Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Chromatographic instrument

Another useful standard Is SRM 1647, priority pollutant polynuclear aromatic hydrocarbons (in acetonitrile). It can be used to calibrate liquid chromatographic Instruments (retention times. Instrument response), to determine percent recoveries, and to fortify aqueous samples with known PAH concentrations. Figure 8 Illustrates the HPLC separation and UV detection (fluorescence is also used extensively) for the 16 priority pollutants. [Pg.117]

For detection, MS is rapidly becoming the method of choice for multiclass, multiresidue analysis owing to its many advantages, recent improvements in technology, and availability of cost-effective commercial instrumentation. Detection systems in general are continually being improved, and in combination with the improvements in chromatographic instruments and techniques, an exceptionally low limit of detection (LOD) is possible for pesticide residues. [Pg.762]

The most common types of preparative-scale gas chromatographic instruments are based on pacXed column technology [489-491]. The primary objective in preparative-scale gas chromatography is to obtain a high sample throughput. An inevitable result of this goal is that either resolution or separation time, or both, must be compromised. The primary method... [Pg.109]

Experimentally there are two methods of determining the ] extracolumn band broadening of a chromatographic instrument. The linear extrapolation method, discussed above, is relatively straightforward to perform and interpret but rests on the validity.. of equation (5.1) and (5.3). The assu itlon that the individual contributions to the extracolumn variance are independent, may not be true in practice, and it may be necessary to couple some of the individual contributions to obtain the most accurate values for the extracolumn variance [20]. It is assumed in equation (5.3) ... [Pg.280]

Solubilizing all or part of a sample matrix by contacting with liquids is one of the most widely used sample preparation techniques for gases, vapors, liquids or solids. Additional selectivity is possible by distributing the sample between pairs of immiscible liquids in which the analyte and its matrix have different solubilities. Equipment requirements are generally very simple for solvent extraction techniques. Table 8.2 [4,10], and solutions are easy to manipulate, convenient to inject into chromatographic instruments, and even small volumes of liquids can be measured accurately. Solids can be recovered from volatile solvents by evaporation. Since relatively large solvent volumes are used in most extraction procedures, solvent impurities, contaminants, etc., are always a common cause for concern [65,66]. [Pg.891]

Guillemin, C. L., New concept in chromatographic instrumentation. The probe-process liquid chromatograph, ]. Chromatogr., 441, 1, 1988. [Pg.95]

One of the attractive features of SFE with CO2 as the extracting fluid is the ability to directly couple the extraction method with subsequent analytical methods (both chromatographic and spectroscopic). Various modes of on-line analyses have been reported, and include continuous monitoring of the total SFE effluent by MS [6,7], SFE-GC [8-11], SFE-HPLC [12,13], SFE-SFC [14,15] and SFE-TLC [16]. However, interfacing of SFE with other techniques is not without problems. The required purity of the CO2 for extraction depends entirely on the analytical technique used. In the off-line mode SFE takes place as a separate and isolated process to chromatography extracted solutes are trapped or collected, often in a suitable solvent for later injection on to chromatographic instrumentation. Off-line SFE is inherently simpler to perform, since only the extraction parameters need to be understood, and several analyses can be performed on a single extract. Off-line SFE still dominates over on-line determinations of additives-an... [Pg.429]

The concept of an integrated micro-fluidic-based system has now been developed (Zhang et al. 2006), with an example shown in Fig. 7. This particular system is based on conventional chromatographic instrumentation and employs a multi-valving system, located between two syringe pumps, shown in the foreground, to enable the introduction of multiple reagents from an auto-sampler to be loaded onto the micro-reactor. Because of the low diffussional distances obtained in this sys-... [Pg.33]

The basic components of an LC-NMR system are some form of chromatographic instrument and an NMR spectrometer equipped with a flow-probe, as shown in Fig. 19.17. In terms of the chromatography of choice, there are many examples in the literature of a wide array of separation instruments employed, from SFC to capillary electrophoresis (CE) [87,88]. By far the most common method (not necessarily the best choice from a separation point of view) of achieving the desired separation is through HPLC. There are many commercial... [Pg.734]

Both gas chromatographic instruments were connected with a PDP 11/45 computer via an analog-to-digital converter. The peak areas were calculated from the digitalized chromatographic data by means of software developed at Delft University of Technology. [Pg.299]

Another important physical property of liquids is the refractive index. Since the refractive index is a constant for a particular liquid at a given temperature, it can be used to help identify substances, check for purity, and measure concentrations. One type of detector found in some liquid chromatograph instruments (Chapter 13) uses refractive index. [Pg.427]

In the late 1970s, Hewlett-Packard introduced the HP-3300 series data-acquisition system, which was able to connect to 60 chromatographic instruments through an A/D converter. This was the beginning of what would become a revolution in CDS development within the analytical instrument industry. By the mid-1980s, all of the major analytical instrument manufacturers offered network-based data-acquisition systems Beckman, HP, PE, VG, and Waters. These were multi-user, time-sharing systems that used A/D converters to acquire data from the instruments. Instrument control, both HPLC and GC, was a capability that would soon follow. Several CDS manufacturers offered serial control of the HP 5890 GC while Waters also offered instrument control for their own HPLCs. [Pg.584]

Instrumentation for laboratory automation is increasingly becoming a plug and play operation. Many ancillary pieces of equipment, such as autosamplers, plate handlers, chromatographic instruments, and many classes of spectrometers, have... [Pg.329]

Fig. 31. HPLC chromatogram of the mixture obtained when using Ratz s procedure for the synthesis of NjPsCMeAzlg (Waters Liquid Chromatograph Instrument, (t Bondapack Cie reverse phase, pure methanol, RI detection, speed rate = 1 cm/mn) A = NaPjfMeAzIg, B = N3P3(MeAz)sCl, C = NaPjCMeAzl Oj... Fig. 31. HPLC chromatogram of the mixture obtained when using Ratz s procedure for the synthesis of NjPsCMeAzlg (Waters Liquid Chromatograph Instrument, (t Bondapack Cie reverse phase, pure methanol, RI detection, speed rate = 1 cm/mn) A = NaPjfMeAzIg, B = N3P3(MeAz)sCl, C = NaPjCMeAzl Oj...
In order for mass spectrometry to be used as an effective tool in LC, the analyte must be ionized in the interface region (source) since MS can only detect charged species. Consequently, the role of the ionization source is twofold (1) it is responsible for interfacing the chromatographic instrument with the detector and (2) it provides an effective and efficient means for sample ionization. It should be pointed out that source ionization efficiency plays a significant role in sample sensitivity for LC-MS analysis. The application of LC is ideally... [Pg.693]

It is beyond the scope of this chapter to discuss and explain how the requirements can be implemented in analytical laboratories. This has been described in a six-article series published in Biopharm [16-21]. We elaborate here on the validation aspect of the rule. Part 11 requires that computer systems used to acquire, evaluate, transmit, and store electronic records should be validated. This is not new, as processes and steps to validate such systems were described earlier in the chapter. FDA s expectations for validation have been described in the Part 11 draft guidance on validation [4]. This guidance makes it very clear that functions as required by Part 11 should be validated in addition to functions that are required to perform an application such as chromatographic instrument control, data acquisition, and evaluation. Specific functions as required by Part 11 are as follows ... [Pg.270]

We have developed a rapid and systematic process for isolation and identification of biologically active components from natural products. The process reduces time and cost through application of advanced chromatographic instrumentation. It generates important activity and chemical information and also provides advanced active fraction(s) to accelerate isolation studies. As a result, lead prioritization, project management, and the cycle time of natural product lead discovery have been significantly improved. [Pg.191]

Of all the analytical techniques, chromatography is the one with the widest scope of applicability. The sales of chromatographic instrumentation is a sector that represents at least half of all the sales of analytical equipment and material in the world. [Pg.3]

Figure 2.16—A miniature chromatograph. Instrument using a capillary column and a photoionisation detector. The instrument, weighing 4 kg including the carrier gas (C02), is mainly used for the analysis of volatile organic compounds (VOCs) in air pollution. The photoionisation detector, which is of limited use because of its variable sensitivity, is well suited for the analysis of hydrocarbons. The high powered UV source emits photons that have energies between 10 and 11 eV, ionising the compounds that exit the column, with the exception of the carrier gas. The ionic current generated is amplified using an electrometer and is proportional to the concentration of analytes (reproduced by permission of Photovac). Figure 2.16—A miniature chromatograph. Instrument using a capillary column and a photoionisation detector. The instrument, weighing 4 kg including the carrier gas (C02), is mainly used for the analysis of volatile organic compounds (VOCs) in air pollution. The photoionisation detector, which is of limited use because of its variable sensitivity, is well suited for the analysis of hydrocarbons. The high powered UV source emits photons that have energies between 10 and 11 eV, ionising the compounds that exit the column, with the exception of the carrier gas. The ionic current generated is amplified using an electrometer and is proportional to the concentration of analytes (reproduced by permission of Photovac).
Some altitude effects on the operation of chromatographic instruments are anticipated. To achieve reproducible retention times for identifying compounds, mobile-phase flows need to be controlled so that they are independent of ambient pressure. Detectors may also respond to changes in pressure. For example, the electron capture detector is a concentration-sensitive sensor and exhibits diminished signal as the pressure decreases. Other detectors, such as the flame ionization detector, respond to the mass of the sample and are insensitive to altitude as long as the mass flow is controlled. [Pg.130]

After evaporation of the alcohol, the trimethylsilyl ether was prepared by dissolving the residue or the standard acid (after methylation) in 0.2 cm3 of dioxane, 0.2 cm3 of hexamethyldisila-zane, and 0.1 cm3 of trimethylchlorosilane and mixing thoroughly. The preparation in a 2-cm3 test tube with Teflon lined cap was left for 30 min at room temperature. After centrifugation, 0.5 pi of the supernatant can be injected into the gas chromatographic instrument. The derivatives are stable if kept in a desiccator. It is recommended that the analysis be carried out the same day. [Pg.519]

On the other hand, the design of most gas chromatographic instruments requires that pyrolysis occur rapidly in order to prevent peak... [Pg.77]

Verify the calibration of all instrumentation involved in monitoring the EtO cycle. Examples include thermocouple and pressure gauge calibration, gas leak testing equipment, relative humidity sensors, and gas chromatographic instrumentation. [Pg.151]

Actinide determinations in urine matrix using column separation automated in ion chromatographic instrument ICP-MS 60... [Pg.527]


See other pages where Chromatographic instrument is mentioned: [Pg.634]    [Pg.56]    [Pg.72]    [Pg.246]    [Pg.210]    [Pg.85]    [Pg.429]    [Pg.410]    [Pg.791]    [Pg.896]    [Pg.64]    [Pg.61]    [Pg.3]    [Pg.153]    [Pg.9]    [Pg.268]    [Pg.97]    [Pg.130]    [Pg.122]    [Pg.64]    [Pg.226]    [Pg.7]    [Pg.48]    [Pg.60]   
See also in sourсe #XX -- [ Pg.6 ]




SEARCH



© 2024 chempedia.info