Atomic absorption spectral interferences

Minimizing Spectral Interference A spectral interference occurs when an analyte s absorption line overlaps with an interferant s absorption line or band. As noted previously, the overlap of two atomic absorption lines is seldom a problem. On the other hand, a molecule s broad absorption band or the scattering of source radiation is a potentially serious spectral interference. [Pg.418]

An important question to consider when using a flame as an atomization source, is how to correct for the absorption of radiation by the flame. The products of combustion consist of molecular species that may exhibit broad-band absorption, as well as particulate material that may scatter radiation from the source. If this spectral interference is not corrected, then the intensity of the transmitted radiation decreases. The result is an apparent increase in the sam-... [Pg.418]

Atomic emission is used for the analysis of the same types of samples that may be analyzed by atomic absorption. The development of a quantitative atomic emission method requires several considerations, including choosing a source for atomization and excitation, selecting a wavelength and slit width, preparing the sample for analysis, minimizing spectral and chemical interferences, and selecting a method of standardization. [Pg.437]

BeryUium aUoys ate usuaUy analyzed by optical emission or atomic absorption spectrophotometry. Low voltage spark emission spectrometry is used for the analysis of most copper-beryUium aUoys. Spectral interferences, other inter-element effects, metaUurgical effects, and sample inhomogeneity can degrade accuracy and precision and must be considered when constmcting a method (17). [Pg.68]

Spectral interferences in AAS arise mainly from overlap between the frequencies of a selected resonance line with lines emitted by some other element this arises because in practice a chosen line has in fact a finite bandwidth . Since in fact the line width of an absorption line is about 0.005 nm, only a few cases of spectral overlap between the emitted lines of a hollow cathode lamp and the absorption lines of metal atoms in flames have been reported. Table 21.3 includes some typical examples of spectral interferences which have been observed.47-50 However, most of these data relate to relatively minor resonance lines and the only interferences which occur with preferred resonance lines are with copper where europium at a concentration of about 150mgL 1 would interfere, and mercury where concentrations of cobalt higher than 200 mg L 1 would cause interference. [Pg.792]

With flame emission spectroscopy, there is greater likelihood of spectral interferences when the line emission of the element to be determined and those due to interfering substances are of similar wavelength, than with atomic absorption spectroscopy. Obviously some of such interferences may be eliminated by improved resolution of the instrument, e.g. by use of a prism rather than a filter, but in certain cases it may be necessary to select other, non-interfering, lines for the determination. In some cases it may even be necessary to separate the element to be determined from interfering elements by a separation process such as ion exchange or solvent extraction (see Chapters 6, 7). [Pg.792]

Spectral overlap of emission and absorption wavelengths Is a potential cause of Interference In atomic absorption spectrometry (57) Thus, (a) the emission line of Fe at 352.424 nm Is close to the resonance line of N1 at 352.454, (b) the emission line of Sb at 217.023 nm Is close to the resonance line of Pb at 216.999 nm, and (c) the emission line of As at 228.812 nm Is close to the resonance line of Cd at 228.802 (57). To date, these practically coincident spectral lines have not been reported to be of practical Importance as sources of analytical Interference In atomic absorption analyses of biological materials. [Pg.258]

Practically all classical methods of atomic spectroscopy are strongly influenced by interferences and matrix effects. Actually, very few analytical techniques are completely free of interferences. However, with atomic spectroscopy techniques, most of the common interferences have been studied and documented. Interferences are classified conveniently into four categories chemical, physical, background (scattering, absorption) and spectral. There are virtually no spectral interferences in FAAS some form of background correction is required. Matrix effects are more serious. Also GFAAS shows virtually no spectral interferences, but... [Pg.606]

Any difference in the behaviour of the analyte atoms in the sample and in the standard implies an interference. AAS using a line source for excitation suffers little spectral interference. Background interference in AAS is more important. This nonspecific absorption is caused by ... [Pg.609]

Spectral interferences are not common in atomic absorption but can occur. An element with an absorption line sufficiently close to the one of the test element that it overlaps would cause a positive interference. Fassel et a/.20) have discussed the problems of spectral interference. This type of interference, especially in biological samples, occurs only rarely, but the analyst should be aware of it. It is more serious if a continuous source is used. Molecular absorption is a more common spectral interference and occurs when a molecular absorption band overlaps with the atomic absorption line. For example, the CaOH species absorbs in the region of the barium 5535.5 A line. A 1 % calcium solution gives an absorption equivalent to what is expected from about 75 ppm barium21). [Pg.86]

Flame emission spectrometry is used extensively for the determination of trace metals in solution and in particular the alkali and alkaline earth metals. The most notable applications are the determinations of Na, K, Ca and Mg in body fluids and other biological samples for clinical diagnosis. Simple filter instruments generally provide adequate resolution for this type of analysis. The same elements, together with B, Fe, Cu and Mn, are important constituents of soils and fertilizers and the technique is therefore also useful for the analysis of agricultural materials. Although many other trace metals can be determined in a variety of matrices, there has been a preference for the use of atomic absorption spectrometry because variations in flame temperature are much less critical and spectral interference is negligible. Detection limits for flame emission techniques are comparable to those for atomic absorption, i.e. from < 0.01 to 10 ppm (Table 8.6). Flame emission spectrometry complements atomic absorption spectrometry because it operates most effectively for elements which are easily ionized, whilst atomic absorption methods demand a minimum of ionization (Table 8.7). [Pg.319]

Interferences in atomic absorption measurements can arise from spectral, chemical and physical sources. Spectral interference resulting from the overlap of absorption lines is rare because of the simplicity of the absorption spectrum and the sharpness of the lines. However, broad band absorption by molecular species can lead to significant background interference. Correction for this may be made by matrix matching of samples and standards, or by use of a standard addition method (p. 30 et seq.). [Pg.331]

Why are spectral interferences less important in atomic absorption spectroscopy and atomic fluorescence spectroscopy than atomic emission spectroscopy ... [Pg.252]

The presence and concentration of various metallic elements in petroleum coke are major factors in the suitability of the coke for various uses. In the test method (ASTM D5056), a sample of petroleum coke is ashed (thermally decomposed to leave only the ash of the inorganic constituents) at 525°C (977°F). The ash is fused with lithium tetraborate or lithium metaborate. The melt is then dissolved in dilute hydrochloric acid and the resulting solution is analyzed by atomic absorption spectroscopy to determine the metals in the sample. However, spectral interferences may occur when using wavelengths other than those recommended for analysis or when using multielement hollow cathode lamps. [Pg.300]

The atomic absorption characteristics of technetium have been investigated with a technetium hollow-cathode lamp as a spectral line source. The sensitivity for technetium in aqueous solution is 3.0 /ig/ml in a fuel-rich acetylene-air flame for the unresolved 2614.23-2615.87 A doublet under the optimum operating conditions. Only calcium, strontium, and barium cause severe technetium absorption suppression. Cationic interferences are eliminated by adding aluminum to the test solutions. The atomic absorption spectroscopy can be applied to the determination of technetium in uranium and its alloys and also successfully to the analysis of multicomponent samples. [Pg.134]

A convenient method is the spectrometric determination of Li in aqueous solution by atomic absorption spectrometry (AAS), using an acetylene flame—the most common technique for this analyte. The instrument has an emission lamp containing Li, and one of the spectral lines of the emission spectrum is chosen, according to the concentration of the sample, as shown in Table 2. The solution is fed by a nebuhzer into the flame and the absorption caused by the Li atoms in the sample is recorded and converted to a concentration aided by a calibration standard. Possible interference can be expected from alkali metal atoms, for example, airborne trace impurities, that ionize in the flame. These effects are canceled by adding 2000 mg of K per hter of sample matrix. The method covers a wide range of concentrations, from trace analysis at about 20 xg L to brines at about 32 g L as summarized in Table 2. Organic samples have to be mineralized and the inorganic residue dissolved in water. The AAS method for determination of Li in biomedical applications has been reviewed . [Pg.324]

D. C. Baxter and J. Ohman, Multi-component standard additions and partial least squares modelling, a multivariate calibration approach to the resolution of spectral interferences in graphite furnace atomic absorption spectrometry, Spectrochim. Acta, Part B, 45(4 5), 1990, 481 491. [Pg.240]

In atomic spectroscopy, absorption, emission, or fluorescence from gaseous atoms is measured. Liquids may be atomized by a plasma, a furnace, or a flame. Flame temperatures are usually in the range 2 300-3 400 K. The choice of fuel and oxidant determines the temperature of the flame and affects the extent of spectral, chemical, or ionization interference that will be encountered. Temperature instability affects atomization in atomic absorption and has an even larger effect on atomic emission, because the excited-state popula-... [Pg.470]

D. The measurement of Li in brine (salt water) is used by geochemists to help determine the origin of this fluid in oil fields. Flame atomic emission and absorption of Li are subject to interference by scattering, ionization, and overlapping spectral emission from other elements. Atomic absorption analysis of replicate samples of a marine sediment gave the results in the table below. [Pg.471]

The formed free atoms absorb the light at a characteristic wavelength from a hollow cathode lamp that is positioned on one side of the flame. A spectrophotometer with a grating monochromator located on the other side of the flame measures the intensity of the light beam. Because absorption is proportional to the number of free atoms that are produced in the flame, the light energy absorbed by the flame is a measure of the element s concentration. The FLAA technique is relatively free of interelement spectral interferences, but it has the sensitivity that is inferior to ICP-AES or GFAA. [Pg.233]

The importance of Walsh s ideas should not be underestimated. Not only had he suggested a potentially highly sensitive method of analysis which would prove eventually to be suitable for the determination of many elements in the periodic table, but at the same time he had suggested a development which, theoretically at least, should lead to virtually specific analysis. The very narrowness of the absorption lines which had hitherto held back progress in AAS suddenly became its most powerful asset. It meant that the chances of spectral overlap of the absorption line of one element with the emission line of another were extremely small. Thus atomic spectral interferences should be, and indeed are, rare in AAS. [Pg.6]

At high concentrations especially, a number of elements produce significant concentrations of polyatomic species in flames. Such species absorb, and may therefore cause spectral interference. However the molecular absorption spectra are very wide compared with the atomic spectral lines. Figure 6, for example, shows how the presence of CaOH species in flames may interfere in the determination of barium by AAS. The formation of any solid particles in the flame causes scatter, which also causes an apparent broad band absorption, especially at lower wavelengths. [Pg.38]


See other pages where Atomic absorption spectral interferences is mentioned: [Pg.112]    [Pg.416]    [Pg.446]    [Pg.609]    [Pg.611]    [Pg.292]    [Pg.305]    [Pg.235]    [Pg.236]    [Pg.241]    [Pg.245]    [Pg.245]    [Pg.246]    [Pg.58]    [Pg.105]    [Pg.95]    [Pg.11]    [Pg.182]    [Pg.356]    [Pg.467]    [Pg.702]    [Pg.292]    [Pg.305]    [Pg.472]    [Pg.135]    [Pg.9]   
See also in sourсe #XX -- [ Pg.186 ]