Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

High-concentration samples

It is important to note that with hydrodynamic injection a volume of sample is injected, which has representative amounts (concentration) of the sample constituents. This characteristic, together with a better precision, makes it the most widely used injection technique in CE. However, owing to the poor detectability in terms of concentration of the widely used detectors (see further), its application is rather limited to the analysis of high-concentration samples. In trace analysis, for example, electromigration is favored over hydrodynamic injection [42],... [Pg.602]

A further advantage of this system is that it can easily he integrated with a commercial ICP-AES or ICP MS instrument without software or electronic interface modifications. It will operate rehably for a wide range of viscosities and has apphcations not only in the oil industry but also for highly concentrated samples from other areas. It uses standard autoanalyser principles similar to the Technicon Auto Analyzers. [Pg.159]

The collected sediment and water samples were centrifuged for 0.5 hour at 200 rpm. The water was decanted and the volume measured prior to transfer to a 250 mL separatory funnel where it was extracted four times with four 50 mL volumes of diethyl ether for the high concentration samples and with 50 mL followed by two 25 mL volumes for the low concentration samples. The 2,4-D samples only were acidified to a pH of 2 with H2S0j, to aid in the solvent extraction. [Pg.41]

For the four herbicides, the soil fraction in the sample bottle was extracted by adding 50 mL of diethyl ether followed by agitation for 15 minutes on a wrist-action shaker. The diethyl ether was decanted and the high concentration samples were extracted three more times with 50 mL of diethyl ether by hand shaking the capped bottles for 2 to 3 minutes. The low concentration samples were extracted two additional times with 25 mL of diethyl ether. [Pg.41]

The volume of acetone is dependent on the original concentration of chlorophyll in the sample, the volume of the dried extract, and the desired concentration range for the analysis. Highly concentrated samples such as spinach may need to be diluted to ensure adherence to the linear range as prescribed by the response curves for each specific chlorophyll. [Pg.949]

Record the potentials in millivolts for samples (using 100-mL aliquots or portions diluted to 100 mL for highly concentrated samples) using the above procedure. Determine the concentration of NH3-N in the samples from the calibration curve. If samples were diluted, multiply the results by the dilution factors. [Pg.178]

Highly concentrated sample must be diluted such that MBAS concentration falls within the linear calibration range. Such concentrations, however, in routine environmental analyses are rarely encountered. [Pg.265]

Because memory effects (carryover) from high concentration samples may take place in the nebulizer and in the plasma torch, the system must be flushed between samples. To recognize and prevent memory effects, instrument blanks are analyzed after high concentration samples additional frequent instrument blanks intersperse analytical sequences as a QC measure. [Pg.232]

Determines whether memory effects (carryover) Every 10 samples in trace element analysis are present in the analytical system After high concentration samples in organic... [Pg.254]

VS and QC concentrations should reflect the expected study sample. For example, the plasma sample concentrations from intravenous or inhaler dosing would be very different. The concentration of samples around the Cmax region may exceed the ULOQ. In that case, a VS/QC should be prepared to mimic the high concentration samples and tested for dilution linearity to extend the assay range. [Pg.153]

Method validation was carried out in six (monkey) or seven (human) validation batches. Each batch included eight levels of standard calibrators, four levels of VS at concentrations at the LLOQ, low, mid, and high concentrations within the calibrator range, and two additional concentrations higher than the ULOQ. The >ULOQ VS were run at dilution factors of 20 and 400 to mimic the expected high concentration samples. The accuracy and precision data of the VS in human plasma are listed in Table 6.3. [Pg.168]

To understand the mechanism of water oxidation, it is necessary to characterize each of the S states. A variety of spectroscopic methods have been brought to bear on this problem. Electron paramagnetic resonance (EPR) and X-ray spectroscopies have been especially useful because these techniques allow the Mn complex to be probed directly. EPR spectroscopy has the restriction that the Mn complex must be paramagnetic to be studied. The S2 state is an odd-electron state, and EPR spectroscopy has been used extensively to study the Mn complex in the S2 state. X-ray spectroscopy has the advantage that any state of the Mn complex is observable. However, the successful application of EPR and X-ray spectroscopies requires that a specific S state be prepared in high yield in highly concentrated samples. [Pg.258]


See other pages where High-concentration samples is mentioned: [Pg.225]    [Pg.182]    [Pg.192]    [Pg.606]    [Pg.90]    [Pg.1249]    [Pg.292]    [Pg.51]    [Pg.123]    [Pg.123]    [Pg.127]    [Pg.312]    [Pg.407]    [Pg.128]    [Pg.129]    [Pg.205]    [Pg.56]    [Pg.17]    [Pg.1249]    [Pg.36]    [Pg.30]    [Pg.31]    [Pg.421]    [Pg.242]    [Pg.1079]    [Pg.36]    [Pg.739]    [Pg.46]    [Pg.156]    [Pg.166]    [Pg.35]    [Pg.146]    [Pg.152]    [Pg.204]    [Pg.76]    [Pg.291]    [Pg.196]    [Pg.256]   
See also in sourсe #XX -- [ Pg.288 ]




SEARCH



High Concentration

Sample concentration

Sampling concentration

© 2024 chempedia.info