Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Calibration quality

Test and calibration Quality procedures Software development life cycle Documentation requirements Training O M manuals... [Pg.587]

Keywords Traceability Gas standards Calibration Quality assurance Proficiency testing International comparisons... [Pg.212]

Commercial spectrometers now come with powerful computers and software. Most of the newer ICP emission systems provide software that can assist in wavelength selection, calibration, background correction, interelement correction, spectral deconvolution, standard additions calibration, quality control charts, and report generation. [Pg.855]

To obtain hyperpolarizabilities of calibrational quality, a number of standards must be met. The wavefunctions used must be of the highest quality and include electronic correlation. The frequency dependence of the property must be taken into account from the start and not be simply treated as an ad hoc add-on quantity. Zero-point vibrational averaging coupled with consideration of the Maxwell-Boltzmann distribution of populations amongst the rotational states must also be included. The effects of the electric fields (static and dynamic) on nuclear motion must likewise be brought into play (the results given in this section include these effects, but exactly how will be left until Section 3.2.). All this is obviously a tall order and can (and has) only been achieved for the simplest of species He, H2, and D2. Comparison with dilute gas-phase dc-SHG experiments on H2 and D2 (with the helium theoretical values as the standard) shows the challenge to have been met. [Pg.11]

I believe that this account has shown the enormous interplay between quantum-chemical computation and experiment in the science of non-linear optics over the past few years. It is heartening to realize that calibrational-quality results have led to the un-masking of quite obscure technical problems in some of the electric-field-induced experiments. This has still to happen for the magnetic-field-induced experiments, but I am sure that it will. [Pg.40]

Finally, a validation set with known concentrations has to be prepared that represents the investigated parameter field. It allows the final evaluation of the calibration quality described by values such as the correlation of predicted and measured concentrations and the root mean square error of prediction (RMSEP), which is comparable to the standard deviation of the predicted values (see Figure 6.6d) [20]. Subsequent to this calibration and validation procedure, a realtime spectroscopic measurement of analyte concentrations within complex reaction mixtures generated in a microreaction process can be accomplished. [Pg.1132]

Wang et al proposed a multivariate dominant factor based non-linearized PLS model for LIBS measurements. In constructing such a multivariate model, non-linear transformation of multi-characteristic line intensities according to the physical mechanisms of a laser-induced plasma spectrum were made, combined with a linear-correlation-based PLS method, to model the non-linear self-absorption and inter-element interference effects. Moreover, a secondary PLS was applied, utilizing information from the whole spectrum to correct the model results further. The proposed method showed a significant improvement when compared with a conventional PLS model. Even compared with the already improved baseline dominant-factor-based PLS model, the PLS model based on the multivariate dominant factor yielded the same calibration quality while decreasing the RMSEP. [Pg.354]

I-Particle Calibration Quality of the Electron Correlation Method... [Pg.115]

To evaluate the image quality of the processing system, one can determine classical parameters like spatial resolution, contrast resolution, dynamic range, local and global distortion. Guidelines for film digitization procedures have been well described now. Furthermore, a physical standard film for both equipment assessment and digitization calibration and control, will be available in a next future (4). [Pg.501]

Quality in NDT depends upon a number of factors. Qualification of NDT personnel, technical state and correctness of choice of testing equipment, availability of approved working procedures of examination, calibration of NDT equipment have decisive importance among those factors of an NDT laboratory. Assessment of NDT laboratory competence is provided through accreditation in compliance with the EN 45000 series standards. [Pg.953]

The main task of the non-destructive testing is availability of objective and reliable information about the object being examined in assessment of its quality. Calibration of NDT equipment before the testing realization, adjustment of... [Pg.958]

If the task is multivariate calibration, for example, the proper choice of a pre-processing method will essentially aflFect the quality of the resultant model. For more details about the use of these techniques together with PCA and PLS, readers are advised to consider the fundamental monograph by Erikson et al [8]. [Pg.215]

Quality control elements required by the instrumental analyzer method include analyzer calibration error ( 2 percent of instrument span allowed) verifying the absence of bias introduced by the sampling system (less than 5 percent of span for zero and upscale cah-bration gases) and verification of zero and calibration drift over the test period (less than 3 percent of span of the period of each rim). [Pg.2200]

Known samples should also be run to verify the accuracy and precision of the routine methods to be used during the unit test. Poor quality will manifest itself as poor precision, measurements inconsistent with plant experience or laboratory history, and disagreement among methods. Plotting of laboratory analysis trends wiU help to determine whether calibrations are drifting with time or changing significantly. Repeated laboratory analyses will establish the confidence that can be placed in the results. [Pg.2558]

A solvent free, fast and environmentally friendly near infrared-based methodology was developed for the determination and quality control of 11 pesticides in commercially available formulations. This methodology was based on the direct measurement of the diffuse reflectance spectra of solid samples inside glass vials and a multivariate calibration model to determine the active principle concentration in agrochemicals. The proposed PLS model was made using 11 known commercial and 22 doped samples (11 under and 11 over dosed) for calibration and 22 different formulations as the validation set. For Buprofezin, Chlorsulfuron, Cyromazine, Daminozide, Diuron and Iprodione determination, the information in the spectral range between 1618 and 2630 nm of the reflectance spectra was employed. On the other hand, for Bensulfuron, Fenoxycarb, Metalaxyl, Procymidone and Tricyclazole determination, the first order derivative spectra in the range between 1618 and 2630 nm was used. In both cases, a linear remove correction was applied. Mean accuracy errors between 0.5 and 3.1% were obtained for the validation set. [Pg.92]

Two separate flowmeter differentials should be read. These will probably have to be read with DP transmitter. It is suggested both DP units have one quality calibrated 6-in. gauge on transmitter output to directly read differential in %FS, FR, or psi. DP transmitters should be carefully bench-checked/calibrated, retaining indicated versus actual calibration data. [Pg.325]

Hundreds of chemical species are present in urban atmospheres. The gaseous air pollutants most commonly monitored are CO, O3, NO2, SO2, and nonmethane volatile organic compounds (NMVOCs), Measurement of specific hydrocarbon compounds is becoming routine in the United States for two reasons (1) their potential role as air toxics and (2) the need for detailed hydrocarbon data for control of urban ozone concentrations. Hydrochloric acid (HCl), ammonia (NH3), and hydrogen fluoride (HF) are occasionally measured. Calibration standards and procedures are available for all of these analytic techniques, ensuring the quality of the analytical results... [Pg.196]

U.S. Environmental Protection Agency, "Transfer Standards for Calibration of Air Monitoring Analyzers for Ozone," EPA-600/4-79-056. Office of Air Quality Planning and Standards, Research Triangle Park, NC, 1979. [Pg.213]

The components of a quality assurance program are designed to serve the two functions just mentioned—control and assessment. Quality control operations are defined by operational procedures, specifications, calibration procedures, and standards and contain the following components ... [Pg.223]

In addition to fulfilling the in-house requirements for quality control, state and local air monitoring networks which are collecting data for compliance purposes are required to have an external performance audit on an annual basis. Under this program, an independent organization supplies externally calibrated sources of air pollutant gases to be measured by the instrumentation undergoing audit. An audit report summarizes the performance of the instruments. If necessary, further action must be taken to eliminate any major discrepancies between the internal and external calibration results. [Pg.224]

Documentation Data volume includes all quality control forms, e.g., zero/span control charts and multipoint calibration results... [Pg.224]

Good-quality Bourdon tube test gauges are highly suitable for pressure measurements of more than 20psi. They should be calibrated against a... [Pg.695]

The elements of a PM plan include periodic inspection, cleaning, and service as warranted, adjustment and calibration of control system components, maintenance equipment and replacement parts that are of good quality and properly selected for the intended function. Critical HVAC system components that require PM in order to maintain comfort and deliver adequate ventilation air include a outdoor air intake opening, damper controls, air filters, drip pans, cooling and heating coils, fan belts, humidification equipment and controls, distribution systems, exhaust fans. [Pg.211]


See other pages where Calibration quality is mentioned: [Pg.79]    [Pg.407]    [Pg.281]    [Pg.355]    [Pg.170]    [Pg.141]    [Pg.315]    [Pg.11]    [Pg.169]    [Pg.280]    [Pg.290]    [Pg.341]    [Pg.535]    [Pg.116]    [Pg.79]    [Pg.407]    [Pg.281]    [Pg.355]    [Pg.170]    [Pg.141]    [Pg.315]    [Pg.11]    [Pg.169]    [Pg.280]    [Pg.290]    [Pg.341]    [Pg.535]    [Pg.116]    [Pg.508]    [Pg.706]    [Pg.707]    [Pg.714]    [Pg.325]    [Pg.312]    [Pg.410]    [Pg.435]    [Pg.190]    [Pg.494]   
See also in sourсe #XX -- [ Pg.495 , Pg.495 ]




SEARCH



Calibrate devices affecting quality

Calibration Strategies and Quality Assurance

Calibration instrument quality

Calibration, Quality Control Materials, and Procedures

Permeation devices, containing dangerous goods, for calibrating air quality monitoring

Quality control calibration

© 2024 chempedia.info