [Pg.17]    [Pg.108]    [Pg.1021]    [Pg.90]    [Pg.1575]    [Pg.348]    [Pg.906]    [Pg.63]    [Pg.136]    [Pg.621]    [Pg.621]    [Pg.622]    [Pg.622]    [Pg.628]    [Pg.632]    [Pg.66]    [Pg.68]    [Pg.103]   
See also in sourсe #XX -- [ Pg.82 , Pg.83 , Pg.84 ]




SEARCH



© 2024 chempedia.info