Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration standards suppliers

The revised database holds over 23 000 analyte values for 660 measurands and 1670 reference materials produced by 56 different producers, from 22 countries. The database is restricted to natural matrix materials (i.e. made from naturally occurring materials, excluding calibration standards manufactured from pure chemicals). Information has been extracted from the relevant certificates of analysis, information sheets, and other reports provided by the reference material producers. As a general rule, the authors have only included in the compilation reference materials for which a certificate of analysis or similar documentation is on file. Information included in the survey is on values for measurands determined in reference materials, producers, suppliers, the cost of the materials, the unit size supplied, and the recommended minimum weight of material for analysis, if available. The new searchable database has been designed to help analysts to select reference materials for quality assurance purposes that match as closely as possible, with respect to matrix type and concentrations of the measurands of interest and their samples to be analyzed see Table 8.3. [Pg.264]

Spherical particles of known diameter (e.g., 5% to 20% of the diameter of the aperture in the glass tube) are used to calibrate the electrical pulse counting instrument. The particles are suspended to an appropriate concentration in electrolyte solution (see recipe). Monodisperse latex particles are commercially available, which can be used for this purpose. Particle size calibration standards can be obtained from a number of chemical suppliers or from the National Institute of Standards and Technology (e.g., NBS 1003b). Lines (1996) lists a number of standards that are appropriate for this purpose. [Pg.585]

PCBs in soils and wastewaters can be rapidly screened on site or in the laboratory by immunoassay technique (Chapter 1.13). Immunoassay test kits are now commercially available from many suppliers. The samples can be tested at the calibration levels of 1 to 50 ppm. The kit primarily contains antibody-coated test tubes or magnetic particles, assay diluent, PCB-enzyme conjugate, a color-forming substance, and a solution to quench the reaction. The method does not distinguish accurately one Aroclor from another. PCBs can be measured semiquantitatively by comparing the optical density of the color formed in the sample against a set of calibration standards using a spectrophotometer. [Pg.239]

The calibration procedures adopted on-site must be agreed on with the customer and conform to recognized industry instrument calibration standards and the supplier s instructions. These procedures must be applied to all in-line instrumentation, loop instrumentation, local controllers, analyzers, and so on. Where the control and monitoring instrumentation is integrated with a computerized control system and where factory tests have been carried out, the installation calibration procedure should be agreed on with the customer. [Pg.564]

Mp. This is illustrated in Figure 1 in which the response calculated for a LALLS detector (by taking the product of concentration and molecular weight of each slice) is superimposed on the measured MWD (DRI, for the standard labeled Mp == 17,000 by the supplier) from which it was generated. The magnitude of this shift depends both on the MWD of the calibration standard and on the molecular weight dependency of the detector response. [Pg.82]

Typical plastics used in electronic and electrical appliances are polyethylene, polypropylene or polyethene terphthalate, and these are studied here as part of the RoHS requirement for the presence of toxic metals. This method is to show that analysis of these plastics used in electrical and electronic equipment is essential, especially if the origin of the plastic is unknown and the supplier is unable to state whether or not they are free of these metals. The metals are measured against calibration standards curves for each metal and may also include additional attachments for improving limits of detection such as ultrasonic nebulisers for Cd, Pb and Cr and the cold trap method for Hg. [Pg.126]

Calibrate the instrument set the meter to pH mode, if appropriate, and then place the electrode assembly in a standard solution of known pH, usually pH 7.00. This solution may be supplied as a liquid, or may be prepared by dissolving a measured amount of a calibration standard in water calibration standards are often provided in tablet form, to be dissolved in water to give a particular volume of solution. Adjust the calibration control to give the correct reading. Remember that your calibration standards will only give the specified pH at a particular temperature, usually either 20 C or 25 C. If you are working at a different temperature, you must establish the actual pH of your calibration standards, either from the supplier, or from literature information. [Pg.59]

There were also attempts to calibrate the SEC columns with help of broad molar mass dispersity poplymers but this is less lehable. The most common and well credible SEC cahbration standards are linear polystyrenes, PS, which are prepared by the anionic polymerizatioa As indicated in section 11.7, according to lUPAC, the molar mass values determined by means of SEC based on PS calibration standards are to be designated polystyrene equivalent molar masses . Other common SEC calibrants are poly(methyl methaciylate)s, which are important for eluents that do not dissolve polystyrenes, such as hexafluoroisopropanol, further poly(ethylene oxide)s, poly(vinyl acetate)s, polyolefins, dextrans, pullulans, some proteins and few others. The situation is much more complicated with complex polymers such as copolymers. For example, block copolymers often contain their parent homopolymers (see sections 11.8.3, 11.8.6 and 11.9). The latter are hardly detectable by SEC, which is often apphed for copolymer characterization by the suppliers (compare Figure 16). Therefore, it is hardly appropriate to consider them standards. Molecules of statistical copolymers of the same both molar mass and overall chemical composition may well differ in their blockiness and therefore their coils may assume distinct size in solution. In the case of complex polymers and complex polymer systems, the researchers often seek support in other characterization methods such as nuclear magnetic resonance, matrix assisted desorption ionization mass spectrometry and like. [Pg.283]

Other factors complicating the direct SEC analysis of nylons in fluorinated solvents are the selection of calibration standards and the polyelectrolyte effect (7) in fluorinated solvents. Another consideration is column durability under high-temperature conditions and in fluorinated solvents. Regarding commercial nylon standards, broad molecular weight nylon standards have only become available from American Polymer Standard Corporation (Mentor, Ohio) in the 1980s. A series of nylon 6,6 standards with molecular weight ranging from 27,000 to 110,000 is now obtainable from commercial suppliers. [Pg.164]

The standard requires that where inspection, testing, and calibration services are conducted fay a supplier s laboratory facility, the laboratory shall comply with ISO/IEC 17025 including use of a laboratory scope. [Pg.392]

The standard requires the supplier to establish and maintain documented procedures to control, calibrate, and maintain inspection, measuring, and test equipment including test software. [Pg.399]

The standard requires the supplier to document the basis used for calibration where no nationally/ recognized standards exist. [Pg.414]

The standard requires the supplier to ensure that the environmental conditions are suitable for the calibration, inspections, measurements, and tests being carried out. [Pg.419]

For most purposes it is not necessary to follow the procedures given above for the preparation of standard buffer solutions the buffer tablets which are available from laboratory suppliers, when dissolved in the specified volume of distilled (de-ionised) water, produce buffer solutions suitable for the calibration of pH meters. [Pg.570]

The 1000 A column did not show any resolution between 312 nm and 57 nm particle sizes. Shown in Fig.2 are the calibration curves for the 2000 A and 3000 A columns and for their combination. The 57 nm particle standard appears to have been erroneously characterized by the supplier. This was subsequently confirmed by electron microscopy. The 2000 X column exhibited a sharp upturn in its calibration curve close to the exclusion limit. It is to be noted that while data points corresponding to 312 and 275 nm diameter particles appear on individual column calibration curves, they are not indicated for the calibration curve of the combination. This is because these larger diameter particles were completely retained in the packed colimms, generating no detector response. The percentage recovery for these particles from individual columns was considerably less than 100 resulting in their complete retention when the columns were combined in series. [Pg.49]

In fact, one can construct conductance cells with accurately known values of l and A in which the conductivity of standard electrolytes can be calibrated however, in analytical practice cells with less restricted shape requirements are applied as their cell constants can be adequately established by measurements on a standard electrolyte (generally KC1) of known conductivity (see Fig. 2.3). Suppliers of commercial cells usually give the cell constants. [Pg.35]

In the standard calibration technique one adds a few millilitres of a concentrated solution of a non-interfering strength adjuster (ISA) and/or a pH adjuster to both the standard and sample solutions before measurement. Usually ISE suppliers provide a list of ISA/pH adjusters appropriate to their ISEs in order to maintain the activity coefficient, ft, and/or pH at the most suitable fixed value. [Pg.70]


See other pages where Calibration standards suppliers is mentioned: [Pg.930]    [Pg.324]    [Pg.895]    [Pg.199]    [Pg.564]    [Pg.113]    [Pg.59]    [Pg.84]    [Pg.84]    [Pg.223]    [Pg.482]    [Pg.498]    [Pg.539]    [Pg.556]    [Pg.1081]    [Pg.355]    [Pg.868]    [Pg.42]    [Pg.333]    [Pg.341]    [Pg.393]    [Pg.123]    [Pg.514]    [Pg.330]   
See also in sourсe #XX -- [ Pg.333 ]




SEARCH



Standardization calibration standards

© 2024 chempedia.info