Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sample blanks

For trace quantities of less than 100 ppm, the most successful method — and the most costly— is neutron activation. The sample is subjected to neutron bombardment in an accelerator where oxygen 16 is converted to unstable nitrogen 16 having a half-life of seven seconds. This is accompanied by emission of (J and 7 rays which are detected and measured. Oxygen concentrations as low as 10 ppm can be detected. At such levels, the problem is to find an acceptable blank sample. [Pg.30]

Field blanks Sample media that are exposed to the same conditions as the media use for the actual sampling but are not connected to a sampling pump. See also Laboratory blanks. [Pg.1440]

Laboratory blanks Sample media that is not sampled on, but is analyzed by the laboratory to detect contamination or other problems associated with preparation and analysis of the samples. See also Field blanks. [Pg.1454]

First, consider a blank sample in which the concentration of both components is equal to zero. With no absorbing species in the sample, there would be no absorbance at any of the wavelengths, and the spectrum would be plotted at [0, 0, 0], the origin of this absorbance data space. Now, consider the spectrum of a sample that contains 1 concentration unit of Component 1 and none of Component 2. This spectrum will have some absorbance at each of the... [Pg.81]

Recovery data indicate that little nitrosamine was lost in the procedure and none appears to have been formed. Exposure to ultraviolet radiation destroyed the BHP originally present and provided a blank sample which confirmed that no BHP was formed during the column extraction procedure. [Pg.341]

Comparison of at least one blank sample and a sample to which a known amount of pesticide has been added additional determinations in the presence of pesticides suspected of interfering with the analyte... [Pg.123]

Determinations should be made on blank samples to which the analyte has been added... [Pg.123]

Sensitivity is a measure of the smallest concentration that can be either measured [limit of detection (LOD)] or accurately quantitated [limit of quantitation (LOQ)]. In the USA, the method for measuring LOD or LOQ is left up to the method developer. European requirements for determining LOD and LOQ are very specific the LOD is based on the mean plus three standard deviations for 20 control blank samples, and the LOQ is defined as the lowest concentration giving an acceptable CV. [Pg.319]

Initially, this method utilized 5-mL conical centrifuge tubes as the collection device for final elution of the extract from the Cig tubes. In practice, these tubes were found to be very difficult to clean and in few instances were the cause of cross-contamination when low-concentration samples were extracted following samples with very high concentrations. Since no commercial graduated tubes were available, disposable culture tubes are used as the receiver. These tubes are individually calibrated before use. A solvent blank sample may be processed through the method from extraction to quantification to determine if contamination from glassware occurs. [Pg.377]

Recommendation Dilute the standard solutions twice with blank solutions prepared from each of the blank samples. Impurities in the blank samples reduce the thermal decomposition of the target analytes in the injection port and stabilize the profiles of ionization and fragmentation of the target analytes. [Pg.546]

The surface grafting of a commercial sheet of poly(ethylene tere-phtalate) was analyzed by ESCA spectra and dye adsorption. The sheets were washed with acetone to remove impurities and then grafted by irradiation for 2 and 20 min. (Figure 9). The blank sample showed no Nls peak in the ESCA spectrum. After grafting the Nls peak increased with increasing grafting time. [Pg.177]

The more detailed ESCA spectra show resolved peaks (Figure 10). The blank sample (A) has a Cls peak containing three components a main peak at 285 eV referred to carbon bonded to other carbons and hydrogens only, and two minor peaks due to C-0 and 0=C-0 at 287 and 289 eV, respectively. The Ols peak has two components of equal intensity referred to 0(C=0) and O(C-O-C)groups at 533 and 535 eV, respectively. The peak intensities correspond very well to the chemical structure of the polyester -j- OC-CgH -CO-O-iCI ) 2 0 n. [Pg.177]

The performance curve presents graphically the relationship between the probability of obtaining positive results PPRy i.e. x > xLSp on the one hand and the content x within a region around the limit of discrimination xDIS on the other. For its construction there must be carried out a larger number of tests (n > 30) with samples of well-known content (as a rule realized by doped blank samples). As a result, curves such as shown in Fig. 4.10 will be obtained, where Fig. 4.10a shows the ideal shape that can only be imagined theoretically if infinitely exact decisions, corresponding to measured values characterized by an infinitely small confidence interval, exist. [Pg.115]

The critical measurement value (critical value, CV), yc, being the lowest signal which can significantly be distinguished from that of a blank sample... [Pg.226]

As a rule, the average blank is estimated from repetition measurements of a - not too small - number of blank samples as arithmetic mean yBL. If there is information that another than normal distribution applies, then the mean of this other distribution should be estimated (see textbook of applied statistics see Arnold [1990] Davies and Goldsmith [1984] Graf et al. [1987] Huber [1981] Sachs [1992]). [Pg.227]

A value, yn, obtained my measuring a blank sample (in calibration, the intercept of the calibration curve is considered to be equal to the blank). [Pg.309]

In a bioanalytical method, analyses of blank samples (plasma, urine, or other matrix) should be obtained from at least six sources. Each blank sample should be tested for the possible interference of endogenous substances, metabolites, or degradation products. The response of the peaks interfering at the retention time of the analyte should be less than 20% of the response of a lower quantitation limit standard, and should be less than 5% of the response of the internal standard that was used [18, 19]. For dissolution studies, the dissolution media or excipients should not give a peak or spot that has an identical Rt or Rf value with the analyte [20]. [Pg.248]

Due to the possible change in retention time and peak profile that may take place during day-to-day operation, it is necessary to measure peak characteristics every day to verify the status of the method validation. A blank sample should be evaluated for an analysis run, where the resolution is determined. For asymmetric peaks, the Gaussian equation cannot be used, so the modified equation, using an exponentially modified Gaussian (EMG) method has been proposed [21]. [Pg.249]

In the bioanalytical studies, the basic calibration should be prepared in the same biological matrix as in the samples of the intended study, which can be achieved by spiking the matrix with known concentration of the analyte. In this case, a blank sample, a zero sample (blank and internal standard), six to eight nonzero samples covering the expected range (including the anticipated QL) should be evaluated as part of the linearity study [27]. [Pg.249]

FIGURE 2.4 Representative chromatograms of reference and blank (A) reference solution, (B) double blank sample. [Pg.81]

Before beginning the analysis of FMs at low concentrations, the laboratory should analyze several laboratory blank samples to assess the degree to which the laboratory is contaminated. With every set of samples analyzed, the laboratory should also analyze a laboratory and field blank sample. Laboratory workers should be advised to be aware of their personal use of fragrance-enhanced consumer products and the potential for laboratory contamination. [Pg.84]

This extracted and dried blank sample is also evaluated by HPLC to rule out any extraneous bands due to solvent interaction or stationary phase interference. [Pg.436]

Another aspect to take into account is that surfactants are often ingredients in the cleaning products used in chemical laboratories. The sampling (as discussed above) and sample handling to which the samples are subjected always carries with it the risk of contamination. It is therefore necessary to process sufficient numbers of blank samples together with the real samples. Numerous samples of oceanic seawater and fossil seawater (taken from wells) have shown traces of LAS (around 1 ppb). Therefore, environmental concentrations found at similar levels should be regarded with caution. [Pg.425]

Shang et al. [5] used ASE for the extraction of NP and NPEO from estuarine sediments. A sample of 15-25 g was extracted three times using hexane/acetone at 100°C and 103 atm. This was followed by a clean-up step using CN-SPE. A blank sample was extracted between all samples to avoid contamination. Hexane/acetone was also used in the ASE method for alkylphenols and NPEO by Heemken et al. [12]. Extraction conditions for samples of 0.5-1 g were 100°C, 150 atm, with a static extraction step of 15 min, and a rinse step with 20 mL solvent. After a clean-up by HPLC, the analytes were derivatised with heptafluorobutyric acid anhydride for GC analysis. [Pg.452]

For this reason the radiation transmitted by a blank sample is measured and taken to be the effective incident radiation. This blank should be identical to the test sample in all aspects except the presence of the test substance ... [Pg.50]

Double-beam atomic absorption spectrophotometers are designed to control variations which may occur in the radiation source but they are not as effective as double-beam molecular absorption instruments in reducing variation because there is no blank sample in flame techniques. [Pg.81]


See other pages where Sample blanks is mentioned: [Pg.451]    [Pg.710]    [Pg.772]    [Pg.334]    [Pg.592]    [Pg.126]    [Pg.122]    [Pg.124]    [Pg.124]    [Pg.648]    [Pg.693]    [Pg.764]    [Pg.185]    [Pg.76]    [Pg.310]    [Pg.310]    [Pg.311]    [Pg.108]    [Pg.248]    [Pg.506]    [Pg.53]    [Pg.98]   
See also in sourсe #XX -- [ Pg.286 ]

See also in sourсe #XX -- [ Pg.227 ]

See also in sourсe #XX -- [ Pg.285 ]

See also in sourсe #XX -- [ Pg.227 ]

See also in sourсe #XX -- [ Pg.360 ]

See also in sourсe #XX -- [ Pg.286 ]

See also in sourсe #XX -- [ Pg.200 ]

See also in sourсe #XX -- [ Pg.200 ]

See also in sourсe #XX -- [ Pg.32 ]




SEARCH



Blank

Blank, blanking

Blanking

Blanks, field quality control sampling

Sample and Blank Material Safety Data Sheets

Sample preparation blanks

Solvent blank samples

© 2024 chempedia.info