Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Sampling concentration

This method can attain, depending on the sample, concentrations on the order of 10 ppm wt. with an error on the order of 20%. [Pg.33]

Specific Rotation. Optical rotation is caused by individual molecules of the optically active compound. The amount of rotation depends upon how many molecules the light beam encounters in passing through the tube. When allowances are made for the length of the tube that contains the sample and the sample concentration, it is found that the amount of rotation, as well as its direction, is a characteristic of each individual optically active compound. [Pg.46]

Finally, we note that the size and shape of the particles of the packing, the packing technique, and column dimensions and configuration are additional factors which influence a GPC experiment. In addition, the flow rate, the sample size, the sample concentration, the solvent, and the temperature must all be optimized. Details concerning these considerations are found in analytical chemistry references, as well as in the technical literature of instrument manufacturers. [Pg.652]

Rearranging the above expression yields impurity concentration as a function of relative intensity, = Ral where represents both sample concentration and any background effects. The stabiHty of the cahbration must be confirmed at least every two weeks by analysis of a known mixed impurity standard. [Pg.90]

Quantitative mass spectrometry, also used for pharmaceutical appHcations, involves the use of isotopicaHy labeled internal standards for method calibration and the calculation of percent recoveries (9). Maximum sensitivity is obtained when the mass spectrometer is set to monitor only a few ions, which are characteristic of the target compounds to be quantified, a procedure known as the selected ion monitoring mode (sim). When chlorinated species are to be detected, then two ions from the isotopic envelope can be monitored, and confirmation of the target compound can be based not only on the gc retention time and the mass, but on the ratio of the two ion abundances being close to the theoretically expected value. The spectrometer cycles through the ions in the shortest possible time. This avoids compromising the chromatographic resolution of the gc, because even after extraction the sample contains many compounds in addition to the analyte. To increase sensitivity, some methods use sample concentration techniques. [Pg.548]

Colorimetric Method. A finely powdered sample treated with sulfuric acid, hydrobromic acid [10035-10-6] and bromine [7726-95-6] gives a solution that when adjusted to pH 4 may be treated with dithi one [60-10-6] ia / -hexane [110-54-3] to form mercuric dithi2onate [14783-59-6] (20). The resultant amber-colored solution has a color iatensity that can be compared against that of standard solutions to determine the mercury concentration of the sample. Concentrations below 0.02 ppm have been measured by this method. [Pg.108]

Compound Sample concentration, ppm Recovery, % Standard deviation... [Pg.242]

Essential features of an automated method are the specificity, ie, the assay should be free from interference by other semm or urine constituents, and the sensitivity, ie, the detector response for typical sample concentration of the species measured should be large enough compared to the noise level to ensure assay precision. Also important are the speed, ie, the reaction should occur within a convenient time interval (for fast analysis rates), and adequate range, the result for most samples should fall within the allowable range of the assay. [Pg.392]

The possibility of preconcentration of selenium (IV) by coprecipitation with iron (III) hydroxide and lanthanum (III) hydroxide with subsequent determination by flame atomic absorption spectroscopy has been investigated also. The effect of nature and concentration of collector and interfering ions on precision accuracy and reproducibility of analytical signal A has been studied. Application of FefOH) as copreconcentrant leads to small relative error (less than 5%). S, is 0.1-0.2 for 5-100 p.g Se in the sample. Concentration factor is 6. The effect of concentration of hydrochloric acid on precision and accuracy of AAS determination of Se has been studied. The best results were obtained with HCl (1 1). [Pg.293]

Equation (33) shows that the maximum capacity ratio of the last eluted solute is inversely proportional to the detector sensitivity or minimum detectable concentration. Consequently, it is the detector sensitivity that determines the maximum peak capacity attainable from the column. Using equation (33), the peak capacity was calculated for three different detector sensitivities for a column having an efficiency of 10,000 theoretical plates, a dead volume of 6.7 ml and a sample concentration of l%v/v. The results are shown in Table 1, and it is seen that the limiting peak capacity is fairly large. [Pg.208]

Figure 4.19 demonstrates the effect of sample concentration on the separation of polyethylene oxide (PEO). At a concentration of 1.6 mg/ml of each... [Pg.112]

Small particle size resins provide higher resolution, as demonstrated in Fig. 4.41. Low molecular weight polystyrene standards are better separated on a GIOOOHxl column packed with 5 /u,m resin than a GlOOOHg column packed with 10 /Ltm resin when compared in the same analysis time. Therefore, smaller particle size resins generally attain a better required resolution in a shorter time. In this context, SuperH columns are best, and Hhr and Hxl columns are second best. Most analyses have been carried out on these three series of H type columns. However, the performance of columns packed with smaller particle size resins is susceptible to some experimental conditions such as the sample concentration of solution, injection volume, and detector cell volume. They must be kept as low as possible to obtain the maximum resolution. Chain scissions of polymer molecules are also easier to occur in columns packed with smaller particle size resins. The flow rate should be kept low in order to prevent this problem, particularly in the analyses of high molecular weight polymers. [Pg.143]

Larger injection volumes, e.g., 2% of the total column volume, are sometimes advantageous in the preparative fractionation of polymers (33). More samples can be injected using larger injection volumes with a slight decrease in resolution. When the same amount of sample is injected with a smaller injection volume and a higher sample concentration, the resolution decreases more significantly. [Pg.145]

The sample concentration also should be kept as low as possible, particularly in analyses of polymers on columns packed with small particle size resins. The maximum sample concentration to achieve maximum resolution decreases as the sample molecular weight becomes higher and the resin particle size becomes smaller. It is usually in the range of 0.05-5 mg/ml, depending on the sample molecular weight and resin particle size. [Pg.145]

FIGURE 7.11 Influence of the protein concentration on the resolution. The standard separation from Fig. 7.2A was investigated with an increasing protein concentration (6.5, 17.2, and 32.5 mg protein/ run). Nearly the same resolution was obtained independent of the sample concentration. [Pg.235]

The dissolution of high molecular weight analytes increases the viscosity of the samples. Samples with a high viscosity can exhibit viscous fingering on the column top. Because this phenomenon can impair resolution, it is recommended to keep the sample concentration below certain limits. These limits are shown in Table 11.5. [Pg.343]

TABLE 11.5 Maximum Allowable Sample Concentrations as a Function of Molecular Weight Range... [Pg.343]

Sample loading must be reduced in accordance with the column inside diameter. Polymers exhibit high solution viscosity, and in order to avoid band broadening due to viscous streaming the sample concentration must be reduced for narrow-bore columns. Overloading effects become noticeable at much lower concentrations using 4.6-mm columns compared to 7.5-mm columns because of the effective sample concentration in a smaller volume column. [Pg.365]

SEC measurements were made using a Waters Alliance 2690 separation module with a 410 differential refractometer. Typical chromatographic conditions were 30°C, a 0.5-ml/min flow rate, and a detector sensitivity at 4 with a sample injection volume of 80 fil, respectively, for a sample concentration of 0.075%. All or a combination of PEO standards at 0.05% concentration each were used to generate a linear first-order polynomial fit for each run throughout this work. Polymer Laboratories Caliber GPC/SEC software version 6.0 was used for all SEC collection, analysis, and molecular weight distribution overlays. [Pg.502]

Sample concentration, and hence enrichment, is certainly a key issue in this area of analysis, since complementary information obtained from NMR or IR spectroscopic detection is often desirable in conjunction with mass spectrometry. Detection methods such as these have far higher concentration thresholds than MS and obtaining adequate quantities of material for detection becomes a significant challenge. [Pg.63]

Figure 6.8 Schematic diagram of a typical interface used for on-line SFE-SFC coupling (from ref. 42) 1, pump 2, heated transfer line 3, valve 4, sample concentrator 5, valve 6, SFC unit. Figure 6.8 Schematic diagram of a typical interface used for on-line SFE-SFC coupling (from ref. 42) 1, pump 2, heated transfer line 3, valve 4, sample concentrator 5, valve 6, SFC unit.
The concentration mode is used when the extract from the SEE module is not concentrated enough to be directly analyzed by the CE instrument, and thus requires a concentration step, which is carried out in the concentrator device. In this mode, during the concentration step the extract from the SEE cell enters the first valve (positioned in the concentration mode (Eigure 6.11(a)), and is then directed to and adsorbed into the sample concentrator which contains an SPE cartridge. While the... [Pg.144]

The direct mode is used when the concentration of the SEE extract is enough for direct analysis in the CE instrument without the need for a pre-concentration step. In this case, the sample concentrator is by-passed and the SEE extract goes directly to the CE instrument. The extract is collected in a CE vial containing an appropriate solvent and is thus ready for the CE analysis (Eigure 6.12). [Pg.145]


See other pages where Sampling concentration is mentioned: [Pg.28]    [Pg.251]    [Pg.401]    [Pg.57]    [Pg.124]    [Pg.165]    [Pg.174]    [Pg.235]    [Pg.378]    [Pg.196]    [Pg.38]    [Pg.70]    [Pg.70]    [Pg.84]    [Pg.84]    [Pg.84]    [Pg.100]    [Pg.144]    [Pg.145]    [Pg.154]    [Pg.234]    [Pg.343]    [Pg.365]    [Pg.462]    [Pg.490]    [Pg.144]    [Pg.145]   


SEARCH



Airborne asbestos sampling analysis concentration

Analytical methods sample concentration

Arsenic Concentrations in Marine Samples

Capillary electrophoresis sample volume/concentration

Column operation sample concentration

Concentration from environmental water samples

Concentration of sample extract

Determination of sample concentrations

Effect of Sample Concentration on Response

Effect of sample concentration

Elemental concentration, pond samples

Elemental concentrations artifacts samples

Elemental concentrations source samples

Femtomole range, sample concentrations

High-concentration samples

Isokinetic sampling of particle concentration

Kuderna-Danish sample concentrator

Large-volume samples concentration methods

Nuclear magnetic resonance sample concentration

On-capillary sample concentration

Online sample concentration

Picomole range, sample concentrations

Reversed-phase HPLC sample concentration effects

Sample Gathering for Oil Concentration Measurement

Sample Storage for Oil Concentration Measurement

Sample application concentrating zones

Sample concentration

Sample concentration

Sample concentration and clean-up solid phase extraction

Sample concentrations, picomole

Sample extraction, filtration and concentration

Sample pre-concentration

Sample preparation concentration

Sample preparation solute concentration

Sample requirements concentration

Sampling average concentrations

Sampling concentration measurement, horizontal

Sampling times, drug concentrations

Solids concentration sampling methods

Sorption sample concentration

© 2024 chempedia.info