Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sensitivity instrumental

ICP-SFMS (Thermo Finnigan, Flement) with cold vapour generation was developed with a guard electrode and a gold amalgamation device using an Au-sorbent for sample pre-concentration to improve the sensitivity. Instrumental parameters of ICP-SFMS such as take-up time, heating temperature of Au-sorbent, additional gas flow, and sample gas flow were optimized. Detection limit calculated as 3 times the standard deviation of 10 blanks was 0,05 ng/1, RSD = 7-9 %. [Pg.171]

U2 = 0.1 V) and are suitable for current measurement. For smaller currents, sensitive instruments with 5 kQ per fiPc (Uj = 5 mV) are used. Small currents are usually measured by the voltage drop across a fixed resistance (calibrated shunt) using an electronic amplifier-voltmeter. This method has the advantage that the circuit does not have to be interrupted to measure the current. [Pg.85]

On the simplest type of instrument (an explosimeter) only one scale is provided, usually with readings from 0 to 100% LEL. However, the detectable changes produced by combustion are too small to be measured accurately in the presence of the low concentrations of contaminants usually encountered in evaluating potential health hazards. For example, the LEL of even the most explosive gas is of the order of 1 %, or 10,000 ppm, which is well in excess of the toxic limit for any gas. Therefore, explosimeters or combustible gas indicators which have only a 0-to-l(X)% LEL explosive scale are not suitable for environmental health testing in the ppm range. More sensitive instruments, including the type used in sampling for environmental health purposes, have a dual scale, in which the second, more... [Pg.270]

The sulfur dioxide analyzer based on the ultraviolet principle is a sensitive instrument. Its detection limit can be less than one ppbv (parts per billion by volume). When used in emission measurements, the sample gas IS normally diluted prior to the measurement using a diluting stack sampler. [Pg.1300]

Radiometer A sensitive instrument for the measurement of heat radiation. [Pg.1471]

EPR methods that allow a more direct determination of kv have been developed. These enable absolute radical concentrations to be determined as a function of conversion. With especially sensitive instrumentation, this can be done by direct measurement/57 160 An alternative method, applicable at high conversions, involves trapping the propagating species in a frozen matrix361 362 by rapid cooling of the sample to liquid nitrogen temperatures. [Pg.217]

A sensitive instrument containing a single large magnet, weighing about six pounds, has also been designed and built (Fig. 5). [Pg.671]

A janitor surrounded by sensitive instruments may be like a bull in a china shop. In one laboratory, the janitors were not allowed to clean anything above floor level, including windows, without supervision by laboratory personnel. Janitors should not be allowed to handle containers of flammable or hazardous waste. Laboratory personnel should clear the floors completely when the janitor announces that it is time for floor waxing. [Pg.116]

Aldehydes Glutaraldehyde Slight pH8 Respiratory complaints and contact dermatitis reported Eyes, sensitivity OES 0.2 ppm/0.7 mg rrr, tOmin only Non-corrosive, useful for heat sensitive instruments... [Pg.209]

The experimental designs discussed in Chapters 24-26 for optimization can be used also for finding the product composition or processing condition that is optimal in terms of sensory properties. In particular, central composite designs and mixture designs are much used. The analysis of the sensory response is usually in the form of a fully quadratic function of the experimental factors. The sensory response itself may be the mean score of a panel of trained panellists. One may consider such a trained panel as a sensitive instrument to measure the perceived intensity useful in describing the sensory characteristics of a food product. [Pg.444]

Environmental monitoring of chloroacetanilides requires methods that have the capability to distinguish between complex arrays of related residues. The two example methods detailed here for water monitoring meet this requirement, but the method for metabolites requires sophisticated mass spectral equipment for the detection of directly injected water samples. In the near term, some laboratories may need to modify this method by incorporation of an extraction/concentration step, such as SPE, that would allow for concentration of the sample, so that a less sensitive and, correspondingly, less expensive, mass spectral detector can be used. However, laboratories may want to consider purchasing a sensitive instrument rather than spending time on additional wet chemistry procedures. In the future, sensitive instrumentation may be less expensive and available to all laboratories. Work is under way to expand the existing multi-residue methods to include determination of additional chloroacetanilides and their metabolites in both water and soil samples. [Pg.387]

The nature of water residue analysis has changed dramatically over the past 60 years. Advances in SPE media have made the isolation of hard to extract residues more attainable. The use of specific and sensitive instrumentation as in HPLC/MS/MS has even precluded the need for extraction in many cases, since the samples can be directly analyzed as they are received from the Held. The future holds even more improvements in sensitivity with rugged new API interfaces and high-resolution mass spectrometers that will dramatically increase the specificity of detection and ease of analysis. [Pg.837]

Later work by several research workers, and most particularly Greef (1969), showed that the very early stages of oxide film formation could be detected using more sensitive instrumentation and the changes in A, 4 and the light intensity correlated very well with the charge passed both in the anodic and cathodic sweeps (see Figure 3.13). [Pg.256]

By knowing the history of the sample it is possible to act correctly during all of the sampling steps, in order to avoid contamination of the material either from the utensils used to collect the sample, or from the reagents, the laboratory atmosphere and even the laboratory personnel. The contamination risk is greatest in analysis for trace components. Trace analysis requires the use of specially acclimatized, sometimes over-pressurized laboratories, of very pure reagents and sensitive instruments, and of specialized personnel, who possess a broad range of... [Pg.433]

The Accelerating Rate Calorimeter (ARC ) is another adiabatic test instrument that can be used to test small samples. The ARC with the clamshell containment design can handle explosive compounds. It is a sensitive instrument that can indicate the onset of exothermicity where the reaction mixture can be accurately simulated (HSE 2000). ARC testing results can be used in determining a time to maximum rate of decomposition, as well as in calculating a temperature of no return for a container or vessel with specific heat removal characteristics. Further information and references related to the ARC are given in CCPS (1995a) and Urben (1999). [Pg.100]

Thermoanalytical methods, including thermogravimetry, have been used originally in the earth sciences and in chemistry. With todays multifunctional and highly sensitive instruments, these methods are becoming important for many other, new fields of science and technology. [Pg.142]

The difficulties include the inconvenience of handling radioactivity and the necessity for obtaining an accurate radiochemical analysis of two phases containing several elements (which often involves complicated spectra). Highly sensitive instrumentation is required for the analysis e.g. a Li-Si surface barrier detector for a particles, a 2 r gas counter for (3-radiation and a Li-Ge detector for 7-radiation. Great care is required during source preparation, which is best done by electrodeposition. [Pg.19]

For this method, the drug candidate is labeled with a radioisotope, such as carbon-14. The AD ME of the compound within the body can be monitored by analyzing samples using high sensitivity instrumentation, for example, accelerator mass spectroscopy. [Pg.181]

Before moving on to excerpt 4E, we call your attention to two ways in which the concept of zero is addressed in excerpt 4D. First, we consider the concept of zero in measured concentrations (i.e., the concentrations reported in the last column of Table 1). Recall that no chromium oxalate was detected in the cells however, the authors do not report this with a zero. Rather, they use the phrase below the detection limit in the text and the less-than symbol (e.g., <0.025 mg/g) in Table 1, which puts an upper limit on the amount of chromium oxalate present. Novice writers might (incorrectly) suggest that no chromium was present in the text and use a zero in the table (0 mg/g). Such uses of zero, however, are incorrect, because (for measured concentrations) zero varies with the sensitivity of the detecting instrument. For example, on one instrument, zero will be less than one part per million on a more sensitive instrument, zero will be less than one part per billion. Instead of reporting zero, authors report that the measurement was below the detection limit for that instrument. Some common ways to express this concept in the text and table are as follows ... [Pg.132]

Measured concentrations should not be reported as zero, even if the substance is not detected. A more sensitive instrument (one with a lower detection limit) might be able to detect it. In other words, even if you don t "see" the substance, it might be there ... [Pg.133]

TURBOVAC pumps are precisely balanced and may generally be connected directly to the apparatus. Only in the case of highly sensitive instruments, such as electron microscopes, is it recommended to install vibration absorbers which reduce the present vibrations te a minimum. For magnetically suspended pumps a direct connection to the vacuum apparatus will usually do because of the extremely low vibrations produced by such pumps. [Pg.50]

Of course assay sensitivity applies in the same way to trials which evaluate superiority, but in those cases things take care of themselves. A conclusion of superiority by definition implies that the trial is a sensitive instrument, otherwise superiority would not have been detected. [Pg.180]

Some of the energy returns to the surface where it is detected by sensitive instruments. When these instruments are placed near the source, so as to record the energy traveling in an essentially vertical path, the technique is known as the reflection method. When the instruments are extended over large distances from the source, as compared to the depths of interest, the technique is known as the refraction method pp351—52)... [Pg.501]


See other pages where Sensitivity instrumental is mentioned: [Pg.82]    [Pg.625]    [Pg.361]    [Pg.32]    [Pg.216]    [Pg.648]    [Pg.89]    [Pg.22]    [Pg.342]    [Pg.473]    [Pg.115]    [Pg.108]    [Pg.40]    [Pg.328]    [Pg.183]    [Pg.2]    [Pg.264]    [Pg.108]    [Pg.18]    [Pg.158]    [Pg.321]    [Pg.193]    [Pg.16]    [Pg.82]    [Pg.200]    [Pg.204]    [Pg.131]   
See also in sourсe #XX -- [ Pg.75 , Pg.76 ]




SEARCH



Analytical Instrumentation for Sensitive Detection and Spectral Reproducibility

Instrumental neutron activation analysis sensitivity

Instrumentation amplifier sensitivity

Instrumentation concepts sensitivity

Sensitivity heat flow instruments

Sensitivity instrument

Sensitivity instrument

Sensitivity of instruments

Sensitivity portable instrumentation

Sensitivity, measurement instruments

Stress instrumentation, sensitivities

© 2024 chempedia.info