Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Response blank

Optimization of System Variables. The dependence of the blank level and the total signal (blank + analyte response) on the liquid flow rate is shown in Figure 10. The conditions are the same as those for Figure 9 except that 10"6 M Hg2(N03)2 at pH 4 is used. Down to the lowest flow rate studied (1500 / L/min), the net response to 5 ppbv S02 is essentially constant. Unfortunately, this flow rate dependence was examined fairly late in the study and the other data reported here were obtained with a liquid flow rate of 2600 pL/min. It is clear, however, that down to at least 1500 nL/min, the response/blank ratio improves it may be advantageous to use a lower flow rate. This behavior also strongly suggests that the transport of mercury from the bulk solution (liberated due to the intrinsic disproprotionation equilibrium) to the carrier air stream is controlled by liquid phase mass transfer. [Pg.393]

The effect of the membrane length upon the blank and the calibration slope are shown for a 1 nM Hg2(N03)2 solution in Figure 13. Both blank and response increase with increasing membrane length butjplateaus after about 0.5 m. The response/blank ratios for the 0.1, 0.2, 0.3, 0.5 and 1.0 m lengths are 0.44, 0.53, 0.61, 0.73 and 0.73 in arbitrary units respectively. Since the blank noise was found to be approximately related to the absolute value of the blank, we chose a length of 0.5 m. [Pg.396]

In preparation for a mass immunization and prophylactic treatment clinic, health departments around the country at the state and local levels have a variety of responsibilities (Blank, Moskin, Zucker, 2003). The following is a listing of those responsibilities that relate to preparedness functions ... [Pg.463]

This record can appear only once in a file and it may occupy more than one line. The second and subsequent lines contain a continuation field, which is a right-justified integer followed by a blank character. This number is incremented by one for each additional line of the record. Several records belong to this categoi y, e.g., AUTHOR (contains names of the people responsible for the contents of the flic), KEYWDS (contains a list of keywords describing the macromolcculc),... [Pg.113]

For a chemical operator, knowledge in the world is the upper and lower limits for a pressure reading shown on the log sheet, along with the response to take. A log sheet with only a blank to fill in the reading requires knowledge in the head. This contrast is shown in Table 6.1. [Pg.103]

On emergency shutdowns or any outage with less than five days planning time, planning should prepare a rough estimate (RE) for the job and issue a blank work plan with the appropriate work orders. Planning will not order miscellaneous parts, materials, tools, etc. this will be the responsibility of the assigned supervisor and area planner. [Pg.832]

It is crucial in quantitative GC to obtain a good separation of the components of interest. Although this is not critical when a mass spectrometer is used as the detector (because ions for identification can be mass selected), it is nevertheless good practice. If the GC effluent is split between the mass spectrometer and FID detector, either detector can be used for quantitation. Because the response for any individual compound will differ, it is necessary to obtain relative response factors for those compounds for which quantitation is needed. Care should be taken to prevent contamination of the sample with the reference standards. This is a major source of error in trace quantitative analysis. To prevent such contamination, a method blank should be run, following all steps in the method of preparation of a sample except the addition of the sample. To ensure that there is no contamination or carryover in the GC column or the ion source, the method blank should be run prior to each sample. [Pg.215]

To determine die diffusion current, it is necessary to subtract the residual current. This can be achieved by extrapolating the residual current prior to the wave or by recording die response of the deaerated supporting electrolyte (blank) solution. Addition of a standard or a calibration curve are often used for quantitation. Polarograms to be compared for this purpose must be recorded in the same way. [Pg.63]

Pectolytic activity was also studied in batch reactors, following the reaction progress in thermostated quartz cuvettes. The reaction medium (3 cm ) was prepared with 1.5 g/L pectin in the standard buffer and 0.063 mg of enzyme. The absorbance of the reaction mixture against the substrate blank was continuously recorded at the spectrophotometer (Perkin Elmer Lambda 2, USA). Typical reaction time was 15 minutes, but initial reaction rates were estimated considering only the absorbances recorded during the first 200 seconds, range of totally linear response. [Pg.443]

Ensure measured response is due solely to the analyte. One analysis of each of >5 separate blanks of each representative commodity... [Pg.117]

Linear response, controls, reagent blanks, study samples... [Pg.1324]

Standards and blanks are the usual controls used in analytical HPLC. Standards are usually interspersed with samples to demonstrate system performance over the course of a batch run. The successful run of standards before beginning analysis demonstrates that the system is suitable to use. In this way, no samples are run until the system is working well. Typically, standards are used to calculate column plate heights, capacity factors, and relative response factors. If day-to-day variability has been established by validation, the chromatographic system can be demonstrated to be within established control limits. One characteristic of good science is that samples... [Pg.44]

The MS response after Sn(IV) addition is given in Fig. 4.5 for two oxidation potentials Eox. The C02 signal grows and passes through a maximum after some minutes. A more pronounced response is observed at higher potentials. The lowest potential at which this effect can be observed is ca. 0.425 V. Blank experiments (without addition of tin, dashed curve in Fig. 4.5) show a C02 production at potentials above 0.4 V, but this was always lower than in the presence of tin. [Pg.164]

The current-time response of the system during Sn(Il) addition presents the same features as the mass intensity-time curve. For comparison the i-t curve for a blank experiment (only adsorbed methanol being present, no tin addition) is also shown in Fig. 4.6a. The observed response is not simply the sum of the individual responses of Sn(II) (Fig. 4.2b) and adsorbed methanol (see dashed curve in Fig. 4.6a), to the applied potential step. [Pg.164]

It is always important to run a blank gradient, ie a record of the detector response during the generation of a gradient. Fig. 4.3c shows... [Pg.153]

In a bioanalytical method, analyses of blank samples (plasma, urine, or other matrix) should be obtained from at least six sources. Each blank sample should be tested for the possible interference of endogenous substances, metabolites, or degradation products. The response of the peaks interfering at the retention time of the analyte should be less than 20% of the response of a lower quantitation limit standard, and should be less than 5% of the response of the internal standard that was used [18, 19]. For dissolution studies, the dissolution media or excipients should not give a peak or spot that has an identical Rt or Rf value with the analyte [20]. [Pg.248]

The DL and QL for chromatographic analytical methods can be defined in terms of the signal-to-noise ratio, with values of 2 1-3 1 defining the DL and a value of 10 1 defining the QL. Alternatively, in terms of the ratio of the standard deviation of the blank response, the residual standard deviation of the calibration line, or the standard deviation of intercept (s) and slope (5) can be used [40, 42], where ... [Pg.255]

After a carrier protein has been activated with sulfo-SMCC, it is often useful to measure the degree of maleimide incorporation prior to coupling an expensive hapten. Ellman s reagent may be used in an indirect method to assess the level of maleimide activity of sulfo-SMCC-activated proteins and other carriers. First, a sulfhydryl-containing compound such as 2-mercaptoethanol or cysteine is reacted in excess with the activated protein. The amount of unreacted sulfhydryls remaining in solution is then determined using the Ellman s reaction (Chapter 1, Section 4.1). Comparison of the response of the sample to a blank reaction using... [Pg.768]

A comparison of the blank corrected values before and after conjugation should give an indication of the percent of peptide coupled. To be more quantitative, a standard curve must be run to focus in on the linear response range of the peptide-Ellman s reaction. Using cysteine as a representative sulfhydryl compound (similar in Ellman s response to a peptide having one free sulfhydryl), it is possible to obtain very accurate determinations of the amount which coupled to the activated carrier. Figure 19.20, discussed previously in this section, shows the results of this type of assay. [Pg.773]


See other pages where Response blank is mentioned: [Pg.106]    [Pg.106]    [Pg.390]    [Pg.772]    [Pg.813]    [Pg.393]    [Pg.393]    [Pg.66]    [Pg.79]    [Pg.4]    [Pg.106]    [Pg.303]    [Pg.63]    [Pg.334]    [Pg.17]    [Pg.122]    [Pg.313]    [Pg.410]    [Pg.619]    [Pg.648]    [Pg.697]    [Pg.720]    [Pg.721]    [Pg.358]    [Pg.224]    [Pg.255]    [Pg.249]    [Pg.211]    [Pg.590]    [Pg.132]    [Pg.335]   
See also in sourсe #XX -- [ Pg.160 , Pg.164 ]




SEARCH



Blank

Blank responses, variability

Blank, blanking

Blanking

Response of the blank

© 2024 chempedia.info