Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Assay calibration

The assay was carried out using a Varian gas chromatograph (model 5000 LC) under the following experimental condition. The oven injector and flame ionization detector temperatures were 125°C and 225°C respectively. A Porapak column was used, the eluent was N2 at a flow rate of 30 ml/min and the injected volume 2 pi. Various concentrations of purified methylene chloride in purified methanol were injected (both solvents were distilled to discard any impurity which might interfere with the sensitive assay). Calibration curves were linear in the range 50-500 ppm (the limit of detection was 10 ppm). Methylene chloride detection in the microspheres was performed by dissolving various amounts (20-200 mg) of microspheres in 220 ml of purified methanol prior to the injection. [Pg.105]

As with any analytical technique, generation of a reproducible standard curve with minimal error is critical. An assay calibration consists of several steps during which the value of the primary standard is transferred to the calibrators used in the final assay [22]. Immunoassay optimization is usually difficult due to protein heterogeneity and matrix effects and these factors, heterogeneity and matrix effects, will also affect MIP based assays [22]. [Pg.130]

Quantitative LC/MS assays generally involve four steps (1) sample preparation, (2) assay calibration, (3) sample analysis, and (4) data management. In this method, the human serum samples are prepared with a liquid-liquid extraction procedure. The internal... [Pg.148]

Assay calibration involves the use of human serum samples fortified with CP-80,794 at 11 concentrations (six replicates per concentration) ranging from 0.05 to lOng/mL. In this particular case, due to a narrow linear dynamic range, two standard curves, ranging from 0.05 to lOng/mL, are constructed to provide the best accuracy. Serum blanks and an 11-point standard curve (two samples per concentration) are analyzed with each set of unknown samples. [Pg.149]

In order to obtain calibration curves for other ratios of parent drug and metabolites, the ratio can be modified as well. In many cases, all analytes are used in equimolar ratios in order to obtain the same calibration range for all of the analytes. However, the assay calibration procedure has to cover the whole calibration range that is going to be used for each and every analyte. [Pg.622]

Murthy K, Stevens LA, Stark PC, Levey AS. Variation in serum creatinine assay calibration A practical application to glomerular filtration rate estimation. Kidney Int 2005 68 1884-1887. [Pg.118]

Analytical toxicology is the use of qualitative and quantitative chemical and physical techniques used in sample preparation, separation, assay calibration, detection and identification, and quantification for the purposes of toxicological research and testing. Examples of the objectives of such analysis include ... [Pg.113]

RETINOIC ACID ASSAY. Calibration curves were obtained by adding known amounts of 13-cis (40 /il) and all trans (40 /il) retinoic acid in ethanol to 1 ml of phosphate buffered saline... [Pg.167]

Immunoassays are used to measure PSA and are commercially available. Most of them use nonisotopic labels, such as enzyme, fluorescence, or chemiluminescence. The majority of these assays are automated on an immunoassay system. Different assays and even the same assay with different lots of reagent may produce different results. The reasons for such differences are due to changes in assay calibration, production lot variation, assay reaction time, reagent matrices, assay sensitivity, and imprecision. Antibodies react with different PSA epitopes therefore, some antibodies react dissimilarly with the various molecular forms of PSA. Assays are classified as equimolar if they bind to free and cPSA equally and nonequimolar if they bind to free or cPSA differently. Examples of equunolar assays are the ACCESS... [Pg.760]

Plasma creatinine is an imperfect marker of GFR and therefore it is not altogether surprising that formulas based predommantly upon it are imperfect. Their use cannot circumvent the very significant spectral interferences affecting plasma creatinine measurement (i.e., hemolysis, icterus, and lipemia) and the formulas are unsuitable for use in patients with acute renal failure, in whom plasma creatinine concentrations are changing rapidly. Additionally, the formulas are critically susceptible to variations in creatinine assay calibration and specificity. Notwithstanding the MDRD formula is thought to improve the estimation of GFR compared with plasma creatinine alone. [Pg.823]

Because the 3rd and 4th IS contain about 5% nicked CG and other CG contaminants, highly pure CG materials have been prepared.These preparations are designated reference reagents (RR) and contain CG (99/688), CGn (99/642), CGp (99/650), CGpn (99/692), CGpcf, and CGa (99/720). The impact of using RR for assay calibration and harmonization is under study. [Pg.2181]

In spite of a common calibrator, assays calibrated using the 1st IRP/3rd IS still yield significant differences in CG values on the same specimens. Therefore CG median values calculated for second trimester serum screening are not transferable and should be considered method-specific, CG measurements in early pregnancy are generally expressed as lU/L or international units per miUifiter. A typical CG concentration at 16 weeks is approximately 30,000 lU/L, and many screening laboratories express CG concentrations in international units per milliliter (e.g., 30,000 lU/L is... [Pg.2181]

Fig 3 shows the kinetics for the ATP assays. Fig. 4 shows assay calibration curves. The estimated detection Umit for each assay was as follows 0.51 picomoles ATP 19%, 5.1 picomoles NADH 21%, 5.1 picomoles Galactose 26%, and 51 picomoles Lactate 22%. [Pg.235]

Once a well-characterized, pure reference standard of known concentration is available, one must be aware of the fact that the actual preparation of the assay calibrators can also influence the PK standard curve. The physical treatment and handling of concentrated protein solutions can be very challenging. Some reference standards require specific handling techniques to prevent aggregation or denaturation of the protein. The effects of vortexing can be disastrous as can the process used to thaw frozen reference standard or calibrator solutions. Details such as the dilution pattern that is used in spiking the calibration solutions can contribute to assay bias and between-laboratory discrepancies. [Pg.257]

Catechin is generally used as a standard in the DMACA assay. Calibration plots are linear and therefore, calibration factors are also applied [148]. Like in all conventional methods the use of purified procyanidins as a standard will give the correctest estimates. A blank, composed of the samples without the chromogen reagent, is also... [Pg.522]

Biomarker assays (as other bioassays) may be classified into definitive quantitative assays, relative quantitative assays, quasi-quantitative assays, and qualitative assays with varying degrees of validation requirements (Table 5.5-3) [14, 15]. For definitive quantitative assays, a well-defined or characterized standard of the biomarker is available. In the case of relative quantitative assays, calibration is performed with a standard that is not well characterized, not available in pure form, or not representative of the endogenous biomarker. Results from these assays are... [Pg.623]

Pre-study validation defines that the method produces reliable results. During pre-study validation, fundamental parameters such as selectivity, assay calibration, accuracy, precision, linearity, and stability are evaluated. [Pg.110]

Quantitative analyses can be performed either isothermally or with a temperature program. For analytes that are liquids at room temperature, assay calibration is by analysis of standard solutions prepared in analyte-free human blood. The same calibration solutions are used in the analysis of tissue digests. Analyte concentrations in the range 0.1-10 or 0.5-50 mg are usually adequate in acute poisoning. Portions of the standards are transferred to headspace vials for analysis as described above, and a... [Pg.1754]

For analytes that are gases at room temperature, assay calibration is much more complicated and involves preparation of dilutions of gaseous calibration mixtures. [Pg.1755]

No standard reference material and procedure is actually available for molal ion activity assays. Calibration of the electrode is required. The calibration solution should match the mean sample composition as closely as possible. Addition of albumin to the calibration standard does not guarantee high accuracy of the results due to a shift of the assay standard potential that is different for plasma, serum, and albumin. Reports on comparisons of clinical analyzers show that interinstrumental deviations may be considerable. Recommendations for calibration and standardization by the IFCC Committee on Electrolytes and Blood Gases are in preparation. [Pg.2374]

The enantiomers of azepinoindole and its 7V-desmethyl metabolites were well resolved on a 8-cyclodextrin-modified silica column (A = 231 run). A 1/99 r-butyl alcohol/water (10 mM sodium phosphate buffer at pH 7 with 15 mM jS-cyclodextrin and 2 mM triethylamine) mobile phase was used to generate the chiral separation. The assay calibration range covered 100-500 ng/mL. Peaks were somewhat tailed. Elution was complete in 25 min [584]. [Pg.211]

HoloTC RIA and holoTC enzyme immunoassay (EIA) were evaluated in a multicentre study involving four European laboratories (Nexo and Hoffmann-Lucke 2011) with similar results calibration curve 1.36-24.5 pmol/L, imprecision 6% for EIA and 10% for RIA. Moreover, RIA results were also similar to those of the microbiological assay (calibration curve 1.36-24.5 pmol/L imprecision 4-7%) (Refsum et al. 2006). [Pg.494]


See other pages where Assay calibration is mentioned: [Pg.317]    [Pg.153]    [Pg.49]    [Pg.219]    [Pg.114]    [Pg.448]    [Pg.3430]    [Pg.98]    [Pg.4]    [Pg.826]    [Pg.960]    [Pg.2080]    [Pg.2177]    [Pg.2181]    [Pg.2]    [Pg.247]    [Pg.252]    [Pg.253]    [Pg.110]    [Pg.300]    [Pg.167]    [Pg.240]    [Pg.326]    [Pg.115]   
See also in sourсe #XX -- [ Pg.69 ]




SEARCH



Analytical performance assays, calibration curves

Calibration curve Bradford protein assay

Calibration curve Lowry protein assay

Calibration curves biomarker assays

Calibration standards commercial assay kits

Commercial assay kits calibration validation

Dilution linearity assays, calibration curves

Enzyme linked immunosorbent assays calibration curves

Ligand binding assay calibration curves

Linearity assays, calibration curves

© 2024 chempedia.info