Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Column operation sample concentration

After drying (removal of water), the extract is quantitatively transferred into a Kudema-Danish flask equipped with a concentrator tube and a Snyder column for sample concentration. The apparatus is placed in a water bath and ether is evaporated out. Use boiling chips in all heating operations. The volume of the extract is concentrated down to 1 to 2 mL. [Pg.157]

SFE-SFC operation consists of extraction, trapping, and transfer to the column for chromatography. Concentration of the solutes of interest before sample introduction to SFC may be achieved by ... [Pg.438]

In the purge-and-trap procedure, vials filled to the brim with the water samples are loaded into an auto-sampler, and then when the unit is operating, samples are drawn, one by one, into a tube where helium sparging occurs. Because the THMs are volatile, the helium sparging draws them out of the samples. The helium-THM gaseous mixture then flows through a trap in which the THMs are adsorbed and concentrated. This is followed by a desorption step in which the desorbed THMs are guided to the GC column. A Hall detector is used. [Pg.342]

Gel permeation chromatograms were generated from a Waters Associates, Inc. GPC equipped with a refractive index detector. The following operating conditions were employed mobile phase, THF flow rate 1 ml/min., columns ICP, 10, 500, 100 A . Sample concentrations were prepared at 0.2% (w/w) a 100 microliter aliquot was used for molecular weight analysis. Standard polystyrene samples (Polymer Laboratories, Inc.) were used to create a calibration curve. [Pg.232]

The instruments for polymer HPLC except for the columns (Section 16.8.1) and for some detectors are in principle the same as for the HPLC of small molecules. Due to sensitivity of particular detectors to the pressure variations (Section 16.9.1) the pumping systems should be equipped with the efficient dampeners to suppress the rest pulsation of pressure and flow rate of mobile phase. In most methods of polymer HPLC, and especially in SEC, the retention volume of sample (fraction) is the parameter of the same importance as the sample concentration. The conventional volumeters— siphons, drop counters, heat pulse counters—do not exhibit necessary robustness and precision [270]. Therefore the timescale is utilized and the eluent flow rate has to be very constant even when rather viscous samples are introduced into column. The problems with the constant eluent flow rate may be caused by the poor resettability of some pumping systems. Therefore, it is advisable to carefully check the actual flow rate after each restarting of instrument and in the course of the long-time experiments. A continuous operation— 24h a day and 7 days a week—is advisable for the high-precision SEC measurements. THE or other eluent is continuously distilled and recycled. [Pg.492]

Operating Condition of GPC. The operating conditions were the same as before (2) except that the sample concentration was 0.5%. Solvent was 1,2,4-trichlorobenzene, temperature was 137°C, injection time was 120 sec, and the flow rate was 1 cc per minute. Four columns having nominal capacities of 7 X 106, 3 X 106, 105, and 103 were used. [Pg.105]

To optimize the operating conditions for the detoxification of the hydrolysates with resins, experiments were carried out with synthetic solutions. Table 3 shows how the detoxifying efficiency of the anionic resin Dowex 1 diminished with increasing treated volume. Aliquots of the solution eluted through the column were sampled after 120,230, and 250 mL. As a whole, after the early 600 mL, the column still had a fair cleaning efficiency for acids but a reduced removal efficiency toward the furan compounds the data show that the cleaning efficiency of the column was 99% for formic acid, 87% for acetic acid, 40% for HMF, and 46% for furfural. An optimized resin/solution ratio of 0.14 g/g was thus extrapolated. The addition of a subsequent detoxification step with the cationic resin Dowex-50W left the concentrations of the acids almost unchanged whereas it further reduced those of HMF and furfural by 14 and 26%, respectively. The detoxification of the hydrolysate was performed on the basis of this optimized detoxification setup. [Pg.546]

The advantage of the selective adsorption of a particular element oxidation state has been exploited for on-line element preconcentration and speciation analysis of Cr by FAAS. Cespon Romero et al. [21] described an FIA system employing a minicolumn made of a chelating resin containing poly(aminopho-sphonic) acid groups, able to selectively retain Cr(III) ions. An FIA manifold was employed for efficient preconcentration and subsequent elution of Cr(III) with a small volume of 0.5 M HC1. The original sample was also treated with ascorbic acid to reduce Cr(VI) to Cr(III) and total Cr is determined as Cr(III) after appropriate retention and elution. Eluates are introduced into an N20-acetylene flame connected to the column outlet. The concentration of Cr (VI) is obtained by difference. Employing a sample volume of 6.6 mL, LoD for total Cr is 0.2 pg l-1. A study of FI operational variables, interferences, and precision is reported for the analysis of tap, mineral, and river waters. [Pg.460]

Fused silica capillary columns are required. The columns shall demonstrate the required separation of all 2378-specific isomers whether a dual column or a single column analysis is chosen. Column operating conditions shall be evaluated at the beginning and end of each 12-hour period during which samples or concentration calibration solutions are analyzed (see Section 7.4). [Pg.444]

One of the most important decisions that is left to the analyst when operating a liquid chromatograph is the choice of detector sensitivity. In some instruments the output from the sensor is monitored continuously over its entire dynamic range and so sensitivity is not an optional experimental parameter. Nevertheless, in this case, the sample size determines the concentration range over which the eluted solutes are monitored and thus an optimum sample size must be chosen. The detector should never be operated at its maximum sensitivity unless such conditions are enjoined by limited sample size or column geometry. Provided that there is adequate sample available, and the sample concentration when eluted is within the linear dynamic range of the detector, the maximum sample size that the column can tolerate should be used. This ensures that the detector noise is always minimal... [Pg.440]


See other pages where Column operation sample concentration is mentioned: [Pg.1530]    [Pg.432]    [Pg.70]    [Pg.432]    [Pg.257]    [Pg.266]    [Pg.382]    [Pg.1011]    [Pg.295]    [Pg.39]    [Pg.164]    [Pg.490]    [Pg.78]    [Pg.53]    [Pg.178]    [Pg.53]    [Pg.10]    [Pg.262]    [Pg.164]    [Pg.237]    [Pg.352]    [Pg.25]    [Pg.275]    [Pg.79]    [Pg.24]    [Pg.677]    [Pg.50]    [Pg.528]    [Pg.4687]    [Pg.279]    [Pg.311]    [Pg.1834]    [Pg.36]    [Pg.541]    [Pg.316]    [Pg.153]    [Pg.363]    [Pg.1826]    [Pg.4686]    [Pg.473]    [Pg.745]   
See also in sourсe #XX -- [ Pg.230 ]




SEARCH



Sample concentration

Sampling concentration

Sampling operations

© 2024 chempedia.info